K111017 · Ultradent Products, Inc. · EMA · Jun 6, 2011 · Dental
Device Facts
Record ID
K111017
Device Name
CLEARTEMP
Applicant
Ultradent Products, Inc.
Product Code
EMA · Dental
Decision Date
Jun 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3275
Device Class
Class 2
Attributes
Pediatric
Indications for Use
ClearTemp is a provisional cement used for luting of thin translucent veneer restorations.
Device Story
ClearTemp is a light-curable, temporary resin-based cement; used by dental professionals for luting thin translucent veneer restorations. Device functions as a bonding agent; applied during clinical procedures to secure provisional veneers. Performance verified via bench testing including flexural strength, hardness, compressive strength, depth of cure, ambient light sensitivity, sorption, film thickness, and radiopacity. Biocompatibility assessed per ISO 10993-1:2009. Output is a cured cement layer; aids in temporary restoration stability. Benefits include secure temporary fixation of veneers; clinical safety supported by predicate history and bench performance data.
Clinical Evidence
No clinical trials were conducted. Evidence consists of a literature study supporting safety and effectiveness, combined with bench testing comparing ClearTemp to the predicate device (TempBond Clear). Bench tests included flexural strength, hardness, compressive strength, depth of cure, ambient light sensitivity, sorption, film thickness, and radiopacity. Biocompatibility testing (cytotoxicity, sensitization, irritation, and genotoxicity) was performed according to ISO 10993-1:2009.
Technological Characteristics
Light-curable, resin-based dental cement. Technological characteristics include flexural strength, hardness, compressive strength, depth of cure, ambient light sensitivity, sorption, film thickness, and radiopacity. Biocompatibility testing performed per ISO 10993-1:2009. No software or electronic components.
Indications for Use
Indicated for luting of thin translucent veneer restorations in patients of any age who have had veneers professionally installed.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
Predicate Devices
TempBond Clear by Sybron Dental Specialties (K053565)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a sequence of characters that appear to be handwritten. The sequence starts with the letter 'K', followed by three instances of the number '1'. After that, there is a '0', another '1', and finally a '7'. The characters are written in a simple, clear style, making them easily readable.
# Section 5
JUN - 6 2011
## Traditional 510(k) Summary
This summary of the Traditional 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR 807 .92.
## Applicant's Name and Address
Ultradent Products Inc. 505 West 10200 South South Jordan, UT 84095
| Contact Person: | |
|------------------------|--|
| Title: | |
| Telephone: | |
| FAX: | |
| Date Summary Prepared: | |
Diane Rogers Regulatory Affairs Manager 800-552-5512 x4491, 801-553-4491 801-553-4609 May 31, 2011
## Name of the Device
Trade Name: Common Name: Device Classification: Classification Product Code: ClearTemp Dental Cement = EMA
#### Legally Marketed Predicate Devices to Which Equivalence is Claimed
The predicate device is: TempBond Clear by Sybron Dental Specialties (K053565)
Description of Device: ClearTemp is a light curable, temporary cement used specifically for luting thin translucent veneers on teeth.
Intended Use of the Device: ClearTemp is a provisional cement used for luting of thin translucent veneer restorations.
The following table shows that each of these devices are similar as test results comparing these devices show similar results. The devices are so equivalent that both of these companies follow standards that
4 Page
{1}------------------------------------------------
show us how to test this type of a device and in Section 18 "Performance and Bench", you will see that test results show the devices have the same technological characteristics.
| Function | ClearTemp | TempBond Clear<br>(K6053565) |
|-------------------|-----------|------------------------------|
| Flexural Strength | × | × |
| Hardness | × | × |
| Compressives | × | × |
| Cure Time | × | × |
| Depth of Cure | × | × |
| Sorption Solution | × | × |
| Radiopacity | × | × |
| Ambient Light | × | × |
| Sensitivity | | |
| Film Thickness | × | × |
## Similarities in the Indications for Use:
| ClearTemp | | EMA | Provisional cement used<br>for luting of thin<br>translucent veneer<br>restorations. |
|----------------|---------|-----|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| TempBond Clear | K053565 | EMA | TempBond Clear with<br>Triclosan is a dual cured,<br>temporary, eugenol free,<br>transparent, resin based<br>cement with triclosan<br>designed to be used in<br>conjunction with<br>temporary restorations<br>such as crowns, bridges,<br>inlays, onlays and splints. |
{2}------------------------------------------------
#### Brief Description of Testing Performed
The following bench tests were conducted during the R & D phase on ClearTemp and compared to Sybron Dental Specialties TempBond Clear (K053565)
- a. Flexural Strength This test will show the strength of the bond during stress. A higher number than our competitors is good. The modulus side of this test shows the strength at which flexing the bond occurs. We prefer the product to be comparable to most of our competitors.
- b. Hardness This test shows the hardness of the bond. We prefer to stay within our competitors range.
- Compressives This test shows different forces on the resin. High numbers according to our ن competitors is acceptable.
- d. Depth of Cure This test shows how far a curing light penetrates into the cure. We want to stay at the high end of our competitors.
- e. Ambient Light Sensitivity This test shows the that the product will cure in ambient light. It shows working time with the product and cure time of the product. We want low times in this category.
- ﻨﺖ Sorption – This test shows how much water the resin absorbs. We want low readings on this test.
- Film thickness testing bond strength at a defined thickness. g.
- h. Radiopacity verifying how clear the product shows up during an Xray.
- i. Sorption Soluability- testing whether the product degrades in solutions or saliva
#### Clinical Summary
A complete Clinical Summary of ClearTemp is included in Section 20. We conducted a literature study to show safety and effectiveness of ClearTemp. ClearTemp is a provisional cement used for luting of thin translucent veneer restorations. The product can be used on any age patient when they have had veneers professionally installed. The device has the same technological characteristics.compared to Sybron Dental Specialities TempBond Clear. These materials have been widely used by numerous manufacturers in the dental industry.
The efficacy or suitability to the intended purpose of ClearTemp has been demonstrated by a combination of in-house testing and side-by-side comparisons to predicate devices currently on the market. Results of our bench testing indicates that ClearTemp performs as well or better than the predicate device currently on the market.
## Summary Risk/Benefit Review
Considering the safe history of our predicate, TempBond Clear by Sybron Dental Specialities (K053565) ClearTemp is considered a safe medical device. Our research indicates that our predicate has been used by many dentists and large group practices in the United States and purchased by a large
6 | Paul C
{3}------------------------------------------------
number of international distributors. To date, there have been no reported complaints of local or systemic adverse effects associated with the use of these products.
ClearTemp was tested for biocompatibility in Cytotoxicity, Sensitization, Irritation and Genotoxicity tests according to ISO 10993-1:2009. An abstract of the testing.along with signed test reports are included in Section 15 "Biocompatibility" of this submission.
In conclusion, ClearTemp has been designed and manufactured with the intended use and claims for the product in mind. Scientific literature, etc. has been collected and evaluated to determine safety and efficacy of similar products used for the same indication. Following the clinical review as documented above, Ultradent Products, Inc. deems that when this device is used under the conditions and for the purposes intended, it will not compromise the clinical condition or the safety of the patient and the association with its use constitutes acceptable risks when weighed against the benefits to the patient. Therefore, the product is compatible with a high level of protection of health and safety and may be released to the market.
## 7 Page
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Diane Rogers Regulatory Affairs Manager Ultradent Products, Incorporation 505 West 10200 South South Jordan, Utah 84095
Re: K111017
Trade/Device Name: ClearTemp Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: April 1, 2011 Received: April 12, 2011
JUN - 6 2011
Dear Ms. Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2- Ms. Rogers
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Th for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{6}------------------------------------------------
## Statement of Indications for Use
510(k) Number (if known):
Device Name: ClearTemp
Indications for Use:
ClearTemp is a provisional cement used for luting of thin translucent veneer restorations.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
510(k) Number:
Susan Ruane
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
(Posted November 13, 2003)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.