The Injection Gold Probe™ Bipolar Electrohemostasis Catheter is indicated for use in endoscopic injection therapy (to deliver pharmacological injection agents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. These sites include the following: Peptic Ulcers, Dieulafoy Lesions, Mallory-Weiss Tears, Bleeding Polyp Stalks, Arteriovenous Malformations (AVMs), Angiomata. The Injection Gold Probe Bipolar Electrohemostasis Catheter also has irrigation capability. Any other use is not recommended.
Device Story
Injection Gold Probe Catheter is a multi-functional endoscopic tool for injection therapy, bipolar electrohemostasis, and irrigation. Device features a 25-gauge needle for delivering pharmacological agents (e.g., vasoconstrictors) and a gold spiral electrode at the tip for tissue cauterization and blood coagulation. Irrigation is performed via saline or sterile water delivery. Used in clinical settings by physicians during endoscopic procedures. Device connects to compatible bipolar electrosurgical generators (Symmetry Endostat, Endostat II, Bicap, or Valleylab) via standard or banana plugs. Output is controlled by the physician to manage GI bleeding sites, potentially reducing hemorrhage and improving patient outcomes. Modifications in this version include updated gold thick film conductor paste, replacement of Medi-Glide coating with Teflon on the needle hub/hypo tube, new ceramic tip material, and adjusted needle assembly dimensions.
Clinical Evidence
No clinical data. Substantial equivalence supported by biocompatibility testing and bench testing demonstrating that the modified device meets all design specifications.
Technological Characteristics
Catheter with 25-gauge injection needle and gold spiral electrode for bipolar electrohemostasis. Materials: gold thick film conductor, ceramic tip, Teflon-coated needle hub/hypo tube. Available in 7Fr (2.3mm) and 10Fr (3.3mm) diameters, 210cm length. Energy source: external bipolar electrosurgical generator. Connectivity: standard or banana plug electrical connectors. Irrigation port for saline/water delivery.
Indications for Use
Indicated for endoscopic injection therapy and electrohemostasis of actual or potential GI bleeding sites (peptic ulcers, Dieulafoy lesions, Mallory-Weiss tears, bleeding polyp stalks, AVMs, angiomata) in patients requiring endoscopic intervention.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
# APR 1 4 2011
# SECTION 6 510(k) SUMMARY
#### 510(k) SUMMARY
#### 1. Submitter:
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Telephone: 508-683-4359 Fax: 508-683-5939
Contact: Ashley Pyle Sr. Regulatory Affairs Specialist Date Prepared: March 29, 2011
#### 2. Proposed Device:
Trade Name: Injection Gold ProbeTM Bipolar Electrohemostatis Catheter Classification Name: Unit. Electrosurgical, Endoscopic (with Accessories) Regulation Number: 876.4300 Product Code: KNS Classification: Class II
#### 3. Predicate Device:
Trade Name: Injection Gold ProbeTM Bipolar Electrohemostatis Catheter Manufacturer and Clearance Number: Boston Scientific Corporation, K961349 Classification Name: Unit, Electrosurgical, Endoscopic (with Accessories) Regulation Number: 876.4300 Product Code: KNS Classification: Class II
### 4. Proposed Device Description:
The Injection Gold Probe Catheter is an injection therapy and bipolar electrohemostasis catheter with irrigation capabilities. The Injection Gold Probe Catheter is available with two different style electrical connectors (plugs), Standard and Banana. The Injection Gold Probe Catheter is available in a working length of 210cm with outer diameter of 7Fr (2.3mm) and 10Fr (3.3mm) compatible with a minimum working channel of 2.8mm and 3.7mm.
- I. Injection Therapy
Each Injection Gold Probe Catheter contains a 25gauge injection needle for injection therapy. The catheter handle contains the injection needle hub. Injection is performed by attaching a luerlock syringe to the "Injection" hub and advancing the needle.
#### II. Bipolar Electrohemostasis
The gold spiral electrode on the probe tip provides bipolar electrohemostasis. The Injection Gold Probe Catheter has a standard plug which adapts to Symmetry Endostat, Endostat II and Bicap bipolar electrosurgical generators or Banana plugs which are recommended for use with Valleylab bipolar electrosurgical generator containing female
{1}------------------------------------------------
Banana plug receptacles. The Valleylab generators should be used with the appropriate Valleylab accessories, reference the operating manuals for specifications.
- III. Irrigation
Irrigation with saline or sterile water is performed by connecting the irrigation hub to the irrigation port of the bipolar generator for the Standard plug or to a 30ml (30cc) or 60ml (60cc) luer-lock syringe for the Banana plug.
#### 5. Intended Use/Indications for Use:
The Injection Gold Probe™ Bipolar Electrohemostasis Catheter is indicated for use in endoscopic injection therapy (to deliver pharmacological injection agents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. These sites include the following: Peptic Ulcers, Dieulafoy Lesions, Mallory-Weiss Tears, Bleeding Polyp Stalks, Arteriovenous Malformations (AVMs), Angiomata. The Injection Gold Probe Bipolar Electrohemostasis Catheter also has irrigation capability. Any other use is not recommended.
## 6. Description of Change for this Submission:
The following changes are being made to the proposed Injection Gold Probe Bipolar Electrohemostasis Catheter:
- The current gold thick film conductor paste is being replaced with a new gold thick film . conductor paste.
- The Medi-Glide™ coating will be eliminated from the inside of the catheter and outside o of the needle hub/hypo tube assembly of the device and replaced with Teflon coating on the outside of the needle hub/hypo tube assembly only.
- The current ceramic tip material is being replaced with a new ceramic tip material. .
- The tolerance of the "radius" component has been slightly adjusted. o
- The length of the hypotube component of the needle assembly was reduced. o
## 7. Technological Characteristics:
The proposed Injection Gold Probe™ Bipolar Electrohemostasis Catheter has the same fundamental design, operating principal and intended use as the predicate Injection Gold Probe™ Bipolar Electrohemostasis Catheter (K961349). The proposed device has material differences and some modifications to needle subassembly when compared to the currently marketed Injection Gold ProbeTM Bipolar Electrohemostasis Catheter (K961349).
#### 8. Performance Data:
Biocompatibility Testing has been performed to demonstrate that the material changes made to the proposed device are biocompatible.
Bench Testing has been performed on the proposed Injection Gold Probe™ Bipolar Electrohemostasis Catheter to demonstrate that the modified device meets the required specifications of the completed tests.
## 9. Conclusion:
All biocompatibility tests conducted on the proposed Injection Gold Probe™ Bipolar Electrohemostasis Catheter passed. Therefore, the Injection Gold Probe™ Bipolar Electrohemostasis Catheter is considered biocompatible.
11
{2}------------------------------------------------
All device bench tests results were acceptable. The data demonstrate that the Injection Gold ProbeTM Bipolar Electrohemostasis Catheter meets design specifications and is suitable for its intended use.
Boston Scientific Corporation has demonstrated that the proposed Injection Gold Probe™ Bipolar Electrohemostasis Catheter is substantially equivalent to Boston Scientific Corporation's currently marketed Injection Gold Probe™ Bipolar Electrohemostasis (K961349).
000027
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three lines representing its wings and tail feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Ashley Pyle Sr. Regulatory Affairs Specialist Boston Scientific Corporation Endoscopy Division 100 Boston Scientific Way MARLBOROUGH MA 01752 .
APR 1 4 2011
Re: K110887
Trade/Device Name: Injection Gold Probe™ Bipolar Electrohemostasis Catheter Regulation Number: 21 CFR §876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Code: KNS Dated: March 29, 2011 Received: March 30, 2011
Dear Ms. Pyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
{4}------------------------------------------------
## Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Huckel Lemur. MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# SECTION 5 INDICATIONS FOR USE STATEMENT
| 510(k) Number (if known): | To Be Determined <span style="text-decoration: overline;">K110887</span> |
|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Injection Gold Probe™ Bipolar Electrohemostasis Catheter |
| Indications For Use: | The Injection Gold Probe™ Bipolar Electrohemostasis Catheter is indicated for use in endoscopic injection therapy (to deliver pharmacological injection agents, such as vasoconstrictors) and endoscopic electrohemostasis (cauterization of tissue and coagulation of blood) of actual or potential bleeding sites in the gastrointestinal tract. These sites include the following: |
| | <ul><li>Peptic Ulcers</li><li>Dieulafoy Lesions</li><li>Mallory-Weiss Tears</li><li>Bleeding Polyp Stalks</li><li>Arteriovenous Malformations (AVMs)</li><li>Angiomata</li></ul> |
| | The Injection Gold Probe Bipolar Electrohemostasis Catheter also has irrigation capability. Any other use is not recommended. |
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;">X</span> AND/OR Over-The-Counter Use<br>(21 CFR 807 Subpart C) |
| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | |
| <div> <img alt="Signature" src="signature.png" width="200"/> </div> Concurrence of CDRH, Office of Device Evaluation (ODE) | |
| (Division Sign-Off) | |
| Division of Reproductive, Gastro-Renal, and Urological Devices | <span style="text-decoration: overline;">K110887</span> |
| 510(k) Number | Special 510(k) Premarket Notification, Injection Gold Probe™ Bipolar Electrohemostasis Catheter 000014 |
. -*
:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.