K110510 · Sonoscape Company Limited · IYN · May 16, 2011 · Radiology
Device Facts
Record ID
K110510
Device Name
S20 DIGITAL COLOR DOPPLER ULTRASOUND SYSTEM
Applicant
Sonoscape Company Limited
Product Code
IYN · Radiology
Decision Date
May 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), OB/Gyn and Urology.
Device Story
S20 Digital Color Doppler Ultrasound System is an integrated, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers (phased, micro-curved, and linear arrays) and displays images in B-Mode, M-Mode, Tissue Harmonic Imaging (THI), Color-Flow Doppler, Pulsed Doppler, Power Doppler, and 3D/4D modes. The system features an all-digital architecture with progressive dynamic receive focusing to enhance diagnostic utility. It is intended for use by qualified physicians in clinical settings. The system provides exam-dependent default settings for ease of use, with advanced soft-menu controls for customization. Output is displayed on the system monitor, allowing clinicians to evaluate anatomy and fluid flow, aiding in diagnostic decision-making and patient management.
Clinical Evidence
Bench testing only. The device was evaluated for design specifications, acoustic output (NEMA UD 2:2004, NEMA UD 3:2004), thermal/mechanical/electrical safety (IEC 60601-1, IEC 60601-2-37), and biocompatibility (ISO 10993-5, ISO 10993-10). No clinical data was required for this 510(k) clearance.
Technological Characteristics
Digital architecture with progressive dynamic receive focusing. Supports B, M, PWD, CWD, Color Doppler, Power Doppler, and 3D/4D modes. Transducers: Phased, Micro-curved, and Linear arrays (2.0-12.0 MHz). Conforms to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-5, and ISO 10993-10. Software-controlled system.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal/adult cephalic, trans-rectal, trans-vaginal, peripheral vascular, musculoskeletal (conventional/superficial), cardiac (neonatal/adult), OB/GYN, and urology applications. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
SonoScape S8 Diagnostic Ultrasound System and Transducers (K092922)
Submission Summary (Full Text)
{0}------------------------------------------------
SonoScape Company LTD
S20 Digital Color Doppier Ultrasound System
# Tab 21 PREMARKET NOTIFICATION 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is :
#### Manufacturer:
SonoScape Company Limited
Address: 4/F., Yizhe Building, Yuquan Road,Nanshan, Shenzhen 518051,
P.R.China
Tel: (86) 755-26722890
Fax: (86) 755-26722850
Contact Person: Zhiqiang Chen
Date Prepared: Feb 12, 2011
Name of the device:
* Trade/Proprietary Name:
S20 Digital Color Doppler Ultrasound System
* Common Name: Diagnostic Ultrasound System and Transducers
* Classification:
Regulatory Class: II
Review Category: Tier II
21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (90-IYN)
21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (90-IYO)
21 CFR 892.1570 Diagnostic Ultrasound Transducer (90-ITX)
#### Legally Marketed Predicate Device:
Premarket Notification 510(k) Summary
{1}------------------------------------------------
SonoScape S8 Diagnostic Ultrasound System and Transducers - K092922
#### Device Description:
The SonoScape S20 ultrasound system is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.
The all digital architecture with progressive dynamic receive focusing allows the system to maximize the utility of all imaging transducers to enhance the diagnostic utility and confidence provided by the system. The exam dependent default setting allows the user to have minimum adjustment for imaging the patient, while the in-depth soft-menu control allows the advanced user to set the system for different situations. The architecture allows cost-effective system integration to a variety of upgrade-able options and features.
This SonoScape system is a general purpose, software controlled, diagnostic ultrasound system. Its basic function is to acquire ultrasound data and display the image in B-Mode (including Tissue Harmonic Image), M-Mode,TDI, Color-Flow Doppler, Pulsed Doppler and Power Doppler, or a combination of these modes, 3D/4D.
#### Intended Use:
The device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), OB/Gyn and Urology.
{2}------------------------------------------------
### Probe Information:
| No. | Probe | Type | Frequency Range | Intended Use |
|-----|-------|--------------------|-----------------|----------------------------------------------------------------------------------------------------------------------------------|
| 1 | 2P1 | Phased Array | 2.0-4.0 MHz | Abdominal<br>Neonatal Cephalic<br>Adult Cephalic<br>Cardiac Adult<br>Cardiac Pediatric |
| 2 | 5P1 | Phased Array | 3.0-7.0 MHz | Pediatric<br>Neonatal Cephalic<br>Cardiac Pediatric |
| 3 | 6V1 | Micro-curved Array | 4.0-8.0 MHz | Trans-rectal<br>Trans-vaginal |
| 4 | 6V3 | Micro-curved Array | 5.0-9.0 MHz | Trans-rectal<br>Trans-vaginal |
| 5 | EC9-5 | Micro-curved Array | 5.0-9.0 MHz | Trans-rectal<br>Trans-vaginal |
| 6 | C611 | Micro-curved Array | 4.0-8.0 MHz | Abdominal<br>Pediatric<br>Neonatal Cephalic<br>Cardiac Pediatric |
| 7 | C344 | Curved Array | 2.0-5.0 MHz | Fetal / Abdominal/ Ob/GYN |
| 8 | C362 | Curved Array | 2.0-6.0 MHz | Fetal / Abdominal/ Ob/GYN |
| 9 | VC6-2 | Curved Array | 2.0-6.0 MHz | Fetal / Abdominal/ Ob/GYN |
| 10 | L741 | Linear Array | 5.0-10.0 MHz | Small Organ (breast, thyroid, testes )<br>Musculo-skeletal (Conventional)<br>Peripheral vessel |
| 11 | L742 | Linear Array | 5.0-12.0 MHz | Small Organ (breast, thyroid, testes )<br>Musculo-skeletal (Conventional)<br>Musculo-skeletal (Superficial)<br>Peripheral vessel |
| 12 | L743 | Linear Array | 5.0-10.0 MHz | Small Organ (breast, thyroid, testes )<br>Musculo-skeletal (Conventional)<br>Musculo-skeletal (Superficial)<br>Peripheral vessel |
Tab 21.1 Probe information
### Safety Considerations:
The S20 Diagnostic Ultrasound System with added transducer incorporates the same fundamental technology as the predicate device. The device has been tested as Track 3 Device per the FDA Guidance document "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems Premarket Notification 510(k) Summary 21-3
{3}------------------------------------------------
SonoScape Company LTD
and Transducers" issued September 9, 2008. The acoustic output is measured and calculated per NEMA UD 2: 2004 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment and NEMA UD3: 2004 Standards for Real-time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment. The device conforms to applicable medical device safety standards, such as IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-5 and ISO 10993-10.
#### Testing:
Laboratory testing was conducted to verify that the S20 Digital Color Doppler Ultrasound System with added transducer met all design specification and was substantially equivalent to the currently marketed Predicate Device as above. The device has been found to conform to applicable medical device safety standards in regards to thermal, mechanical and electrical safety as well as biocompatibility. Acoustic output is measured and calculated according to "Acoustic Output Measuring Standard for Diagnostic Ultrasound Equipment".
| Standards No. | Standards Title | Version | Date |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|------------|
| IEC 60601-1 | IEC 60601-1, Medical Electrical Equipment - Part 1:<br>General Requirements for Safety, 1988;<br>Amendment 1, 1991-11, Amendment 2, 1995. | 1988 | 10/31/2005 |
| IEC 60601-1-2 | IEC 60601-1-2, (Second Edition, 2001), Medical<br>Electrical Equipment - Part 1-2: General<br>Requirements for Safety - Collateral Standard:<br>Electromagnetic Compatibility -- Requirements and<br>Tests. | 2007 | 07/31/2008 |
| IEC<br>60601-2-37 | IEC 60601-2-37 (2004) (2005) Amendment 2,<br>Medical electrical equipment - Part 2-37: Particular<br>requirements for the basic safety and essential<br>performance of ultrasonic medical diagnostic and<br>monitoring equipment. | 2007 | 09/08/2009 |
Tab 21.2 Applicable Safety Standards
Premarket Notification 510(k) Summary
{4}------------------------------------------------
| NEMA UD 2 | NEMA UD 2-2004, Acoustic Output Measurement<br>Standard for Diagnostic Ultrasound Equipment<br>Version 3. | 2004 | 09/08/2009 |
|-------------|--------------------------------------------------------------------------------------------------------------------------------------|------|------------|
| ISO 10993-5 | ISO 10993-5:1999, Biological evaluation of medical<br>devices -- Part 5: Tests for In Vitro cytotoxicity. | 2009 | 09/12/2007 |
| 10993-10 | ISO 10993-10:2002, Biological evaluation of<br>medical devices - Part 10: Tests for irritation and<br>delayed-type hypersensitivity. | 2002 | 09/12/2007 |
## Conclusion:
The conclusions drawn from testing of the S20 Diagnostic Ultrasound System with added transducer demonstrate that the device is as safe and effective as the legally marketed predicate devices.
Premarket Notification 510(k) Summary
21-5
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling the image of an eagle. The eagle is stylized with three curved lines representing its body, wings, and head. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993
SonoScape Company Limited % Ms. Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 237-023 Shanghai, 200237 CHINA
MAY 16 2011
Re: K110510
Trade/Device Name: S20 Digital Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: February 18, 2011 Received: February 22, 2011
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the S20 Digital Doppler Ultrasound System, as described in your premarket notification:
Transducer Model Number
2P1 Phased Array 5P1 Phased Array 6V1 Micro-curved Array 6V3 Micro-curved Array EC9-5 Micro-curved Array C611 Micro-curved Array C362 Curved Array C344 Curved Array VC6-2 Curved Array
L743 Linear Array L741 Linear Array L742 Linear Array
{6}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely Yours,
Michael D. O'Keeffe
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{7}------------------------------------------------
## 4.4 Tab 3 Indications For Use
510(k) Number:
Device Name: S20 Digital Doppler Ultrasound System
Indications for Use:
The SonoScape S20 device is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and aduit), OB/Gyn and Urology.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Maitha D. Juma
(Division Sign Off)
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) Number K116510
Indications For Use
{8}------------------------------------------------
#### System: Sonoscape S20 Diagnostic Ultrasound Pulsed Echo System Diagnostic Ultrasound Pulsed Doppler Imaging System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | Other* | |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--|----------|-------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | | Combined | Specify |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| Fetal<br>Imaging&<br>Other | Fetal | N | N | N | | N | N | | Note 1 | Notes 2,4,5 |
| | Abdominal | N | N | N | | N | N | | Note 1 | Notes 2,4,5 |
| | Intra-operative Specify | | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | N | N | N | | N | N | | Note 1 | Notes 2,4 |
| | Small Organ (specify) | N | N | N | | N | N | | Note 1 | Notes 2,4,6 |
| | Neonatal Cephalic | N | N | N | N | N | N | | Note 1 | Notes 2,3,4 |
| | Adult Cephalic | N | N | N | N | N | N | | Note 1 | Notes 2,3,4 |
| | Trans-rectal | N | N | N | | N | N | | Note 1 | Notes 2,4 |
| | Trans-vaginal | N | N | N | | N | N | | Note 1 | Notes 2,4 |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph.(non-Card) | | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | N | N | N | | N | N | | Note 1 | Notes 2,4 |
| | Musculo-skeletal<br>(Superficial) | N | N | N | | N | N | | Note 1 | Notes 2,4 |
| | Intravascular | | | | | | | | | |
| | Other (Ob/GYN) | N | N | N | | N | N | | Note 1 | Notes 2,4,5 |
| Cardiac | Cardiac Adult | N | N | N | N | N | N | | Note 1 | Notes 2,3,4 |
| | Cardiac Pediatric | N | N | N | N | N | N | | Note 1 | Notes 2,3,4 |
| | Intravascular(Cardiac) | | | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | |
| | Other (specify) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | | N | N | | Note 1 | Notes 2,4 |
| | Other (specify) | | | | | | | | | |
N = new indication; E = added under this appendix P = previously cleared by FDA; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Note 5: 4I)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Indications For Use
Mhnd D. O'Khm
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
K110510
510K
{9}------------------------------------------------
Transducer: 2P1 Phase Array
Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal<br>Imaging&<br>Other | Fetal | | | | | | | | |
| | Abdominal | N | N | N | | N | N | Note 1 | Notes 2,4 |
| | Intra-operative Specify | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 |
| | Adult Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph.(non-Card) | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (Ob/GYN) | | | | | | | | |
| Cardiac | Cardiac Adult | N | N | N | N | N | N | Note 1 | Notes 2,3,4 |
| | Cardiac Pediatric | N | N | N | N | N | N | Note 1 | Notes 2,3,4 |
| | Intravascular(Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (specify) | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/TH; M/Color M ; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D Note 5: 4D
Note 6: Small Organ: breast, thyroid, testes
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of In Vitro Diagnostic Devices (OIVD)
Indications For Use
Michael O'Kee
(Division Sign-Off)
Division o Office of In Vitro
510K K116510
{10}------------------------------------------------
Transducer: 5P1 Phase Array Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|----------------------|-------------------------|-------------------|---|-----|-----|---------|-------------|----------|-------------|
| General | Specific | | | | | Color | Power | Other* | Other* |
| (TRACK 1 | (TRACKS 1 & 3) | B | M | PWD | CWD | Doppler | (Amplitude) | Combined | Specify |
| ONLY) | | | | | | | Doppler | | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal | Fetal | | | | | | | | |
| Imaging& | | | | | | | | | |
| Other | Abdominal | | | | | | | | |
| | Intra-operative Specify | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | N | N | N | | N | N | Note 1 | Notes 2,4 |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card) | | | | | | | | |
| | Musculo-skeletal | | | | | | | | |
| | (Conventional) | | | | | | | | |
| | Musculo-skeletal | | | | | | | | |
| | (Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (Ob/GYN) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | N | N | N | N | N | N | Note 1 | Notes 2,3,4 |
| | Intravascular(Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | | |
| Vessel | Other (specify) | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: BM: B/PWD: B/THI: M/Color Doppler: B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Note 5: 4D
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Indications For Use
(Division Sian-Off
Division of Radiological Devices Office of In
510K K//65/0
{11}------------------------------------------------
#### Transducer: 6V1 Micro-curved Array Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal<br>Imaging&<br>Other | Fetal | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative Specify | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | N | N | N | | N | N | Note 1 | Notes 2,4 |
| | Trans-vaginal | N | N | N | | N | N | Note 1 | Notes 2.4 |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph.(non-Card) | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (Ob/GYN) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular(Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (specify) | | | | | | | | |
N = new indication; E = added under this appendix P = previously cleared by FDA; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Note 5: 4D
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Indications For Use
---
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagu
510K K110510
{12}------------------------------------------------
Transducer: 6V3 Micro-curved Array Diagnostic Ultrasound Transducer .
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | | Other* | |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|----------|-----------|--------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Combined | Specify | | |
| Ophthalmic | Ophthalmic | | | | | | | | | | |
| Fetal<br>Imaging&<br>Other | Fetal | | | | | | | | | | |
| | Abdominal | | | | | | | | | | |
| | Intra-operative Specify | | | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | | | |
| | Laparoscopic | | | | | | | | | | |
| | Pediatric | | | | | | | | | | |
| | Small Organ (specify) | | | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | | |
| | Trans-rectal | N | N | N | N | N | N | Note 1 | Notes 2,4 | | |
| | Trans-vaginal | N | N | N | N | N | N | Note 1 | Notes 2.4 | | |
| | Trans-urethral | | | | | | | | | | |
| | Trans-esoph. (non-Card) | | | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | | | |
| | Intravascular | | | | | | | | | | |
| | Other (Ob/GYN) | | | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | | |
| | Intravascular(Cardiac) | | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | | |
| | Other (specify) | | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | | |
| | Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR
Note 5: 4D
Over-The-Counter Use (21 CFR 807 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Indications For Use
*(Division Sign-Off)*
Division of Radiological Devic Office of In e Evaluation and Safety
510K K110510
{13}------------------------------------------------
Transducer: EC9-5 Micro-curved Array Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | | Mode of Operation | | | | | | | |
|----------------------------------------------------------------------------------|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| | General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| | Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal<br>Imaging&<br>Other | Fetal | | | | | | | | |
| | | Abdominal | | | | | | | | |
| | | Intra-operative Specify | | | | | | | | |
| | | Intra-operative Neuro | | | | | | | | |
| | | Laparoscopic | | | | | | | | |
| | | Pediatric | | | | | | | | |
| | | Small Organ (specify) | | | | | | | | |
| | | Neonatal Cephalic | | | | | | | | |
| | | Adult Cephalic | | | | | | | | |
| | | Trans-rectal | N | N | N | | N | N | Note 1 | Notes 2,4 |
| | | Trans-vaginal | N | N | N | | N | N | Note 1 | Notes 2,4 |
| | | Trans-urethral | | | | | | | | |
| | | Trans-esoph.(non-Card) | | | | | | | | |
| | | Musculo-skeletal<br>(Conventional) | | | | | | | | |
| | | Musculo-skeletal<br>(Superficial) | | | | | | | | |
| | | Intravascular | | | | | | | | |
| | | Other (Ob/GYN) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | | |
| | | Cardiac Pediatric | | | | | | | | |
| | | Intravascular(Cardiac) | | | | | | | | |
| | | Trans-esoph.(Cardiac) | | | | | | | | |
| | | Intra-cardiac | | | | | | | | |
| | | Other (specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | |
| | | Other (specify) | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added under this appendix | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Note 5: 4D
Over-The-Counter Use (21 CFR 807 Subpart C)
3-7
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
510K
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Indications For Use
Nerd D. Oth
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
{14}------------------------------------------------
Transducer: C611 Micro-curved Array Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | Other* | |
|------------------------------|------------------------------------|---|-------------------|-----|-----|------------------|---------------------------------|----------|-------------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Combined | Specify | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| Fetal<br>Imaging&<br>Other | Fetal | | | | | | | | | |
| | Abdominal | N | N | N | | N | N | Note 1 | Notes 2,4 | |
| | Intra-operative Specify | | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | N | N | N | | N | N | Note 1 | Notes 2,4 | |
| | Small Organ (specify) | | | | | | | | | |
| | Neonatal Cephalic | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph.(non-Card) | | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| | Intravascular | | | | | | | | | |
| | Other (Ob/GYN) | | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | N | N | N | N | N | N | Note 1 | Notes 2,3,4 | |
| | Intravascular(Cardiac) | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | |
| | Other (specify) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | |
| | Other (specify) | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWI)
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Note 5: 4D
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Indications For Use

(Division Sign-Off)
Division of Radiological Devices ce Evaluation and Safety Office of In Vitro
K110510
510K
{15}------------------------------------------------
Transducer: C362 Curved Array Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | | Other* | |
|------------------------------|------------------------------------|---|-------------------|-----|-----|------------------|---------------------------------|----------|-----------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Combined | Specify | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| Fetal<br>Imaging &<br>Other | Fetal | N | N | N | | N | N | Note 1 | Notes 2,4 | |
| | Abdominal | N | N | N | | N | N | Note 1 | Notes 2,4 | |
| | Intra-operative Specify | | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | | | | | | | | | |
| | Small Organ (specify) | | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card) | | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| | Intravascular | | | | | | | | | |
| | Other (Ob/GYN) | N | N | N | | N | N | Note 1 | Notes 2.4 | |
| Cardiac | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | |
| | Intravascular(Cardiac) | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | |
| | Other (specify) | | | | | | | | | |
| Peripheral | Peripheral vessel | | | | | | | | | |
| Vessel | Other (specify) | | | | | | | | | |
N = new indication; E = added under this appendix P = previously cleared by FDA; Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color M ; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Note 5: 4D
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of In Vitro Diagnostic Devices (OIVD)
Indications For Use
*(Division Sign-Off)*
Division of Radiologi uation and Safety
510
{16}------------------------------------------------
Transducer: C344 Curved Array Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Clinical Application | Mode of Operation | | | | | | | | |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| Fetal<br>Imaging&<br>Other | Fetal | N | N | N | N | N | N | Note 1 | Notes 2, 4 | |
| | Abdominal | N | N | N | N | N | N | Note 1 | Notes 2, 4 | |
| | Intra-operative Specify | | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | | | | | | | | | |
| | Small Organ (specify) | | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph.(non-Card) | | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | | |
| | Intravascular | | | | | | | | | |
| | Other (Ob/GYN) | N | N | N | N | N | N | Note 1 | Notes 2, 4 | |
| Cardiac | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | |
| | Intravascular(Cardiac) | | | | | | | | | |
| | Trans-esoph.(Cardiac) | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | |
| | Other (specify) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | |
| | Other (specify) | | | | | | | | | |
Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD
Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents
Note 3: TDI Note 4: 3D
Note 6: Small Organ: breast, thyroid, testes
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Note 5: 4D
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Indications For Use
(Division Sign
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
510K K110510
{17}------------------------------------------------
#### Sonoscape Company LTD
### Diagnostic Ultrasound Indications for Use Form
Transducer: VC6-2 Curved Array Diagnostic Ultrasound Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | | |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal<br>Imaging &<br>Other | Fetal | N | N | N | N | N | N | Note 1 | Notes 2,4,5 |
| | Abdominal | N | N | N | N | N | N | Note 1 | Notes 2,4,5 |
| | Intra-operative Specify | | | | | | | | |
| | Intra-operative Neuro | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (specify) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph.(non-Card) | | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | | |
| | Intravascular | | | | | | | | |
| | Other (Ob/GYN) | N | N | N | N | N | N…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.