K110599 · Tornier, Inc. · PHX · Jun 14, 2011 · Orthopedic
Device Facts
Record ID
K110599
Device Name
AEQUALIS ASCEND MODULAR REVERSE SHOULDER SYSTEM
Applicant
Tornier, Inc.
Product Code
PHX · Orthopedic
Decision Date
Jun 14, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Aequalis Ascend Modular Reverse Shoulder System is indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by: Rheumatoid arthritis Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) Correction of functional deformity Fractures of the humeral head Traumatic arthritis Revision of the devices if sufficient bone stock remains The Reverse Metaphysis and Assembly Screw or the Reverse Metaphysis, Reverse Spacer, and Reverse Spacer Assembly Screw are indicated for use as components of the Aequalis Ascend Modular Reverse total shoulder replacement and for transformation of the Aequalis Ascend Anatomic shoulder into a reverse shoulder prosthesis without the removal of the Diaplyseal stem during revision surgery for patients with a functional deltoid muscle. The components are pennitted to be used in the transformation from anatomic to reverse if: The Diaphyseal stem is well fixed The patient has a functional deltoid muscle The arthropathy is associated with a massive and non-repairable rotator cuff-tear. Notes: All components are single use The humeral stem is intended for cemented or cementless use The glenoid implant is anchored to the bone with 4 screws and is for non- cemented fixation.
Device Story
Modular reverse shoulder prosthesis; replaces shoulder joint in patients with functional deltoid and massive, non-repairable rotator cuff tears. System components (reverse metaphysis, assembly screws, inserts, optional spacers) assemble in OR onto existing well-fixed Aequalis Ascend Anatomic humeral stems or as primary total shoulder replacement. Metal metaphysis impacts onto stem taper; secured by assembly screw. Polyethylene inserts provide articulation. Used by orthopedic surgeons in clinical settings. Enables conversion from anatomic to reverse configuration without removing well-fixed diaphyseal stems. Benefits include restoration of joint function and pain relief in complex shoulder pathologies.
Clinical Evidence
Bench testing only. Testing included fatigue testing, fretting corrosion, and static disassembly testing of the reverse insert.
Technological Characteristics
Modular shoulder prosthesis. Materials: Titanium (ASTM F136) for metaphysis/spacer; Cobalt-Chromium-Molybdenum (ASTM F1537) for assembly screws; UHMWPE (ASTM F648/ISO 5834-2) for screw plugs/inserts. Fixation: Cemented or cementless humeral stem; non-cemented glenoid fixation via 4 screws. Single-use.
Indications for Use
Indicated for patients with functional deltoid muscle and massive, non-repairable rotator cuff tears causing pain and disability due to rheumatoid arthritis, osteoarthritis, avascular necrosis, functional deformity, humeral head fractures, traumatic arthritis, or revision surgery with sufficient bone stock.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Aequalis Reversed or Aequalis Reversed II glenoid implants (K081059)
Submission Summary (Full Text)
{0}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three parallel lines forming its body and head. The bird is positioned above a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 2, 2016
Tornier, Incorporated Mr. Brahim Hadri Sr. Regulatory Affairs Specialist 100 Cummings Center, Suite 444C Beverly, Massachusetts 01915
Re: K110599
Trade/Device Name: Aequalis® Ascend™ Modular Reverse Shoulder System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, HSD, KWT Dated: February 28, 2011 Received: March 17, 2011
Dear Mr. Hadri:
This letter corrects our substantially equivalent letter of June 14, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r load be act not a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
#### K110599 510(k) Number (if known):
# Device Name: Aequalis® Ascend™ Modular Reverse Shoulder System
#### Indications for Use
- The Aequalis Ascend Modular Reverse Shoulder System is indicated for use ● as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by:
- o Rheumatoid arthritis
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis and O avascular necrosis)
- Correction of functional deformity o
- o Fractures of the humeral head
- Traumatic arthritis o
- Revision of the devices if sufficient bone stock remains o
- The Reverse Metaphysis and Assembly Screw or the Reverse Metaphysis, o Reverse Spacer, and Reverse Spacer Assembly Screw are indicated for use as components of the Aequalis Ascend Modular Reverse total shoulder replacement and for transformation of the Aequalis Ascend Anatomic shoulder into a reverse shoulder prosthesis without the removal of the Diaplyseal stem during revision surgery for patients with a functional deltoid muscle. The components are pennitted to be used in the transformation from anatomic to reverse if:
- The Diaphyseal stem is well fixed o
- The patient has a functional deltoid muscle o
- The arthropathy is associated with a massive and non-repairable rotator o cuff-tear.
- Notes: o
- o All components are single use
- The humeral stem is intended for cemented or cementless use O
- The glenoid implant is anchored to the bone with 4 screws and is for non- O cemented fixation.
| Prescription Use | X (Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use (21 CFR 807 Subpart C) |
|------------------|-------------------------------|--------|---------------------------------------------|
|------------------|-------------------------------|--------|---------------------------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
El Keith
| 510(k) Submission: | | |
|------------------------------|-----------------------------------------------------------|---------------|
| Tornier Inc. Aequalis Ascend | (Division Sign-Off) | Page 11 of 76 |
| | Division of Surgical, Orthopedic, and Restorative Devices | |
| 510(k) Number | K110599 | |
{3}------------------------------------------------
K110599 (144)
JUN 1 4 2011
## SECTION 5
#### 510(k) SUMMARY
#### Summary of Safety and Effectiveness information
#### Tornier Inc. Aequalis Ascend Modular Reverse Shoulder System
Regulatory authority: Safe Medical Devices Act of 1990, 21 CRF 807.92
- 1) Device name
Trade name: Acqualis Ascend Modular Reverse Shoulder System Common name: Shoulder Prosthesis
Classification Number/ Classification name/Product code:
- Shoulder joint metal/polymer non-constrained cemented prosthesis are class II devices under 21 CFR 888.3650 (product code KWT) and are classified by the Orthopedic Devices Panel.
- . Shoulder joint metal/polymer semi-constrained cemented prosthesis are class II devices under 21 CFR 888.3660 (product code KWS) and are classified by the Orthopedic Devices Panel.
- Shoulder joint humeral (hemi-shoulder) metallic uncemented ● prosthesis are class II devices under CFR 888.3690 (product code HSD) and are classified by the Orthopedic Devices Panel.
- 2) Submitter
Tornier Inc.
7701 France Avenue South; Suite 600 Edina, MN 55435 Registration Number: 9100540
#### 3) Company contact
Brahim Hadri
Sr. Regulatory affairs Specialist 100 Cummings Center, Suite 444C. Beverly, MA 01915, U.S.A Phone: 1 978 232-9997 ext: 617 1 978-232-9998 Fax:
bhadri@tornier.com
S 10(k) Submission
Tornier Inc. Aequalis Ascend Modular Reverse Shoulder System
Page 12 of 76
{4}------------------------------------------------
K110599 (2 of 4)
- 4) Classification
| Device class: | Class II |
|-----------------------|---------------|
| Classification panel: | Orthopedic |
| Product code: | KWT; KWS; HSD |
- 5) Legally Marketed Device to which Equivalence is Claimed:
- · Tornier Aequalis Ascend Modular Anatomic Shoulder Prosthesis: K102924
- · Tornier Aequalis Reversed Shoulder Prosthesis: K100142
- · Tornier Aequalis Reversed Fracture Shoulder Prosthesis: K082120
- · Encore Reverse Shoulder Prosthesis: K051075
- · Biomet Comprehensive Reverse Shoulder: K080642
- · Depuy Delta Xtend Reverse Shoulder System: K062250, K073676, K081620
- 6) Device description
The Aequalis Ascend Modular Reverse Shoulder system is a modular Reverse version of the Aequalis Ascend Modular Anatomic Shoulder System (K102924) with Reverse adapters.
The Aequalis Ascend Modular Reverse Shoulder system is supplied in separate sterile packages which will be assembled in the operating room. The components provided will be:
- The Modular Reverse Metaphysis and Modular Reverse Assembly Screw .
- � The Modular Reverse Inserts
- . The Modular Reverse Angled Inserts
- An optional Modular Reverse Spacer and Modular Reverse Spacer Assembly o Screw can be used when maximum insert thickness does not achieve sufficient tension.
- When the Modular Reverse Spacer is implanted, the Modular Reverse O Spacer Assembly Screw will be used in place of the Modular Reverse Assembly Screw.
{5}------------------------------------------------
#### Primary Reverse Shoulder:
The Aequalis Ascend Modular Reverse Metaphysis and Aequalis Ascend Modular Reverse Assembly Screw will mate with the existing; FDA cleared Tornier Aequalis Ascend Modular Anatomic Distal Stem (K102924) which when assembled together with the Aequalis Ascend Modular Reverse Insert form complete reverse prostheses. The reverse assembly must be used in association with the Aequalis Reversed or Aequalis Reversed II glenoid implants (K081059).
### Conversion Reverse Shoulder:
The Aequalis Ascend Reverse Metaphysis and Aequalis Ascend Modular Reverse Assembly Screw are designed to allow the transformation of the Aequalis Ascend Anatomic shoulder (K102924) into an Aequalis Ascend Modular Reverse Shoulder without removal of the Diaphyseal Stem if the stem is well fixed during revision surgery. The Aequalis Ascend Reverse Metaphysis is a metal component aimed to articulate as a reverse prosthesis with a specific polyethylene insert. The metal metaphysis is impacted onto the taper of the Diaphyseal Stem and then secured with the Ascend Modular Reverse Assembly Screw. The screw is tightened until it is firmly seated on the Reverse Metaphysis. A Reverse Spacer can be used when the maximum insert thickness does not achieve sufficient tension. During these situations, after impaction of the Reverse Metaphysis on the Diaphyseal stem, the metal spacer is impacted on the taper of the Reverse Metaplıysis and then secured with the Ascend Modular Reverse Spacer Assembly Screw. The screw is tightened until it is firmly seated on the Reverse Spacer.
#### 7) Materials
- · The Modular Reverse Metaphysis and Modular Reverse Spacer components are manufactured from Titanium according to ASTM F 136 Standard Specification for Wrought Titanium - 6Aluminum - 4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications
- The Modular Reverse Assembly Screw and Modular Reverse Spacer Assembly Screw o are manufactured from Cobali Chromium Molybdenum alloy according to ASTM F-1537 Standard Specification for Wrought Cobalt-28-Chromium-6-Molybdenum Alloy for Surgical Implants
- . The Modular Reverse Assembly Screw and Modular Reverse Spacer Assembly Screw include an Ultra High Molecular Weight Polyethylene (UHMWPE) plug. The material of the screws is accordance to ASTM F648 Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants or ISO-5834-2. Implants for surgery -- Ultra-high-molecular-weight polyethylene -- Part 2: Moulded forms
510(k) Submission Tornier Inc. Aequalis Ascend Modular Reverse Shoulder System
Page 14 of 76
{6}------------------------------------------------
#### 8) Indications for Use
- The Aequalis Ascend Modular Reverse Shoulder System is indicated for use . as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by:
- Rheumatoid arthritis o
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis and o avascular necrosis)
- Correction of functional deformity o
- Fractures of the humeral head o
- Traumatic arthritis o
- Revision of the devices if sufficient bone stock remains o
- The Reverse Metaphysis and Assembly Screw or the Reverse Metaphysis, . Reverse Spacer, and Reverse Spacer Assembly Screw are indicated for use as components of the Aequalis Ascend Modular Reverse total shoulder replacement and for transformation of the Aequalis Ascend Anatomic shoulder into a reverse shoulder prosthesis without the removal of the Diaphyseal stem during revision surgery for patients with a functional deltoid muscle. The components are permitted to be used in the transformation from anatomic to reverse if:
- The Diaphyseal stem is well fixed O
- The patient has a functional deltoid muscle 0
- The arthronathy is associated with a massive and non-repairable rotator O cuff-tear.
- . Notes:
- All components are single use o
- The humeral stem is intended for cemented or cementless use o
- The glenoid implant is anchored to the bone with 4 screws and is for nono cemented fixation.
#### 9) Summary of technologics
The Tornier Aequalis Ascend Modular Reverse Shoulder System was subjected to non-clinical testing such as Fatigue Testing, Fretting Corrosion, Static Disassembly Testing of Reverse Insert. The results of those non-clinical tests allow us to conclude that the Tornier Aequalis Ascend Modular Reverse Shoulder System described in this submission is substantially equivalent and as safe and effective as the cited predicate devices.
Page 15 of 76
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.