ASAHI TREASURE FLOPPY PERIPHERAL GUIDE WIRE

K110553 · Asahi Intecc USA, Inc. · DQX · Mar 18, 2011 · Cardiovascular

Device Facts

Record IDK110553
Device NameASAHI TREASURE FLOPPY PERIPHERAL GUIDE WIRE
ApplicantAsahi Intecc USA, Inc.
Product CodeDQX · Cardiovascular
Decision DateMar 18, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

The ASAHI Treasure Floppy Peripheral Guide Wire is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only.

Device Story

Steerable guide wire; 0.018-inch diameter; 190 cm or 300 cm length. Stainless steel core; stainless steel and platinum-nickel distal coil; radiopaque tip. Hydrophilic coating on distal portion; PTFE coating (type I or II) on proximal section. Used by physicians during peripheral vascular procedures to facilitate placement/exchange of diagnostic/therapeutic devices. Provides navigation through vessel curvature via bendable distal tip. Benefits patient by enabling precise delivery of interventional tools.

Clinical Evidence

Bench testing only. No clinical data. Testing included tensile strength, torque strength, torqueability, tip flexibility, coating adhesion, catheter compatibility, and particulate testing. Biocompatibility evaluation included systemic toxicity, hemolysis, intracutaneous, cytotoxicity, sensitization, pyrogen, plasma recalcification, coagulation, thromboresistance, and complement activation studies.

Technological Characteristics

Stainless steel core; stainless steel and platinum-nickel distal coil. Hydrophilic distal coating; PTFE proximal coating. 0.018-inch diameter; 190/300 cm lengths. Radiopaque distal tip. Mechanical navigation principle. Sterilized device.

Indications for Use

Indicated for patients undergoing intravascular procedures requiring the placement and exchange of diagnostic and therapeutic devices within the peripheral vasculature.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KIIOS MAR 1 8 2011 510(K) SUMMARY #### 9.0 510(K) SUMMARY ு. This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | APPLICANT | Asahi Intecc Co., Ltd.<br>1703 Wakita-cho, Moriyama-ku<br>Nagoya, Aichi 463-0024<br>Japan | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | OFFICIAL<br>CORRESPONDENT | Yoshi Terai<br>President, CEO<br>Asahi Intecc USA, Inc.<br>2500 Red Hill Avenue, Suite 210<br>Santa Ana, CA 92705<br>Tel: (949) 756-8252<br>FAX (949) 756-8165<br>e-mail: yoshi.terai@asahi-intecc.com | | TRADE NAME: | ASAHI Treasure Floppy Peripheral Guide Wire | | COMMON NAME: | Guide Wire | | CLASSIFICATION<br>NAME: | Wire, Guide, Catheter | | DEVICE<br>CLASSIFICATION: | Class 2 per 21 CFR §870.1330 | | PRODUCT CODE | DQX | | PREDICATE DEVICE: | Asahi - ASAHI Treasure 12 Peripheral Guide Wire - 510(k) K061984<br>Asahi - ASAHI Astato XS 20 Peripheral Guide Wire - 510(k) K103057<br>Asahi - JoWire Neo's PTCA Guide Wire – 510(k) K022762<br>Asahi - ASAHI PTCA Guide Wire - 510(k) K070945 | DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION: The ASAHI Treasure Floppy Peripheral Guide Wire is a steerable guide wire with a maximum diameter of 0.018 inches (0.45mm) and available in 190 cm and 300 cm length. The guide wire is constructed from a stainless steel core wire with a stainless steel and platinum-nickel coil. The coil part (distal end) of the guide wire is radiopaque to achieve visibility. The distal end of the coil part is a straight configuration and is easily bendable with the vessel curvature. A hydrophilic coating is applied to the coil part (distal portion) of the guide wire. The proximal section of the guide wire is coated with PTFE, and the PTFE in the proximal coating is available in two types - PTFE type I and type II. CONFIDENTIAL Asahi Intecc February 24, 2011 のお気になると、それではなくてしました。 この時間では、 この時間ではない {1}------------------------------------------------ ### 510(K) Summary ### Indication for Use: The ASAHI Treasure Floppy Peripheral Guide Wire is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only. ### TECHNICAL CHARACTERISTICS: This submission represents minor dimensional specifications and the addition of a PTFE coating. All other specifications are equivalent to those listed for the currently cleared predicate devices. Comparisons of the ASAHI Treasure Floppy Peripheral Guide Wire and predicate devices show that the technological characteristics such as product performance, design and intended use are substantially equivalent to the current marketed predicate devices. The ASAHI Treasure Floppy Peripheral Guide Wire is similar in design - device dimensional specifications, and intended use, manufacturing process, operating principle. shelf life and sterilization process are the same and materials that have been used in other predicate devices in that its core wire, tip coils and solders remain the same. ### PERFORMANCE DATA: Enclosed within this submission is performance data that demonstrates that the ASAHI Treasure Floppy Peripheral Guide Wire meets all predetermined performance criteria. All components that come in direct contact with the patient have a long history of use in medical devices and are proven to be biocompatible for use in the vasculature. Furthermore, this submission contains reference to predicate ASAHI devices that use the same materials as used in the subject device. And In vitro bench testing and shelf-life testing, including tensile strength, torque strength, torqueability, tip flexibility, coating adherence and catheter compatibility as listed below were conducted on the ASAHI Treasure Floppy Peripheral Guide Wire. This 510(k) notice includes mechanical and functional bench testing that demonstrates that the ASAHI Treasure Floppy Peripheral Guide Wire performs as intended. The biocompatibility testing as listed below was leveraged from predicate devices with identical materials and manufacturing process. Performance test/evaluation summary: Device performance: Tensile Strength Turns to Failure (Torque Strength) Torqueability (Torque Response) Tip Flexibility Coating Adhesion Slipping Ability of Guide Wire in PTA Balloon Catheter Particulate testing CONFIDENTIAL Asahi Intecc February 24, 2011 9. {2}------------------------------------------------ ## 510(K) SUMMARY Biocompatibility/evaluation: Systemic Toxicity Study In Vitro Hemolysis Study Intracutaneous Study Cytotoxicity Study Sensitization Study Pyrogen Study Plasma Recalcification Time Coagulation Study In Vivo Thromboresistance Study C3a Complement Activation Study SC5b-9 Complement Activation Study ### SUMMARY/CONCLUSION: ਰੇ. The ASAHI Treasure Floppy Peripheral Guide Wire characteristics are substantially equivalent to the specified predicate device and other currently marketed devices for the same indication for use. CONFIDENTIAL Asahi Intecc February 24, 2011 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or feathers. To the left of the graphic, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the edge of the logo. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Asahi Intecc USA, Inc. c/o Mr. Yoshi Terai President, CEO 2500 Red Hill Ave. Suite 210 Santa Ana, CA 92705 MAR 1 8 2011 Re: K110553 Trade/Device Name: Treasure Floppy Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II (two) Product Code: DQX Dated: February 24 2011 Received: February 28, 2011 Dear Mr. Terai: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Yoshi Terai Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE STATEMENT #### 2.0 INDICATIONS FOR USE STATEMENT 510(k) Number (if known): K1 0553 Device Name: ASAHI Treasure Floppy Peripheral Guide Wire Indications for Use: 2. The ASAHI Treasure Floppy Peripheral Guide Wire is intended to facilitate the placement and exchange of diagnostic and therapeutic devices during intravascular procedures. This device is intended for peripheral vascular use only. Prescription Use_X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Image /page/5/Figure/11 description: The image shows a document with the title "Concurrence of CDRH, Office of Device Evaluation (ODE)". There is a signature on the left side of the image. The text "Page of" is on the right side of the image. The number "K110 553" is written near the bottom of the image.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...