← Product Code [HRS](/productcode/HRS) · K110700

# ASCENSION ATLAS HUMERAL PLATING SYSTEM (K110700)

_Ascension Orthopedics, Inc. · HRS · Dec 6, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K110700

## Device Facts

- **Applicant:** Ascension Orthopedics, Inc.
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Dec 6, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Ascension ATLAS® Humeral Fracture Plate system is designed fo fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus. Indications include: Dislocated two-, three-, and four-fragment fractures of the proximal humerus, including fractures involving osteopenic bone Pseudoarthroses in the proximal humerus Osteotomies in the proximal humerus

## Device Story

System consists of stainless steel (SS 316L) humeral reconstruction plates (4-hole and 7-hole) and associated 3.5mm/2.7mm locking, non-locking, and lag screws. Used by orthopedic surgeons in clinical settings for internal fixation of proximal humerus fractures, dislocations, osteotomies, and non-unions. Device provides mechanical stabilization of bone segments to facilitate healing. Surgeon selects appropriate plate size and screw configuration based on fracture pattern; implants are secured to bone to maintain alignment during the healing process.

## Clinical Evidence

Bench testing only. Device met test specifications for ASTM F382-99, part A1.

## Technological Characteristics

Materials: Stainless steel (SS 316L). Standards: ASTM F382-99. Components: 4-hole and 7-hole plates, 3.5mm/2.7mm locking/non-locking/lag screws. Mechanical fixation principle.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Synthes LCP Proximal Humerus Plate ([K011815](/device/K011815.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) Summary

| SUBMITTER NAME:                                 | Ascension Orthopedics, Inc.<br>8700 Cameron Road<br>Austin, TX 78754-3832                                                                                                                                                                                                                                                                                                                                                        |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) CONTACT:                                 | Susan Walton                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                 | swalton@ascensionortho.com<br>Phone: (512) 836-5001 x1591<br>Toll Free FAX (888) 508-8081<br>Alternate FAX (512) 836-6933                                                                                                                                                                                                                                                                                                        |
| TRADE NAME:                                     | Ascension® ATLAS® Humeral Fracture Plate System                                                                                                                                                                                                                                                                                                                                                                                  |
| COMMON NAME:                                    | Plate, Fixation, Bone                                                                                                                                                                                                                                                                                                                                                                                                            |
| CLASSIFICATION:                                 | 21 CFR 888.3030 - Single/Multiple component metallic bone<br>fixation appliances and accessories                                                                                                                                                                                                                                                                                                                                 |
| PRODUCT CODE:                                   | HRS                                                                                                                                                                                                                                                                                                                                                                                                                              |
| PANEL:                                          | Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                                       |
| PREDICATE DEVICE:                               | K011815 - Synthes LCP Proximal Humerus Plate                                                                                                                                                                                                                                                                                                                                                                                     |
| DEVICE<br>DESCRIPTION:                          | The Ascension® ATLAS® Humeral Plate System will be composed<br>of left and right humeral reconstruction plate implants in four-hole<br>(HFP-0930-004LS, HFP-0930-004RS) and seven-hole (HFP-0930-<br>007LS, HFP-0930-007RS) lengths or sizes. The system will also<br>feature 3.5mm locking, non-locking and lag screws and 2.7mm<br>locking and non-locking screws.                                                             |
|                                                 | The ATLAS® Humeral Fracture Plate will be a single component<br>made from stainless steel (SS 316L). The non-locking, locking and<br>lag screws will be made from stainless steel (SS 316L).                                                                                                                                                                                                                                     |
| INTENDED USE:                                   | The Ascension ATLAS® Humeral Fracture Plate system is designed fo<br>fractures and fracture dislocations, osteotomies and non-unions of the<br>proximal humerus. Indications include:                                                                                                                                                                                                                                            |
|                                                 | <ul><li>Dislocated two-, three-, and four-fragment fractures of the proximal humerus, including fractures involving osteopenic bone</li><li>Pseudoarthroses in the proximal humerus</li><li>Osteotomies in the proximal humerus</li></ul>                                                                                                                                                                                        |
| SUMMARY OF<br>TECHNOLOGICAL<br>CHARACTERISTICS: | The technological characteristics for the Ascension ATLAS<br>Humeral Fracture Plate System were compared to the predicate<br>device, the Synthes LCP Proximal Humerus Plate (K011815). The<br>technological characteristics were defined to be plate length, number<br>of holes and location of the holes as well as screw sized. These<br>characteristics were determined to be the same or similar to the<br>predicate device. |

|( || 0700 ( || 2)

--

{1}------------------------------------------------

# K 110700(zlz)

#### NONCLINICAL TESTING

### BASIS OF SUBSTANTIAL EQUIVALENCE:

The ATLAS Humeral Fracture Plate meets the test specifications for ASTM F382-99, part A1.

Ascension Orthopedics believes that this system is substantially equivalent to the legally marketed predicate device based on similarities in design, materials and indications.

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

l'ood and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Ascension Orthopedics, Inc. % Ms. Susan Walton 8700 Cameron Road Austin, Texas 78754

DEC - 6 2011

Rc: K110700

Trade/Device Name: Ascension® ATLAS® Humeral Fracture Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: October 20, 2011 Received: October 21, 2011

Dear Ms. Walton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Susan Walton

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely your

Por D.m
Dep con Dip

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# KII0700 (11)

### Indications for Use Statement

510(K) Number:

Ascension® ATLAS® Humeral Fracture Plate System Device Name:

Indications for Use:

The Ascension ATLAS® Humeral Fracture Plate system is designed for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus. Indications Include:

- Dislocated two-, three-, and four-fragment fractures of the proximal humerus, including A fractures involving osteopenic bone
- A Pseudoarthroses in the proximal humerus
- A Osteotomies in the proximal humerus

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart B) OR

Over-The-Counter Use (Part 21 CFR 801Subpart C)

### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Michael Ores for MXM

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number

---

**Source:** [https://fda-staging.innolitics.com/device/K110700](https://fda-staging.innolitics.com/device/K110700)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).