FUJIREBIO DIAGNOSTICS VITAMIN D CONTROL

K110641 · Fujirebio Diagnostics,Inc. · JJX · May 27, 2011 · Clinical Chemistry

Device Facts

Record IDK110641
Device NameFUJIREBIO DIAGNOSTICS VITAMIN D CONTROL
ApplicantFujirebio Diagnostics,Inc.
Product CodeJJX · Clinical Chemistry
Decision DateMay 27, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Indications for Use

Fujirebio Diagnostics Vitamin D Control is intended for use as an assayed control serum to monitor the precision of laboratory testing procedures for the analysis of Vitamin D.

Device Story

Fujirebio Diagnostics Vitamin D Control is a quality control material used in clinical laboratories to monitor the precision of Vitamin D assays. The device consists of human serum, bovine protein, biochemical materials, and chemical preservatives (Proclin 300, Gentamicin), provided in lyophilized form. It contains specific target concentrations of 25(OH) Vitamin D2 and 25(OH) Vitamin D3. Laboratory personnel reconstitute the control and use it alongside patient samples during testing procedures. By comparing the measured values of the control against established target ranges, clinicians verify the accuracy and precision of their diagnostic assays. This monitoring process ensures the reliability of patient Vitamin D test results, which are essential for clinical decision-making regarding bone health and metabolic status.

Clinical Evidence

No clinical data. Bench testing only. Stability studies (shelf-life and reconstituted) were performed and found acceptable. Value assignment was conducted using multiple assays (DiaSorin LIAISON, DiaSorin RIA, IDS EIA, and LC-MS/MS) with a minimum of 20 measurements per assay.

Technological Characteristics

Lyophilized human serum matrix containing bovine protein, biochemicals, and preservatives (Proclin 300, Gentamicin). Contains 25(OH) Vitamin D2 and 25(OH) Vitamin D3. Stability testing performed per EN 13640. Storage: 12 months at 2-8°C unopened; 14 days at 2-8°C or 63 days at -10°C reconstituted.

Indications for Use

Indicated for use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the measurement of 25-OH Vitamin D in clinical laboratory settings.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k110641 B. Purpose for Submission: New device C. Measurand: Quality control materials for Vitamin D assays D. Type of Test: Quality Control Materials E. Applicant: Fujirebio Diagnostics, Inc. F. Proprietary and Established Names: Fujirebio Diagnostics Vitamin D Control G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JJX | Class I, reserved | 21 CFR 862.1660 Quality Control Material | Clinical Chemistry (75) | H. Intended Use: 1. Intended use(s): Refer to indication for use below. 2. Indication(s) for use: Fujirebio Diagnostics Vitamin D Control is intended for use as an assayed control serum to monitor the precision of laboratory testing procedures for the analysis of Vitamin D. 3. Special conditions for use statement(s): {1} For in vitro diagnostic use only. Fujirebio Diagnostics Vitamin D Control is made from human source material, and therefore should be treated as potentially infectious. Each human donor unit used to manufacture this control was tested by FDA accepted methods and found non-reactive for Hepatitis B Surface Antigen (HBsAg), antibody to Hepatitis C (HCV), antibody to HIB-1/HIV-2, and antibody to Treponema Pallidum (Syphilis). This product may also contain other human pathogens for which there are no approved tests. All human source material should be considered potentially infectious and handled with the same precautions used with patient specimens. 4. Special instrument requirements: None I. Device Description: The Fujirebio Diagnostics Vitamin D Control is prepared from human serum, protein (bovine), purified biochemical materials, and chemicals. It also contains Proclin 300® and Gentamicin as preservatives. The controls are provided in lyophilized form. The control levels contain Vitamin D at the corresponding target concentrations. | Analyte | Analyte Form | Target Concentrations (ng/mL) | | | | --- | --- | --- | --- | --- | | | | Level 1 | Level 2 | Level 3 | | 25 (OH) Vitamin D | 25 (OH) D2 | 0 | 10 | 10 | | | 25 (OH) D3 | 10 | 20 | 60 | J. Substantial Equivalence Information: 1. Predicate device name(s): Bio-Rad Liquichek™ Specialty Immunoassay Control – 25-OH Vitamin D 2. Predicate 510(k) number(s): k043108 3. Comparison with predicate: | Items | Fujirebio Diagnostics Vitamin D Control (Candidate Device) | Bio-Rad Liquichek™ Specialty Immunoassay Control – 25-OH Vitamin D (Predicate Device) | | --- | --- | --- | {2} | Similarity | | | | --- | --- | --- | | Intended Use | Same | For use as an assayed quality control serum to monitor the precision of laboratory testing procedures for the measurement of 25-OH Vitamin D | | Analyte(s) | Same | 25-OH Vitamin D | | Levels of Controls | Same | 3 | | Matrix | Same | Human Serum with additives | | Differences | | | | Vitamin D Analyte Forms | 25(OH) Vitamin D2 25(OH) Vitamin D3 | 25(OH) Vitamin D3 | | Other Analytes | None | Contains also: Anti-Tg Anti-TPO C-peptide Erythropoietin (EPO) Intact PTH (iPTH) IGF-I Osteocalcin | | Storage (unopened) | 12 months at 2 to 8°C | 2 years at -20°C to -70°C | | Form | Lyophilized | Liquid | # K. Standard/Guidance Document Referenced (if applicable): EN standard, 13640 – Stability testing of In Vitro diagnostic reagents. # L. Test Principle: Not applicable # M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Not applicable b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability The Vitamin D2 and D3 antigen material were obtained from commercial vendors. The {3} Vitamin D2 and D3 antigen material are supplied as a white solid. The Vitamin D2 and Vitamin D3 antigens are solubilized in 95% Ethyl Alcohol and subsequently spiked into the 25-hydroxy Vitamin D2/D3 Control Matrix to manufacture the Vitamin D Level 1, 2, and 3 controls. ## Value Assignment: - The control ranges as determined using DiaSorin LIAISON, DiaSorin RIA, IDS EIA and LC-MS/MS assays are provided in the assigned value sheet for each lot release. Each control range was determined from a minimum of 20 measurements per assay using 2 reagent/calibrator lots, 2 instruments, in 2 runs with 5 replicates each run. The ranges were determined as mean +/- 2 SD for DiaSorin LIAISON, DiaSorin RIA assays or mean +/-30% for IDS EIA and LC-MS/MS assays. - In the labeling the sponsor recommends that each end user laboratory establish its own means and acceptable ranges and use the assigned value sheet as guidance only. ## Stability: - Shelf life stability: Real-time testing at 2-8°C were conducted and is still on-going. The stability study protocol and acceptance criteria have been reviewed and found to be acceptable. The current test results support a shelf life of 12 months at 2-8°C. - Reconstituted Stability: The stability study protocol and acceptance criteria to determine the reconstituted stability and freeze/thaw cycle of the controls have been reviewed and found to be acceptable. The control solutions are stable for 14 days when stored tightly capped at 2-8°C, and stable for 63 days when stored at -10°C. Controls may be frozen and thawed repeatedly for up to 9 cycles d. Detection limit: Not applicable e. Analytical specificity: Not applicable f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: {4} Not applicable b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: The control ranges as determined using DiaSorin LIAISON, DiaSorin RIA, IDS EIA and LC-MS/MS assays are provided in the assigned value sheet for each lot release. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...