The Digital Ultrasonic Diagnostic Imaging System DUS 60 is intended for diagnostic ultrasound imaging analysis in gynecology rooms. obstetrics rooms, examination rooms, intensive care units, and emergency rooms. The system is intended for use by or on the order of a physician or similarly qualified health care professional for ultrasound evaluation of Fetus, Abdomen, Pediatrics, Small Organ, Neonatal Cephalic, Cardiology, Peripheral Vessel, Musculo-skeleton (both Conventional and Superficial). Urology (including prostate). Transrecta and Transvagina.
Device Story
Portable digital ultrasonic diagnostic imaging system; utilizes array of probes (linear, convex, microconvex, transrectal, transvaginal) with 2.0-10.0 MHz frequency range. Transforms ultrasound waves reflected from body tissue into visual information; displays in B-Mode, M-Mode, B+M Mode, and PW Mode. Used in clinical settings (gynecology, obstetrics, ICU, emergency rooms) by physicians or qualified healthcare professionals. Provides real-time diagnostic imaging for clinical evaluation; assists in diagnosis of various anatomical structures and conditions. Benefits patients through non-invasive diagnostic imaging.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1, IEC 60601-2-37), electromagnetic compatibility (IEC 60601-1-2), acoustic output testing per FDA guidelines, and biocompatibility (ISO 10993-1, -5, -10).
Technological Characteristics
Portable ultrasound system; frequency range 2.0-10.0 MHz. Probes: linear, convex, microconvex, transrectal, transvaginal. Modes: B, M, B+M, PW. Biocompatibility per ISO 10993. Electrical safety per IEC 60601-1/2-37. EMC per IEC 60601-1-2.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in patients requiring evaluation of fetus, abdomen, pediatrics, small organs, neonatal cephalic, cardiology, peripheral vessels, musculoskeletal (conventional/superficial), and urology (including prostate, transrectal, and transvaginal).
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
DUS 6 Digital Ultrasonic Diagnostic Imaging System (K091680)
{0}------------------------------------------------
. ▼ - ..
ﺘﺮ
# Section 5 - 510(k) Summary
# APR 2 6 2011
9
K110999
### Prepared in accordance with the requirements of 21 CFR Part 807.92
. . . . .
| 1. Submitter: | Edan Instruments, Inc.<br>3/F-B, Nanshan Medical Equipments Park, Nanhai Rd 1019#, Shekou,<br>Nanshan Shenzhen, 518067 P.R. China<br>Tel.: (0755) 26858736<br>Fax: (0755) 26882223 |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Randy Jiang |
| Prepare date: | Feb 06, 2011 |
| 2. Device name<br>and classification: | Device Name:<br>Digital Ultrasonic Diagnostic Imaging System, Model DUS 60<br>Classification Name:<br>892.1560 System, Imaging, Pulsed echo, Ultrasonic<br>Product code: IYO<br>892.1570 Transducer, Ultrasonic, Diagnostic<br>Product code: ITX<br>Regulatory Class: Class II |
| 3. Predicate<br>Device: | DUS 6 Digital Ultrasonic Diagnostic Imaging System / K091680 / Edan Instruments, Inc.<br>DC-3 Diagnostic Ultrasound System / K081320 / Shenzhen Mindray Bio-medical Electronics Co., Ltd. |
| 4. Device<br>Description: | DUS 60 Digital Ultrasound Diagnostic Imaging System is a portable digital ultrasonic diagnostic system applied in ultrasound diagnostic examination of abdominal, obstetrical, small parts, gynecological, orthopedic, cardiac, and urological applications.<br>It is designed to produce ultrasound waves into body tissue and to present the returned echo information on the monitor; the resulting information is displayed in the following display modes: B-Mode, M-Mode, B+M Mode and PW Mode. This system controlled by software is a Track 3 device that employs an array of probes that include linear array, convex linear array, microconvex linear array, transrectal and transvaginal with a frequency range of approximately 2.0MHz-10.0MHz. |
{1}------------------------------------------------
### 5. Intended Use:
The Digital Ultrasonic Diagnostic Imaging System DUS 60 is intended for diagnostic ultrasound imaging analysis in gynecology rooms. obstetrics rooms, examination rooms, intensive care units, and emergency rooms. The system is intended for use by or on the order of a physician or similarly qualified health care professional for ultrasound evaluation of Fetus, Abdomen, Pediatrics, Small Organ, Neonatal Cephalic, Cardiology, Peripheral Vessel, Musculo-skeleton (both Conventional and Superficial). Urology (including prostate). Transrecta and Transvagina.
### 6. Effectiveness and Safety Considerations:
Clinical Test: Clinical testing is not required.
#### Non-clinical Test:
The following safety standards are conducted on the subject device:
- (1) IEC 60601-1 and IEC 60601-2-37 Electrical Safety
(2) IEC 60601-1-2 Electromagnetic Compatibility
(3) Acoustic output testing as per the guideline "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" dated September 9, 2008.
(4) ISO 10993-1, ISO 10993-5 and ISO 10993-10
### 7. Comparison to the Predicate Device
Comparison to the predicate devices, the subject device has the similar technology characteristics and has the same intended use, same design principle, same electrical classification and same accuracy. The mainly difference between the subject device and predicate devices primarily includes physical specifications, display monitor and display mode. All above differences do not affect the usage, safety and effectiveness, and no new question is raised regarding the safety and effectiveness.
### 8. Substantially Equivalent Determination
The related tests (including safety, EMC, performance and biocompatibility) of the entire device were conducted on the DUS 60 Digital Ultrasonic Diagnostic Imaging System. This premarket notification submission demonstrates that DUS 60 Digital Ultrasonic Diagnostic Imaging System is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
APR 2 6 2011
Edan Instruments % Mr. Marc M. Mouser Engineering Leader & FDA Office Coordinator Underwriters Laboratories, Inc. 2600 N.W. Lake Road CAMAS WA 98607-8542
Re: K110999
Trade/Device Name: Digital Ultrasonic Imaging Systems, Model DUS 60 Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO and ITX Dated: March 8, 2011 Received: April 11, 2011
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the Digital Ultrasonic Imaging Systems, Model DUS 60, as described in your premarket notification:
Transducer Model Number
| C363UA | L742UA |
|--------|--------|
| C362UA | L763UA |
| C343UA | E743UA |
| C321UA | E613UA |
| L743UA | C613UA |
{3}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Paul Hardy at (301) 796-6542.
Sincerely Yours,
Mary S Pastil
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
Digital Ultrasonic Diagnostic Imaging System 510K Submission
# Section 6 - Indications for Use
510(k) Number (if known):
Device Name: Digital Ultrasonic Diagnostic Imaging System, Model DUS 60
The Digital Ultrasonic Diagnostic Imaging System DUS 60 is intended for diagnostic ultrasound imaging analysis in gynecology rooms, obstetrics rooms, examination rooms, intensive care units, and emergency rooms. The system is intended for use by or on the order of a physician or similarly qualified health care professional for ultrasound evaluation of Fetus, Abdomen, Pediatrics, Small Organ, Neonatal Cephalic, Cardiology, Peripheral Vessel, Musculo-skeleton (both Conventional and Superficial), Urology (including prostate), Transrecta and Transvagina.
| Prescription Use | X |
|-----------------------------|---|
| (21 CFR Part 801 Subpart D) | |
| Or Over the Counter Use | |
| (21 CFR Part 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IFNEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Mary S Postl
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number K110999
Page 1 of 12
{5}------------------------------------------------
#### DUS 60 Digital Ultrasonic Diagnostic Imaging System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------------|--------------------|------------------|
| | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | N | N | N | | | N | Note 1 |
| Abdominal | N | N | N | | | N | Note 1 |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | N | | | N | Note 1 |
| Small Organ (Specify) | N | N | N | | | N | Note 1 |
| Neonatal Cephalic | N | N | N | | | N | Note 1 |
| Adult Cephalic | | | | | | | |
| Transrectal | N | N | N | | | N | Note 1 |
| Transvaginal | N | N | N | | | N | Note 1 |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | N | N | N | | | N | Note 1 |
| Musculo-skeletal (Superficial) | N | N | N | | | N | Note 1 |
| Intravascular | | | | | | | |
| Cardiac | N | N | N | | | N | Note 1 |
| Intravascular | | | | | | | |
| Peripheral vascular | N | N | N | | | N | Note 1 |
| Other (Urology) | N | N | N | | | N | Note 1 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M;
Note 1: This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Mary S. Castle
(Division Sign Off)
Orvision Sinr Division of Radiological De Office of In Evaluation and Safety 510K
Page 2 of 12 6-2
{6}------------------------------------------------
#### DUS 60 with C363UA Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------------|--------------------|------------------|
| Clinical Application | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | N | N | N | | | N | Note 1 |
| Abdominal | N | N | N | | | N | Note 1 |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | N | | | N | Note 1 |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | N | N | N | | | N | Note 1 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M;
Note 1: This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Prescription Use (Per 21 CFR 801.109)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Mary Slatu
(Division Sign-Off)
Division of Radiological Review
Page 3 of 12
Division of Radiological Device Office of In Diagnostic Device Evaluation and Safety 510K
6-3
{7}------------------------------------------------
### DUS 60 with C362UA Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------------|--|-------------------|---|-----|-----|---------------|--------------------|------------------|
| | | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | | | | | | | | |
| Fetal / Obstetrics | | N | N | N | | | N | Note 1 |
| Abdominal | | N | N | N | | | N | Note 1 |
| Intra-operative (Specify) | | | | | | | | |
| Intra-operative (Neurological) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | | N | N | N | | | N | Note 1 |
| Small Organ (Specify) | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Transrectal | | | | | | | | |
| Transvaginal | | | | | | | | |
| Transurethral | | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac | | | | | | | | |
| Intravascular | | | | | | | | |
| Peripheral vascular | | | | | | | | |
| Other (Urology) | | N | N | N | | | N | Note 1 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M;
Note 1: This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)
Prescription Use (Per 21 CFR 801.109)
Mary S. Poty
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety . E10K
Page 4 of 12
{8}------------------------------------------------
#### DUS 60 with C343UA Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------------|--------------------|------------------|
| | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | N | N | N | | | N | Note 1 |
| Abdominal | N | N | N | | | N | Note 1 |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | N | | | N | Note 1 |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | N | N | N | | | N | Note 1 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M;
Note 1: This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)
810K
Prescription Use (Per 21 CFR 801.109)
Mary Stotel
(Division Sign Off)
Office of in
Page 5 of 12
{9}------------------------------------------------
#### DUS 60 with C321UA Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| | | | | | | | |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | | | | | | | |
| Abdominal | | | | | | | |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | N | | | N | Note 1 |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | N | N | N | | | N | Note 1 |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Cardiac | N | N | N | | | N | Note 1 |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | N | N | N | | | N | Note 1 |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M;
Note 1: This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
510K
Prescription Use (Per 21 CFR 801.109)
Mary S. Postl
(Division Sign Off)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 6 of 12
6-6
{10}------------------------------------------------
### DUS 60 with L743UA Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------------|--------------------|------------------|
| | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | | | | | | | |
| Abdominal | | | | | | | |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | | | | | | | |
| Small Organ (Specify) | N | N | N | | | N | Note 1 |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | N | N | N | | | N | Note 1 |
| Musculo-skeletal (Superficial) | N | N | N | | | N | Note 1 |
| Intravascular | | | | | | | |
| Cardiac | | | | | | | |
| Peripheral vascular | N | N | N | | | N | Note 1 |
| Other (Urology) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Additional comments: Combined mode: B+M;
Note 1: This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Prescription Use (Per 21 CFR 801.109)
Mary S Patel
(Division Sign Off)
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K. K110999
Page 7 of 12 6-7
{11}------------------------------------------------
### DUS 60 with L742UA Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | Combined<br>(Specify) | Other*<br>(Specify) |
|---------------------------------|-------------------|---|-----|-----|------------------|---|-----------------------|---------------------|
| | B | M | PWD | CWD | Color<br>Doppler | | | |
| Ophthalmic | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | |
| Abdominal | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | |
| Intra-operative (Neurological) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | | | | | | | | |
| Small Organ (Specify) | N | N | N | | | N | Note 1 | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Transrectal | | | | | | | | |
| Transvaginal | | | | | | | | |
| Transurethral | | | | | | | | |
| Musculo-skeletal (Conventional) | N | N | N | | | N | Note 1 | |
| Musculo-skeletal (Superficial) | N | N | N | | | N | Note 1 | |
| Intravascular | | | | | | | | |
| Cardiac | | | | | | | | |
| Intravascular | | | | | | | | |
| Peripheral vascular | N | N | N | | | N | Note 1 | |
| Other (Urology) | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Additional comments: Combined mode: B+M;
Note 1: This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Prescription Use (Per 21 CFR 801.109)
Mary Shishl
---
(Division Sign-Off)
on Sion-Division of Rediological Devices Office of In Vitro D tic Device Evaluation and Sately 510K
Page 8 of 12 6-8
{12}------------------------------------------------
Digital Ultrasonic Diagnostic Imaging System 510K Submission
# Diagnostic Ultrasound Indications for Use Form
### DUS 60 with L763UA Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------------|--------------------|------------------|
| | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | | | | | | | |
| Abdominal | | | | | | | |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | | | | | | | |
| Small Organ (Specify) | N | N | N | | | N | Note 1 |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | N | N | N | | | N | Note 1 |
| Musculo-skeletal (Superficial) | N | N | N | | | N | Note 1 |
| Intravascular | | | | | | | |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | N | N | N | | | N | Note 1 |
| Other (Urology) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M;
Note 1: This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
510K
Prescription Use (Per 21 CFR 801.109)
Division Sign-Off
Office of In ce Evaluation and Safety Page 9 of 12
6-9
{13}------------------------------------------------
#### DUS 60 with E743UA Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
1
| Clinical Application | Mode of Operation | | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------|-----------|-----------|--|
| | ਤਿ | M | PWD | CWD | Color | Combined | Other* | |
| | | | | | Doppler | (Specify) | (Specify) | |
| Ophthalmic | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | |
| Abdominal | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | |
| Intra-operative (Neurological) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | | | | | | | | |
| Small Organ (Specify) | | | | | | | | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Transrectal | N | N | N | | | N | Note l | |
| Transvaginal | | | | | | | | |
| Transurethral | | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | | |
| Intravascular | | | | | | | | |
| Cardiac | | | | | | | | |
| Intravascular | | | | | | | | |
| Peripheral vascular | | | | | | | | |
| Other (Urology) | N | N | N | | | N | Note 1 | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M;
Note 1: This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
610K
Prescription Use (Per 21 CFR 801.109)
______________________________________________________________________________________________________________________________________________________________________________
Mary Stotler
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 10 of 12
6-10
{14}------------------------------------------------
### DUS 60 with E613UA Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------------|--------------------|------------------|
| | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | | | | | | | |
| Abdominal | | | | | | | |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | | | | | | | |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | N | N | N | | | N | Note 1 |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Cardiac | | | | | | | |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M;
Note 1: This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Prescription Use (Per 21 CFR 801.109)
Mary Pashl
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K
K110999
Page 11 of 12
<sup>6-11</sup>
{15}------------------------------------------------
Digital Ultrasonic Diagnostic Imaging System 510K Submission
## Diagnostic Ultrasound Indications for Use Form
#### DUS 60 with C613UA Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | |
|---------------------------------|-------------------|---|-----|-----|---------------|--------------------|------------------|
| | B | M | PWD | CWD | Color Doppler | Combined (Specify) | Other* (Specify) |
| Ophthalmic | | | | | | | |
| Fetal / Obstetrics | | | | | | | |
| Abdominal | | | | | | | |
| Intra-operative (Specify) | | | | | | | |
| Intra-operative (Neurological) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | N | | | N | Note 1 |
| Small Organ (Specify) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Transrectal | | | | | | | |
| Transvaginal | | | | | | | |
| Transurethral | | | | | | | |
| Musculo-skeletal (Conventional) | | | | | | | |
| Musculo-skeletal (Superficial) | | | | | | | |
| Intravascular | | | | | | | |
| Cardiac | N | N | N | | | N | Note 1 |
| Intravascular | | | | | | | |
| Peripheral vascular | | | | | | | |
| Other (Urology) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix
Additional comments: Combined mode: B+M;
Note 1: This feature does not use contrast agent.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)
Prescription Use (Per 21 CFR 801.109)
Mary Statel
(Division Sign-Off)
Division of Office of In ation and Safety BIOK
Page 12 of 12
6-12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.