The Ambu® aScope™ 2 is intended for use as an aid in the placement of an Endotracheal Tube (ETT) directly or through an intubating laryngeal mask during non- difficult and difficult intubating procedures or for visualization of the airway during Percutaneous Tracheostomy (PT) procedures. The Ambu® aScope™ 2 achieves its purpose by providing the user with a visual confirmation of where the tip of the Ambu® aScope™ 2 is in the human anatomy. The flexible tip of the Ambu® aScope™ 2 allows the user to guide the ETT in the desired direction. The system is for use in a hospital environment. The target population is adults/children that have been clinically evaluated for ETT size 6 or larger.
Device Story
Ambu aScope 2 system comprises a sterile, single-use flexible endoscope and a reusable monitor. Endoscope features a CMOS camera and LED light source at the distal tip; manoeuvrable tip controlled by clinician via handle. System inputs visual data from airway; displays real-time video on monitor. Clinician uses visual feedback to guide ETT placement or visualize airway during percutaneous tracheostomy. Includes Luer connector for topical anesthetic administration or oxygen flow. Monitor can be mounted to IV pole; supports external video recording via cable. Device aids clinicians in airway management; provides visual confirmation of scope tip location; facilitates ETT guidance. Benefits include reduced risk of cross-contamination due to single-use scope design.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ISO 8600-1/3/4, ISO 594-1, and ISO 10993-1/5/10. Tests included tip manoeuvrability, bending section endurance, Luer channel functionality (anesthetic delivery and oxygen flow), distal tip temperature (<41°C), monitor imaging performance, cleaning/disinfection endurance (150 cycles), battery capacity, and sterile packaging integrity (ASTM F88). Environmental testing (shock, vibration, drop) confirmed durability.
Technological Characteristics
Flexible endoscope with manoeuvrable tip; CMOS camera and LED light source. Single-use scope; reusable monitor. Luer connector for auxiliary access. Sterilized via Ethylene Oxide. Connectivity: Video Out cable for external recording. Electrical safety per IEC 60601-1/2-18; EMC per IEC 60601-1-2. Biocompatibility per ISO 10993.
Indications for Use
Indicated for adults and children requiring endotracheal tube (ETT) placement (size 6 or larger) or airway visualization during percutaneous tracheostomy. Used in hospital settings for both non-difficult and difficult intubations.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}------------------------------------------------
#### Section 5: 510(k) Summary K-110962
This 510(k) summary has been prepared in accordance with 21 CFR 807.87(h) and the 510(k) has been prepared in accordance with 21 CFR 807.92
| Submitter | Ambu A/S<br>Baltorpbakken 13<br>DK-2750 Ballerup<br>Denmark<br>tel.: +45 7225 2000<br>fax.: +45 7225 2055 | | |
|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|-------------|
| Contact Person | Name: Kaja Tengbjerg<br>Job Title: Regulatory Affairs Professional<br>Address: Ambu A/S, Baltorpbakken 13, DK-2750 Ballerup<br>Telephone number: +45 7225 2215<br>Fax number: +45 7225 2055 | | |
| Date Summary<br>Prepared | March, 2011 | | |
| Device Trade Name | Ambu® aScope™ 2<br>Ambu® aScope™ Monitor | | |
| Device Common Name | Endoscope for endotracheal intubation | | |
| Device Classification | Tracheal Tube<br>Product Codes: BTR<br>21 CFR 868.5730<br>Class II | | |
| Legally Marketed<br>devices to which the<br>device is substantially<br>equivalent | Manufacturer | Trade Name | 510k number |
| | Ambu A/S | Ambu® aScope™ and<br>Ambu® aScope™ Monitor | K093186 |
{1}------------------------------------------------
:
・・・ :
| Description of the<br>Device | Ambu aScope System consist of Ambu aScope 2 and Ambu<br>aScope Monitor.<br>Ambu aScope 2 is for viewing anatomical structures in the upper<br>airways, and as an aid in placement of an endotracheal tube<br>(ETT), an ETT size 6 or larger can be used. A camera at the<br>distal tip of the aScope provides the user with an indication of<br>the placement of the aScope. The manoeuvrable tip allows the<br>user to guide the ETT in the desired direction. Ambu aScope is<br>for single patient use and it is sterile.<br>The Ambu aScope 2 must be connected to Ambu aScope<br>Monitor. The monitor displays the image and it is reusable. |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Ambu aScope 2 has the following physical and performance<br>characteristics:<br>- Manoeuvrable tip controlled by the clinician<br>- Flexible insertion cord<br>- Camera and LED light source at the distal tip<br>- Topical Anaesthetics can be administered to the patient via<br>a channel, with standard Luer connector<br>- Oxygen flow can be applied via a channel using a flow<br>connector attached to the Luer connector<br>- Sterile by Ethylene Oxide sterilisation<br>- For Single Patient Use<br><br>Ambu aScope Monitor has the following physical and<br>performance characteristics:<br>- Displays the image from Ambu aScope 2 on the screen.<br>- Can be fixed to e.g. an IV pole.<br>- By connecting a Video Out Cable to the Ambu aScope<br>Monitor the image can be displayed and/or recorded on an<br>external monitor and/or video recorder.<br>- Reusable device. |
| Indications for Use | The Ambu® aScope™ 2 is intended for use as an aid in the<br>placement of an Endotracheal Tube (ETT) directly or through an<br>intubating laryngeal mask during non- difficult and difficult<br>intubating procedures or for visualization of the airway during<br>Percutaneous Tracheostomy (PT) procedures. The Ambu®<br>aScope™ 2 achieves its purpose by providing the user with a<br>visual confirmation of where the tip of the Ambu® aScope™ 2 is<br>in the human anatomy.<br>The flexible tip of the Ambu® aScope™ 2 allows the user to<br>guide the ETT in the desired direction. The system is for use in a<br>hospital environment.<br>The target population is adults/children that have been clinically<br>evaluated for ETT size 6 or larger. |
.
{2}------------------------------------------------
# 510(k) Application – Ambu® aScope™ 2 and Ambu® aScope™ Monitor
Summary of the technological characteristics in comparison to the predicate devices
Ambu aScope 2 is similar to predicate device:
- They possess a similar Intended Use. ●
- They both are flexible scopes with a manoeuvrable tip. .
- They are single-use devices, which are delivered sterile. ●
- They both use a LED-light source located at the tip of . the scopes
- They possess a CMOS camera located at the distal tip to . provide an image
- They have a handle with a control button giving the . operator the ability to steer the tip of the scope up and down
- In both devices, an image is provided on a separate . monitor
Ambu concludes that the Ambu aScope 2 and Ambu aScope Monitor is substantially equivalent to the predicate device.
{3}------------------------------------------------
| Performance Data -<br>Bench | The following data has been submitted in the premarket notification: |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Ambu has provided declaration of conformity to the following<br>recognized consensus standards applicable for this type of device:<br>ISO 8600-1, ISO 8600-3 and ISO 8600-4 Optics and optical<br>instruments - medical endoscopes and certain accessories.<br>ISO 594-1 Conical fittings with a 6% (luer) taper for syringes,<br>needles and certain other medical equipment.<br>The declaration of conformity is based on test data.<br>Result: All tests were passed. |
| | Performance test report was submitted to document the following<br>properties of the Ambu aScope 2 (single use):<br>Manoeuvrability of tip (accept criteria: 120+/- 10 ° to each side)<br>Endurance of the bending section (accept criteria: 50 bends to<br>each side of minimum 90° and 10 bends to each side of 120+/-<br>10 °)<br>Luer channel when used for topical anaesthetics (accept criteria: |
| | verification of ability to administer topical anaesthetics through<br>the channel – when 1 ml is injected at least 0.8 ml was expelled<br>in the distal end)<br>Luer channel when used for Oxygen (accept criteria: Luer<br>channel is compatible with 10 seconds oxygen flow at 5 bars and<br>3 L/min. followed by a 3 seconds blocking of working channel at<br>a pressure of 5 bars)<br>Temperature at distal end of Ambu aScope (accept criteria:<br>temperature below 41 °C)<br>Result: All tests were passed. |
| | Performance test report was submitted to document the following<br>properties of the Ambu aScope Monitor (reusable):<br>Imaging performance; evaluation of colours, flickering, contrast<br>and brightness, rated on a scale from 1-3, where 1 is best.<br>(accept criteria: rating 1 and max 2 ratings of 2, when 32 scopes<br>are evaluated for all 4 properties (128 evaluations))<br>Cleaning endurance of Ambu aScope Monitor (accept criteria:<br>monitor can withstand the prescribed cleaning and disinfection<br>method for 150 times reprocessing)<br>Battery capacity of Ambu aScope Monitor (accept criteria: at<br>least 70% battery capacity after 150 times charging and de-<br>charging)<br>Result: All tests were passed. |
| | Performance test report was submitted to document Shelf life of<br>Ambu aScope 2. Testing was done on finished, sterilized, shipped<br>and aged products:<br>Performance test of the Ambu aScope 2. Test according to Final<br>Quality Inspection Procedure of Ambu aScope 2.<br>Accept criteria: Product specifications fulfilled<br>Sterile packaging integrity of the Ambu aScope 2 pouch.<br>Accept criteria: The seal strength must be greater than 0.4 N<br>when tested according to ASTM F88.<br>Result: All tests were passed |
| | Since the device is in compliance with the listed standards and have<br>passed the listed performance tests, it is concluded that<br>technological characteristics of Ambu aScope 2 and Ambu aScope<br>Monitor is as safe and effective and performs as well as or better<br>than the chosen legally marketed predicate device. |
Section 5 - Page 4 of 6
{4}------------------------------------------------
Environmental tests to demonstrate the compliance to the following standards:
Transportation in designated packaging:
- EN 60068-2-27 Basic environmental testing procedures l Part 2: Tests - Test Ea and guidance: Shock: 500 repetitive shocks (bump) in each of 6 directions 400 m/s2 (40g)
- EN 60068-2-64 Environmental testing Part 2-64: Tests -Test Fh: Vibration, broadband random and guidance: Random vibration 1.6 grms, 10-150Hz, 30 min/axis
- EN 60068-2-31 Environmental testing Part 2-31: Tests -ー Test Ec: Rough handling shocks, primarily for equipmenttype specimens: 12 falls from 1.2m height
Tests performed with the product without its packaging: Bounce
- EN 60608-2-6 Environmental testing Part 2-6: Tests Test l Fc: Vibration (sinusoidal): Sinus vibration 5Hz, 1grms, 1 hour.
Free fall
- EN 60068-2-31 Environmental testing Part 2-31: Tests l Test Ec: Rough handling shocks, primarily for equipmenttype specimens:
Ambu aScope: 2 falls per relevant orientation from 1.2m height to a smooth concrete surface.
Ambu aScope Monitor: 2 falls per relevant orientation from 0.2m height to a smooth concrete surface.
After each of the above environmental tests, the packaging integrity and the device were inspected, and performance test of the device was performed.
Result: All products and packaging passed the tests.
Based on the above environmental testing Ambu has concluded that Ambu aScope 2 and Ambu aScope Monitor, can withstand the stresses applied to the product during transport and handling prior to the use of the device, and is as safe and effective and performs as well as or better than the chosen legally marketed predicate device.
{5}------------------------------------------------
| | Data for compliance to the general requirements for the device were submitted: |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Biocompatibility tests shows that the device complies with the requirements of ISO 10993-1: |
| | - Cytotoxicity (ISO 10993-5) |
| | - Sensitization (ISO 10993-10) |
| | - Intracutaneous reactivity test (ISO 10993-10) |
| | Result: All tests were passed. |
| | Tests that verifies the following properties: |
| | - Cleaning validation and Liquid Chemical Sterilization and Disinfection Validation of the Ambu aScope Monitor according to AAMI TIR12 and AAMI TIR30, to validate the prescribed method of cleaning and disinfection. |
| | - Electro Magnetic Compatibility in compliance with IEC 60601-1-2. |
| | - Electrical Safety in compliance with IEC 60601-1 and IEC 60601-2-18. |
| | Result: All tests were passed. |
| | Since the device passed all the tests to demonstrate compliance to the general requirements for this kind of device, it is concluded that Ambu aScope 2 and Ambu aScope monitor is as safe and effective and performs as well as or better than the chosen legally marketed predicate device. |
| Performance Data - Clinical | Not applicable. |
| Conclusion | Based on the indication for use, technological characteristics, performance data and comparison to predicate devices it has been concluded that the Ambu aScope 2 and Ambu aScope Monitor has equivalent functionality and intended use as the predicate device. It is concluded that Ambu aScope 2 and Ambu aScope Monitor are as safe and effective and performs as well as or better than the chosen legally marketed predicate device. |
.
.
:
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. Sanjay H. Parikh Vice President Operations US Agent Ambu Incorporated 6740 Baymeadow Drive Glen Burnie, Maryland 21060
NOV 1 8 2011
Re: K110962
Trade/Device Name: Ambu aScope 2 Ambu aScope Monitor Regulation Number: 21 CFR 868.5730 Regulation Name: Tracheal Tube Regulatory Class: II Product Code: BTR Dated: November 14, 2011 Received: November 15, 2011
Dear Mr. Parikh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
### Page 2 – Mr. Parikh
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
510(k) Application - Ambu® aScope™ 2 and Ambu® aScope™ Monitor
# Section 4: Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name:
#### Indications for Use:
The Ambu@ aScope™ 2 is intended for use as an ald in the placement of an Endotracheal Tube (ETT) directly or through an intubating laryngeal mask during non- difficult intubating procedures or for visualization of the airway during Percutaneous Tracheostomy (PT) procedures.
The Ambu® aScope™ 2 achieves its purpose by providing the user with a visual confirmation of where the tip of the Ambu@ aScope™ 2 is in the human anatomy. The flexible tip of the Ambu® aScope™ 2 allows the user to guide the ETT in the desired direction.
The system is for use in a hospital environment. The target population is adults/children that have been clinically evaluated for ETT size 6 or larger.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER
## PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
y Archittin
(Division Sian-Off) Division of Anesthesiology, General Hospital Infection Control. Dental Devices
510(k) Number: k110962
Section 4 - Page 1 of 1
Version 1.0
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.