COMPUTED DENTAL RADIOGRAPHY

K110768 · Schick Technologies, Inc. · MUH · Apr 20, 2011 · Dental

Device Facts

Record IDK110768
Device NameCOMPUTED DENTAL RADIOGRAPHY
ApplicantSchick Technologies, Inc.
Product CodeMUH · Dental
Decision DateApr 20, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.1800
Device ClassClass 2

Indications for Use

The Computed Oral Radiology System is indicated for patients undergoing an intra-oral dental x-ray examination. It produces instant, digital, intra-oral images of a patient's mouth while reducing the necessary x-ray dosage.

Device Story

System functions as digital imaging receptor for dental x-ray examinations; replaces traditional film. Input: X-ray radiation captured by digital sensor. Transformation: Sensor converts X-ray energy to digital image data; wireless transmission via IEEE 802.11 b/g protocol from Power and Transceiver (PAT) unit to host computer. Output: Digital intra-oral images displayed on computer workstation. Used in dental clinical settings; operated by dental professionals. Includes Counter Top Dock (CTD) for battery charging and initial wired USB configuration to wireless network. Benefits: Instant image availability; reduced X-ray dosage compared to film.

Clinical Evidence

Bench testing only. No clinical data provided. Risk mitigation and design validation studies performed to verify system performance and safety following the transition to wireless connectivity.

Technological Characteristics

Solid state X-ray imager; IEEE 802.11 b/g wireless protocol for data/control signals; rechargeable battery power source in Power and Transceiver (PAT) unit; Counter Top Dock (CTD) for charging and USB configuration. Digital imaging receptor form factor.

Indications for Use

Indicated for patients undergoing intra-oral dental x-ray examinations.

Regulatory Classification

Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K110768 APR 2 0 2011 # 510(k) Summary # Schick Computed Oral Radiology System ### Common/Classification Name: Solid State X-ray Imager 21CFR892.1650 #### Schick Technologies 30-30 47th Avenue Long Island City, NY 11101 718-937-5765 718-937-5962 (fax) Contact: Scott Doyle sdoyle@schicktech.com #### A. Legally Marketed Predicate Devices The computer oral radiology system was most recently cleared under K072134: November 2, 2007. The device and its predicates are small digital imaging receptors that may be used in place of dental x-ray film. The images are displayed on a computer workstation. # B. Modification Description Modifications to the device from its immediate predicate (K072134) are related to making it a wireless (WiFi) device vice a USB connected device. The modified device uses wireless IEEE 802.11 b/g protocol for image data transfer and control signal transfer to and from the Power and Transceiver (PAT) and the host computer. A rechargeable battery power source is also included in the PAT. A Counter Top Dock (CTD) has been added to this system. The CTD is a support device that provides charging power to the PAT and is used to provide a temporary wired connection, via USB, from the host computer to the PAT to allow initial configuration to the host wireless network. The modification does not alter the fundamental technology or the intended use. - C. Intended Use The operational environment remains unchanged from the predicate. There are no changes to the indications for use from the predicate devices. #### D. Substantial Equivalence Summary The modified system has had risks evaluated and mitigated as necessary. Testing and design validation have been used to verify risk mitigation. The principal risk is imaging system failure causing a retake which may include additional X-ray exposure. #### E. Conclusion Schick Technologies has concluded the modified system is substantially equivalent to its predicates. Risk analysis, testing and validation studies support this conclusion {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized depiction of an eagle or bird in flight, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The logo is black and white. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Mr. Scott Doyle Project Manager Schick Technologies, Inc. 30-30 47th Avenue LONG ISLAND CITY NY 11101 APR 2 0 2011 Re: K110768 Trade/Device Name: Computed Oral Radiology System Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: MUH Dated: March 17, 2011 Received: March 22, 2011 . Dear Mr. Doyle: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {2}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours. Mary Pastel Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known): K110768 Computed Oral Radiology System Device Name: Indications for Use: The Computed Oral Radiology System is indicated for patients undergoing an intra-oral dental x-ray examination. It produces instant, digital, intra-oral images of a patient's mouth while reducing the necessary x-ray dosage. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mary Patel (Division Sign-Off) (Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety 510K. K110768 Page 1_of 1 5
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...