← Product Code [MBI](/productcode/MBI) · K110800

# META-LOCK SUTURE ANCHOR SYSTEM (K110800)

_Tarsus Medical, Inc. · MBI · Sep 27, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K110800

## Device Facts

- **Applicant:** Tarsus Medical, Inc.
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Sep 27, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Tarsus Medical Meta-Lock Suture Anchor System is intended for fixation of suture to bone for the indications listed below: SHOULDER: Bankart repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsulo-labral reconstruction, biceps tenodesis, Deltoid repair - Lateral instability, medial instability, achilles tendon repair/reconstruction, Midfoot ANKLE: reconstructions - FOOT: Hallux valgus reconstruction - WRIST: Scapholunate ligament - HAND: Ulnar or lateral collateral ligament reconstruction - ELBOW: Biceps tendon reattachment - . KNEE: Extra capsular repairs; reattachment of : medial collateral ligament, lateral collateral ligament, posterior oblique ligament or joint capsule to tibia and joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tendonesis; patellar ligament and tendon avulsions.

## Device Story

Meta-Lock Suture Anchor System; mechanical device for reattaching soft tissue to bone. Comprised of titanium and stainless steel anchor with polyester suture. Used by surgeons in clinical/OR settings for orthopedic repairs. Device provides secure fixation point in bone; allows surgeon to anchor sutures for tissue reattachment. Benefits patient by restoring anatomical integrity of ligaments/tendons. No electronic or software components.

## Clinical Evidence

Bench testing only. Non-clinical tests included anchor tensile testing, system tensile testing, cyclic testing, magnetic resonance (MR) testing, galvanic corrosion testing, simulated movement testing, and simulated use testing. No clinical data presented.

## Technological Characteristics

Materials: Commercially Pure Titanium (Grade 2) per ASTM F67-06; Stainless Steel (Biodur 108 Alloy) per ASTM F2229; polyester suture. Mechanical fixation device. No energy source, connectivity, or software.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- GII QuickAnchor Plus ([K051989](/device/K051989.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

1/2

K//0800

# SEP 2 7 2011

## 5. 510(k) Summary

| General Information |                                                                                                      |
|---------------------|------------------------------------------------------------------------------------------------------|
| Date Compiled       | September 21, 2011                                                                                   |
| Classification      | Class II, 21 CFR § 888.3040, Smooth or threaded metallic bone fixation fastener.<br>Product code MBI |
| Trade Name          | Meta-Lock™ Suture Anchor System                                                                      |
| Submitter           | Tarsus Medical, Inc.<br>465 Fairchild Drive<br>Suite 230<br>Mountain View, CA 94043                  |
| Contact             | Nicholas Mourlas<br>President and CEO<br>Tel: (650) 237-0070<br>Fax: (650) 237-0071                  |
| Intended Use        |                                                                                                      |

The Tarsus Medical Meta-Lock Suture Anchor System is intended for fixation of suture to bone for the indications listed below:

| SHOULDER: Bankart repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, |
|---------------------------------------------------------------------------------------------------|
| capsule shift/capsulo-labral reconstruction, biceps tenodesis, Deltoid repair                     |

- Lateral instability, medial instability, achilles tendon repair/reconstruction, Midfoot ANKLE: reconstructions
- FOOT: Hallux valgus reconstruction
- WRIST: Scapholunate ligament
- HAND: Ulnar or lateral collateral ligament reconstruction
- ELBOW: Biceps tendon reattachment
- . KNEE: Extra capsular repairs; reattachment of : medial collateral ligament, lateral collateral ligament, posterior oblique ligament or joint capsule to tibia and joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tendonesis; patellar ligament and tendon avulsions.

Predicate Device GII QuickAnchor Plus Manufactured by Depuy Mitek

K051989

Device Description

The Meta-Lock™ Suture Anchor System is a suture anchor used for the reattachment of soft tissue to bone.

Materials

The Meta-Lock is comprised of Commercially Pure Titanium (Grade 2) conforming to ASTM F67-06, Stainless Steel (Biodur 108 Alloy) conforming to ASTM F2229, and polyester suture.

{1}------------------------------------------------

110800 2/2

### Testing

The non-clinical tests performed by the company include anchor tensile testing, system tensile testing, cyclic testing, magnetic resonance (MR) testing, galvanic corrosion testing, simulated movement testing, and simulated use testing. The test results demonstrate that the Meta-Lock is substantially equivalent to the legally marketed predicate device.

#### Summary of Substantial Equivalence

Tarsus Medical, Inc. believes the Meta-Lock Suture Anchor System is substantially equivalent to the predicate product. The intended use, method of operation, methods of construction and materials used, are either identical or substantially equivalent to the existing legally marketed predicate product.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features three diagonal lines that curve and merge into a single line at the bottom. The symbol is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the border of the circle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Tarsus Medical, Inc. % Nicholas Mourlas 465 Fairchild Drive, Suite 230 Mountain View, CA 94043

SEP 2 7 2011

Re: K110800

Trade/Device Name: Meta-Lock™ Suture Anchor System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: MBI Dated: August 16", 2011 Received: August 17", 2011

Dear Dr. Mourlas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Paris 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

### Page 2 - Dr. Nicholas Mourlas

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Eimel Keith

for Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## 4. Indications for Use Statement

| 510(k) Number<br>(if known): | This application | K110800 |
|------------------------------|------------------|---------|
|------------------------------|------------------|---------|

Device Name:

Meta-Lock™ Suture Anchor System

| Indications for Use: | The Tarsus Medical Meta-Lock Suture Anchor System is intended for fixation of suture to bone for the indications listed below:                                                                                                                                                                       |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SHOULDER:            | Bankart repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsulo-labral reconstruction, biceps tenodesis; deltoid repair                                                                                                                                |
| ANKLE:               | Lateral instability, medial instability, achilles tendon repair/reconstruction, midfoot reconstructions                                                                                                                                                                                              |
| FOOT:                | Hallux valgus reconstruction                                                                                                                                                                                                                                                                         |
| WRIST:               | Scapholunate ligament                                                                                                                                                                                                                                                                                |
| HAND:                | Ulnar or lateral collateral ligament reconstruction                                                                                                                                                                                                                                                  |
| ELBOW:               | Biceps tendon reattachment                                                                                                                                                                                                                                                                           |
| KNEE:                | Extra capsular repairs; reattachment of : medial collateral ligament, lateral collateral ligament, posterior oblique ligament or joint capsule to tibia and joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tendonesis; patellar ligament and tendon avulsions. |

Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Per 21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Michael Olsen for MXM
(Division Sign-Off)

Division of Surgical. Orthopedic, and Restorative Dev ترك

**510(k) Number** K110806

011

---

**Source:** [https://fda-staging.innolitics.com/device/K110800](https://fda-staging.innolitics.com/device/K110800)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).