K110572 · Philips Medical Systems · MUE · Sep 23, 2011 · Radiology
Device Facts
Record ID
K110572
Device Name
MAMMODIAGNOST DR
Applicant
Philips Medical Systems
Product Code
MUE · Radiology
Decision Date
Sep 23, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1715
Device Class
Class 2
Indications for Use
The Philips MammoDiagnost DR is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The MammoDiagnost DR is intended to be used in the same clinical applications as traditional film/screen systems.
Device Story
Full-field digital mammography system for screening and diagnostic procedures. System comprises gantry with rotating X-ray tube (Tungsten/Molybdenum anode, 0.3/0.15 focal spots), motorized compression paddles, and solid-state amorphous selenium detector (3584 x 2816 matrix). Operated by clinicians in clinical settings via Eleva Workspot (AWS) with 19-inch touch monitor. System captures X-ray projections; processes data into DICOM MG format for review on diagnostic workstations, CAD systems, or PACS. Automatic Exposure Control (AEC) optimizes image quality and radiation dose based on breast thickness. Provides high-resolution images to assist radiologists in detecting breast cancer; supports clinical decision-making through standardized image quality metrics (MTF, DQE, NPS).
Clinical Evidence
Clinical image evaluation performed on 6 cases (screening and diagnostic) from non-pregnant women over 40. Two expert breast imaging subspecialist radiologists reviewed images using MQSA accreditation criteria. Attributes scored 1-5; all cases passed. Bench testing included sensitometric response, spatial resolution (MTF), noise power spectrum (NPS), detective quantum efficiency (DQE), dynamic range, and phantom testing (ACR and CDMAM). AEC performance and radiation dose (mean glandular dose) measured across 2cm, 4cm, and 6cm breast-equivalent phantoms.
Technological Characteristics
Full-field digital mammography system. X-ray source: Rotating tube, 0.3/0.15 focal spots, Tungsten/Molybdenum anode, Rhodium/Molybdenum filters. Detector: Solid-state amorphous selenium, 3584 x 2816 matrix. Connectivity: DICOM MG output. AEC: Automatic exposure control for dose optimization. Software: Eleva Workspot (AWS) for image acquisition and processing.
Indications for Use
Indicated for generating mammographic images for screening and diagnosis of breast cancer in women over age 40. Contraindicated for pregnant women and women with breast implants.
Regulatory Classification
Identification
A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.
Siemens Mammomat Inspiration (Supplement to P030010/S)
Submission Summary (Full Text)
{0}------------------------------------------------
Summary of Safety and Effectiveness
SEP 2 3 2011
# 510(k) Summary
In accordance with the requirements of the Safe Medical Device Act, Philips Medical Systems herewith submits a 510(k) Summary.
### Name and Address of manufacturer:
Philips Medical Systems DMC GmbH Roentgenstrasse 24 22335 Hamburg Germany Establishment Registration No .: 3003768251 Owner/Operator Number: 1217116
### Name, title and phone number of official correspondent:
Linda Jalbert Women's Health Care Philips Healthcare 3000 Minuteman Road Andover, MA 01810 Phone 978 659-7434
#### Device Identification:
Device Trade Name: Philips MammoDiagnost DR Common Name: Full Field Mammography System
### Classification of the device:
Device Classification Name: Product Code: Device Classification No .: Panel: Regulatory Status:
Full Field Digital,System,X-ray,Mammographic MUE Part 892.1715 Radiology Class II
### Device(s) Identification:
Device Trade Name: Common Name:
Philips MammoDiagnost DR Full Field Digital Mammography System
{1}------------------------------------------------
6)
| Predicate devices: | |
|--------------------|----------------------------------|
| Device Trade Name: | Siemens Mammomat Novation |
| Applicant: | Siemens Medical Solution |
| 510(k) No.: | P030010 |
| Device Trade Name: | Siemens Mammomat Inspiration |
| Applicant: | Siemens Medical Solution |
| 510(k) No.: | Supplement to P030010 (P030010/S |
The Philips MammoDiagnost DR Full Field Digital Mammography System is considered substantially equivalent to Siemens Mammomat Novation and Inspiration Full Field Digital Mammography Systems. There is no significant difference in intended use or technology.
### Device Description:
The Philips MammoDiagnost DR digital mammography system is designed to perform screening and diagnostic mammography procedures on standing, seated or recumbent patients. The system consists of a gantry with an operator console including the Eleva Workspot (EWS), consisting of an acquisition workstation (AWS) and a 19" touch screen monitor. The system's rotating X-ray tube (0.3 / 0.15 Focal Spot, Anode: Tungsten/Molybdenum) offers 4 anode tracks. The gantry provides motorized elevation, rapid rotation and compression via two foot switches. An operator radiation shield and a removable recessed patient face shield are included in the system as well as the following paddles: Compression paddles (7" x 9,5" x 12"), Axilla and Spot compression. The system is equipped with a solid state amorphous selenium detector, providing a matrix size of 3584 x 2816.
The acquisition workstation provides a high performance central processing unit with a minimum of 500GB hard disk, and a 4GB memory. A 19" flat panel touch screen color monitor on pivot mount allows the users a selectable viewing position. Acquired images are provided in DICOM MG format which allows its integration into a digital mammography environment including diagnostic review workstations, CAD Systems and PACS.
### Indications for Use:
The MammoDiagnost DR is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The MammoDiagnost DR is intended to be used in the same clinical applications as traditional film/screen systems.
{2}------------------------------------------------
### Summary of Non-Clinical Performance Testing:
### Sensitometric Response:
Sensitometric Response is a measure of the sensitivity of the image acquisition system to different levels of x-ray exposures.
Image /page/2/Figure/6 description: This image is a graph that plots the median LSB against the detector entrance dose in micrograys. The x and y axes are on a logarithmic scale. The graph shows a linear relationship between the two variables, with the equation y = 3.5877x and an R-squared value of 1. The plot includes a scatter plot of the median and a line of best fit.
### Spatial Resolution:
Image sharpness of an image detector can be quantified by its modulation transfer function (MTF). The spatial resolution was measured according to the IEC 62220-1-2.
Image /page/2/Figure/9 description: The image is a plot of Binned MTF vs Frequency. The x-axis is Frequency in LP/mm from 0 to 6. The y-axis is Binned MTF from 0 to 1.2. There are three lines plotted: Binned Vertical, Binned Horizontal, and Binned Average.
{3}------------------------------------------------
### Noise Power Spectrum (NPS)
Noise power spectrum (NPS) is a characterization of the noise distribution over the spatial frequency which is an important factor contributing to image quality. The NPS (noise power spectra) was measured at 8 dose levels, and was calculated according to the IEC 62220-1-2.
510(K)
Image /page/3/Figure/4 description: The image contains two plots showing horizontal and vertical noise power spectrum (NPS) as a function of frequency. The top plot displays horizontal NPS, while the bottom plot displays vertical NPS, both measured in mm²LSB². Each plot contains multiple curves representing different exposure levels, ranging from 14.3 uGy to 836.7 uGy, with frequency on the x-axis in LP/mm, ranging from 0.1 to 10.
{4}------------------------------------------------
### Signal-to-Noise Ratio (SNR):
The efficiency of SNR transfer of the image acquisition system is measured by the DQE (detective quantum efficiency) as a function of spatial frequency. DQE provides a measure of the combined effect of the noise and contrast performance of an imaging system, expressed as a function of object detail. The DQE was measured according to the IEC 62220-1-2.
510(k)
Image /page/4/Figure/4 description: The image is a graph that shows the horizontal DQE (Detective Quantum Efficiency) as a function of frequency. The x-axis represents the frequency in LP/mm (line pairs per millimeter), ranging from 0 to 6. The y-axis represents the horizontal DQE, ranging from 0.0% to 80.0%. There are multiple lines on the graph, each representing a different xSpec value in uGy (microGray), including 14.3, 27.3, 53.9, 106.4, 212.1, 420, and 836.7.
Image /page/4/Figure/5 description: The image is a graph that shows the vertical DQE (detective quantum efficiency) as a function of frequency. The x-axis represents the frequency in LP/mm, ranging from 0 to 6. The y-axis represents the vertical DQE, ranging from 0.0% to 80.0%. There are multiple lines on the graph, each representing a different ySpec value, including 14.3uGy, 27.3uGy, 53.9uGy, 106.4uGy, 212.1uGy, 420uGy, and 836.7uGy.
{5}------------------------------------------------
### Dynamic Range
Dynamic range is a measure of the signal response of a detector exposed to x-rays. The dynamic range was characterized by plotting the DQE over dose at low and mid frequencies.
Image /page/5/Figure/5 description: The figure is a graph comparing xSpec and ySpec with respect to Entrance Dose in microGrays. The x-axis shows the Entrance Dose from 0 to 900 microGrays, and the y-axis shows DQE @ 3lp/mm from 0% to 70%. Both xSpec and ySpec increase as the Entrance Dose increases. At 800 microGrays, xSpec is at about 63% and ySpec is at about 56%.
#### Repeated Exposure Test
The purpose of this measurement is to qualify the lag behavior of the digital x-ray detector. The measurement was repeated 100 times, the time period between each image was set to one minute.
Image /page/5/Figure/8 description: The image is a graph that shows the ghost image factor over time. The x-axis represents time in minutes, ranging from 0 to 100. The y-axis represents the ghost image factor, ranging from 0.00 to 0.35. There are three lines on the graph: one representing the ghost image factor using the mean, one representing the ghost image factor using the average, and one representing the limitation, which is a straight line at 0.30.
#### Phantom Image Tests
ﮯ ﺩ
Image quality is also demonstrated by analysis of phantom images. Philips evaluated the visibility of various features of the American College of Radiology (ACR) accreditation phantom
{6}------------------------------------------------
and the CDMAM contrast-detail mammography phantom. Scoring of the ACR phantom and analysis of the CDMAM phantom images were used to assess the MammoDiagnost DR image quality. CDMAM image acquisition and analysis were performed following EUREF guidelines (Bosmans H et al., 2009) and the published scientific literature (Young KC et al., 2008). The table below presents the results for normal and high dose using a 4 mm thickness ACR MAP phantom. Both phantom tests demonstrate the excellent detection capabilities of the MammoDiagnost DR System.
510(k)
| Softcopy | | | | | | | | | |
|----------------|--------|-----------|-------|-----------------|-----------|-------|--------|-----------|-------|
| Mode | Fibers | Artifacts | Score | Speck<br>Groups | Artifacts | Score | Masses | Artifacts | Score |
| Normal<br>Dose | 5.5 | 0 | 5.5 | 4.0 | 0 | 4.0 | 4.0 | 0 | 4.0 |
| High<br>Dose | 6.0 | 0 | 6.0 | 4.0 | 0 | 4.0 | 4.0 | 0 | 4.0 |
| Hardcopy | | | | | | | | | |
| Mode | Fibers | Artifacts | Score | Speck<br>Groups | Artifacts | Score | Masses | Artifacts | Score |
| Normal<br>Dose | 5.5 | 0 | 5.5 | 4.0 | 0 | 4.0 | 4.0 | 0 | 4.0 |
| High<br>Dose | 6.0 | 0 | 6.0 | 4.0 | 0 | 4.0 | 4.0 | 0 | 4.0 |
### Automatic Exposure Control (AEC)
Automatic exposure control (AEC) is a technology that is widely used in standard x-rav imaging and digital imaging systems. The objective of an AEC system is to optimize image quality while minimizing patient dose in an effort to produce consistent radiology images. The performance of the AEC was evaluated by acquiring data sets of varying breast equivalent thicknesses in both normal and high dose modes. An Aluminum target of 0.2 mm was used for the CNR measurement. CNR in both normal and high dose contact modes changed by less than 10% per centimeter of PMMA thickness.
### Patient Radiation Dose
Breast equivalent phantom blocks were used to simulate fibroglandular/adipose tissue composition. Phantom thicknesses of 2 cm, 4 cm, and 6 cm were tested. For each phantom the system AEC was used to select x-ray exposure parameters. Results for the 50/50 fibroglandular/adipose tissue composition in contact mode are given in the table below.
| Thickness<br>(cm) | mAs | kV | Target | Filter | Entrance<br>exposure<br>(mR) | Mean<br>Glandular<br>Dose (mGy) |
|-------------------|-------|----|--------|--------|------------------------------|---------------------------------|
| 2 | 35.2 | 25 | W | Rh | 98.1 | 0.44 |
| 4 | 86.7 | 26 | W | Rh | 289.9 | 0.87 |
| 6 | 194.3 | 28 | W | Rh | 854.0 | 1.91 |
{7}------------------------------------------------
/
.
### Summary of Clinical Image Evaluation:
In accordance to the Class II Special Controls Guidance Document: Full-Field Digital Mammography System, an image attribute review was completed on 6 mammographic image cases that were acquired from the Philips FFDM system, MammoDiagnost DR. The purpose of the image attribute review was to determine if images from the Philips FFDM device had acceptable quality for mammographic screening and diagnostic usage and were substantially equivalent to a predicate device as determined by a MQSA review of the investigational FFDM images.
Six screening and diagnostic mammography cases consisting of 4 standard views (RCC. RMLO, LCC, LMLO) including magnification or spot views from the diagnostic cases were reviewed by two expert Radiologist's meeting the radiologists' qualifications specified in the Class II Special Controls Guidance Document: Full-Field Digital Mammography System. The mammography cases were acquired from non-pregnant women over the age of 40. Women with breast implants were not included in the study.
The image sets for this evaluation were reviewed in the same manner as clinical images submitted by a mammography facility for MQSA accreditation. The reviewers evaluated the mammographic attributes for each case, scored the attributes with a 1 (worse)-5 (best) score and determined an overall pass/fail for each case. The two expert breast imaging subspecialist radiologists agreed that the six image sets from the Philips MammoDiagnost DR system are of sufficiently acceptable quality for clinical mammographic usage.
### Substantial Equivalence Discussion:
The Philips MammoDiagnost DR Full Field Digital Mammography System is considered substantially equivalent to Siemens Mammomat Novation and Inspiration Full Field Digital Mammography Systems. The MammoDiagnost DR has the same intended use and indications for use as the predicate devices. In addition there is no significant difference in technology. Like the predicate Inspiration System, the MammoDiagnost has an X-ray system with 0.3 / 0.15 Focal Spots and Tungsten/Molybdenum anodes, Rhodium//Molybdenum filters, Both systems use a solid state amorphous selenium detector with a matrix size of 3584 x 2816. The image quality performance characterized by DQE and MTF are very similar.
### Conclusion:
Philips believes that the Philips MammoDiagnost DR Full Field Digital Mammography System is substantially equivalent to the currently legally marketed devices. It has the same intended use, does not introduce new indications for use, has the same technological characteristics and does not introduce new potential hazards or safety risks.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or forms, stacked on top of each other. The figures are depicted with flowing, curved lines, giving the impression of movement or fluidity.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Linda Jalbert Director, Quality and Regulatory Philips Medical Systems 3000 Minuteman Road ANDOVER MA 01845
Re: K110572
Trade/Device Name: Philips MammoDiagnostic DR Regulation Number: 21 CFR 892.1715 Regulation Name: Full-field mammography system Regulatory Class: II Product Code: MUE Dated: August 15, 2011 Received: August 16, 2011
SEP 2 3 2011
Dear Ms. Jalbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{9}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to hitp://www.fda.gov/Medicall.Devices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
## Indications for Use
510(k) number (if known):
Device Name:
Philips MammoDiagnost DR
Indications For Use:
The Philips MammoDiagnost DR is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The MammoDiagnost DR is intended to be used in the same clinical applications as traditional film/screen systems.
Prescription Use AND/OR Over-The-Counter Use X (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
uly D. O'h
(Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K110572
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.