K110993 · Femasys, Inc. · LKF · Oct 12, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K110993
Device Name
FEMCHEC PRESSURE MANAGEMET DEVICE
Applicant
Femasys, Inc.
Product Code
LKF · Obstetrics/Gynecology
Decision Date
Oct 12, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Indications for Use
The FemChec™ Pressure Management Device limits the maximum applied pressure while instilling contrast media into the uterus and fallopian tubes. It is to be used in conjunction with an intrauterine catheter for performance of hysterosalpingogram (HSG), such as tubal occlusion confirmation tests in women who have had tubal procedures for permanent female sterilization. The FemChec limits the applied intrauterine pressure to 200 mmHg. The FemChec is provided with a fluid collector, stopcock, and intrauterine catheter for performing the HSG procedure.
Device Story
FemChec is a manually operated, pressure-limiting contrast media syringe system used during hysterosalpingogram (HSG) procedures. The device includes a syringe, fluid collector, stopcock, and intrauterine catheter. It functions by limiting the maximum intrauterine pressure applied during the instillation of contrast media to 200 mmHg. Used in clinical settings by healthcare providers, the device ensures controlled fluid delivery to the uterus and fallopian tubes. By preventing excessive pressure, the device assists in confirming tubal occlusion, potentially reducing risks associated with over-pressurization during the procedure.
Clinical Evidence
Bench testing only. Testing included fluid instillation function, attachment functionality, single-hand actuation, pressure testing (limit < 200 mmHg), and biocompatibility (cytotoxicity, irritation, sensitization, and acute systemic toxicity) per ISO 10993 standards.
Technological Characteristics
Manually operated piston syringe system; includes fluid collector, stopcock, and intrauterine catheter. Pressure management component limits output to 200 mmHg. Biocompatibility per ISO 10993. Sterile.
Indications for Use
Indicated for women undergoing hysterosalpingogram (HSG) procedures, including tubal occlusion confirmation tests following permanent female sterilization.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
K110993.
## OCT 1 2 2011
## 510(k) Summary for the FemChec™ Pressure Management Device
Date of Summary: May 31, 2011
510(k) Submitter and Primary Contact:
Lisa Peacock Vice President, Regulatory Affairs Femasys Inc. 5000 Research Court Suite 100 Suwanee, GA 30024 Tel: 770-500-3910 Fax: 770-500-3980 LPEACOCK@FEMASYS.COM
Device Common
contrast media syringe
LKF
unassigned
unclassified
Obstetrics/Gynecology
Name:
FDA Device
cannula, manipulator/injector, uterine Classification Name:
Product Code:
Classification Regulation:
Device Class:
Panel:
Indication for Use:
Device Description:
The FemChec™ Pressure Management Device limits the maximum applied pressure while instilling contrast media into the uterus and fallopian tubes. It is to be used in conjunction with an intrauterine catheter for performance of hysterosalpingogram (HSG), such as tubal occlusion confirmation tests in women who have had tubal procedures for permanent female sterilization. The FemChec limits the applied intrauterine pressure to 200 mmHq.
The FemChec is provided with a fluid collector, stopcock, and intrauterine catheter for performing the HSG procedure.
The FemChec Pressure Management Device (FemChec) is a pressure-limiting contrast media syringe that is connected to an intrauterine catheter for low-pressure instillation of contrast media during HSG procedures.
{1}------------------------------------------------
# KII0993
#### 510(k) Summary for the FemChec™ Pressure Management Device
#### Predicate Devices:
K020954 H/S Elliptosphere Procedure Tray K792134 Bonchek Vein Distention System. 200 mm Ha K082725 VasoShield Pressure Controlling Syringe
The FemChec is similar to the H/S Elliptosphere Procedure Tray in that FemChec is a manually operated piston syringe for contrast media instillation, as an accessory to an intrauterine catheter for HSG procedures. Likewise, the predicate tray contains accessories to intrauterine catheters for HSG procedures, including a manually operated piston syringe for contrast media instillation.
The FemChec is similar to the Bonchek Vein Distention System, 200 mm Hq, in that both devices: 1) are svringes for instilling fluids, 2) contain pressure management components, and 3) limit applied pressures to a similar value. The devices differ in the mechanisms of operation and materials of the pressure management component and in their labeled anatomical locations of use, as Bonchek is labeled for use in cardiovascular applications. Although, the clinical use of the Bonchek to limit applied intrauterine pressure during HSG tubal occlusion confirmation tests was reported in the PMA Summary of Safety and Effectiveness Data for the Adiana Permanent Contraception System.
The FemChec is similar to the VasoShield Pressure Controlling Syringe in that both devices: 1) are syringes for instilling fluids, 2) contain pressure management components with similar mechanisms of operation and with similar materials, and 3) limit applied pressures to a similar value. The devices differ in the anatomical locations of use, as VasoShield is indicated for use in cardiovascular applications. ·
#### Summary of Testing:
The FemChec was tested by the following non-clinical methods to demonstrate that the device is substantially equivalent to the predicate devices in functionality, safety, and effectiveness:
- · fluid instilling function of syringe
- · attachment functionality
- · single-hand actuation
- · pressure testing < 200 mm Hg
- · biocompatibility of components according to ISO 10993 standards: cytotoxicity, irritation, sensitization, and acute systemic toxicity
{2}------------------------------------------------
#### 510(k) Summary for the FemChec™ Pressure Management Device
### Conclusions of Substantial Equivalence Demonstrations
The FemChec was compared to the predicate H/S Elliptosphere Procedure Tray for the following aspects, and found to have similar technological characteristics and to be substantially equivalent:
- · Indications for use, regarding instillation of contrast media during HSG procedures
- · Syringe design
- · Syringe materials
- Instillation principles of operation
- · Sterility
The FemChec was compared to the predicates Bonchek Vein Distention System, 200 mm Hg and the VasoShield Pressure Controlling Syringe for the following aspects, and found to have similar technological characteristics and to be substantially equivalent:
- · Indications for use, regarding limiting pressure
- · Pressure management components and materials (compared to VasoShield) ·
- · Pressure management principles of operation
- · Sterility
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Lisa Peacock V.P. Regulatory Affairs Femasys Inc. 5000 Research Court, Ste. 100 SUWANEE GA 30024
0CT 1 2 2011
Re: K110993
Trade/Device Name: FemChec™ Pressure Management Device Regulation Number: None Regulation Name: None Regulatory Class: Unclassified Product Code: LKF Dated: September 23, 2011 Received: October 6, 2011
Dear Ms. Peacock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
#### Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Huker Lemur MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health.
Enclosure
{5}------------------------------------------------
Femasys Inc.
Page 1 of 1
#### Indications for Use
510(k) Number (if known): K110993
Device Name: FemChec™ Pressure Management Device
Indications for Use:
The FemChec™ Pressure Management Device limits the maximum applied pressure while instilling contrast media into the uterus and fallopian tubes. It is to be used in conjunction with an intrauterine catheter for performance of hysterosalpingogram (HSG), such as tubal occlusion confirmation tests in women who have had tubal procedures for permanent female sterilization. The FemChec limits the applied intrauterine pressure to 200 mmHg.
The FemChec is provided with a fluid collector, stopcock, and intrauterine catheter for performing the HSG procedure.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE.ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
*_*on Sign Off
( vision Sign-Off Lision of Reproductive. Gast Unclogical Devices 515(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.