K110643 · Merit Medical Systems, Inc. · DQO · Jul 1, 2011 · Cardiovascular
Device Facts
Record ID
K110643
Device Name
MERIT MARQUIS FLOW SWITCH
Applicant
Merit Medical Systems, Inc.
Product Code
DQO · Cardiovascular
Decision Date
Jul 1, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The Flow Switch is an angiographic accessory intended for use as an on-off device for angiography and other high pressure applications.
Device Story
The Merit Marquis Flow Switch is an angiographic accessory used to control liquid flow in high-pressure applications. It connects in-line via a female Luer lock inlet and male Luer lock outlet. The device features a non-slip sliding handle that allows for one-handed operation; it provides an audible click when moved to the OFF position. Flow direction is indicated by arrows on the handle. It is used by clinicians in clinical settings to manage fluid delivery during angiography. By providing a reliable, easy-to-operate shut-off mechanism, the device assists in controlling contrast or saline delivery, potentially improving procedural efficiency and safety for the patient.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included surface inspection, Luer fitting verification, pressure testing (ON/OFF positions), slider motion/force testing, flow testing, and comprehensive biocompatibility (L929 MEM, Kligman Maximization, Intracutaneous/Systemic Injection, Rabbit Pyrogen, Hemocompatibility, Heavy Metals). Packaging integrity was validated via bubble emission, dye penetration, seal strength, and burst strength tests following accelerated aging and simulated shipping.
Technological Characteristics
Materials: Plastics compliant with USP <661>. Sensing/Actuation: Mechanical sliding handle with audible click. Energy Source: Manual. Dimensions: Standard Luer lock (ISO 594-1/594-2). Connectivity: None. Sterilization: Ethylene oxide (ISO 11135). Standards: ISO 11070, ISO 594-1, ISO 594-2, ASTM F2096, ASTM F1929, ASTM F88, ASTM D4169, ISO 10993-1.
Indications for Use
Indicated for use as an on-off device for angiography and other high pressure applications in patients undergoing angiographic procedures.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles an abstract bird or human figure with outstretched arms.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 3, 2016
Merit Medical Ireland Ltd. C/O Ms. Stephanie Erskine 1600 West Merit Parkway South Jordan, UT 84095
Re: K110643
Trade/Device Name: Merit Marquis Flow Switch Regulation Number: 21 CFR 870.4290 Regulation Name: Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass Regulatory Class: Class II Product Code: DTL Dated: April 20, 2011 Received: April 22, 2011
Dear Ms. Erskine:
This letter corrects our substantially equivalent letter of July 1, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 – Stephanie Erskine
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
## Eric E. Richardson -S
for Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Section 4
## Indications for Use Statement
K110643
510(k) Number (if known):
Device Name: Merit Marquis® Flow Switch
Indications for Use:
The Flow Switch is an anglographic accessory intended for use as an on-off device for angiography and other high pressure applications.
Prescription Use ਮ (Part 21 CFR 801 Subpart D) :
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Fli. Gual ho BDZ
ion Sidh-Off n of Cardiovascular Devices
510(k) Number_K110643
{3}------------------------------------------------
## Section 5
| 510(k) Summary | | |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| General<br>Provisions | Submitter Name: | Merit Medical Systems, Inc. |
| | Address: | 1600 West Merit Parkway<br>South Jordan, UT 84095 |
| | Telephone Number: | (801) 208-4349 |
| | Fax Number: | (801) 253-6967 |
| | Contact Person: | Stephanie Erskine |
| | Registration Number: | 1721504 |
| | Correspondent Name: | Merit Medical Ireland Ltd. |
| | Address: | Parkmore Business Park,<br>Galway, Ireland |
| | Telephone Number: | (353) 91 703 761 |
| | Fax Number: | (353) 91 771 888 |
| | Contact Person: | Mark Mullaney |
| | Date of Preparation: | 21-Apr-2011 |
| | Registration Number: | 9616662 |
| Subject<br>Device | Trade Name: | Merit Marquis® Flow Switch |
| | Common/Usual Name: | High pressure angiographic flow control<br>switch / stopcock |
| | Classification Name: | Adaptor, Stopcock, Manifold, Fitting,<br>Cardiopulmonary Bypass |
| Predicate<br>Device | Trade Name: | FLOSWITCH(R) HP |
| | Classification Name: | Catheter, intravascular, diagnostic |
| | Premarket Notification: | K913871 |
| | Manufacturer: | BOSTON SCIENTIFIC CORP. |
| Classification | Class II | |
| | FDA Product Code: | DTL |
| | 21 CFR § 870.4290 | |
| | Division of Cardiovascular Devices | |
| Intended Use | The Flow Switch is an angiographic accessory intended for use as<br>an on-off device for angiography and other high pressure<br>applications. | |
| Device<br>Description | The Flow Switch is a device used to connect in line to liquid flow.<br>This uses a Female Luer lock (inlet) and Male Luer lock (outlet)<br>connection. Shut off is by means of a non-slip sliding handle that<br>can be operated one-handed and gives an audible click into the OFF<br>position. Flow directions are indicated by arrows on the handle. | |
| Technological<br>Characteristics | Technological characteristics of the subject Merit Marquis® Flow<br>Switch are substantially equivalent to those of the predicate, the<br>currently marketed BOSTON SCIENTIFIC CORP. FLOSWITCH(R)<br>,HP [K913871]. | |
| Safety &<br>Performance<br>Tests | No performance standards have been established under Section 514<br>of the Food, Drug and Cosmetic Act for these devices. However, a<br>battery of tests performed on the Merit Marquis® Flow Switch were<br>designed to demonstrate that the device meets critical design<br>specifications as well as clinical performance attributes for its<br>intended use. Where appropriate, the tests were based on the<br>requirements of the following documents:<br>ISO 11070: 1998, Sterile Single-Use Intravascular Catheter<br>Introducers. ISO 594-1:1996, Conical fittings with a 6 % (Luer) taper for<br>syringes, needles and certain other medical equipment –<br>Part 1: General requirements ISO 594-2:1998, Conical fittings with 6% (Luer) taper for<br>syringes, needles and certain other medical equipment –<br>Part 2: Lock fittings ASTM F 2096-04; Standard Test Method for Detecting Gross<br>Leaks in Medical Packaging by Internal Pressurization<br>(Bubble Test) ASTM F 1929-98 (Reapproved 2004); Standard Test Method<br>for Detecting Seal Leaks in porous Medical Packaging by<br>Dye Penetration ASTM F 88-07; Standard Test Method for Seal Strength of<br>Flexible Barrier Materials. ASTM D4169 – 09 Standard Practice for Performance<br>Testing of Shipping Containers and Systems1 ISO 11135:2007 – Sterilization of health care products –<br>Ethylene oxide – Part 1: Requirements for development,<br>validation and routine control of a sterilization process for<br>medical devices ISO 10993-1: 2009, Biological Evaluation of medical Devices<br>Part 1: Evaluation and Testing within a risk management<br>process, and the FDA Modified ISO 10993 Test Profile | |
{4}------------------------------------------------
・
{5}------------------------------------------------
The Merit Marquis® Flow Switch was compared to the predicate device for various performance attributes that demonstrate the safety or efficacy of the device. The following is a list of all significant testing that was successfully completed:
- . Surface inspection
- Male and Female Luer conical fittings with a 6% (Luer) .
- Pressure in OFF position .
- Pressure in open/ON position .
- Slider Handle motion test ●
- Slider force test ●
- . Flow Test
Biocompatibility testing included:
- L929 MEM Elution: MEM extraction .
- Kligman Maximization: Saline and cottonseed oil extractions .
- Intracutaneous Injection: Saline and cottonseed oil . extractions
- . Systemic Injection: Saline and cottonseed oil extractions
- . Material Medicated Rabbit Pyrogen Test: Saline extraction
- Hemocompatibility - In-Vitro Hemocompatibility - Indirect
- Hemocompatibilitv Hemolysis Indirect ●
- Hemocompatibility Prothrombin Time Assay (PT) Indirect .
- Hemocompatibility Unactivated Partial Thromboplastin . Time Assay (UPTT) Indirect
- . Heavy Metals Content: USP Physicochemical Tests for Plastics USP <661>
Packaging performance before and after exposure to accelerated aging and simulated shipping and handling conditions:
- Bubble emission .
- Dye penetration ●
- Seal peel tensile strength ●
- Burst strength ●
- Visual inspection .
Summary of Substantial Equivalence Based on the indications for use, design, safety, and performance testing, the subject Merit Marquis® Flow Switch meets the requirements that are considered essential for its intended use and is substantially equivalent to the predicate device. the currently marketed FLOSWITCH(R) HP manufactured by currently marketed BOSTON SCIENTIFIC CORP.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.