K110712 · Neurovision Medical Products, Inc. · ETN · Jun 28, 2011 · Ear, Nose, Throat
Device Facts
Record ID
K110712
Device Name
DRYTOUCH SUCTION STIMULATOR PROBE
Applicant
Neurovision Medical Products, Inc.
Product Code
ETN · Ear, Nose, Throat
Decision Date
Jun 28, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1820
Device Class
Class 2
Indications for Use
The Suction Stimulator Probe is a dedicated manual surgical instrument that allows the surgeon to clear secretions and test surgical tissue with nerve stimulation at the same time and with the same instrument. It is intended for use only by a licensed physician and in conjunction with the Neurovision SE (Nerveana) Nerve locator Monitor System.
Device Story
Disposable, single-use, sterile surgical instrument; combines suction and nerve stimulation. Consists of stainless steel shaft, PTFE insulation, ball-tip suction, and proximal connector for nerve stimulator. Surgeon operates device manually during tissue dissection; finger-hole allows manual suction adjustment. Connects to Neurovision SE (Nerveana) Nerve locator Monitor System. Ball-tip reduces current density, enables stimulation at right angles (e.g., drill holes), and prevents bone chip clogging. Provides concurrent suction and stimulation; assists surgeon in identifying nerves during surgery; improves surgical efficiency by combining two functions into one instrument.
Clinical Evidence
Bench testing only. Testing included micrometer measurement of ball tip diameter, electrical continuity testing, and suction testing with applied air. Sterilization validated per ISO 11135-1:2007; EO residuals per ISO 10993-7:2008; shelf life confirmed via package integrity testing per ISO 11607.
Technological Characteristics
Stainless steel shaft, PTFE insulation. Ball-tip design. 2mm diameter; 13cm or 26cm lengths. Manual suction control via finger-hole. Connects to external nerve stimulator. Sterilized via Ethylene Oxide (EO).
Indications for Use
Indicated for use by licensed physicians during surgery to clear secretions and perform nerve stimulation in conjunction with the Neurovision SE (Nerveana) Nerve locator Monitor System.
Regulatory Classification
Identification
A surgical nerve stimulator/locator is a device that is intended to provide electrical stimulation to the body to locate and identify nerves and to test their excitability.
{0}------------------------------------------------
K 110712
P. 1 of 2
Image /page/0/Picture/1 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in large, bold, black letters. Underneath the word is a line with a heartbeat symbol in the middle, and the word "MEDICAL" is in smaller, bold, black letters to the right of the line.
Traditional 510k - Suction Stimulator Probe
JUN 2 8 2011
#### 5. 510K Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807 and in particular 21 CFR §807.92, the following summary of information is provided:
### Applicant Information
Christine Vergelv Regulatory Affairs Manager Neurovision Medical Products, Inc. 2225 Sperry Ave., Suite 1000 Ventura, CA 93003 PH 805-866-6999 Fax: 413-410-4579 christie@neurovisionmedical.com
| Date of Summary | June 28, 2011 |
|-----------------|---------------|
|-----------------|---------------|
| Device Identification | Trade Name: DryTouch Suction Stimulator Probe<br>Common Name: Suction Stimulator Probe |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------|
| Classification | Class II<br>Number 874.1820, Nerve Stimulator / Locator<br>Product code: ETN |
| Predicate Devices | Manufacturer: Elmed Inc.<br>Device Name: Frazier Suction Coagulator (part #23-5001)<br>510k Number: Pre-amendment device |
| | Manufacturer: Neurovision Medical Products Inc.<br>Device Name: Hemostat/Stimulator Probe (SHEM-s)<br>510K Number: K895676 |
| | Manufacturer: Axon Systems, Inc.<br>Device Name: XPAK II<br>510K Number: K090838 |
#### Device Description
The Suction Stimulator Probe is a disposable, single use, sterile instrument that combines concurrent suction and stimulation into one device for use during tissue dissection and nerve stimulation during surgery.
The instrument consists of a stainless steel shaft, partially insulated with biocompatible PTFE and ending with a ball-tip suction. Proximal connectors attach the instrument to a nerve stimulator. A tapered type adapter attaches the device to a standard suction catheter and a finger-hole allows for adjusting the suction strength manually by the surgeon. The insulated shaft has a 2mm diameter and 2 length options of 13 or 26 cm. The ball tip feature reduces the current density of stimulation, allows stimulation along right angles, such as drill holes, and prevents bone chip clogging of the instrument.
{1}------------------------------------------------
K110712
P. 2 of 2
Image /page/1/Picture/1 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in large, bold, black letters. Underneath the word "neurovision" are three horizontal lines, followed by a graphic of a heartbeat, and then the word "MEDICAL" in smaller, black letters.
Traditional 510k - Suction Stimulator Probe
## Intended Use
The Suction Stimulator Probe is a dedicated manual surgical instrument that allows the surgeon to clear secretions and test surgical tissue with nerve stimulation at the same time and with the same instrument. It is intended for use only by a licensed physician and in conjunction with the Neurovision SE (Nerveäna) Nerve locator Monitor System.
## Technological Characteristics of Device in Relation to Predicate Devices
The Suction Stimulator Probe combines technology from several predicate devices: the Frazier Suction Coagulator (part #23-5001), the Hemostat/Stimulator Probe (SHEM-s) K895676, and the XPAX II Nerve Stimulation Probe (K090838).
- The Suction Stimulator Probe is substantially equivalent to the Frazier Suction Coagulator with an extended shaft length: the body of the Suction Stimulator Probe is a Frazier Suction Coagulator with its shaft extended from 17.7cm to 26 cm or shortened to 13 cm.
- The Suction Stimulator Probe's stimulating capability is substantially equivalent to the stimulating capability of the Hemostat/Stimulator Probe (SHEM-s) K895676: the device is coated with PTFE insulating material and a dedicated wire is added for connection to an approved nerve stimulator.
- The Suction Stimulator probe is substantially equivalent to the Axon Systems XPAK II (K090838) in that both are stimulating, dissecting instruments with biocompatible electrical insulation, which attach to a nerve stimulator for nerve stimulation during surgery.
## Assessment of Performance Data Used to Justify Substantial Equivalence
Bench testing to assess performance and safety included micrometer measurement of the ball tip diameter, electrical continuity testing, and suction testing with applied air. These testing criteria were selected to test the quality of the manufacturing, the reliability of the surface area of the ball as indicating the current density of delivered stimulation, the continuity of the device as reflecting the reliability of current delivered to the tip, and finally, the gross delivery of air/suction as reflecting the suction power of the device.
### Sterilization:
- EO validation is compliant with ISO 11135-1.2007. ●
- EO residuals analysis is compliant with ISO 10993-7:2008. .
- The Sterility Assurance Level is 10-6. .
- . Device is not labeled non-pyrogenic.
### Shelf Life Testing:
- Shelf life is 1 year and was confirmed by package integrity testing compliant with ISO 11607. .
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
Neurovision Medical Products, Inc. c/o Ms. Christine Vergely Regulatory Manager 2225 Sperry Avenue, Suite 1000 Ventura, CA 93003
JUN 2 8 2011
Re: K110712
Trade/Device Name: DryTouch Suction Stimulator Probe Regulation Number: 21 CFR 874.1820 Regulation Name: Surgical Nerve Stimulator/Locator Regulatory Class: Class II Product Code: ETN Dated: May 22, 2011 Received: May 24, 2011
Dear Ms. Vergely:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Ms. Christine Vergely
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Melvin R. Fahlman, Ph.D.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Neurovision Medical. The word "neurovision" is in bold, black letters on the top line. Underneath the word is a line with a graphic and the word "MEDICAL" in smaller, black letters. The text "Traditional 510k - Suction Stimulator Probe" is at the bottom of the image.
# Indications for Use
510(k): K110712
Device Name: Suction Stimulator Probe
## Statement of Indications for Use for the Suction Stimulator Probe:
The Suction Stimulator Probe is a dedicated manual surgical instrument that allows the surgeon to clear secretions and test surgical tissue with nerve stimulation at the same time and with the same instrument. It is intended for use only by a licensed physician and in conjunction with the Neurovision SE (Nerveana) Nerve locator Monitor System.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Aper Douch
(Division SignsOff) urrence of CDRH, Office of Device Evaluation (ODE) Division of Ophthalmic, Neurological Nose and Throat Devices
**Prescription Use**
(Per 21 CFR 801.109)
510(k) Number k110712
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.