← Product Code [JJW](/productcode/JJW) · K110846

# RANDOX URINALYSIS CONTROLS, LEVELS 1 AND 2 (K110846)

_Randox Laboratories Limited · JJW · Sep 30, 2011 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K110846

## Device Facts

- **Applicant:** Randox Laboratories Limited
- **Product Code:** [JJW](/productcode/JJW.md)
- **Decision Date:** Sep 30, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Randox Urinalysis Controls (URNAL Control 1 and URNAL Control 2) are urine controls containing Bilirubin, Blood, Creatinine, Crystals, Galactose, Glucose, hCG, Ketones, Leukocytes, Microalbumin, Nitrite, pH, Protein, Red Blood Cells, Specific Gravity, Urobilinogen and White Blood Cells. The Randox Urinalysis controls (URNAL Control 1 and URNAL Control 2) are intended for in vitro diagnostic use in the quality control of urine test strips for the analytes; Bilirubin, Blood, Creatinine, Glucose, Ketones, Leukocytes, Microalbumin, Nitrite, pH, Protein, Specific Gravity and Urobilinogen. This control is also intended for the evaluation of microscopic test procedures for Crystals, Red Blood Cells and White Blood Cells and also for the confirmatory tests hCG and Galactose. This in vitro diagnostic device is intended for prescription use only.

## Device Story

Randox Urinalysis Controls (URNAL Control 1 and URNAL Control 2) are liquid-stable, assayed quality control materials; contain Bilirubin, Blood, Creatinine, Crystals, Galactose, Glucose, hCG, Ketones, Leukocytes, Microalbumin, Nitrite, pH, Protein, Red Blood Cells, Specific Gravity, Urobilinogen, and White Blood Cells. Used in clinical laboratories to verify performance of urine test strips, microscopic examination procedures, and confirmatory tests. Healthcare providers use control results to validate accuracy and precision of patient sample testing; ensures reliability of diagnostic urinalysis results.

## Clinical Evidence

No clinical data. Bench testing only. Stability studies (real-time) confirmed 18-month shelf-life at 2-8°C and 30-day open-vial stability. Value assignment performed via in-house analysis and confirmed through external studies across 5 instrument families in 113 laboratories.

## Technological Characteristics

Liquid-stable, assayed urine control material. Contains stabilized human-derived or synthetic components for analytes including Bilirubin, Blood, Creatinine, Crystals, Galactose, Glucose, hCG, Ketones, Leukocytes, Microalbumin, Nitrite, pH, Protein, Specific Gravity, Urobilinogen, and White Blood Cells. Designed for use with standard urine test strips and microscopic urinalysis protocols. No electronic, software, or energy-based components.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k110846

B. Purpose for Submission:
New device

C. Measurand:
Quality control materials for multiple constituents in urine: glucose, bilirubin, ketones, specific gravity, blood, pH, protein, urobilinogen, nitrite, leukocytes, hCG, microalbumin, creatinine, galactose, red blood cells, white blood cells, and crystals

D. Type of Test:
Qualitative and semi-quantitative urinalysis controls for urinalysis reagent test strips, tablet tests, and manual cell counts with hemocytometer

E. Applicant:
Randox Laboratories Limited

F. Proprietary and Established Names:
Randox Liquid Urinalysis Controls, Level 1 and Level 2

G. Regulatory Information:
1. Regulation section:
862.1660, Quality control material, (assayed, and unassayed)
2. Classification:
I, reserved
3. Product code:
JJW
4. Panel:

{1}

75, Chemistry

## H. Intended Use:

1. Intended use(s):

See below

2. Indication(s) for use:

The Randox Urinalysis Controls (URNAL Control 1 and URNAL Control 2) are urine controls containing Bilirubin, Blood, Creatinine, Crystals, Galactose, Glucose, hCG, Ketones, Leukocytes, Microalbumin, Nitrite, pH, Protein, Red Blood Cells, Specific Gravity, Urobilinogen and White Blood Cells. The Randox Urinalysis controls (URNAL Control 1 and URNAL Control 2) are intended for in vitro diagnostic use in the quality control of urine test strips for the analytes; Bilirubin, Blood, Creatinine, Glucose, Ketones, Leukocytes, Microalbumin, Nitrite, pH, Protein, Specific Gravity and Urobilinogen. This control is also intended for the evaluation of microscopic test procedures for Crystals, Red Blood Cells and White Blood Cells and also for the confirmatory tests hCG and Galactose. This in vitro diagnostic device is intended for prescription use only.

3. Special conditions for use statement(s):

The labeling contains a statement that all human source material was tested by FDA-approved methods and found to be non-reactive for the presence of HBsAg, HCV and antibody to HIV 1/2.

In-vitro diagnostic use. Prescription use.

4. Special instrument requirements:

For use with the urine analyzers listed in the labeling.

## I. Device Description:

The Randox Urinalysis Controls URNAL Control 1 and URNAL Control 2 are ready-to-use, liquid controls derived from human urine. The source material has been tested by FDA approved methods and found non-reactive for Hepatitis B surface antigen, Hepatitis C and the antibody to HIV-1 and HIV-2. Each control level is purchased separately and is packaged in 12x12 mL bottles intended to be stored at 2-8°C.

{2}

J. Substantial Equivalence Information:

|  1. Predicate Device Name | 2. 510(k)  |
| --- | --- |
|  Quantimetrix Urine Dipstick Control | k874890  |
|  Quantimetrix Urinalysis Microscopics Control | k925256  |

3. Comparison with predicate:

COMPARISON OF RANDOX URINALYSIS CONTROLS WITH THE PREDICATE DEVICE

|  CHARACTERISTICS | RANDOX URINALYSIS CONTROLS LEVELS 1 & 2 | QUANTIMETRIX URINALYSIS MICROSCOPICS CONTROL k925256 | QUANTIMETRIX URINE DIPSTICK CONTROL k874890  |
| --- | --- | --- | --- |
|  INTENDED USE | The Randox Urinalysis Controls Level 1 and 2 are intended for in vitro diagnostic use in the quality control of the urinalysis test | Same | Same  |
|  TYPE OF TEST | Reagent strip, sediment examination, confirmatory tests | Sediment examination | Reagent strip, confirmatory tests  |
|  SIZE | 12ml | 120ml | 120ml  |
|  FORMAT | Liquid | Liquid | Liquid  |
|  MATRIX | Human Urine | Human Urine | Human Urine  |
|  STORAGE (Unopened) | 2 to 8°C until expiration date | 2 to 8°C until expiration date | 2 to 8°C until expiration date  |
|  OPEN VIAL CLAIM | 30 days at 2°C to 25°C | 2 to 8°C until expiration date | 2 to 8°C until expiration date  |
|  SHIPPING TEMPERATURE | 2 to 8°C | 2 to 8°C | 2 to 8°C  |
|  ANALYTES | Bilirubin, Blood, Creatinine, Crystals, Galactose, Glucose, hCG, Ketones, Leukocytes, Microalbumin, Nitrite, pH, Protein, Red Blood Cells, | Red Blood Cells, White Blood Cells, Casts, Crystals and Bacteria | Glucose, Bilirubin, Ketones, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes, Microalbumin, Creatinine, Total protein  |

{3}

Specific Gravity, Urobilinogen and White Blood Cells

K. Standard/Guidance Document Referenced (if applicable):

None were referenced.

L. Test Principle:

Not applicable

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Not applicable

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability and Value Assignment: Control solutions are derived from gravimetrically prepared stock solutions and analyzed in-house. Value assignment was determined through multiple analyses in-house for visual reading and on instrumentation. Urinalysis instrumentation calibration is factory preset. Control values are confirmed in external studies with 5 different instrument families in 113 laboratories. Assigned control ranges are listed in the labeling by manufacturer and method.

Stability: Real time stability studies were conducted for closed and open vials. Shelf-Life and Open Vial Stability testing protocols and acceptance criteria were described and found to be adequate.

Shelf-Life – The sponsor provided real time stability data to demonstrate that the Randox Urinalysis Controls, Levels 1 and 2, can be stored for 18 months at 2-8 °C.

Open-vial – The sponsor provided data demonstrating that upon opening a vial, the material is stable for 30 days at 2-8 °C.

d. Detection limit:

Not applicable

{4}

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
Not applicable

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

{5}

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

6

---

**Source:** [https://fda-staging.innolitics.com/device/K110846](https://fda-staging.innolitics.com/device/K110846)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
