← Product Code [MCB](/productcode/MCB) · K110620

# PREMIER C. DIFFICILE GDH (K110620)

_Meridian Bioscience, Inc. · MCB · May 3, 2011 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K110620

## Device Facts

- **Applicant:** Meridian Bioscience, Inc.
- **Product Code:** [MCB](/productcode/MCB.md)
- **Decision Date:** May 3, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.2660
- **Device Class:** Class 1
- **Review Panel:** Microbiology
- **Attributes:** Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K110620 · May 3, 2011 | PREMIER C. DIFFICILE GDH | Meridian Bioscience, Inc. | Clinical patient samples from independent clinical test sites | The clinical performance of the Premier C. difficile GDH assay was evaluated by comparing results against bacterial culture in 733 patient samples collected from symptomatic patients at multiple clinical sites. | Clinical performance; Sensitivity; Specificity; Symptomatic patients |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical performance evaluation compared to bacterial culture; Study Period: November 2011 | Symptomatic persons suspected of having C. difficile infection (CDI); age range 22 days to 99 years; Sample Size: 733; Number of Sites: 5 | Bacterial C. difficile culture | Sensitivity (92.3%) and Specificity (95.8%) |

## Indications for Use

Premier™ C. difficile GDH is a qualitative enzyme immunoassay screening test to detect Clostridium difficile antigen, glutamate dehydrogenase, in fecal specimens from symptomatic persons suspected of having C. difficile infection (CDI). This test does not distinguish between toxigenic and non-toxigenic strains of C. difficile. Samples from symptomatic patients that produce positive results with this test must be further tested with an assay designed to detect toxigenic C. difficile strains and assist with the diagnosis of CDI.

## Device Story

Premier™ C. difficile GDH is a qualitative enzyme immunoassay (EIA) for detecting glutamate dehydrogenase (GDH) antigen in fecal specimens. The device uses breakaway microwells coated with polyclonal rabbit anti-GDH antibodies. The process involves adding diluted patient stool and HRP-conjugated anti-GDH antibody to the wells; incubating; washing to remove unbound material; and adding a chromogenic substrate. The presence of GDH forms an antibody-enzyme complex, resulting in a blue color that turns yellow upon addition of stop solution. Results are interpreted spectrophotometrically. The test is performed in clinical laboratories by professional staff. It serves as a screening tool; positive results require follow-up testing for toxigenic strains to assist in CDI diagnosis. The device aids clinicians in identifying patients requiring further diagnostic workup for C. difficile infection.

## Clinical Evidence

Clinical trial evaluated 733 patient samples (ages 22 days to 99 years) across multiple US sites. Compared to bacterial culture, the assay demonstrated an overall sensitivity of 92.3% (95% CI: 86.0-95.9%) and specificity of 95.8% (95% CI: 93.9-97.1%). No performance differences were observed based on age, gender, or geographic location.

## Technological Characteristics

Qualitative enzyme immunoassay; breakaway microwells coated with polyclonal rabbit anti-GDH antibodies; HRP-conjugated detection antibody; chromogenic substrate (urea peroxide/tetramethylbenzidine); stop solution (1M phosphoric acid). Spectrophotometric readout (OD450 or OD450/630). Preservatives: ProClin® and gentamicin. Manual or automated microplate washer/incubator compatible.

## Regulatory Identification

A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.

## Predicate Devices

- TechLab C diff. Chek-60 ELISA ([K030992](/device/K030992.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
K110620

B. Purpose for Submission:
To seek clearance for the Premier™ C. difficile GDH Assay

C. Measurand:
Clostridium difficile antigen, glutamate dehydrogenase (GDH)

D. Type of Test:
Qualitative enzyme immunoassay

E. Applicant:
Meridian Bioscience, Inc.

F. Proprietary and Established Names:
Premier™ C. difficile GDH Assay

G. Regulatory Information:
1. Regulation section: 21 CFR 866.2660- Microorganism Differentiation and Identification Device
2. Classification: Class I
3. Product code: MCB- Antigen, C. difficile
4. Panel: 83, Microbiology

1

{1}

H. Intended Use:

1. Intended use(s):

Premier™ C. difficile GDH is a qualitative enzyme immunoassay screening test to detect Clostridium difficile antigen, glutamate dehydrogenase, in fecal specimens from symptomatic persons suspected of having C. difficile infection (CDI). This test does not distinguish between toxigenic and non-toxigenic strains of C. difficile. Samples from symptomatic patients that produce positive results with this test must be further tested with an assay designed to detect toxigenic C. difficile strains and assist with the diagnosis of CDI.

2. Indication(s) for use:

Premier™ C. difficile GDH is a qualitative enzyme immunoassay screening test to detect Clostridium difficile antigen, glutamate dehydrogenase, in fecal specimens from symptomatic persons suspected of having C. difficile infection (CDI). This test does not distinguish between toxigenic and non-toxigenic strains of C. difficile. Samples from symptomatic patients that produce positive results with this test must be further tested with an assay designed to detect toxigenic C. difficile strains and assist with the diagnosis of CDI.

3. Special conditions for use statement(s):

For prescription use only

4. Special instrument requirements:

Not Applicable

I. Device Description:

Premier™ C. difficile GDH is a qualitative enzyme immunoassay screening test to detect Clostridium difficile antigen, glutamate dehydrogenase, in fecal specimens from symptomatic persons suspected of having C. difficile infection (CDI). Breakaway microwells are coated with polyclonal rabbit anti-GDH antibodies. Diluted patient specimen and a horseradish peroxidase conjugated (HRP) anti-GDH antibody are added to the microwells and incubated to allow any antigen in the patient sample to bind the antibody conjugate and immobilized microwell antibody. Upon completion of the incubation, a wash step is performed to remove unbound material. If C. difficile GDH is present, an antibody-enzyme complex is formed. A chromogenic substrate is added to the microwells and incubated. A blue color develops in the presence of bound enzyme. Stop solution is added, changing the initial blue reaction to yellow. Test results are interpreted spectrophotometrically.

2

{2}

Reagents and Test Components provided:

- Premier™ C. difficile GDH Microwells: Polyclonal antibody-coated Microwells, specific to C. difficile GDH
- Premier™ C. difficile GDH Enzyme Conjugate: HRP-conjugated polyclonal antibodies specific to C. difficile GDH in a buffered protein solution containing 0.1% ProClin® and 0.03% gentamicin as preservatives
- Premier™ 20X Wash Buffer II: Concentrated wash buffer containing 0.2% thimerosal as a preservative. Reagent is diluted before use
- Premier™ Substrate I: Buffered solution containing urea peroxide and tetramethylbenzidine at pH 5.0
- Premier™ Stop Solution I: 1M Phosphoric acid
- Premier™ C. difficile GDH Sample Diluent/Negative Control: Buffered protein solution containing 0.1% ProClin® and 0.03% gentamicin as preservatives
- Premier™ C. difficile GDH Positive Control: C. difficile GDH in a buffered protein solution containing 0.1% ProClin® and 0.03% gentamicin as preservatives
- Transfer pipettes
- Microwell strip holder
- Microwell plate sealer
- General laboratory microplate washers and the StatFax™ incubator/shaker can be used with this assay. Use of this equipment is optional.

J. Substantial Equivalence Information:

1. Predicate device name(s):

TechLab C diff. Chek-60 ELISA

2. Predicate 510(k) number(s):

K030992

{3}

3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | K110620 | K030992  |
|  Intended Use | Qualitative enzyme immunoassay screening test to detect Clostridium difficile glutamate dehydrogenase in fecal specimens from symptomatic persons suspected of having C. difficile infection (CDI). For professional use in a clinical laboratory setting. For use as an aid in diagnostic testing of samples from patients suspected of having C. difficile infection. | Same  |
|  Capture Reagents | Polyclonal antibodies specific to C. difficile GDH | Same  |
|  Specimen Preparation | 1. Add 200 uL of Sample Diluent/Negative Control to a test tube 2. Add 50 uL of thoroughly mixed stool to the Sample Diluent/Negative Control test tube. 3. Vortex tube. Add 100uL of diluted stool to microwell. | Same  |
|  Test Procedure | Add 1 drop (50 uL) Conjugate to each well Add 2 drops (100 uL) Substrate to each well. 2 drops of Positive Control and 100 uL of Sample Diluent/Negative Control are added to the microwells | Same  |
|  Interpretation of Results | Assay defines OD readings for positive and negative interpretations | Same  |

{4}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | K110620 | K030992  |
|  Detection Reagents | Rabbit polyclonal antibodies specific to C. difficile GDH conjugated to horseradish peroxidase (HRP) | Antigen-specific mouse monoclonal antibody conjugated to HRP.  |
|  Test Procedure | Microwells are washed 5-7 times with diluted Wash Buffer
2 drops (100 uL) of Stop Solution are added to the Microwells
Reaction must be read within 30 minutes of addition of Stop Solution | Microwells are washed 4 times with diluted Wash Buffer
1 drop (50 uL) of Stop Solution is added to the Microwells
Reaction must be read within 10 minutes of adding Stop Solution  |
|  OD Reader | Automatic | Automatic with Visual Read option  |

## K. Standard/Guidance Document Referenced (if applicable):

1. Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover
Human Specimens that are Not Individually Identifiable - Guidance for Sponsors, Institutional Review Boards., Clinical Investigators and FDA Staff
http://www.fda.gov/cdrh/oivd/guidance/1588.html

2. Guidance for Industry and FDA Staff - Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests.
http://www.fda.gov/cdrh/osb/guidance/1620.html

## L. Test Principle:

Premier C. difficile GDH is a qualitative enzyme immunoassay screening test to detect Clostridium difficile antigen, glutamate dehydrogenase, in fecal specimens from symptomatic persons suspected of having C. difficile infection (CDI). Breakaway microwells are coated with polyclonal rabbit anti-GDH antibodies. Diluted patient specimen and a horseradish peroxidase conjugated anti-GDH antibody are added to the microwells and incubated to allow any antigen in the patient sample to bind the antibody conjugate and the immobilized microwell antibody. Upon completion of the incubation, a wash step is performed to remove unbound material. If C. difficile GDH is present, an antibody-enzyme complex is formed. A chromogenic substrate is added to the microwells and incubated. A blue color develops in the

{5}

presence of bound enzyme. Stop solution is added, changing the initial blue reaction to yellow. Test results are interpreted spectrophotometrically.

Premier C. difficile GDH is based on the well-established, widely used microplate enzyme immunoassay platform. The predicate device, TECHLAB C. DIFF CHEK-60, is also built on this platform.

Premier C. difficile GDH Positive Control: Non-toxigenic C. difficile strain 11186 is grown in liquid media, harvested, and filtered to remove organisms. Bulk manufacture of Positive Control includes dilution of C. difficile GDH antigen in a buffered protein solution containing Gentamicin and ProClin™ as preservatives.

Rabbit polyclonal anti-C. difficile Common Antigen antibody is used in the manufacture of Premier C. difficile GDH Enzyme Conjugate and Premier C. difficile GDH Microwells. The antibody is specific for C. difficile Common Antigen (GDH) and is internally manufactured at Meridian Bioscience. The antibody is purified using Protein-A, and is qualified for use by Meridian using internally qualified enzyme immunoassay (EIA) methods.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Reproducibility:

Reproducibility studies were performed by three clinical laboratories using blind coded panels. Samples were randomly sorted within each panel to mask identities. Each panel consisted of three contrived moderately positive specimens, three contrived low-positive specimens, three contrived high-negative specimens, and one natural negative specimen. The moderately-positive, low-positive, and high-negative samples were prepared from negative stool spiked with C. difficile GDH antigen. Panels were tested at three independent laboratories by two operators at each laboratory, twice each day over 5 non-consecutive days. Results were read at single (OD₄₅₀) and dual (OD₄₅₀/₆₃₀) wavelengths. The expected results were obtained with all samples by all technologists at all test sites at each time interval. Reproducibility of the assay is 100% for moderate positives, low-positives, high-negatives, and weak negatives.

#### b. Linearity/assay reportable range:

Not applicable

#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):

No calibrators are supplied or used with this assay.

6

{6}

7

External Controls:

Two External Control reagents are included with the Premier C. difficile GDH assay. External Controls aid the user in detection of reagent deterioration, adverse environmental or test conditions, washing failures, or variance in operator performance that can cause test errors. External Controls are required for routine Quality Control and should be included with each test run.

Positive Control:

C. difficile GDH antigen is in a buffered protein solution containing 0.1% ProClin®

and

0.03% gentamicin as preservatives. Non-toxigenic C. difficile 11186 is grown in liquid media, harvested, and filtered to remove organisms. Bulk manufacture of Positive Control includes dilution of C. difficile antigen in a buffered protein solution containing Gentamicin and ProClin® as preservatives. Negative Control: Buffered protein solution containing 0.1% ProClin® and 0.03% Gentamicin as preservatives.

Storage Conditions:

A Sample Storage study was designed to verify the adequacy of the sample age, preservation, handling, and storage criteria included in the Premier C. difficile GDH package insert. Aliquots of unpreserved contrived positive and negative samples were stored at the following temperatures to determine the worst case storage time for each temperature range: room (19-30 C), refrigerated (2-8 C), conventional freezer (-16 to -28 C), and ultralow freezer (-66 to -84 C). A total of 10 samples were tested, with the Limit of Detect (LoD), Low Positive (LP), and High Negative (HN) samples consisting of manufactured specimens. Results suggest that the Control reagents should optimally be stored at 2-8 C when not in use.

Stability:

Dating is provisionally set with data derived from accelerated stability studies. Results are confirmed with real-time shelf-life studies. Accelerated Stability performed with three lots of Positive and Negative Controls has verified the stability for at least 18 months under normal conditions of use and storage. Real-time studies are ongoing.

Expected Values:

Expected Values for OD₄₅₀:

Positive Control: ≥ 0.600

Negative Control: &lt; 0.150

Expected Values for OD₄₅₀/₆₃₀:

Positive Control: ≥ 0.600

{7}

Negative Control: &lt; 0.100

**Validation:**
The performance of the Positive and Negative Controls was validated as part of clinical trials.

**d. Detection limit:**

In order to determine the analytical limit of detection (LoD) of *C. difficile* GDH antigen diluted in a human stool matrix, purified GDH antigen was spiked into negative stool matrix and diluted serially two-fold. A minimum of 45 replicates of each concentration were tested to determine the concentration that represents the Limit of Blank, or the dilution that produces approximately 50% positive and 50% negative results. The next higher two-fold concentration from the Limit of Blank that produced a minimum of 95% positive results was determined to be the Limit of Detection. The next lower two-fold concentration from the Limit of Blank that produced a minimum of 95% negative results was determined to be the High Negative value.

Results: Limit of Blank = 4 ng/mL
Limit of Detection = 8 ng/mL
High Negative = 2 ng/mL

**e. Analytical specificity:**

**Interference Studies:**

Selected drugs and other non-microbial substances that might be present in stool samples from healthy persons or patients suspected of having *C. difficile* associated disease were added to three negative stool samples and three positive stool samples. The contrived positive specimens were prepared from a pool of donor stools that were confirmed as negative. The samples were inoculated with *C. difficile* strain 11186 at 8 ng/mL, which is the limit of detection for this assay. Potentially interfering substances were added at final concentrations of 5% V/V or greater. Dilution Controls for each sample were prepared by adding a phosphate-buffered saline solution in place of the potentially interfering substance. Each sample was tested in triplicate.

The following substances, at the specified saturated solvent/diluent concentrations, do not interfere with Premier *C. difficile* GDH test results in the final concentrations listed: Barium sulfate (5 mg/mL), Fecal fat (2.65 mg stearic acid and 1.3 mg palmitic acid/mL), Hemoglobin (3.2 mg/mL), Imodium AD® (Loperamide HCl) (6.67 x 10-3 mg/mL), Kaopectate® (Bismuth subsalicylate) (0.87 mg/mL), Metronidazole (12.5 mg/mL), Mucin (3.33 mg/mL), Mylanta® (Aluminum hydroxide w/ magnesium hydroxide) (4.2 mg/mL), Pepto-Bismol® (Bismuth subsalicylate) (0.87 mg/mL), Polyethylene glycol (79.05 mg/mL), Prilosec® (Omeprazole) (0.5 mg/mL), Simethicone (0.625 mg/mL), Tagamet® (Cimetidine) (0.5 mg/mL), Tums® (Calcium carbonate) (0.5 mg/mL), Vancomycin HCl (2.5 mg/mL), Whole blood (25%), White

8

{8}

blood cells (5%).

## Cross-reactivity Studies:

Potentially cross-reactive microorganisms that might be present in stool samples from healthy persons or patients suspected of having *C. difficile* associated disease were added to a natural negative and contrived positive sample. The contrived positive specimens were prepared from a pool of donor stools that was confirmed negative. The contrived positive sample was prepared by spiking a confirmed negative sample with *C. difficile* strain 11186 at 8 ng/mL, which is the limit of detection for this assay. Potentially cross-reactive microorganisms were added at final concentrations of 1.2 × 10⁸ CFU/mL (bacteria or fungi) or final concentrations greater than 1 × 10⁵ TCID₅₀/mL (viruses). Dilution controls for each sample were prepared by adding a 0.85% saline solution in place of the potentially cross-reactive organisms.

The following microorganisms, at the indicated concentrations, do not interfere with Premier *C. difficile* GDH test results: *Aeromonas hydrophila*, *Bacillus cereus*, *Bacillus subtilis*, *Bacteroides fragilis*, *Campylobacter coli*, *Campylobacter fetus*, *Campylobacter jejuni*, *Candida albicans*, *Citrobacter freundii*, *Clostridium butyricum*, *Clostridium bifermentans*, *Clostridium histolyticum*, *Clostridium novyi*, *Clostridium perfringens*, *Clostridium septicum*, *Clostridium sordellii*, *Clostridium tetani*, *Enterobacter aerogenes*, *Enterobacter cloacae*, *Enterococcus faecalis*, *Escherichia coli*, *Escherichia coli O157:H7*, *Escherichia hermannii*, *Escherichia fergusonii*, *Helicobacter pylori*, *Klebsiella pneumoniae*, *Lactococcus lactis*, *Listeria monocytogenes*, *Peptostreptococcus anaerobius*, *Plesiomonas shigelloides*, *Porphyromonas asaccharolytica*, *Proteus vulgaris*, *Pseudomonas aeruginosa*, *Pseudomonas fluorescens*, *Salmonella Group B*, *Salmonella Group C*, *Salmonella Group D*, *Salmonella Group E*, *Serratia liquefaciens*, *Serratia marcescens*, *Shigella boydii*, *Shigella flexneri*, *Shigella sonnei*, *Staphylococcus aureus*, *Staphylococcus epidermidis*, *Vibrio parahaemolyticus*, *Yersinia enterocolitica*, Adenovirus Type 40, Adenovirus Type 41, Coxsackievirus Strain B4, Echovirus Strain 11, Rotavirus Strain WA.

Stools spiked with *Staphylococcus aureus* (Cowan strain I) and *Clostridium sporogenes* were found to be cross-reactive with Premier *C. difficile* GDH.

## Strain Reactivity Studies:

*C. difficile* stock cultures from different sources were tested and produced positive reactions at a concentration of 5.7 × 10⁷ cells/mL with the Premier *C. difficile* GDH assay. Strains tested were as follows: 8864, 10463, 43598, 2004052, 2004111, 2004118, 2004205, 2004206, 2005070, 2005257, 2005325, 2006240, 2007431, 2007435, 2007858, 2008016, 2008029, 2008162, 2008188, 2008341, 2008351, 2009018, 2009065, 2009066, 2009099, 2009132, 2009155, 2009277, 11186, B1, BI8, BI17, BK6, CF1, G1, J7, K12, Y1, 234, 586, 611, 620, 2C62, 2C165, C122, UNC 19904, X15076.

9

{9}

f. Assay cut-off:

Premier C. difficile GDH is manufactured with a fixed cut-off with cut-off values of 0.200 at A450 or 0.150 at A450/630. This is accomplished on a lot to lot basis with specific procedures for optimizing solid phase and conjugate reagents. The validity of this cutoff was proven in verification studies on three lots of fully manufactured product and clinical trials performed on three lots of fully manufactured product.

2. Comparison studies:

a. Method comparison with predicate device:

Not applicable

b. Matrix comparison:

Not applicable

3. Clinical studies:

a. Clinical Sensitivity:

Clinical trials for the Premier C. difficile GDH assay were conducted from November 2010 to February 2011. Performance characteristics of the Premier C. difficile assay were determined by comparison to bacterial C. difficile culture. Independent clinical test sites located in the Midwestern, Southeastern and Southwestern regions of the United States evaluated a total of 733 qualified patient samples. Samples were prospectively collected from 337 (46.0%) males and 390 (53.2%) females. Gender was not defined for 6 (1%) patients. The age groups of patients ranged from 22 days to 99 years. No differences in test performance were observed based on patient age, gender, or geographic location. Overall sensitivity was determined to be 92.3% (95% CI: 86.0 - 95.9%). Overall specificity was determined to be 95.8% (95% CI: 93.9 - 97.1%).

Performance Characteristics for Premier C. difficile GDH (by Site)

|   | Positive Samples |   |   | Negative Samples  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Site | Premier GDH/ Culture | Sensitivity % | 95% CI | Premier GDH/ Culture | Specificity % | 95% CI  |
|  Total Sites | 108/117 | 92.3% | 86.0 – 95.9% | 590/616 | 95.8% | 93.9 – 97.1%  |
|  Site 1 | 13/16 | 81.3% | 57.0 – 93.4% | 84/87 | 96.6% | 90.3 – 98.8%  |
|  Site 2 | 28/30 | 93.3% | 78.7 – 98.2% | 132/140 | 94.3% | 89.1 – 97.1%  |
|  Site 3 | 44/46 | 95.7% | 85.5 – 98.8% | 147/153 | 96.1% | 91.7 – 98.2%  |

{10}

11

|  Site 4 | 15/15 | 100.0% | 79.6 – 100.0% | 169/175 | 96.6% | 92.7 – 98.4%  |
| --- | --- | --- | --- | --- | --- | --- |
|  Site 5 | 8/10 | 80.0% | 49.0 – 94.3% | 58/61 | 95.1% | 86.5 – 98.4%  |

- Discrepant specimens were evaluated using an FDA-cleared ELISA test for the detection of *C. difficile* GDH. This evaluation was not used to change the initial study results.
- Sixteen of the 26 false positive specimens were positive when tested with another FDA cleared GDH assay.
- Eight of the nine false negative specimens were negative when tested with another FDA cleared GDH assay

Performance Characteristics for Premier *C. difficile* GDH (by Patient Age)

|   | Positive Samples |   |   | Negative Samples  |   |   |
| --- | --- | --- | --- | --- | --- | --- |
|  Patient Age | Premier GDH/ Culture | Sensitivity % | 95% CI | Premier GDH/ Culture | Specificity % | 95% CI  |
|  0-28 days | 0/0 | N/A | N/A | 1/1 | 100.0% | 20.7 – 100.0%  |
|  29 days to 2 years | 22/24 | 91.7% | 74.2 – 97.7% | 56/60 | 93.3% | 84.1 – 97.4%  |
|  > 2 years to < 12 years | 24/25 | 96.0% | 80.5 – 99.3% | 106/109 | 97.2% | 92.2 – 99.1%  |
|  12 years to < 18 years | 9/11 | 81.8% | 52.3 – 94.9% | 59/61 | 96.7% | 88.8 – 99.1%  |
|  18 years to 21 years | 3/4 | 75.0% | 30.1 – 95.4% | 22/23 | 95.7% | 79.0 – 99.2%  |
|  > 21 years | 50/53 | 94.3% | 84.6 – 98.1% | 345/361 | 95.6% | 92.9 – 97.3%  |
|  Not Defined | 0/0 | N/A | N/A | 1/1 | 100.0% | 20.7 – 100.0%  |

b. Clinical specificity:

See 3a) above

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values/Reference range:

{11}

The frequency of antibiotic-associated diarrhea caused by *C. difficile* is dependent on several factors including: patient population, type of institution and epidemiology. The reported incidence of *C. difficile* infection in patients suspected of having antibiotic-associated diarrhea is 15-25% although different facilities may find positivity rates outside this range.

**N. Proposed Labeling:**

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

**O. Conclusion:**

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

12

---

**Source:** [https://fda-staging.innolitics.com/device/K110620](https://fda-staging.innolitics.com/device/K110620)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
