The Communify PACS Viewer is an image display software application that is intended for use by qualified physicians and other personnel for reading, diagnostic review, and analysis of digital images acquired from imaging devices such as CT, MR, CR, DX, MG, US, NM, PET, and other devices. The Communify PACS Viewer is a Picture Archiving and Communications System (PACS) viewer designed to be used to view Digital Imaging and Communications in Media (DICOM) and non-DICOM information and data. The Communify PACS Viewer is a software application that runs on standard "off-the-shelf" personal computers, business computers, and servers running standard operating systems. Communify PACS Viewer is an image and display software that accepts DICOM data from any OEM modality which supports DICOM standard imaging data; the system provides the capability to organize images generated by OEM vendor equipment, perform digital manipulations, create graphical representations of anatomical areas, and perform quantitative measurements. Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images.
Device Story
Communify PACS Viewer is a software application for viewing, manipulating, annotating, and analyzing medical images. Input: DICOM and non-DICOM data from OEM imaging modalities (CR, CT, DX, MR, NM, PT, RF, US, XA). Operation: Runs on standard off-the-shelf PCs/servers; provides tools for zoom, pan, triangulation, window/level, and measurement. Features include an Annotation Overlay Template Wizard for metadata display and a Hanging Protocol editor for custom screen layouts. Output: Visual display of images and quantitative measurements for diagnostic review by physicians. Benefit: Facilitates efficient case study review and clinical decision-making through organized image presentation and manipulation tools.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Software-only PACS viewer application. Operates on standard off-the-shelf PCs and servers. Supports DICOM standard imaging data. Features include image manipulation (zoom, pan, window/level), annotation overlays, and hanging protocol configuration. No specific hardware or materials of construction.
Indications for Use
Indicated for qualified physicians and personnel for reading, diagnostic review, and analysis of digital images (DICOM/non-DICOM) from modalities including CT, MR, CR, DX, MG, US, NM, and PET. Mammographic images require FDA-approved monitors.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
## 5. 510(k) Summary (21 CFR 807.92)
JUL 20 2011
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: _K)//0 717_
| Submitter's name: | Christopher Relyea |
|--------------------------------|-------------------------------------------------------------|
| Submitter's address: | 160 West 71st Street, 18th Floor, New York, NY 10023<br>USA |
| Phone number: | 646-526-7111 |
| Fax number: | 203-621-3198 |
| Name of contact person: | Christopher Relyea |
| Date the summary was prepared: | 02/23/2011 |
| Device Name/Trade Name: | Communify PACS Viewer |
| Common Name: | Medical Image Workstation |
| Classification Name: | 21 CFR 892.2050 |
| Product Code | LLZ |
{1}------------------------------------------------
Substantially equivalent to the devices are the following:
| Manufacturer | Device | 510(k) |
|----------------------------------|----------------------|---------|
| Viztek, Inc. | Opal-RAD | K063337 |
| RamSoft, Inc. | RamSoft PACS | KO31562 |
| General Electric Medical Systems | Centricity PACS Plus | KO23557 |
### Comparison to Predicate Devices
The similarity of Communify PACS Viewer with the predicate devices listed above is that they all are or contain PACS Viewer systems. They can all process and display medical images from DICOM compliant modalities such as CR, CT, DX, MR, NM, PT, RF, US, XA, and others.
Most of the tools, features, and settings in the PACS Viewers are commonly available across Communify PACS Viewer and the predicate devices listed above. Each device, including Communify PACS Viewer, has its own slightly unique interface, but someone familiar with any of the devices should be able to use any other device without difficulty.
#### Conclusion
The Communify PACS Viewer meets its intended use and design specifications for a picture archiving and communications system viewer. It has the required capabilities relating to the transfer, display, and digital processing of medical images. It is our conclusion that Communify PACS Viewer is significantly comparable to the PACS Viewers within Opal-RAD, RamSoft PACS, and Centricity PACS Plus.
#### Device Description
The Communify PACS Viewer is an application which allows users to view, manipulate,
2
{2}------------------------------------------------
annotate, transmit to other facilities, print, and animate all manner of DICOM images and modalities. These modalities include, but are not limited to, CR, CT, DX, MR, NM, PT, RF, US, and XA.
The Communify PACS Viewer contains common image manipulation functions (such as zoom, pan, triangulation, and window/level) and common image labeling tools (including measurements tools, drawing tools, and annotation overlays). The annotation overlay displays all the important metadata (as configured by the user) for each displayed series study. Although annotation fields depend on the modality and the patient study, the Annotation Overlay Template Wizard provides a full list of annotation fields that the user can assign into the image display.
For easy access to tools and features, there are shortcut keys, toolbars, and right-click menus.
The Hanging Protocol editor of the Communify PACS Viewer configures the presentation layout of images on the screen when a study is loaded. This allows commonly used display formats and presets to be saved and easily accessed to allow for faster case study reviews.
#### Intended Use Statement
The Communify PACS Viewer is an image display software application that is intended for use by qualified physicians and other personnel for reading, diagnostic review, and analysis of digital images acquired from imaging devices such as CT, MR, CR, DX, MG, US, NM, PET, and other devices. The Communify PACS Viewer is a Picture Archiving and Communications System (PACS) viewer designed to be used to view Digital Imaging and Communications in Media (DICOM) and non-DICOM information and data.
3
{3}------------------------------------------------
The Communify PACS Viewer is an image display software application that accepts DICOM data from any OEM modality which supports DICOM standard imaging data; the system provides the capability to organize images generated by OEM vendor equipment, perform digital manipulation, create graphical representations of anatomical areas, and perform quantitative measurements.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for the Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings and body. The text "DEPARTMENT OF HEALTH &" is vertically oriented and curves around the left side of the logo.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Chris Relyca CEO Flying Squirrel Storage LLC 160 West 71st Street, 18th Floor NEW YORK NY 10023
JUL 20 2011
Re: K110797
Trade/Device Name: Communify PACS Viewer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 2, 2011 Received: July 7, 2011
Dear Mr. Relyca:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary S Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 4. Indications for Use Statement
510(k) Number (if known): K II0 797 Device Name: Communify PACS Viewer
Indications for Use:
The Communify PACS Viewer is a Picture Archiving and Communications System (PACS) viewer designed to be used to view Digital Imaging and Communications in Media (DICOM) and non-DICOM information and data. The Communify PACS Viewer is a software application that runs on standard "offthe-shelf" personal computers, business computers, and servers running . standard operating systems. Communify PACS Viewer is an image and display software that accepts DICOM data from any OEM modality which supports DICOM standard imaging data; the system provides the capability to organize images generated by OEM vendor equipment, perform digital manipulations, create graphical representations of anatomical areas, and perform quantitative measurements.
Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images.
Prescription Use _X
Mary S. Patel
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
K110797
510K
、 Over-The-Counter Use
{7}------------------------------------------------
(Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.