Archived clinical cases were used in simulated comparison testing to validate the performance and functionality of the EV1000 Clinical Platform against predicate devices.
Simulated comparison testing of archived clinical cases; Retrospective comparative analysis
Critical care patients
Predicate devices (K100709, K082308, K072735)
Performance and functionality of the EV1000 Clinical Platform
Indications for Use
The EV1000 Clinical Platform is indicated for use primarily for critical care patients in which the balance between cardiac function, fluid status and vascular resistance needs continuous or intermittent assessment. Analysis of the thermodilution curve in terms of mean transit time and the shape is used to determine intravascular and extravascular fluid volumes. When connected to an Edwards oximetry catheter, the monitor measures oximetry in adults and pediatrics. The EV1000 Clinical Platform may be used in all settings in which critical care is provided.
Device Story
EV1000 Clinical Platform comprises Databox and touchscreen Monitor; mounted on IV pole. System integrates with Edwards pressure transducers, FloTrac sensors, VolumeView components, and oximetry catheters. Databox receives physiologic signals from patient-connected accessories; embedded algorithms process signals to calculate hemodynamic parameters (e.g., cardiac output, stroke volume, systemic vascular resistance, extravascular lung water). Monitor displays processed data via GUI. Used in OR and ICU by clinicians for continuous or intermittent hemodynamic assessment. Output informs clinical decision-making regarding fluid management and cardiac function. Benefits include real-time monitoring of critical physiologic parameters to guide therapy.
Clinical Evidence
Evidence includes bench testing, pre-clinical animal studies, simulated comparison testing of archived clinical cases, clinical usability/human factor assessments, and a multi-center clinical trial. Testing confirmed performance and functionality equivalent to predicate devices.
Technological Characteristics
System consists of Databox and touchscreen panel PC monitor connected via ethernet. Measures physiologic parameters via integrated sensors/transducers. Software-based signal processing and parameter calculation. Designed for OR and ICU environments.
Indications for Use
Indicated for critical care patients requiring assessment of cardiac function, fluid status, and vascular resistance. Oximetry monitoring indicated for adults and pediatrics. Used in all critical care settings.
Regulatory Classification
Identification
A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.
{0}------------------------------------------------
JUN 1 4 2011
:
## 5. 510(k) Summary
| EV1000 CLINICAL PLATFORM 510(k) SUMMARY | |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Edwards Lifesciences LLC<br>One Edwards Way<br>Irvine, CA 92614-5686 |
| Contact Person: | Marguerite Thomlinson, Sr. Manager, Regulatory Affairs<br>Edwards Lifesciences, LLC, Critical Care<br>Phone: (949) 756-4386<br>Fax: (949) 809-5676 |
| Date Prepared: | March 1, 2011 |
| Trade name: | EV1000 Clinical Platform |
| Classification<br>Name: | Single-Function, Preprogrammed Diagnostic Computer<br>(21 CFR 870.1435) |
| Predicate<br>Devices: | • K100709, Edwards Lifesciences LLC, EV1000 Platform<br>• K082308, Edwards Lifesciences LLC, Vigileo Arterial<br>Pressure Cardiac Output/Oximetry Monitor<br>• K072735, Pulsion Medical Systems, PiCCO-2 |
| Device<br>Description: | The EV1000 Clinical Platform consists of a Databox and<br>Monitor components, which can be mounted to an IV pole. The<br>EV1000 Clinical Platform measures patient physiologic<br>parameters when it is used as a system with various Edwards<br>components, including the Edwards pressure transducers, the<br>FloTrac sensor, the components of the VolumeView System,<br>oximetry catheters/sensors, and the corresponding accessories<br>applied to the patient.<br><br>The EV1000 Databox receives incoming signals from the patient<br>through the connections provided by the accessories applied to the patient. The algorithms embedded in the Databox process the signals and provide parameter calculations.<br><br>The EV1000 Monitor is connected to the Databox via an<br>ethernet cable. The Monitor is a touchscreen, panel PC with a<br>graphical user interface (GUI). The Monitor displays the<br>measured and calculated parameter values from the Databox.<br><br>The EV1000 Clinical Platform, when used with the<br>VolumeView System, measures and/or calculates hemodynamic<br>parameters such as: |
| | EV1000 CLINICAL PLATFORM 510(k) SUMMARY |
| • | Auto-calibrated continuous parameters: cardiac output, cardiac index, stroke volume, stroke volume index, systemic vascular resistance, systemic vascular resistance index, and stroke volume variation; |
| • | Manual-calibrated continuous parameters: cardiac output, cardiac index, stroke volume, stroke volume index, systemic vascular resistance, systemic vascular resistance index, and stroke volume variation; and, |
| • | Manual-calibrated intermittent parameters: cardiac output , cardiac index, extravascular lung water, extravascular lung water index, global ejection fraction, global end-diastolic volume, global end-diastolic volume index, intrathoracic blood volume, pulmonary vascular permeability index, stroke volume, stroke volume index, systemic vascular resistance, and systemic vascular resistance index. |
| | When connected to a FloTrac sensor, the EV1000 Clinical Platform continuously measures/calculates arterial pressure cardiac output, cardiac index, stroke volume, stroke volume index, stroke volume variation, systemic vascular resistance, and systemic vascular resistance index. |
| | When connected to Edwards oximetry sensors, the EV1000 Clinical Platform continuously measures/calculates oximetry parameters. |
| Indications for Use/ Intended Use | The EV1000 Clinical Platform is indicated for use primarily for critical care patients in which the balance between cardiac function, fluid status and vascular resistance needs continuous or intermittent assessment. Analysis of the thermodilution curve in terms of mean transit time and the shape is used to determine intravascular and extravascular fluid volumes. When connected to an Edwards oximetry catheter, the monitor measures oximetry in adults and pediatrics. The EV1000 Clinical Platform may be used in all settings in which critical care is provided. |
| EV1000 CLINICAL PLATFORM 510(k) SUMMARY | |
| Comparative<br>Analysis: | Verification and validation testing was conducted to compare<br>the performance and functionality of the EV1000 Clinical<br>Platform to the predicate devices. This testing regimen included<br>head-to-head bench and pre-clinical studies, and comparative<br>analysis of archived clinical data and clinical data obtained<br>during a multi-center clinical trial of the EV1000 Clinical<br>Platform. The EV1000 Clinical Platform has been shown to be<br>safe and effective and substantially equivalent to the cited<br>predicate devices for their intended use in the OR and ICU<br>environments. |
| Functional/<br>Safety Testing: | The EV1000 Clinical Platform has successfully undergone<br>functional and performance testing, including software<br>verification and validation, mechanical and electrical testing,<br>bench studies, pre-clinical animal studies, simulated comparison<br>testing of archived clinical cases, clinical usability and human<br>factor assessment and a multi-center clinical study. The EV1000<br>Clinical Platform has been shown to be safe and effective and<br>substantially equivalent to the cited predicate devices for their<br>intended use in the OR and ICU environments. |
| Conclusion: | The proposed EV1000 Clinical Platform is safe and effective<br>and is substantially equivalent to the predicate devices. |
{1}------------------------------------------------
{2}------------------------------------------------
·
.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized depiction of an eagle or bird with three wing-like strokes, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Edwards Lifesciences, LLC Edwards Enescions Thomlinson, J.D. c/o Ms. Marguethe Thommers)
Sr. Manager of Regulatory Affairs, Critical Care One Edwards Way Irvine, CA 92614
JUN 1 4 2011
K110597 Re: K110597
Trade/Device Name: EV1000 Clinical Platform
Trade/Device School Co., 01 CFR 870 1435 Trade/Device Name: 21 CFR 870.1435 Regulatory Number: 21 CFR 870.1435
Regulation Name: Single-function, Preprogrammed Diagnostic Computer
Single-Single-Sunction, Hunce) Regulatory Class: II (two) Product Code: DXG Dated: May 13, 2011 Received: May 16, 2011
Dear Ms. Thomlinson:
Deal interest We have reviewed your Section 510(k) premarket notifically equivalent (for the indications)
referenced above and have determined the device is substantially equivales of the we nave and have determined the devices is substantially cqurvalisms.
for use stated in the enclosure) to legally marketed predicate devices market din interstate for use stated in the encloser to tegally marketed prodical Device arminents, or to
for use stated in the enclosed in accordent date of the Fedical Device of Duris, or BMA for use stated in according in accordines with the provisions of the Federal Food, Drux).
commerce prior o May 28, 1976, the encordines with the provisions of the Pederal Foo comments in to may 20, 2011 accordance with the provisions of the Peach of PMA).
devices that have been require approval of a premarket approval one of Act. The Act. The A devices market the device, subject to the general ontrols provisions of the Act. The Act. The Act. The Act. The Act. The Act. The and Cosments ( the Act the Act include requirements for annual registration, listing of
general controls provisions of the Act include requirements for annual misbranding an I ou film, woreated include requirements for ammun registered, whiles against misbranding and
devices, good manufacturing practice, labeling, and prohibitions against misbran adulteration.
adulteration:
If your device is classified (see above) into either class III (PMA), etc.
If your device and liting al controls . Existing major regulations affective and If your device is classified (see above) into either class II (Special Collicols) of Children)
it may be subject to additions. Existing major regulations affecting your devi If your device is externed in the States on the St. In additions article for the may be
it may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may
foun found in the Code of Federal Regulations, Title 21, Farst 800 to College of Register.
{4}------------------------------------------------
Page 2 - Ms. Marguerite Thomlinson, J.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## K110597
## Indications for Use 4.
510(k) Number (if known):
EV1000 Clinical Platform Device Name:
Indications for Use:
The EV1000 Clinical Platform is indicated for use primarily for critical care patients in which the balance between cardiac function, fluid status and vascular resistance needs continuous or intermittent assessment. Analysis of the thermodilution curve in terms of mean transit time and the shape is used to determine intravascular and extravascular fluid volumes. When connected to an Edwards oximetry catheter, the monitor measures oximetry in adults and pediatrics. The EV1000 Clinical Platform may be used in all settings in which critical care is provided.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ﻟﻤ
Division of Cardiovascular Devices
**510(k) Number** K110597
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.