← Product Code [NBW](/productcode/NBW) · K110948

# HEALTH CARE SYSTEM SOFTWARE (K110948)

_Taidoc Technology Corporation · NBW · May 13, 2011 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K110948

## Device Facts

- **Applicant:** Taidoc Technology Corporation
- **Product Code:** [NBW](/productcode/NBW.md)
- **Decision Date:** May 13, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 862.1345
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry
- **Attributes:** Software as a Medical Device

## Indications for Use

The Health Care System Software is an optional software accessory for use with the following models with data management capabilities: a) Clever Chek blood glucose meters, b) Clever Chek blood glucose plus blood pressure monitors, and c) Clever blood pressure monitors. When use with one of these meters, Health Care System Software transfers data from the device's memory into a computer for enhanced data management. The Health Care System Software is intended for use in home and clinical settings as an aid for users and their health care professionals to review, analyze and evaluate the historical test results to support health management effectively.

## Device Story

Software accessory for data management; transfers stored test results from blood glucose meters and blood pressure monitors to computer. Used in home and clinical settings by patients and healthcare professionals. Enables review, analysis, and evaluation of historical test data via tabular and graphical displays. Does not modify transferred data. Facilitates health management by providing longitudinal view of patient test results.

## Clinical Evidence

Bench testing only. Validation included hardware data transfer verification and software validation. No clinical data presented.

## Technological Characteristics

Software-based data management system. Connectivity via cable for data transfer from compatible meters. Features include tabular and graphical data display and printer settings. No specific materials or energy sources applicable as it is a software accessory.

## Regulatory Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

## Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Clever Chek Health Care System Software ([K070941](/device/K070941.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K110948

# 8. 510 (k) Summary

# MAY 1 3 2011

1. Submitter Information Company name . Contact person Teling Hsu Address Phone

FAX E-mail Date Prepared

2. Name of Device Trade Names Product code Classification Names and Regulations

TaiDoc Technology Corporation 6F, No. 127, Wugong 2nd Rd. Wuqu Township, Taipei County 24888, Taiwan (+886-2) 6625-8188 ext.1176 (+886-2) 6625-0288 teling.hsu@taidoc.com.tw March 30th, 2011

Health Care System Software NBW; DXN; FLL

a) Calculator/data processing module for clinical use, Class I, 21 CFR 862.2100 b) Glucose Test System, Class II, 21 CFR 862.1345 c) Noninvasive Blood Pressure Measurement System, Class II, 21 CFR 870.1130, d) General Hospital, Class II, 21 CFR 880.2910

3. Predicate Device Trade/Proprietary Name: Common/Usual Name: Submitter 510 (k) Number

Clever Chek Health Care System Software Data management software TaiDoc Technology Corporation K070941

{1}------------------------------------------------

### 4. Device Description

The Health Care System Software is an optional software accessory for use with the following models with data management capabilities: a) blood glucose meters, b) blood glucose plus blood pressure monitors, and c) blood pressure monitors. When use with one of these devices, Health Care System Software transfers data from the device's memory into a computer for enhanced data management.

#### 5. Intended Use

The Health Care System Software is an optional software accessory for use with the following models with data management capabilities: a) Clever Chek blood glucose meters, b) Clever Chek blood glucose plus blood pressure monitors, and c) Clever blood pressure monitors. When use with one of these meters, Health Care System Software transfers data from the device's memory into a computer for enhanced data management.

The Health Care System Software is intended for use in home and clinical settings as an aid for users and their health care professionals to review, analyze and evaluate the historical test results to support health management effectively.

### 6. Comparison to Predicate Device

The Health Care System Software is substantially equivalent to the Clever Chek Health Care System Software (K070941). Both management software programs can be described as follows:

- have the same intended use and intended users
- have the same data presentation
- have same programming language
- data transferred from the device cannot be changed or modified in any way.

The modifications include:

- -Added temperature data tabular display setting
- -Added temperature graph display setting
- -Added temperature data printer setting

{2}------------------------------------------------

## 7. Performance Studies

Testing of Health Care System Software included validation of hardware (data transfer through the cable) and software validation. Results demonstrate that the system meets its intended use.

# 8. Conclusion

Health Care System Software is substantially equivalent to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol. The logo is black and white.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAY 1 3 2011

Mr. Teling Hsu Regulatory Affairs Specialist TaiDoc Technology Corporation 6F, No, 127, Wugong 2nd Road Wugu Township Taipei County Taiwan 24888

Re: K110948

Trade/Device Name: Health Care System Software Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: II Product Code: NBW, DXN, FLL Dated: May, 4, 2011 Received: May 6, 2011

Dear Mr. Hsu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Nh for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Indications for Use

510(k) Number: k110948

Device Name: Health Care System Software

Indications for Use:

The Health Care System Software is an optional software accessory for use with the following models with data management capabilities: a) Clever Chek blood glucose meters, b) Clever Chek blood glucose plus blood pressure monitors, and c) Clever blood pressure monitors. When use with one of these meters, Health Care System Software transfers data from the device's memory into a computer for enhanced data management.

The Health Care System Software is intended for use in home and clinical settings as an aid for users and their health care professionals to review, analyze and evaluate the historical test results to support health management effectively.

Prescription Use_ _ _ Over-The-Counter Use _ X AND/OR 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Rhit C. Aly 5/12/16
Division Sign-Off

Division Sign-Off Office of In Vitro Diagnostic Device

Evaluation and Safety
510(k) K110948

7.

---

**Source:** [https://fda-staging.innolitics.com/device/K110948](https://fda-staging.innolitics.com/device/K110948)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).