Philips Medizinsysteme Boeblingen GmbH, Cardiac AN
Product Code
LKD · Anesthesiology
Decision Date
Sep 30, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 868.2480
Device Class
Class 2
Attributes
Pediatric
Indications for Use
IntelliVue MP40, MP50, MP60, MP70, MP80, MP90 and MX800 and IntelliBridge EC10 and EC5 IC Module: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. IntelliVue TcG10 Measurement Module: Indicated for use by health care professionals whenever there is a need to support the continuous and noninvasive monitoring of transcutaneous partial pressures of blood gas carbon dioxide and oxygen of patients not under gas anesthesia.
Device Story
The Philips IntelliVue TcG10 Measurement Module provides continuous, noninvasive monitoring of transcutaneous partial pressures of blood gas carbon dioxide (tcpCO2) and oxygen (tcpO2). The device uses a sensor placed on the skin surface; the sensor warms the skin to induce local hyperemia, increasing arterial blood supply to the dermal capillary bed for gas diffusion measurement. The module interfaces with Philips IntelliVue patient monitors (MP40-MX800) via IntelliBridge EC10 and EC5 ID modules. The system is used in hospital environments by healthcare professionals. Data is displayed on the patient monitor to assist clinicians in assessing blood gas partial pressures, which correlate with arterial blood gas values. The TcG10 module uses the same functional measurement and calibration technology as the predicate device, with added internal power and RS232 communication protocols for integration into the Philips monitoring ecosystem.
Clinical Evidence
No clinical data required per Class II Special Controls Guidance Document for Cutaneous Carbon Dioxide and Oxygen Monitors. Evidence consists of bench performance testing and a clinical evaluation including literature review and clinical acceptance studies with experienced nursing and physician staff.
Technological Characteristics
Modular patient monitoring system. TcG10 module uses transcutaneous electrochemical sensing (Radiometer tc Sensor 84). Connectivity via RS232 protocol to IntelliBridge EC10/EC5 modules and IntelliVue patient monitors. Software-based modifications for communication and device identification. System operates on AC/internal power. Materials in contact with patient skin are identical to the predicate device.
Indications for Use
Indicated for health care professionals to monitor physiological parameters in patients. TcG10 module indicated for continuous, noninvasive monitoring of transcutaneous pCO2 and pO2 in pediatric subgroups and adult patients not under gas anesthesia.
Regulatory Classification
Identification
A cutaneous carbon dioxide (PcCO2) monitor is a noninvasive heated sensor and a pH-sensitive glass electrode placed on a patient's skin, which is intended to monitor relative changes in a hemodynamically stable patient's cutaneous carbon dioxide tension as an adjunct to arterial carbon dioxide tension measurement.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Cutaneous Carbon Dioxide (PcCO2) and Oxygen (PcO2) Monitors; Guidance for Industry and FDA.” See § 868.1(e) for the availability of this guidance document.
{0}------------------------------------------------
SEP 3 0 2011
510 (k) Summary
K110474
'
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.
1. The submitter of this premarket notification is:
Dr. Jens-Peter Seher Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany Tel: ++49 7031 463-2086 Fax: ++49 7031 463-2442 e-mail: jens-peter.seher@philips.com
This summary was prepared on February 11, 2011.
2. The names of the devices are:
For the modified devices: Trade Name: Philips MP40, MP50, MP70, MP80, MP90, and MX800 IntelliVue Patient Monitor Common Name: Multiparameter Patient Monitor
and
Trade Name: Philips IntelliBridge EC10 and EC5 ID Module Common Name: Interface Module
and
。
For the new device: Trade name: Philips IntelliVue TcG10 Measurement Module Common name: tcpCO2/tcpO2 Measurement Module
{1}------------------------------------------------
Classification names are as follows:
| Device Panel | Classification | ProCode | Description |
|---------------------------|----------------|---------|------------------------------------------------------------------------------------------|
| Cardiovascular<br>Devices | \$870.1435, II | DXG | Computer, Diagnostic, Pre-<br>Programmed, Single-Function |
| | \$870.2060, II | DRQ | Amplifier and Signal<br>Conditioner, Transducer Signal |
| | \$870.2900, I | DSA | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector |
| Anesthesiology<br>Devices | \$868.2480, II | LKD | Monitor, Carbon Dioxide,<br>Cutaneous |
| | \$868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for<br>Infant not under Gas Anesthesia |
| | \$868.2500, II | LPP | Monitor, Oxygen, Cutaneous, for<br>Use other than for Infant not<br>under Gas Anesthesia |
- 3. The modified Philips MP40, MP50, MP60, MP70, MP80, MP90, and MX800 IntelliVue Patient Monitors are substantially equivalent to the previously cleared Philips IntelliVue Patient Monitors marketed pursuant to K013199, K032858, K040304, K041235, K050762, K053522, K062392, K71426, K081793, K082633, K083517, K091927, K100939, K101449 and K102562.
The modified Philips IntelliBridge EC10 and EC5 ID modules are substantially equivalent to the previously cleared Philips IntelliBridge EC10 and EC5 ID modules marketed pursuant to ко82483.
Module IntelliVue TcG10 Measurement i s Philips The new substantially equivalent to the previously cleared TCM CombiM Monitoring System marketed pursuant to K093154 and in particular to the TCM CombiM Measurement Module which is the measurement unit of the TCM CombiM Monitoring System.
- 4. Description of the devices
The modified Philips MP40, MP50, MP70, MP80, MP90, and MX800 IntelliVue Patient Monitors are a flexible and modular monitoring solution optimized for the surgical, cardiac, general medical and neonatal care environments. The monitors can be connected to the Philips Multi-Measurement Module (MMS) family with its extensions and to the IntelliVue family plug-in measurement modules. The monitors can also be connected to the IntelliVue anesthetic gas modules and to the new Intellivue TcG10 Measurement Module.
The modified Philips IntelliBridge EC10 and EC5 ID modules build an external medical device solution to collect data from external devices at the bedside providing the data to patient monitors, and patient monitoring networks. The modified EC10 and EC5 ID modules support the interface of the IntelliVue patient monitors with the new IntellVue TcG10 Measurement Module.
{2}------------------------------------------------
The new IntellVue TcG10 Measurement Module is a device for the continuous and noninvasive measurement of the transcutaneous partial pressures of blood gas carbon dioxide (pCO2) and oxygen (p02) of patients from all pediatric subgroups and of adult patients not under gas anesthesia in hospital environment. The transcutaneous measurement of pCO2 and pO2 makes use of the fact that carbon dioxide and oxygen gases are able to diffuse through body tissue and skin and can be detected by a sensor placed at the skin surface. By warming up this sensor, a local hyperemia is induced, which increases the supply of arterial blood to the dermal capillary bed below the sensor. The transcutaneous blood gas values (tcpCO2 and tcpO2) have to be interpreted primarily as the blood gas partial pressures prevailing at the level of the arterialized skin tissue. In general, this value correlates well with the corresponding arterial blood gas partial pressure.
- 5. Indications for Use
IntelliVue MP40, MP50, MP60, MP70, MP80, MP90 and MX800 and IntelliBridge EC10 and EC5 IC Module: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
IntelliVue TcG10 Measurement Module:
Indicated for use by health care professionals whenever there is a need to support the continuous and noninvasive monitoring of transcutaneous partial pressures of blood gas carbon dioxide and oxygen of patients not under gas anesthesia.
6. Technological characteristics
modified and new devices have the same substantial The technological characteristics, such as fundamental design, user interface, materials, energy source, and measurement technology, as the legally marketed predicate devices.
The modification of the Philips MP40, MP50, MP60, MP70, MP80, MP90, and MX800 IntelliVue Patient Monitors is solely limited to software changes which are necessary to interface the patient monitors to the new IntelliVue TcG10 Measurement Module via the IntelliBridge EC10 and EC5 ID Module.
The modifications of the IntelliBridge EC10 and EC5 ID Module are solely limited to software changes.
software change for the EC10 is necessary the due to The communication between the new IntelliVue TcG10 Measurement Module and the patient monitors.
The modification for the EC5 ID is necessary to add specific device identifiers for the new IntelliVue TcG10 Measurement Module.
The fundamental communication and identification procedures are not changed.
{3}------------------------------------------------
The new IntelliVue TcG10 Measurement Module uses the identical functional measurement and calibration module as the predicate device. The interface software to external devices has been modified to the general RS232 protocol and an internal power supply has been added. With this software modification and this internal power supply the IntelliVue TcG10 Measurement Module is able to operate independent of the predicate host system but now interfaced to the IntelliVue monitor family. The new IntelliVue TcG10 Measurement Module uses the identical Radiometer tc Sensor 84 and the identical medical accessories, which come into contact with the patient skin, as the predicate device.
#### 7. Brief discussion of non-clinical performance tests
The determination of substantial equivalence of the performance of the cutaneous carbon dioxide (pcCO2) and oxygen (pcO2) measurement of the new IntelliVue TcG10 Measurement Module interfaced to the IntelliVue patient monitors modified via the modified EC10 and EC5 ID Modules was based on bench IntelliBridge performance testing in accordance to the special controls guidance document "Class II Special Controls Guidance Document: Cutaneous Carbon Dioxide (pcCO2) and Oxygen (pcO2) Monitors; Guidance for Industry and FDA", document issued on December 13, 2002. The bench performance testing has demonstrated that the system consisting of modified MP40, MP50, MP60, MP70, MP80, MP90, and MX800 IntelliVue Patient Monitors, modified IntelliBridge EC10 and EC5 ID Module and new IntelliVue TcG10 Measurement Module has within specifications, which have been substantial performed equivalent to those of the predicate device.
## 8. Brief discussion of clinical performance tests
Clinical data are not required in accordance to the special controls guidance document "Class II Special Controls Guidance Document: Cutaneous Carbon Dioxide (PcCO2) and Oxygen (PcO2) Monitors; Guidance for Industry and FDA", document issued on December 13, 2002, to IEC 60601-2-23:1999, and to IEC 60601-3-1:1996.
A clinical evaluation was conducted including literature review clinical acceptance studies with experienced nursing and and physican staff.
- 9. Conclusion nonclinical performance clinical evaluation tests demonstration that the system consisting of the modified The Philips MP40, MP50, MP60, MP70, MP80, MP90, and MX800 IntelliVue Patient Monitors, the modified IntelliBridge EC10 and EC5 ID Module, and the new IntelliVue TcG10 Measurement Module is as safe, as effective, and performs as the predicate devices has been demonstrated by non-clinical performance and clinical evaluation tests
{4}------------------------------------------------
#### 10. Summary and Conclusion
Verification and validation testing activities were conducted to establish the safety, performance, functionality, and reliability characteristics of the modified and news devices with respect to the predicate devices. vav testing were executed electromagnetic environmental testing, regression testing and testing from the hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate devices.
The results demonstrate that the modified Philips MP40, MP50, MP60, MP70, MP80, MP90, and MX800 IntelliVue Patient Monitors, the modified IntelliBridge EC10 and EC5 ID Module, and the new IntelliVue TcG10 Measurement Module are as safe and as effective and perform as the predicate devices.
The modified and the new devices are substantially equivalent in intended use and fundamental technological characteristics compared to the appropriate predicate devices. The modified and new devices introduce no new questions concerning the safety or efficacy and are, therefore, substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with a stylized bird in flight. The bird is composed of three parallel, curved lines that suggest movement and speed. To the left of the bird, there is a partial circular border with text that is difficult to read due to the image quality. The overall design is simple and modern, likely representing themes of freedom, travel, or progress.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring. MD 20993-0002
Dr. Jens-Peter Seher Sr. Regulatory Affairs Engineer Philips Medizin Systeme Boblingen GmbH. Hewlett-Packard - Str. 2 Boblingen Germany D 71034
SEP 3 0 2011
Re: K110474
Trade/Device Name: Philips MP40, MP50, MP70, MP70, MP80, MP90, and MX800 IntelliVue Patient Monitors, Philips IntelliBridge EC10 and EC5 HD Module, Philips IntelliVue TcG10 Measurement Module Regulation Number: 21 CFR 868.2480 Regulation Name: Cutaneous Carbon Dioxide (PcCO2) Monitor Regulatory Class: II Product Code: LKD, KLK, LPP Dated: September 23, 2011 Received: September 28, 2011
Dear Dr. Peter Seher:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications w for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Dr. Peter Seher
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem /default.html for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Nh for
Anthony D. Watson, B.S.. M.S.. M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): K10474
Device Names:
- Philips MP40, MP50, MP60, MP70, MP80, MP90, and MX800 ၊ IntelliVue Patient Monitors
- Philips IntelliBridge EC10 and EC5 ID Module ー
- ﺍ Philips IntelliVue TcG10 Measurement Module
- -
### Indications for Use:
## IntelliVue MP40, MP50, MP60, MP70, MP80, MP90 and MX800 and IntelliBridge EC10 and EC5 IC Module:
Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.
# IntelliVue TcG10 Measurement Module:
Indicated for use by health care professionals whenever there is a need to support the continuous and noninvasive monitoring of transcutaneous partial pressures of blood gas carbon dioxide and oxygen of patients not under gas anesthesia.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X AND/OR Over-<br>(21 C |
|-------------------------------------------------|-------------------------|
|-------------------------------------------------|-------------------------|
-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Schult
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K 110474
Page 14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.