The OSI O-ball Abutment implant is a self-tapping titanium threaded screw indicated for long term intra-bony fixation of upper and lower dentures in edentulous cases. These devices will permit immediate splinting and ability and short-term fixation of failing crown and bridge installations, for full or partial edentulism. They can be used in the anterior regions of the maxillary and mandibular arches and are indicated for immediate loading when there is good primary stability and appropriate occlusal load.
Device Story
Self-tapping titanium threaded screw implant; used for long-term fixation of dentures or short-term stabilization of failing crown/bridge installations. Implant features o-ball head; available in 3.0mm-4.5mm diameters and 10mm-18mm lengths. Manufactured from ASTM F136 Ti6Al4V; threaded surface blasted to increase surface area. Used by dental clinicians in clinical settings. Provides mechanical support for prosthetic devices; enables immediate loading when primary stability and occlusal load conditions are met. Benefits patient by restoring chewing function and stabilizing dental prosthetics.
Clinical Evidence
Bench testing only. Fatigue testing conducted according to a modified ISO 14801 standard on the 3.0mm implant variant. Testing successfully demonstrated mechanical integrity under specified loading conditions.
Indicated for edentulous or partially edentulous patients requiring long-term intra-bony fixation of dentures or short-term fixation of failing crown/bridge installations. Suitable for anterior maxillary and mandibular arches. Contraindicated without good primary stability or appropriate occlusal load.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
K110348
510k Summary
JUL 2 9 2011
# July 22, 2011
# Trade Name: Juell OSI O-ball Abutment Dental Implant
Common Name: o-ball implant
Material: ASTM F136 Ti6Al4V
## Company Contact:
ﻟ
John Roderick, Operations Manager Juell Dental 2401 N. Commerce Ardmore, OK 73401 (580) 798-4414
## Device Description
The JUELL OSI O-Ball Abutment Dental Implant is available with a gingival collar in four diameters: 3.0mm, 3.5mm, 4.0mm and 4.5mm. The JUELL OSI O-Ball Abutment Dental Implant is available in thread lengths ranging from 10mm to 18mm. The implant is manufactured from ASTM F136 Ti6Al4V. The threaded surface is blasted with particulate media to increase the surface area.
## Indications for Use
The OSI O-ball Abutment implant is a self-tapping titanium threaded screw indicated for long term intrabony fixation of upper and lower dentures in edentulous cases. These devices will permit immediate splinting and ability and short-term fixation of failing crown and bridge installations, for full or partial edentulism. They can be used in the anterior regions of the maxillary and mandibular arches and are indicated for immediate loading when there is good primary stability and appropriate occlusal load.
## Testing
Fatigue was conducting according to a mofified ISO 14801 on the 3.0mm implant with the exception of the angle the implant was held being adjusted to 15° and the loading point being on the o-ball causing the length from the bone level to be 5mm and the length from the holding line to be 8mm.
| | OSI o-ball | Nobel Active o-ball<br>(k102436) | Champion<br>(k091182) |
|---------------------|-------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Diameter (mm) | 3.0,3.5,4.0,4.5 | 3.3 | 4, 4.5 |
| Length (mm) | 10,13,15,17,18 | 10, 11.5, 13,15 | 10 to 16 |
| Implant head shape | o-ball | o-ball | o-ball |
| Indications for use | The OSI O-ball Abutment<br>implant is a self-tapping<br>titanium threaded screw<br>indicated for long term intra- | The Nobel Active 3.0<br>implant is indicated for<br>use in the treatment of<br>missing lateral incisors or<br>the mandibular central | The Champions<br>Implants MIMI 1-<br>Piece Precision<br>Implant is intended<br>to support single or<br>multi-unit |
#### Substantial Equivalence
{1}------------------------------------------------
| | bony fixation of upper and<br>lower dentures in edentulous<br>cases. These devices will<br>permit immediate splinting<br>and ability and short-term<br>fixation of failing crown and<br>bridge installations, for full or<br>partial edentulism. They can<br>be used in the anterior<br>regions of the maxillary and<br>mandibular arches and are<br>indicated for immediate<br>loading when there is good<br>primary stability and<br>appropriate occlusal load. | and lateral incisors to<br>support prosthetic<br>devices, such as artificial<br>teeth, in order to restore<br>chewing function in<br>partially edentulous<br>patients. The Nobel<br>Active 3.0 Implants may<br>be put in immediate<br>function provided that<br>stability requirements<br>detailed in the manual<br>are satisfied. | restorations in both<br>long-term and<br>temrporary<br>applications<br>throughout the<br>maxillary and<br>mandibular<br>arches. The MIMI 1-<br>Piece Precision<br>Implant is indicated<br>for immediate<br>loading when there<br>is good<br>primary stability and<br>an appropriate<br>occlusal load. The<br>MIMI 1-Piece<br>Precision Implants<br>have diameters<br>from 4,0 to 4,5mm<br>and are available in<br>length of 10 to<br>16mm. |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | Ti6Al4V | Ti6Al4V | Ti6Al4V |
| Sterilization | Delivered sterile | Delivered sterile | Delivered sterile |
| Procode | Dze | Dze | Dze |
| Screw pitch | Between Imtec fine and<br>coarse | Self-tapping compression<br>threads | Microthreads with<br>cutting flute and<br>coarse compression<br>threads |
| Screw surface | Blasted and clean | Sandblasted and acid<br>etched | Blasted and etched |
| Fatigue testing | Tested to modified ISO 14801<br>successfully | Tested according to the<br>guidance document<br>successfully | No testing<br>information in the<br>510k summary |
The materials, surface treatment, and design are similar and show Juell o-ball abutments to be substantially equivalent to the predicate devices. The indications for use is similar to those of the predicate devices so the Juell o-ball abutments are substantially equivalent to the predicate devices. Both Juell o-ball abutment implants and the predicate devices tested in fatigue successfully, so Juell oball abutment implants are substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol. The text is in all caps and is in a sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Jucil Dental C/O Ms. Angela Blackwell Senior Consultant Biologics Consulting Group 2401 North Commerce Ardmore, Oklahoma 73401
JUL 2 9 2011
Re: K110548
Trade/Device Name: OSI o-Ball Abutment Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous dental Implant Regulatory Class: II Product Code: DZE Dated: July 22, 2011 Received: July 25, 2011
Dear Ms. Blackwell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Ms. Blackwell
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Susan Purnan
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
# Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K | 1 05 48
Device Name: OSI o-ball abutment implants
Indications for Use:
The OSI O-ball Abutment implant is a self-tapping titanium threaded screw indicated for long term intra-bony fixation of upper and lower dentures in edentulous cases. These devices will permit immediate splinting and ability and short-term fixation of failing crown and bridge installations, for full or partial edentulism. They can be used in the anterior regions of the maxillary and mandibular arches and are indicated for immediate loading when there is good primary stability and appropriate occlusal load.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Knox
Page Page of
(Division Sign-Off) Division of Anesthesiology. General Hospital Infection Control. Dental Devices
510(k) Number: K110548
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.