Microware Bone Plates and Bone Screws are provided non-sterile. Microware Bone Plates and Bone Screws are intended to treat fractures of various bones, including the clavicle, pelvis, scapula, long bone (humerus, ulna, radius, femur, tibia and fibula), and small bone (metacarpals, metatarsals and phalanges). Microware Bone Plates and Bone Screws Accessories: 1. The round washer is used to increase bone contact area for distributing the forces/load and prevent the screw head from sinking into the bone. 2. The spherical washer is used to increase bone contact area for distributing the forces/load, and prevent the screw head from sinking into the bone. In addition, it can be applicable on the insertion at acute angles. 3. The spiked washer is used for ligament reattachment or fixation. 4. The support screw is anchored in cancellous bone by its thread. It supports the countersink head of a cannulated screw in cancellous bone.
Device Story
Microware Bone Plates and Bone Screws are orthopedic fixation implants; used for fracture stabilization in various bones. System includes cortex, cancellous, and shaft screws with self-tapping flutes; hexalobular sockets for torque transmission; cruciform recesses; and cannulated screws. Accessories include round, spherical, and spiked washers for load distribution and ligament fixation, plus support screws for cancellous bone. Implants are non-sterile; intended for surgical use by orthopedic surgeons. Device functions as mechanical fixation; provides structural support to bone fragments during healing. Benefits include improved insertion ease, enhanced torque control, and versatile fixation options for diverse fracture types.
Clinical Evidence
Bench testing only. Performance evaluations included axial pullout, self-tapping performance, driving torque, and torsional properties. No clinical data provided.
Technological Characteristics
Materials: stainless steel and titanium alloy. Features: self-tapping flutes, hexalobular sockets, cruciform recesses, cannulated designs. Accessories: round, spherical, and spiked washers; support screws. Non-sterile; requires sterilization by user.
Indications for Use
Indicated for patients requiring fracture treatment of the clavicle, pelvis, scapula, long bones (humerus, ulna, radius, femur, tibia, fibula), and small bones (metacarpals, metatarsals, phalanges). Accessories indicated for load distribution, prevention of screw head sinking, acute angle insertion, and ligament reattachment/fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
#### Special 510(k) Summary of Safety and Effectiveness 3.
Submitter's Name: Microware Precision Co., Ltd.
Address: No. 12, Keyuan 2nd Rd., Situn District, Taichung City 40763, Taiwan Phone No: +886-4-24636275
Fax No: +886-4-24636276
Contact Person: Harrison Du
Preparation date: October/02/2011
Registration Number: 3007738812
Device Name: Microware Bone Plates and Bone Screws
Proprietary Name: Microware Bone Plates and Bone Screws
Classification Name: Class II
CFR 888.3040: screw, fixation, bone- HWC CFR 888.3030: washer, bolt, non-spinal, metallic- NDG CFR 888.3030: plate, fixation, bone-HRS CFR 888.3030: appliance, fixation, nail/blade/plate combination, multiple component-KTT
ﺮﺍ
## Predicate Device Information:
Microware Bone Plates and Bone Screws (K072562) Kens Internal Fixation System (K090786) Synthes 2.4mm Cannulated Screw (K012945) Synthes Metallic Spiked Washers (K013806) Synthes Spherical Washers (K052483)
{1}------------------------------------------------
K111008 #2/3
## Materials:
The added devices are manufactured from stainless steel and titanium alloy that are the same as Microware Bone Plates and Bone Screw (K072562).
## Indication for use:
Microware Bone Plates and Bone Screws are provided non-sterile. Microware Bone Plates and Bone Screws are intended to treat fractures of various bones, including the clavicle, pelvis, scapula, long bone (humerus, ulna, radius, femur, tibia and fibula), and small bone (metacarpals, metatarsals and phalanges).
Microware Bone Plates and Bone Screws Accessories:
- 1. The round washer is used to increase bone contact area for distributing the forces/load and prevent the screw head from sinking into the bone.
- 2. The spherical washer is used to increase bone contact area for distributing the forces/load, and prevent the screw head from sinking into the bone. In addition, it can be applicable on the insertion at acute angles.
- 3. The spiked washer is used for ligament reattachment or fixation.
- 4. The support screw is anchored in cancellous bone by its thread. It supports the countersink head of a cannulated screw in cancellous bone.
#### Device Description:
The added devices do not change the original intended use, material, label, and package. The added devices with changes are described as follows:
- Added cortex, cancellous, and shaft screws are designed with self-tapping flutes and are 1. easier to insert into the bone than non-self-tapping screws.
- Added cortex screws 1.5 and 2.0mm are designed with a cruciform recess and can be applied 2. with a cruciform tool.
- Added screws of all sizes are designed with a hexalobular socket and they can improve torque 3. transmission to facilitate easy insertion and removal.
- Added cannulaed screws extend diameter specification with 6.5, 4.5, 4.0, 3.0 and 2.4mm. 4.
- Added implants include the round washer, spherical washer, spiked washer, and support న్. screw which can be used with or without other implants.
- Added screws extend length specifications. 6.
{2}------------------------------------------------
KIII008 #3/3
In addition, a new set of recommended sterilization parameters are added to the "Instruction for use".
# Substantial Equivalence:
Performance testing included axial pullout, self-tapping performance, driving torque, torsional properties and the results demonstrated substantial equivalence to the predicate devices.
A comparison has shown that the proposed device is very similar or identical in terms of indication for use, material, followed performance and standard, and sterilization to the predicate devices, and no significant difference between the proposed and predicate devices has been found. Thus, the "Microware Bone Plates and Bone Screws" is substantially equivalent to the predicate devices.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
ﮨﯿﮟ ﮨﮯ ۔
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three curved lines representing its wings and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Microware Precision Co., Ltd. % Mr. Harrison Du No. 12, Keyuan 2nd Road Situn District Taichung City, Taiwan 40763
MAY - 2 2012
Re: K111008
Trade/Device Name: Microware Bone Plates and Bone Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC, KTT, NDG Dated: April 02, 2012
Received: April 2, 2012
Dear Mr. Du:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vithe Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical OFF at 807), accimig (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
### Page 2 - Mr. Harrison Du
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to .
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
Eunice Keith
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### Indications for Use 2.
# 510{k) Number (if known):
Device Name: Microware Bone Plates and Bone Screws Indications for Use:
Microware Bone Plates and Bone Screws are provided non-sterile. Microware Bone Plates and Bone Screws are intended to treat fractures of various bones, including the clavicle, pelvis, scapula, long bone (humerus, ulna, radius, femur, tibia and fibula), and small bone (metacarpals, metatarsals and phalanges).
Microware Bone Plates and Bone Screws Accessories:
- 1. The round washer is used to increase bone contact area for distributing the forces/load and prevent the screw head from sinking into the bone.
- 2. The spherical washer is used to increase bone contact area for distributing the forces/load and prevent the screw head from sinking into the bone. In addition, it can be applicable on the insertion at acute angles.
- 3. The spiked washer is used for ligament reattachment or fixation.
- 4. The support screw is anchored in cancellous bone by its thread. It supports the countersink head of a cannulated screw in cancellous bone.
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | |
|------------------------|--|
| (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
4
*(Division Sign-Off)*
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K111008
Page 1 of 1
510(k) Submission ver.S003
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.