EsCom100 is a light curing, radiopaque composite filling material for use in anterior and posterior restorations and is indicated for all carious classes.
Device Story
EsCom100 is a light-cured, radiopaque composite resin used by dental professionals for filling cavities in anterior and posterior teeth. The material is applied to the tooth cavity and cured using a dental light source. It is designed to provide high compressive strength and handling characteristics suitable for restorative procedures. The device is intended for clinical use by dentists to restore tooth structure affected by caries. It benefits patients by providing a durable, aesthetic, and radiopaque filling material for various cavity classes.
Clinical Evidence
Bench testing only. Safety and performance validations were performed, including biocompatibility testing, to ensure compliance with US regulations and ISO 4049 standards. No clinical data was required or provided.
Technological Characteristics
Light-cured, radiopaque composite resin. Available in 18 shades. Complies with ISO 4049. Mechanical properties include compressive strength, flexural strength, and thermal expansion coefficient. Standalone dental material.
Indications for Use
Indicated for patients requiring anterior or posterior dental restorations for all carious classes.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
K110428
SPIDENT
FEB 2 5 2 201 SPIDENT Co., Ltd.
Room 203 & 312. Korea Industrial Complex, 722. Kojan-dong. Namdong-ku. Incheon. Korea 405-821 Tel: +82(32)819-4570 Fax : +82(32)819-4572
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92(c).
Date: __ December. 22, 2010
1. Company and Correspondent making the submission:
| | Company |
|---------|-----------------------------------------------------------------------------------------------------|
| Name | SPIDENT Co., Ltd. |
| Address | Room 203 & 312, Korea Industrial<br>Complex, 722, Kojan-dong, Namdong-ku,<br>Incheon, Korea 405-821 |
| Phone | +82(32)819-4570 |
| Fax | +82(32)819-4572 |
| Contact | J. M. Ahn |
#### 2. Device:
Proprietary Name - EsCom100 Common Name - Dental composite and filling materials Classification Name - Tooth shade resin material
- 3. Predicate Device:
ESTELITE SIGMA QUICK, K080940
- 4. Classifications Names & Citations: EBF, 872.3690
- 5. Description:
The EsCom100 is light-cured composite resin which makes it ideal for small cavities in anterior and posterior teeth. It has the characteristics of good handling, high compressive strength, and radiopaque. It has 18 shades: A1. A2. A3. A3.5. A4, B1, B3, C2, C3, D2, D3, AO2,AO3, PA1, PA2, PA3,PA3.5 (yellow).
The EsCom100 is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. It is substantially equivalent in design, function and intended use to the predicate devices.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "SPIDENT" in a stylized font. The letters are black and bold, and they are enclosed in a square bracket-like design. The top and bottom lines of the bracket extend beyond the letters, creating a frame around the word.
# SPIDENT Co., Ltd.
Room 203 & 312. Korea Industrial Complex. 722. Kojan-dong. Namdong-ku, Incheon, Korea 405-821 Tel: +82(32)819-4570 Fax : +82(32)819-4572
#### 6. Indication for use:
EsCom100 is a light curing, radiopaque composite filling material for both anterior and posterior applications and is indicated for all carious classes.
### 7. Review:
The EsCom100 has the similar device characteristics as the predicate device, the ESTELITE SIGMA QUICK; intended use, material, chemical composition, design and use concept are similar.
The EsCom100 has the similar mechanical properties as the predicate device: compressive strength, flexural strength, flow thickness, wear, polymerization shrinkage and thermal expansion coefficience.
The EsCom100 has been subjected to extensive safety, performance, and product . validations prior to release. Safety tests including biocompatibility have been performed to ensure the devices comply with the US regulations and ISO 4049.
Based on the comparison of intended use and technical features, the EsCom100 is substantially equivalent to the predicate devices.
8. Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, FDA's "Guidance for the Preparation of Premarket notifications for Dental Composite" and based on the information provided in this premarket notification SPIDENT Co., Ltd. concludes that the EsCom100 is safe and effective and substantially equivalent to predicate devices as described herein.
- 9. SPIDENT Co., Ltd. will update and include in this summary any other information deemed reasonably necessary by the FDA.
END
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the department's logo, which consists of three stylized human figures. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the perimeter of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
Spident Company, Limited C/O Mr. Marc M. Mouser Responsible Third Party Official Underwriters Laboratories, Incorporated 2600 NW Lake Road Camas, Washington 98607-9526
FEB 2 5 231
Re: K110428
Trade/Device Name: EsCom 100 Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: January 31, 2011 Received: February 14, 2011
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your (1 Nirs), It may of bagfeel to adalof Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Mouser
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements mean mat 1 DA mas made a deters and regulations administered by other Federal agencies. of the Act of ally I ederal statues and reguirements, including, but not limited to: registration Touring (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting and ilsting (21 CFR 1 at 607), lace.ing (21 CFR 803); good manufacturing (21 CFR 803); good manufacturing (reporting of medical device related ad reror of the may (QS) regulation (21 CFR Part 820);
1 a for a medical as set forth in the quality systems (QS) regulation (21 CFR 820 practice requirements as occession in wet radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1115809.htm for a Illig.//WWW.Ida.gov//100utf Driveral Health's (CDRH's) Office of Compianter for Devices in the Center the Center 101 Devilos and ruallorogy is reference to premarket notification" prease note the regulation entired, "The orders and of adverse events under the MDR regulation (21 CFR Part 803), please go to
MDR regulation (21 CFR Fart 009), prease go wo
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.html for the CDRH's Ifth; www.laa.gom:re and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the You may other butter general informational and Consumer Assistance at its toll-free Division of Sinan Manufacts (101) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
San Rasino
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number K)d4af
Device Name: EsCom100
Indication for use:
EsCom100 is a light curing, radiopaque composite filling material for use in anterior and posterior restorations and is indicated for all carious classes.
AND/OR Prescription Usc______________________________________________________________________________________________________________________________________________________________ (Per 21CFR801 Subpart D)
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Per 21CFR807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| <img alt="Signature" src="signature.png"/> | |
|----------------------------------------------|--|
| (Division Sign-Off) | |
| Division of Anesthesiology, General Hospital | |
| Infection Control, Dental Devices | |
| SPIDENT CO., LTD. | 510(k) Number: K110428 |
|-------------------|------------------------|
|-------------------|------------------------|
00000020
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.