ELITECH CLINICAL SYSTEMS CALCIUM ARSENAZO; ELITECH CLINICAL SYSTEMS ALBUMIN; ELITECH CLINICAL SYSTEMS ELICAL 2; ELITECH
Applicant
Elitechgroup
Product Code
CIX · Clinical Chemistry
Decision Date
Jun 21, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1035
Device Class
Class 2
Indications for Use
ELITech Clinical Systems CALCIUM ARSENAZO is intended for the quantitative in vitro diagnostic determination of total calcium in human serum and plasma on ELITech Clinical Systems Selectra analyzers. It is not intended for use in Point of Care settings. ELITech Clinical Systems ALBUMIN is intended for the quantitative in vitro diagnostic determination of albumin in human serum and plasma on ELITech Clinical Systems Selectra analyzers. It is not intended for use in Point of Care settings. ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers. ELITech Clinical Systems ELITROL I and ELITROL II are multi-parametric control sera for in vitro diagnostic use in quality control of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra Analyzers.
Device Story
The ELITech Clinical Systems suite includes colorimetric reagents for calcium and albumin, a lyophilized multi-analyte calibrator (ELICAL 2), and two-level lyophilized controls (ELITROL I/II). These are intended for use on ELITech Clinical Systems Selectra analyzers in clinical laboratory settings. The calcium assay uses Arsenazo III dye-binding; the albumin assay uses Bromocresol green dye-binding. The system processes human serum or plasma samples. Healthcare providers use the quantitative results to diagnose and monitor parathyroid, bone, renal, and liver/kidney conditions. The calibrator and controls ensure assay accuracy and precision. The device is not for point-of-care use.
Clinical Evidence
Bench testing only. Performance validated per CLSI guidelines (EP5-A2, EP6-A, EP7-A2, EP9-A2, EP17-A). Precision studies showed total CVs of 1.9-2.3%. Linearity confirmed across 1.6-6.0 g/dL (Albumin) and 5.0-15.0 mg/dL (Calcium). Method comparison against predicates yielded R² values ≥ 0.972. Interference testing identified positive bias for Calcium with triglycerides > 1119 mg/dL.
Technological Characteristics
Colorimetric assays using Arsenazo III (calcium) and Bromocresol green (albumin). Reagents are liquid, ready-to-use. Calibrator and controls are lyophilized human serum-based. Analyzers: Selectra ProM. Calibration frequency: 28 days. Storage: 2-8°C (reagents/calibrator/controls).
Indications for Use
Indicated for: Calcium measurements used in diagnosis/treatment of parathyroid disease, bone diseases, chronic renal disease, and tetany. Albumin measurements used in diagnosis/treatment of liver or kidney diseases. Calibrator and controls used for calibration and quality control of quantitative assays on Selectra analyzers.
Regulatory Classification
Identification
An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
Roche Diagnostics Calibrator for Automated Systems (k033501)
Roche Diagnostics Precinorm Universal and Precipath Universal Control Sera (k041227)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k110830
B. Purpose for Submission:
New device
C. Measurand:
Albumin, Calcium
D. Type of Test:
Quantitative, colorimetric chemistry tests
E. Applicant:
Seppim S.A.S.
F. Proprietary and Established Names:
ELITech Clinical Systems Calcium Arsenazo
ELITech Clinical Systems Albumin
ELITech Clinical Systems ELICAL 2
ELITech Clinical Systems ELITROL I
ELITech Clinical Systems ELITROL II
G. Regulatory Information:
{1}
| Measurand | Regulation Section | Classification | Product Code | Panel |
| --- | --- | --- | --- | --- |
| Albumin | 21CFR862.1035 | II | CIX-Bromocresol green dye-binding, albumin | (75) Clinical Chemistry |
| Calcium | 21CFR862.1145 | II | CJY – Azo dye, calcium | (75) Clinical Chemistry |
| Calibrator | 21CFR862.1150 | II | JIX-Calibrator, multi-analyte mixture | (75) Clinical Chemistry |
| Control | 21CFR862.1660 | I, Reserved | JJY – Multi-Analyte Controls | (75) Clinical Chemistry |
H. Intended Use:
1. Intended use(s):
See Indications for use below.
2. Indication(s) for use:
ELITech Clinical Systems Albumin
ELITech Clinical Systems Albumin is intended for the quantitative in vitro diagnostic determination of albumin in human serum and plasma on ELITech Clinical Systems Selectra analyzers. It is not intended for use in Point of Care settings.
Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.
ELITech Clinical Systems Calcium Arsenazo
ELITech Clinical Systems Calcium Arsenazo is intended for the quantitative in vitro diagnostic determination of total calcium in human serum and plasma on ELITech Clinical Systems Selectra analyzers. It is not intended for use in Point of Care settings.
Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms).
{2}
3
ELITech Clinical Systems ELICAL 2
ELITech Clinical Systems ELICAL 2 is a multi-parametric calibrator for in vitro diagnostic use in the calibration of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers.
ELITech Clinical Systems ELITRIOL I and ELITRIOL II
ELITech Clinical Systems ELITRIOL I and ELITRIOL II are multi-parametric control sera for in vitro diagnostic use in quality control of quantitative ELITech Clinical Systems methods on ELITech Clinical Systems Selectra analyzers.
3. Special conditions for use statement(s):
For in vitro diagnostic use only. For prescription use only
4. Special instrument requirements:
For use with the ELITech Clinical Systems Selectra ProM analyzer
I. Device Description:
The ELITech Clinical Systems Calcium Arsenazo and ELITech Clinical Systems Albumin kit reagents are one reagent systems. Reagents are supplied in liquid ready-to-use forms. Each kit is composed of 12 x 20 mL vials of reagent. The calcium reagent contains MES buffer (100 mM) and arsenazo III (200 µM). The albumin reagent contains succinate buffer (87 mM), bromocresol green (0.2 mM), and non-ionic surfactant brij 35 (7.35 mL/L).
ELITech Clinical Systems ELICAL2 is a lyophilized calibrator based on human serum containing constituents to ensure optimal calibration. ELICAL 2 is prepared exclusively from the blood of donors tested individually and found to be negative for HbsAg and to the antibodies to HCV and HIV according to FDA-approved methods.
ELITROL I and ELITROL II are two level quality control products consisting of a lyophilized human serum containing constituents at desired levels. ELITROL I and ELITROL II are prepared exclusively from the blood of donors tested individually and found to be negative for HBsAg and to antibodies to HCV and HIV according to FDA-approved methods.
Substantial Equivalence Information:
1. Predicate Device Name(s):
Stanbio Calcium LiquiColor (Arsenazo III)
Roche Diagnostics Albumin Gen.2
Roche Diagnostics Calibrator for Automated Systems
Roche Diagnostics Precinorm Universal and Precipath Universal Control Sera
{3}
2. Predicate 510(k) number(s): k921625, k063744, k033501, k041227
3. Comparison with predicate:
| Similarities and Differences for Albumin | | |
| --- | --- | --- |
| Item | Candidate Device ELITech Clinical Systems Albumin | Predicate Device (k063744) Roche Diagnostics Albumin Gen.2 |
| Intended Use | Intended for the quantitative in vitro diagnostic determination of albumin in human serum and plasma. | Same |
| Indications For Use | Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys. | Same |
| Specimen Type | Human serum, plasma | Same |
| Assay Type | Liquid ready-for-use | Same |
| Assay Principle | Colorimetric | Same |
| Analytical Range | 1.6 – 6.0 g/dL | 0.2 – 6.0 g/dL |
| Reagent Storage | 2-25°C, stable until expiration date on label | 15-25°C, stable until expiration date on label |
| Stability (on board) | 28 days | 4 weeks |
| Instrument | Selectra ProM Analyzer | Cobas c111 Analyzer |
| Similarities and Differences for Calcium | | |
| --- | --- | --- |
| Item | Candidate Device ELITech Clinical Systems Calcium Arsenazo | Predicate Device (k921625) Stanbio Calcium LiquiColor Arsenazo III |
| Intended Use | Intended for the quantitative in vitro diagnostic determination of total calcium in human serum and plasma. | Same |
{4}
| Similarities and Differences for Calibrator (ELICAL 2) | | |
| --- | --- | --- |
| Item | Candidate Device
ELITech Clinical Systems
ELICAL 2 | Predicate Device
(k033501)
Roche Diagnostics Calibrator
for Automated Systems |
| Intended Use/Indications for Use | For in vitro diagnostic use in the calibration of quantitative methods | Same |
| Format | Lyophilized calibrator based on human sera | Same |
| Levels | One | Same |
| Stability | Lyophilized: 2 to 8°C, stable until expiration date on label
Reconstituted:
15 to 25°C, 8 hours
2 to 8 °C, 2 days
-25 to -15°C, 4 weeks | Same |
{5}
| Similarities and Differences for Controls (ELITROL I and II) | | |
| --- | --- | --- |
| Item | Candidate Device
ELITech Clinical Systems
ELITROL I and ELITROL II | Predicate Device
(k041227)
Roche Diagnostics
Precinorm Universal and
Precipath Universal Control
Sera |
| Intended Use/Indications for Use | For in vitro diagnostic use in quality control of quantitative methods | Same |
| Format | Lyophilized controls based on human sera | Same |
| Levels | Two (ELITROL I and ELITROL II) | Same |
| Stability | Lyophilized: 2 to 8°C, stable until expiration date on label
Reconstituted:
15 to 25°C, 12 hours
2 to 8 °C, 5 days
-25 to -15°C, 4 weeks | Same |
J. Standard/Guidance Document Referenced (if applicable):
- CLSI Guideline, EP5-A2 Evaluation of Precision Performance of Clinical Chemistry Devices – Second Edition
- CLSI Guideline, EP9-A2 Method Comparison and Bias Estimation Using Patient Samples – Second Edition
- CLSI Guideline, EP6-A Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
- CLSI Guideline, EP7-A2 Interference Testing in Clinical Chemistry; Approved Guideline – Second Edition
- CLSI Guideline, EP17-A Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline
- FR EN 13640:2002 Stability Testing of In Vitro Diagnostic Reagents
- In Vitro Diagnostics Devices: Guidance for the preparation of 510(k) submissions (1977)
- Guidance for Industry and FDA Staff: “Format for Traditional and Abbreviated 510(k)s” (2005)
- Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use: Guidance for Industry and FDA Staff (2004)
{6}
7
L. Test Principle:
ELITech Clinical Systems Albumin
This method involves the binding of albumin with bromocresol green. Bromocresol green forms a complex with albumin at pH 4.20. The color intensity at 620 nm is directly proportional to the albumin concentration in the sample.
ELITech Clinical Systems Calcium Arsenazo
This method utilizes Arsenazo III, which has a high affinity for calcium at neutral pH. Arsenazo III forms in neutral medium a blue complex with calcium. The color intensity at 600 nm in directly proportional to the total calcium concentration in the sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Serum - Within run and total precision studies were performed according to CLSI EP5-A2 for albumin and calcium. Studies were performed by conducting two runs per day, two measures per run, for 3 levels of samples, on 2 Selectra ProM instruments for 20 operating days. For albumin, data from level 1 is from a diluted serum sample, level 2 is from a natural human serum pool sample, and level 3 is from a spiked human serum pool sample. For calcium, data from level 1 is from a control serum sample, level 2 is from a natural human serum pool sample, and level 3 is from a spiked human serum pool sample. Results are presented in the tables below.
Albumin
| | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Concentration Levels (g/dL) | 2.50 | 3.50 | 5.00 |
| No. of Samples | 80 | 80 | 80 |
| No. of Replicates | 2 | 2 | 2 |
| | | | |
| Mean | 2.54 | 3.53 | 4.98 |
| Total SD | 0.06 | 0.08 | 0.10 |
| Total %CV | 2.3 | 2.1 | 2.1 |
| Within-Run SD | 0.02 | 0.02 | 0.04 |
| Within-Run %CV | 0.9 | 0.5 | 0.8 |
{7}
8
## Calcium
| | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Concentration Levels (mg/dL) | 8.00 | 10.00 | 12.00 |
| No. of Samples | 80 | 80 | 80 |
| No. of Replicates | 2 | 2 | 2 |
| | | | |
| Mean | 8.75 | 9.68 | 11.97 |
| Total SD | 0.17 | 0.19 | 0.22 |
| Total %CV | 1.9 | 1.9 | 1.9 |
| Within-Run SD | 0.11 | 0.09 | 0.08 |
| Within-Run %CV | 1.3 | 0.9 | 0.7 |
## b. Linearity/assay reportable range:
Serum-Linearity was evaluated for albumin and calcium according to CLSI EP6-A by comparing observed versus expected values for 11 equally-spaced serum samples. Samples were prepared from high and low analyte concentration serum pools. Each sample was evaluated in triplicate on the Selectra ProM instrument. Linear regression results from these studies are found in the table below.
| | Slope | Intercept | R² | Standard Error | Concentration Ranges tested |
| --- | --- | --- | --- | --- | --- |
| ALB | 0.9648 | 0.1082 | 0.9976 | 0.08 | 1.55 – 6.08 g/dL |
| Ca | 1.0235 | -0.1865 | 0.9993 | 0.10 | 4.96 – 15.39 mg/dL |
The results of the study support the sponsor’s claims that albumin assay is linear across the measuring range of $1.6 - 6.0\mathrm{g / dL}$ and the calcium assay is linear across the measuring range of $5.0 - 15.0\mathrm{mg / dL}$.
## c. Traceability, Stability, Expected values (controls, calibrators, or methods):
### Traceability
The sponsor claims traceability of albumin and calcium used in the ELICAL 2 calibrator, ELITROL I, and ELITRIOL II to reference materials, ERM-DA 470 and SRM 909b, respectively.
### Value Assignment
The ELICAL 2 calibrator solution and the ELITROL I and ELITROL II
{8}
control solutions are tested against predetermined values on the Selectra ProM analyzer using the albumin and calcium reagents. Value assignment for the ELICAL 2 calibrator solution was performed by quadruplicate runs on the Selectra ProM analyzer, using 2 reagent lots, and 3 test vials per lot. Value assignment for the ELITROL I and ELITRIOL II control solutions was performed by triplicate runs on the Selectra ProM analyzer, using 2 reagent lots, and 3 test vials per lot. The mean analyte value is calculated and a target value is assigned. The labeling states that obtained control solution values should fall within the specified range provided on lot-specific value sheets and that laboratories should establish appropriate quality control procedures when using this product for its intended use. The labeling also states that laboratories should recalibrate when reagents lots change, when quality control results fall out of range, and after a maintenance operation.
## Stability
Calibrator material is purchased from a commercial vendor (previously cleared under k033501). The sponsor claimed the following for stability: ELICAL 2 is stable until the expiration date printed on the label when stored at 2-8°C prior to reconstitution. After reconstitution the stability is 8 hours when stored at 15-25°C, 2 days at 2-8°C or 4 weeks (when frozen once) at -25° and -15°C.
Control material is purchased from a commercial vendor (previously cleared under k041227). The sponsor claimed the following for stability: Before reconstitution, the shelf-life of the ELITech Clinical Systems Elitrol I and Elitrol II is 30 months at 2-8°C. After reconstitution the stability is 12 hours when stored at 15-25°C, 5 days when stored at 2-8°C or 4 weeks (when frozen once) at -25° and -15°C.
## d. Detection limit:
Serum-Limit of blank (LoB), limit of detection (LoD), and limit of quantification (LoQ) were determined according to CLSI EP17-A with the Selectra ProM instrument for calcium and albumin. Testing was conducted over multiple days, with one run across two instruments, and with two reagent lots. The results are as follows:
| | LOB | LOD | LOQ | Acceptable Total Error of LOQ | Claimed measuring range |
| --- | --- | --- | --- | --- | --- |
| No. Samples tested | 1 | 4 | 4 | | |
| No. | 60 | 15 | 15 | | |
{9}
| Replicates | | | | | |
| --- | --- | --- | --- | --- | --- |
| ALB | 0.002 g/dL | 0.003 g/dL | 0.50 g/dL | ≤ 0.05 g/dL | 1.6 – 6.0 g/dL |
| Ca | 0.34 mg/dL | 0.36 mg/dL | 5.00 mg/dL | ≤ 0.32 mg/dL | 5.0 – 15.0 mg/dL |
# e. Analytical specificity:
Serum-Interference studies were performed according to the CLSI EP7-A2 for albumin and calcium analytes. Different concentrations of potential interference substances were spiked into pooled patient serum samples. The sponsor's definitions of no significant interference is $< 10\%$ difference between the tested and the control samples. Two albumin test concentrations: $3.50~\mathrm{g / dL}$ and $5.00~\mathrm{g / dL}$ , and two calcium test concentrations: $8.00~\mathrm{mg / dL}$ and $12.00~\mathrm{mg / dL}$ were used in these studies. There was no significant interference for albumin and calcium when these analytes and interferents were tested in the concentration ranges indicated in the table below:
| Albumin |
| --- |
| Triglycerides (up to 3000 mg/dL) |
| Unconjugated bilirubin (up to 30.0 mg/dL) |
| Conjugated bilirubin (up to 29.5 mg/dL) |
| Hemoglobin (up to 500 mg/dL) |
| Acetaminophen (up to 30 mg/dL) |
| Ascorbic acid (up to 20mg/dL) |
| Acetylsalicylic acid (up to 200 mg/dL) |
| γ globulin (up to 1500 mg/dL) |
| Calcium |
| --- |
| Triglycerides (up to 1119 mg/dL) |
| Unconjugated bilirubin (up to 30.0 mg/dL) |
| Conjugated bilirubin (up to 29.5 mg/dL) |
| Hemoglobin (up to 500 mg/dL) |
| Magnesium (up to 10.91 mg/dL) |
| Acetaminophen (up to 30 mg/dL) |
| Ascorbic acid (up to 20mg/dL) |
| Acetylsalicylic acid (up to 200 mg/dL) |
Triglyceride concentrations greater than $1119\mathrm{mg / dL}$ were found to interfere with calcium measurement. Therefore the sponsor states in their labeling that a positive bias is observed with triglyceride concentrations above 1119 $\mathrm{mg / dL}$ .
{10}
f. Assay cut-off:
Not applicable.
# 2. Comparison studies:
a. Method comparison with predicate device:
Serum and Plasma- Two sets of method comparison studies, serum and plasma, were performed according to the CLSI EP9-A2 guideline. The serum study was completed with 100 albumin (10 spiked and 90 unaltered samples) and 102 calcium human serum samples (9 spiked and 93 unaltered samples) spanning the linear range of the assay. The plasma study was completed with (40) lithium heparinized human plasma samples spanning the linear range of each assay. Each serum and plasma sample was analyzed in singlet using the Selectra ProM analyzer (test method) and in duplicate on the predicates (k921625 - Stanbio Calcium LiquiColor reagent on the SIRRUS analyzer and k063744 - Roche Diagnostics Albumin Gen.2 on the cobas c11 analyzer). Comparison of individual test values versus mean predicate values yielded the following results:
| Matrix-Analyte | Slope | Intercept | R2 | Conc. Range Tested |
| --- | --- | --- | --- | --- |
| Serum-ALB | 0.961 | 0.122 | 0.994 | 1.6 – 5.54 g/dL |
| Plasma- ALB | 0.952 | 0.251 | 0.997 | 1.71 – 5.26 g/dL |
| Serum- Ca | 1.008 | -0.001 | 0.991 | 5.03 – 14.40 mg/dL |
| Plasma- Ca | 0.992 | 0.246 | 0.972 | 5.47 – 14.76 mg/dL |
b. Matrix comparison:
Matrix comparison studies were not performed. See serum and plasma method comparison data above (Section M.2.a.).
The sponsor claimed that lithium heparin plasma is an acceptable anticoagulant.
# 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
{11}
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not applicable.
5. Expected values/Reference range:
Reference ranges are provided in the labeling from literature as follows:
| | Serum/Plasma |
| --- | --- |
| Albumin | 3.5 – 5.2 g/dL (Adults) |
| | 3.2 – 4.6 g/dL (60-90 years) |
| | 2.9 – 4.5 g/dL (90+ years) |
| Calcium | 8.6 – 10.3 mg/dL |
Wu, A.H.B., Tietz Clinical Guide to Laboratory Test, $4^{\text{th}}$ Ed., (W.B. Saunders Company), (2006), 66.
Endres, D.B., Rude, R.K., Disorders of Bone, Tietz Fundamentals of Clinical Chemistry, $6^{\text{th}}$ Ed., Burtis, C.A., Ashwood, E.R., Bruns, D.E. (Saunders), (2008), 711.
# N. Proposed Labeling:
The labeling is sufficient and does satisfy the requirements of 21 CFR Part 809.10.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.