K110807 · Stryker Corp. · MAY · May 25, 2011 · Orthopedic
Device Facts
Record ID
K110807
Device Name
ABG II MONOLITHIC HIP STEM
Applicant
Stryker Corp.
Product Code
MAY · Orthopedic
Decision Date
May 25, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The indications for use for total hip arthroplasty include: noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis: rheumatoid arthritis: correction of functional deformity; revision procedures where other treatments or devices have failed; and. nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. Additional indication specific to use of ABG™II Monolithic Hip Stem with compatible Howmedica Osteonics Constrained Liners: When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability. Stryker's ABG™ II Monolithic Hip Stem is intended for cementless use only and is intended for total and hemi-arthroplasty procedures.
Device Story
ABG II Monolithic Hip Stem is a sterile, single-use femoral prosthesis for cementless, press-fit total or hemi-arthroplasty. Device consists of TMZF (Ti-11.5 Mo-6Zr-2Fe) alloy stem with proximal hydroxylapatite coating; features 12° anteversion and 130° neck angle. Used by orthopedic surgeons in clinical settings to replace diseased or damaged femoral head/neck, restoring hip function and alleviating pain. Compatible with various Howmedica Osteonics femoral heads, bipolar heads, and acetabular components. Surgeon selects appropriate size (1-8) based on patient anatomy; device is implanted via press-fit technique. Provides structural support for hip joint; enables articulation when paired with compatible acetabular components. Benefits include pain relief and improved mobility for patients with degenerative joint disease or fractures.
Clinical Evidence
No clinical data provided. Substantial equivalence established via non-clinical bench testing, specifically femoral neck and distal stem fatigue testing performed in accordance with FDA guidance 'Non-clinical Information for Femoral Stem Prostheses (17 September 2007)'.
Indicated for patients requiring primary or revision total hip arthroplasty or hemiarthroplasty due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, nonunions, or proximal femur fractures. Constrained liner use indicated for patients at high risk of dislocation (prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intraoperative instability).
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
K110807
## 510(k) Summary of Safety and Effectiveness
MAY 2 5 2011
| Sponsor | Howmedica Osteonics Corp. |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| | 325 Corporate Drive |
| | Mahwah, NJ 07430 |
| Contact Person | Stephanie Fitts |
| | Sr. Director, Regulatory Affairs & Regulatory Compliance |
| | Howmedica Osteonics Corp. |
| | 325 Corporate Drive |
| | Mahwah, NJ 07430 |
| | Phone: (201) 831-5405 |
| Date Prepared: | March 21, 2011 |
| Proprietary Name: | ABG ™ II Monolithic Hip Stem |
| Common Name: | Hip prosthesis |
| Classification Name: | Hip joint metal/ceramic/polymer semi-constrained cemented or |
| | nonporous uncemented prosthesis, 21 CFR §888.3353 (Product<br>Codes MAY, LZO, MEH) |
| | Hip joint metal/polymer/metal semi-constrained porous coated<br>uncemented prosthesis, 21 CFR §888.3358 (Product Codes LPH,<br>MBL) |
Legally Marketed Device to Which Substantial Equivalence is Claimed: ABG TM II Modular Hip Stem K092406.
## Device Description:
Howmedica Osteonics is introducing a monolithic femoral hip prosthesis. The basic design of the ABG™ II Monolithic Hip Stem, referred as the ABG™ II, is similar to other total hip systems commercially distributed such as the Stryker ABG™ II Modular Hip System.
The subject hip stem is a TMZF (Ti-11.5 Mo-6Zr-2Fe) alloy femoral stem with a roughened hydroxylapatite coating in the proximal region. It is intended for cementless, press-fit application and designed for use with the currently available compatible Howmedica Osteonics' femoral heads, bipolar heads, and their compatible acetabular components.
The ABGTM II Monolithic Hip Stem is a sterile, single-use device intended for use in primary and revision total hip arthroplasty to alleviate pain and restore function. The stem is designed for use with the currently available compatible Howmedica Osteonics' femoral heads and their compatible acetabular components. Head compatibility with the stems includes: V40 Biolox Delta, Biolox Delta Universal Taper Heads and Sleeves, V40 CoCr Heads, V40 LFIT CoCr Heads, C-Taper Alumina Heads when used with the V40/C-taper Adaptor, C-Taper Delta Heads when used with C-taper Adaptor, UHR Universal Head, Unitrax Heads when used with the Unitrax V40 Modular Adapter.
{1}------------------------------------------------
The ABGTM II Monolithic Hip Stem will be available in 8 sizes ranging from size 1 through 8 with one neck angle of 130°. The ABG™ II Monolithic Hip Stem anatomic stem is made with 12° of anteversion built into the proximal portion of the stem.
## Indications:
The indications for use for total hip arthroplasty include:
- t noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- 프 rheumatoid arthritis;
- correction of functional deformity;
- .. revision procedures where other treatments or devices have failed; and,
- . nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of ABG™ II Monolithic Hip Stem with compatible Howmedica Osteonics Constrained Liners:
- · When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intraoperative instability.
The Stem is intended for cementless, press-fit use only and is intended for total and hemiarthroplasty procedures'.
Summary of Technologies: Device comparison showed that the proposed device is substantially equivalent in intended, use, materials and performance characteristics to the predicate device.
Non-Clinical Testing: Non-clinical laboratory testing was performed for the hip stem to determine substantial equivalence. Non-clinical testing was provided as outlined in the FDA Guidance Document entitled "Non-clinical Information for Femoral Stem Prostheses (17 September 2007)". Femoral neck fatigues testing and distal stem fatigues testing were conducted on the worst-case sizes determined by Finite Element Analysis. The testing demonstrated that the ABG ™ II Monolithic Hip Stem is substantially equivalent to devices currently cleared for marketing.
Clinical Testing: Clinical testing was not required for this submission.
Conclusion: The ABG ™ II Monolithic Hip Stem is substantially equivalent to the predicate devices identified in this premarket notification.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the staff and a winding shape representing the snakes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Howmedica Osteonics Corp. % Dr. Stephanie Fitts 325 Corporate Drive Mahwah, NJ 07430
MAY 25 2011
Re: K110807
Trade/Device Name: ABG™ II Monolithic Total Hip Stem Regulation Number: 21 CFR §888.3353; 21 CFR §888.3358 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis; Hip joint metal/polymer/metal semi-constrained porous coated uncemented prosthesis Regulatory Class: II Product Code: MAY, LZO, MEH; LPH, MBL Dated: March 21, 2010 Received: March 23, 2010
Dear Dr. Fitts.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), . You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Dr. Stephanie Fitts
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourceforYou/Industry/default.htm
Sincerely yours,
Mark A. Milleson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices _Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Section 4: Indications for Use Statement
510(k) Number (if known):
Device Name: ABGTM II Monolithic Hip Stem
The indications for use for total hip arthroplasty include:
noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis:
rheumatoid arthritis:
correction of functional deformity;
revision procedures where other treatments or devices have failed; and.
nonunions, femoral neck fractures, and trochanteric fractures of the proximal
femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of ABG™II Monolithic Hip Stem with compatible Howmedica Osteonics Constrained Liners:
When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
Stryker's ABG™ II Monolithic Hip Stem is intended for cementless use only and is intended for total and hemi-arthroplasty procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use ----------AND/OR (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (Of
Mark A. Miller
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110807
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.