Patients undergoing mammography (implied by device use)
Not applicable for this study
Clinical acceptance of images processed with the algorithm
Indications for Use
The Senographe DS and Senographe Essential FFDM systems generate digital mammographic images that can be used for screening and in the diagnosis of breast cancer. The Senographe DS and Senographe Essential FFDM systems are intended to be used in the same clinical applications as traditional film-based mammographic systems. Premium View is an image-processing algorithm, which increases the visibility of breast structures. The main advantage is to provide a single breast image, where the contrast in the fatty tissues is similar to that obtained by setting WW (window width) and WL (window level) for optimum visualization of fatty tissues, and the contrast in the fibroglandular tissue is similar to that obtained by setting WW (window width) and WL (window level) for optimal visualization of fibro-glandular tissues. PV i is an option that can simplify the presentation of mammographic images, improve workflow, and streamline the review process of images with very dark or bright areas by presenting the image with the WW and WL optimized for review with minimal need for the user to make adjustments for the various tissue areas. This could be especially useful with patients who have very dense breasts, or implants that currently require multiple adjustments of WW and WL to review.
Device Story
Premium View i (PVi) is a software-based image processing upgrade for GE Senographe DS and Senographe Essential Full Field Digital Mammography (FFDM) systems. The system captures raw digital mammographic data via an amorphous silicon flat panel detector with cesium iodide scintillator. PVi processes these images to enhance breast structure visibility by applying specific Look Up Tables (LUTs) to high-pass and low-pass image extractions. This optimizes contrast for both fatty and fibro-glandular tissues simultaneously, reducing the need for manual window width (WW) and window level (WL) adjustments during clinical review. The device is used in clinical settings by radiologists to streamline image review, particularly for dense breasts or implants. By presenting an optimized image, it improves workflow efficiency and aids in diagnostic decision-making. The system is deployed as a software upgrade or option on existing or new FFDM installations.
Clinical Evidence
Clinical testing was performed to quantify clinical acceptance of images retrospectively processed with the Premium View i algorithm. No specific quantitative metrics (e.g., sensitivity/specificity) were provided in the summary; the conclusion of substantial equivalence was based on clinical acceptance and non-clinical verification/validation.
Technological Characteristics
Software-based image processing algorithm (Premium View i) for FFDM systems. Utilizes high-pass and low-pass image extraction with specific Look Up Tables (LUTs) to manage image dynamic range. Operates on existing GE Senographe DS/Essential hardware (amorphous silicon flat panel detector, dual track Mo/Rh x-ray tube).
Indications for Use
Indicated for screening and diagnosis of breast cancer in patients, including those with dense breasts or implants, using digital mammographic images.
Regulatory Classification
Identification
A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.
{0}------------------------------------------------
DEC 22 2011
KI10798
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | 03/18/2011 |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare<br>3000 N. Grandview Blvd<br>Mail Code W-709<br>Waukesha, WI 53188 |
| Primary Contact Person: | Steven Kachelmeyer, RAC<br>Regulatory Manager<br>GE Healthcare |
| Phone: | 262-548-2432 |
| Fax: | 262-997-1080 |
| Secondary Contact Person: | David Blonski<br>X-ray Regulatory Affairs Director<br>GE Healthcare |
| Phone: | 262-513-4072 |
| Fax: | 262-544-3844 |
| Device:<br>Trade Name:<br>Common/Usual Name:<br>Classification Names: | Premium View i<br>Full Field Digital Mammography System<br>21 CFR 892.1715, Class II |
| Product Code: | MUE |
| Predicate Device(s): | Senographe DS-P990066/S16,<br>Senographe Essential-P990066/S21 |
| Device Description: | The Senographe DS and Senographe Essential are both<br>full field digital mammography systems consisting of a<br>digital detector, a dual track x-ray tube (molybdenum /<br>rhodium) and an x-ray generator with control. The digital<br>detector is a flat panel of amorphous silicon on which<br>cesium iodide is deposited to maximize the detection of x- |
Image /page/0/Picture/9 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized and connected, creating a recognizable monogram.
·
.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. There are small decorative elements resembling droplets or flourishes around the outer edge of the circle. The logo is black and white.
rays. The x-ray filter is a wheel with both a molybdenum and a rhodium filter to allow various combinations of xray tube track and filter when imaging breasts of different radiological densities. The system includes a feature called Automatic Optimization of Parameters (AOP) that automatically selects the kVp, the optimal x-ray tube track and beam filtration and then terminates the exposure based on the breast density to provide consistent image quality for the user across a wide range of breast sizes and densities.
The subject of this submission will introduce a modification to a previously approved (P990066 / S015 and P990066 S020) image processing algorithm called Premium View. Premium View is an image-processing algorithm, which increases the visibility of breast structures. The main advantage is to provide a single breast image, where the contrast in the fatty tissues is similar to that obtained by setting WW (window width) and WL (window level) for optimum visualization of fatty tissues, and the contrast in the fibro-glandular tissue is similar to that obtained by setting WW and WL for optimal visualization of fibro-glandular tissues.
Premium View i, when utilized with the Senographe DS or Senographe Essential introduces a software change that applies different LUT values during the image processing prior to display of a very dense breast, or one with implants. This Premarket Notification will implement this technology on GE Healthcare's existing Full Field Digital Mammography systems as an upgrade to existing systems, or as an option to new installations.
- Intended Use: The Senographe DS and Senographe Essential FFDM systems generate digital mammographic images that can be used for screening and in the diagnosis of breast cancer. The Senographe DS and Senographe Essential FFDM systems are intended to be used in the same clinical applications as traditional film-based mammographic systems.
Premium View is an image-processing algorithm, which increases the visibility of breast structures. The main
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized and appear to be hand-drawn or calligraphic. The logo is black and white.
advantage is to provide a single breast image, where the contrast in the fatty tissues is similar to that obtained by setting WW (window width) and WL (window level) for optimum visualization of fatty tissues, and the contrast in the fibro-glandular tissue is similar to that obtained by setting WW and WL for optimal visualization of fibroglandular tissues.
PV is an option that can simplify the presentation of mammographic images, improve workflow, and streamline the review process of images with very dark or bright areas by presenting the image with the WW and WL optimized for review with minimal need for the user to make adjustments for the various tissue areas. This could be especially useful with patients who have very dense breasts, or implants that currently require multiple adjustments of WW and WL to review.
Technology: Premium View (PV) is image processing software which further improves the efficiency of the existing autocontrast feature; it also includes a modified version of the existing Tissue Equalization (TE) processing.
> Such optimization is achieved by applying different LUTs (Look Up Tables) to a high pass and low pass extractions of the image. In doing so, management of the image dynamic is achieved so that by applying PV it increases the visibility of breast structures. This is done by enhancing the contrast in glandular tissue, while preserving the visibility of the whole breast.
> Premium View i, when utilized with the Senographe DS or Senographe Essential introduces a software change that applies different LUT values during the image processing prior to display of a very dense breast, or one with implants.
Determination of Premium View i when installed on a Senographe DS or Substantial Equivalence: Senographe Essential FFDM system delivers functionality of comparable type that is substantially equivalent to our
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. The letters are stylized and connected, creating a unique and recognizable design. The logo is black and white.
currently marketed systems listed above and complies with the same or equivalent standards and have the same intended uses.
## Summary of Non-Clinical Tests:
The Premium View i option complies with voluntary standards as detailed in Section 9, 11 and 17 of this premarket submission. The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- . Requirements Reviews
- Design Reviews .
- Testing on unit level (Module verification) .
- . Integration testing (System verification)
- . Performance testing (Verification)
- . Safety testing (Verification)
- . Simulated use testing (Validation)
## Summary of Clinical Tests:
The subject of this premarket submission, Premium View i image processing option, included clinical testing to quantify the clinical acceptance of images that had been retrospectively processed with this image processing algorithm.
- Conclusion: GE Healthcare considers the Premium View i option to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Steven Kachelmeyer, RAC Regulatory Manager GE Healthcare 3000 N. Grandview Blvd Mail Code W-709 WAUKESHA WI 53188
DEC 2 2 2011
Re: K110798
Trade/Device Name: Premium View i (PVi) Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field digital mammography system Regulatory Class: II Product Code: MUE Dated: November 16, 2011 Received: November 17, 2011
Dear Mr. Kachelmeyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device . Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a circle. There are decorative swirls or flourishes around the circle, adding to the logo's distinctive appearance.
510(k) Number (if known):
Device Name: Premium View i (PVi)
Indications for Use:
The Senographe DS and Senographe Essential FFDM systems generate digital mammographic images that can be used for screening and in the diagnosis of breast cancer. The Senographe DS and Senographe Essential FFDM systems are intended to be used in the same clinical applications as traditional film-based mammographic systems.
Premium View is an image-processing algorithm, which increases the visibility of breast structures. The main advantage is to provide a single breast image, where the contrast in the fatty tissues is similar to that obtained by setting WW (window width) and WL (window level) for optimum visualization of fatty tissues, and the contrast in the fibroglandular tissue is similar to that obtained by setting WW and WL for optimal visualization of fibro-glandular tissues.
PV i is an option that can simplify the presentation of mammographic images, improve workflow, and streamline the review process of images with very dark or bright areas by presenting the image with the WW and WL optimized for review with minimal need for the user to make adjustments for the various tissue areas. This could be especially useful with patients who have very dense breasts, or implants that currently require multiple adjustments of WW and WL to review.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k)
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.