K111025 · Smith & Nephew, Inc. · HSB · Jul 1, 2011 · Orthopedic
Device Facts
Record ID
K111025
Device Name
TRIGEN LOW PROFILE BONE SCREW
Applicant
Smith & Nephew, Inc.
Product Code
HSB · Orthopedic
Decision Date
Jul 1, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TRIGEN® Low Profile Bone Screw can be used with several types of nails in Smith & Nephew's (TRIGEN) Titanium Nail System. The TRIGEN® Low Profile Bone Screw therefore has the following indications: Indications for interlocking intramedullary nails include simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated. In addition to the indications for intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability (e.g. Femoral Antegrade Nail, Trochanteric Antegrade Nail and Femoral/Recon Antegrade Nail) are indicated for the following: subtrochanteric fractures, intertrochanteric fractures, and ipsilateral femoral shaft/neck fractures. In addition to the indications for interlocking intramedullary nails, devices that use a retrograde femoral surgical approach (e.g. Knee Nail; Retrograde and Supracondylar Nails) are indicated for the following: comminuted supracondylar fractures with or without intra-articular extension; fractures that require opening the knee joint to stabilize the femoral condylar segment; fractures above total knee implants (peri-prosthetic fractures). The TRIGEN® InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft fractures and segmental shaft fractures; subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; intracapsular fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; bone lengthening and shortening. SURESHOT® TAN Nails are indicated for fractures of the femur including simple long bone fractures, severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening; and for fixation of fractures that occur in and between the proximal third and distal fourth of the femur. In addition, SURESHOT TAN Nails contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability and are indicated for the following: subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; and intracapsular fractures. The TRIGEN® Low Profile Bone Screw is intended for single use only.
Device Story
TRIGEN® Low Profile Bone Screw is a titanium alloy (Ti-6Al-4V) orthopedic implant designed for use with Smith & Nephew's TRIGEN® and SURESHOT® intramedullary nail systems. The screw features an internal hex drive and specific thread/tip geometry to provide fixation for long bone fractures. It is intended for single-use by surgeons in clinical/OR settings. The device functions as a mechanical fastener to provide compression and rotational stability when inserted through nail holes/slots into bone. Clinical benefit is derived from stabilizing fractures to facilitate healing. No software or algorithms are involved.
Clinical Evidence
No clinical data provided. Safety and effectiveness supported by bench testing, including static torsional strength, bending fatigue, and axial pullout strength.
Technological Characteristics
Material: Ti-6Al-4V titanium alloy. Dimensions: 4.5 mm and 5.0 mm diameters; lengths 20-110 mm. Features: Internal hex drive, specific thread form and cutting tip geometry. Single-use. No software or energy source.
Indications for Use
Indicated for patients requiring fixation of long bone fractures, including simple, comminuted, spiral, oblique, segmental, nonunions, malunions, polytrauma, pathologic fractures, tumor resection/grafting, bone lengthening/shortening, and peri-prosthetic fractures. Specific applications include femoral shaft/neck, subtrochanteric, intertrochanteric, and supracondylar fractures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
TRIGEN Titanium Intramedullary (IM) Nail System (K981529)
{0}------------------------------------------------
K111025(pg 1/3)
JUL - 1 2011
Smith & Nephew, Inc. Summary of Safety and Effectiveness TRIGEN® Low Profile Bone Screw
Date of Summary: 04/12/2011
Contact Person and Address Natalie P. Williams Regulatory Affairs Specialist Smith & Nephew, Inc. 7135 Goodlett Farms Parkway Cordova, TN 38016 (901) 399-5161
Name of Device: TRIGEN® Low Profile Bone Screw Common Name: Bone Screw Device Classification Name and Reference: 21 CFR 888.3020 Intramedullary Fixation Rod Device Class: II Panel Code: Orthopaedics/87 HSB
#### Device Description
Subject of this Traditional 510(k) Premarket Notification is the TRIGEN® Low Profile Bone Screw. The subject device is a modification of the bone screws included in the (TRIGEN) Titanium Intramedullary (IM) Nail System cleared by premarket notification K981529. The TRIGEN® Low Profile Bone Screw is manufactured from titanium alloy (Ti-6Al-4V), contains an internal hex drive feature in the screw head, and is available in the sizes included in Table 1 below.
Table 1: TRIGEN® Low Profile Screw Sizes
| Diameter | Length Range |
|----------|-----------------------------------|
| 4.5 mm | 20 mm - 65 mm, 2.5 mm increments |
| 5.0 mm | 20 mm – 80 mm, 2.5 mm increments |
| 5.0 mm | 80 mm - 110 mm, 5.0 mm increments |
The subject device is designed to be used with the SURESHOT® TAN Nails cleared under K092748 as well as TRIGEN® titanium intramedullary nail systems designed to use 4.5 mm or 5.0 mm diameter screws.
#### Intended Use
The TRIGEN® Low Profile Bone Screw can be used with several types of nails in Smith & Nephew's (TRIGEN) Titanium Nail System. The TRIGEN® Low Profile Bone Screw therefore has the following indications:
Indications for interlocking intramedullary nails include simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction following tumor resection and grafting; supracondylar fractures; bone lengthening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated.
In addition to the indications for interlocking intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability (e.g. Femoral
{1}------------------------------------------------
Antegrade Nail, Trochanteric Antegrade Nail and Femoral/Recon Antegrade Nail) are indicated for the following: subtrochanteric fractures, intertrochanteric fractures, and ipsilateral femoral shaft/neck fractures.
In addition to the indications for interlocking intramedullary nails, devices that use a retrograde femoral surgical approach (e.g. Knee Nail, Retrograde and Supracondylar Nails) are indicated for the following: comminuted supracondylar fractures with or without intra-articular extension; fractures that require opening the knee joint to stabilize the femoral condylar segment; fractures above total knee implants (peri-prosthetic fractures).
The TRIGEN® InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft fractures and segmental shaft fractures; subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; intracapsular fractures; nonunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; bone lengthening and shortening.
SURESHOT® TAN Nails are indicated for fractures of the femur including simple long bone fractures, severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening; and for fixation of fractures that occur in and between the proximal third and distal fourth of the femur.
In addition, SURESHOT* TAN Nails contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability and are indicated for the following: subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; and intracapsular fractures.
The TRIGEN® Low Profile Bone Screw is intended for single use only.
#### Performance Data
Performance testing has been conducted to ensure the safety and effectiveness of the subject device. Static Torsional Strength, Bending Fatigue, and Axial Pullout Strength of the subject device have been evaluated. A review of the testing has demonstrated that there are no new issues related to the safety or effectiveness of the subject device.
Clinical data was not needed to support the safety and effectiveness of the subject device.
#### Substantial Equivalence Information
The overall design, indications for use, intended use, materials, and sterilization of the TRIGEN Low Profile Bone Screw is substantially equivalent to the bone screws included in the (TRIGENY) Titanium IM Nail System cleared under premarket notification K981529. The indications and intended use of the subject device is also similar to the bone screws included in the (Russell-Taylor) IM Nail System cleared via K983942. Design features of the TRIGEN® Low Profile Bone Screw have been compared to the previously cleared devices in Table 2 below.
{2}------------------------------------------------
| Device Comparison | TRIGEN ® Low Profile Bone Screw<br>(Subject Device) | (TriGen ®) Titanium Nail System<br>(K981529) | (Russell –Taylor ®) IM Nail System (K983942) |
|---------------------------------|-------------------------------------------------------------|------------------------------------------------------------------------|------------------------------------------------------|
| Material | Ti-6Al-4V | Ti-6Al-4V | Stainless Steel |
| Diameter | 4.5 mm and 5.0 mm | 4.5 mm and 5.0 mm | 4.5 mm and 5.0 mm |
| Length | | | |
| 4.5 mm Diameter | 20 - 65 mm,<br>2.5 mm increments | 20 - 65 mm,<br>5.0 mm increments | 20 - 65 mm,<br>5.0 mm increments |
| | 5.0 mm Diameter | 20 - 80 mm,<br>2.5 mm increments;<br>80 - 110 mm,<br>5.0 mm increments | 20 - 110 mm,<br>5.0 mm increments |
| Head Drive | Internal Hex | External Hex | Internal Hex |
| Similar Thread Form | Y | Y | Y |
| Similar Cutting Tip<br>Geometry | Y | N | Y |
Table 2: Comparison of the TRIGEN® Low Profile Bone Screw to Predicate Devices
# Conclusion
.
As previously noted, this premarket notification is being submitted to request clearance for the TRIGEN® Low Profile Bone Screw. Based on the similarities to the predicate devices and a review of the testing, the device is substantially equivalent to the bone screws currently marketed under premarket notifications K981529 and K983942.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Smith & Nephew, Inc. % Natalie P. Williams Regulatory Affairs Specialist 7135 Goodlett Farms Parkway Cordova, TN 38016
Re: K111025
Trade/Device Name: TRIGEN Low Profile Bone Screw Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: April 12, 2011 Received: April 13, 2011
JUL - 1 2011
Dear Ms. Williams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page - 2 - Ms. Natalie P. Williams
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm1 for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/Medical/Devices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
cerely yours,
Fa. Dettman, Jr.
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): Kl(1025 (pg 1/2)
Device Name: TRIGEN® Low Profile Bone Screw
### Indications for Use:
The TRIGEN® Low Profile Bone Screw can be used with several types of nails in Smith & Nephew's (TRIGEN) Titanium Nail System. The TRIGEN® Low Profile Bone Screw therefore has the following indications:
Indications for interlocking intramedullary nails include simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated.
In addition to the indications for intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability (e.g. Femoral Antegrade Nail, Trochanteric Antegrade Nail and Femoral/Recon Antegrade Nail) are indicated for the following: subtrochanteric fractures, intertrochanteric fractures, and ipsilateral femoral shaft/neck fractures.
In addition to the indications for interlocking intramedullary nails, devices that use a retrograde femoral surgical approach (e.g. Knee Nail; Retrograde and Supracondylar Nails) are indicated for the following: comminuted supracondylar fractures with or without intra-articular extension; fractures that require opening the knee joint to stabilize the femoral condylar segment; fractures above total knee implants (peri-prosthetic fractures).
The TRIGEN® InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft fractures and segmental shaft fractures; subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; intracapsular fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; bone lengthening and shortening.
SURESHOT® TAN Nails are indicated for fractures of the femur including simple long bone fractures, severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening; and for fixation of fractures that occur in and between the proximal third and distal fourth of the femur.
In addition, SURESHOT TAN Nails contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability and are indicated for the following:
Continued on next page →
{6}------------------------------------------------
## Indications for Use
subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; and intracapsular fractures.
The TRIGEN® Low Profile Bone Screw is intended for single use only.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ilor M. Melkenon
(Division Sign-Off) (Division of Surgical, Orthopedic, Division of Storative Devices
Page 2 of __
K11025
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.