The CentriMag Drainage Cannula is intended to be used as a venous drainage cannula with an extracorporeal bypass circuit for extracorporeal circulatory support for periods up to six hours.
Device Story
The CentriMag Drainage Cannula Kit is a sterile, single-use, disposable venous drainage cannula designed for use within extracorporeal bypass circuits. The kit includes a PVC cannula and various accessories, including a cap, apical sewing ring handle, apical sewing ring, apical support cuff, and stabilizer rings. The device is intended for short-term extracorporeal circulatory support, specifically for periods up to six hours. It functions as a conduit for venous blood drainage during bypass procedures. The device is intended for clinical use by healthcare professionals in settings where extracorporeal circulatory support is required.
Clinical Evidence
No clinical data was performed for this device. Substantial equivalence was demonstrated through bench testing, including physical testing, sterilization validation, biocompatibility, shelf life, transportation, in vitro hemolysis, and flow versus pressure drop testing.
Technological Characteristics
The device is constructed from Poly Vinyl Chloride (PVC). It is a sterile, single-use, disposable cannula. It functions as a passive venous drainage conduit in an extracorporeal bypass circuit. No software or electronic components are involved.
Indications for Use
Indicated for use as a venous drainage cannula within an extracorporeal bypass circuit for circulatory support for durations up to six hours.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
{0}------------------------------------------------
# K110983
## 510(k) Summary – Levitronix CentriMag Drainage Cannula Kit
JUN 2 0 2011
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807, and in particular §807.92.
| A. | Application Information | | | | | | | | |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|--------------------------------|-----------------------------|--------------------------------------------------------|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| | <table><tr><td>Date Prepared:</td><td>April 6, 2011</td></tr><tr><td>Submitter's Name & Address:</td><td>Levitronix LLC<br/>45 First Avenue<br/>Waltham, MA 02451</td></tr><tr><td>Contact Person:</td><td>Lydia Sakakeeny, Ph.D.<br/>Regulatory Affairs Specialist<br/>Ph: (781) 466-6553<br/>Fax: (781) 622-5090<br/>e-mail: Isakakeeny@levitronix-us.com</td></tr></table> | Date Prepared: | April 6, 2011 | Submitter's Name & Address: | Levitronix LLC<br>45 First Avenue<br>Waltham, MA 02451 | Contact Person: | Lydia Sakakeeny, Ph.D.<br>Regulatory Affairs Specialist<br>Ph: (781) 466-6553<br>Fax: (781) 622-5090<br>e-mail: Isakakeeny@levitronix-us.com | | |
| Date Prepared: | April 6, 2011 | | | | | | | | |
| Submitter's Name & Address: | Levitronix LLC<br>45 First Avenue<br>Waltham, MA 02451 | | | | | | | | |
| Contact Person: | Lydia Sakakeeny, Ph.D.<br>Regulatory Affairs Specialist<br>Ph: (781) 466-6553<br>Fax: (781) 622-5090<br>e-mail: Isakakeeny@levitronix-us.com | | | | | | | | |
| B. | Device Information | | | | | | | | |
| | <table><tr><td>Trade or Proprietary Name:</td><td>CentriMag Drainage Cannula Kit</td></tr><tr><td>Common or Usual Name:</td><td>CentriMag Drainage Cannula Kit</td></tr><tr><td>Classification Name:</td><td>Catheter, cannula and tubing, vascular, cardiopulmonary<br/>bypass (DWF, 870.4210), Class II</td></tr><tr><td>Performance Standard:</td><td>Performance standards do not currently exist for<br/>these devices. None are established under section<br/>514 of the Food, Drug and Cosmetic Act.</td></tr></table> | Trade or Proprietary Name: | CentriMag Drainage Cannula Kit | Common or Usual Name: | CentriMag Drainage Cannula Kit | Classification Name: | Catheter, cannula and tubing, vascular, cardiopulmonary<br>bypass (DWF, 870.4210), Class II | Performance Standard: | Performance standards do not currently exist for<br>these devices. None are established under section<br>514 of the Food, Drug and Cosmetic Act. |
| Trade or Proprietary Name: | CentriMag Drainage Cannula Kit | | | | | | | | |
| Common or Usual Name: | CentriMag Drainage Cannula Kit | | | | | | | | |
| Classification Name: | Catheter, cannula and tubing, vascular, cardiopulmonary<br>bypass (DWF, 870.4210), Class II | | | | | | | | |
| Performance Standard: | Performance standards do not currently exist for<br>these devices. None are established under section<br>514 of the Food, Drug and Cosmetic Act. | | | | | | | | |
#### C. Legally Marketed Predicate Device
| Levitronix Device | Predicate | Predicate 510(k) |
|--------------------------------|-------------------------------------------------|------------------|
| CentriMag Drainage Cannula Kit | Edwards Lifesciences<br>Venous Drainage Cannula | K8701091 |
<sup>1 510</sup>k originally issued to Research Medical, Inc.
{1}------------------------------------------------
#### D. Device Description
### CentriMag Drainage Cannula Kit
The CentriMag Drainage Cannula Kit consists of a sterile, single-use, disposable, noncoated, Poly Vinyl Chloride (PVC) Cannula and the following accessories:
- A) One Cap with Tape
B) One Apical Sewing Ring Handle
C) One Apical Sewing Ring
D) One Apical Support Cuff
E) Two Stabilizer Rings - Large
F) Two Stabilizer Rings - Medium
#### ய் Intended Use
The CentriMag Drainage Cannula is intended to be used as a venous drainage cannula with an extracorporeal bypass circuit for extracorporeal circulatory support for periods up to six hours.
#### F. Comparison to Predicate Device
The CentriMag Drainage Cannula Kit has indications for use, design features, and functional characteristics which are substantially equivalent to the predicate device. The device raises no new safety or effectiveness issues.
#### G. Summary of Performance Data
The CentriMag Drainage Cannula Kit has successfully undergone functional testing demonstrating substantial equivalence to the predicate device. The following functional testing was performed. All met pre-established acceptance criteria.
- Physical Testing .
- . Sterilization Validation
- Biocompatibility .
- Shelf Life Studies .
- Transportation .
- . Hemolysis (in vitro)
- Flow versus Pressure Drop .
#### ાન. Clinical Performance
Clinical testing was not performed for the CentriMag Drainage Cannula Kit.
#### l. Conclusion
The Levitronix CentriMag Drainage Cannula Kit is substantially equivalent to the Edwards Lifesciences Venous Drainage Cannula (K870109).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized, curved lines.
Food and-Drug Administration-10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Levitronix LLC c/o Lydia Sakakeeny, Ph.D. Regulatory Affairs Specialist 45 First Avenue Waltham, MA 02451
JUN 2 0 2011
Re: K110983
CENTRIMAG® Drainage.Cannula Kit Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary bypass vascular catheters, cannula, or tubing Regulatory Class: Class II (two) Product Code: DWF Dated: April 6, 2011 Received: April 7, 2011
Dear Dr. Sakakeeny:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 -- Lydia Sakakeeny, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincefely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use Statement
Applicant:
Levitronix LLC
510(k) Number (if known):
Device Name:
CentriMag Drainage Cannula Kit
110983
### Indications for Use:
The CentriMag Drainage Cannula is intended to be used as a venous drainage cannula with an extracorporeal bypass circuit for extracorporeal circulatory support for periods up to six hours.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
---
Division Sidn-Off
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.