Porous coated components are intended for uncemented biological fixation.
Device Story
Taperloc® Complete Microplasty System is an uncemented modular hip prosthesis designed to replace the natural hip, femoral neck, and head. It combines design features of Taperloc® Complete full-length stems with the shortened design of existing Taperloc® Microplasty stems. Features include reduced neck angle, variable neck lengths, Type 1 taper geometry with neck flats, polished neck, standard/high offset versions, two distal profiles, and an updated insertion hole. The substrate is Ti-6Al-4V alloy with a proximal porous plasma-sprayed coating for biological fixation. Used by orthopedic surgeons in clinical settings for hip arthroplasty. The device provides structural support and joint articulation, aiming to restore hip function and mobility in patients with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including distal and proximal stem fatigue testing (5 million cycles at 67-517 lbs; 10 million cycles at 120 lbs) and range of motion analysis, all meeting acceptance criteria. Coating characterization was performed per FDA guidance for metallic plasma-sprayed coatings.
Technological Characteristics
Substrate: Ti-6Al-4V (ASTM F-136). Coating: Porous plasma-sprayed Ti-alloy (ASTM F1580). Design: Tapered wedge, Type 1 trunnion, reduced neck angle, two distal profiles. Uncemented biological fixation. No software or electronics.
Indications for Use
Indicated for patients with non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, non-union, femoral neck fracture, trochanteric fractures of the proximal femur with head involvement unmanageable by other techniques, and revision procedures where other treatments failed.
Regulatory Classification
Identification
A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).
{0}------------------------------------------------
#### t : 1 4 2 2 - - 1 8 -MANUFACTURING CORP.
| 510(k) Summary<br>SEP 30 2011 | |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Preparation Date: | September 29, 2011 |
| Applicant/Sponsor: | Biomet Manufacturing Corp.<br>56 East Bell Drive<br>P.O. Box 587<br>Warsaw, IN 46581-0587<br>Establishment Registration Number: 1825034 |
| Contact Person: | Becky Earl<br>Regulatory Specialist |
| Proprietary Name: | Taperloc® Complete Microplasty System |
| Common Name: | Uncemented porous modular hip prosthesis |
| Classification Name: | Hip joint metal/metal semi-constrained, with an uncemented acetabular<br>component, prosthesis (21 CFR §888.3330) |
| | LPH- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous<br>Uncemented (21 CFR 888.3358) |
| | KWA-Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular<br>Component) (21 CFR 888.3330) |
| | JDL- Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular<br>Component) (21 CFR 888.3320) |
| | LZO-Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer,<br>Cemented or Non-Porous, Uncemented (21 CFR 888.3353) |
| | KWZ-Prosthesis, Hip, Constrained, Cemented or Uncemented,<br>Metal/Polymer (21 CFR 888.3310) |
| | JDI- Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented (21<br>CFR 888.3350) |
| | KWY-Prosthesis, Hip, Hemi-Femoral, Metal/Polymer, Cemented or<br>Uncemented (21 CFR 888.3390) |
| | MEH-Prosthesis, Hip, Semi-constrained, Uncemented, Metal/Polymer,<br>Non-Porous, Calcium-Phosphate (21 CFR 888.3353) |
| | KWL-Hip Joint Femoral (Hemi-Hip) Metallic Cemented or Uncemented<br>Prosthesis (21 CFR 888.3360) |
| | OQG-Prosthesis, hip, semi-constrained, metal/polymer + additive, porous<br>uncemented (888.3358) |
Malling Address: P.O. Box 587 Warsaw, IN. 46581-0587 800.348.9500 Toll Free: Office: 574.267.6639 574-267-8137 Main Fax: www.biomet.com
Shipping Address: 56 E Bell Drive Warsaw, IN 46582
{1}------------------------------------------------
## Legally Marketed Devices To Which Substantial Equivalence Is Claimed:
Taperloc® Complete Stems-Biomet-(K101086)
Lateralized Taperioc® Microplasty Femoral Components -Biomet-(K062994)
Taperloc® Complete, Sizes 5mm and 6mm-Biomet-(K103755)
Fitmore® Hip Stem-Zimmer-(K071723)
## Device Description:
The TaperIoc® Complete Microplasty Stems are an update to the existing TaperIoc® Microplasty Stems and are designed to replace the patient's natural hip, femoral neck, and head, due to disease or accident. The Taperloc® Complete Microplasty Stems will combine the design features of the Taperioc® Complete full-length stems to the shortened design of the existing Taperloc® Microplasty Stems. These design features include: a reduced neck angle, shorter/longer neck lengths, reduced Type 1 taper geometry with neck flats, polished neck, standard and high offset versions, two distal profiles, and an updated insertion hole. Stem sizes ranges are within the ranges of legally marketed predicates. The substrate material is Ti-6AI-AV, ASTM F-136. The proximal intramedullary region is sprayed with porous plasma spray.
## Intended Use:
Porous coated components are intended for uncemented biological fixation.
#### Indications for Use:
- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis
- Correction of functional deformity 3.
- Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with 4. head involvement, unmanageable by other techniques.
- Revision procedures where other treatment or devices have failed. 5.
## Porous coated components are intended for uncemented biological fixation.
## Summary of Technologies:
The technological characteristics are the same as the predicates identified in the Legally Marketed Devices to which substantial equivalence is claimed. The subject microplasty stems are designed to update the existing Taperloc® Microplasty implants (K062994) to join the Taperloc® Complete family of stems as a microplasty option. The subject microplasty stems combine the design features of the Taperloc® Complete full-length stems (K101086 and K103755) and the shortened design of the Taperloc® Microplasty stems, both standard and high offset, or lateralized (K062994). Technological characteristics include a tapered wedge, a porous plasma sprayed region, a modified shortened trunnion (Type 1), a reduced neck angle, two distal profiles and an updated insertion/extraction hole, identical to the Taperloc® Complete full-length stem predicates (K101086 and K103755). Stem lengths are within the range of the legally marketed predicates.
## Non-Clinical Testing:
Nonclinical performance testing included distal and proximal stem fatigue testing of the worst-case stem, consistent with the "Guidance for Industry and FDA Staff Non-clinical Information for Femoral Stem Prostheses", ISO 7206-4:2002, ASTM F2068-03 and ISO 7206-6:1992, as well as a Range of Motion analysis consistent with ISO 21535:2009. All 6 stems passed distal stem fatigue for 5 million cycles at 67 lbs to 517lbs,
{2}------------------------------------------------
meeting the acceptance criteria. All 6 stems passed 10 million cycles at 120lbs in proximal fatigue test, meeting the acceptance criteria. The minimum Range of Motion passed its simulation, meeting the acceptance criteria.
Additional Ti-alloy (ASTM F1580) plasma sprayed metallic coating characterization data was performed for clarification per the "Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement" and per the "Guidance for Industry on the Testing of Metallic Plasma Sprayed Coatings on Orthopedic Implants to Support Reconsideration of Postmarket Surveillance Requirements." The PPS plasma sprayed coating is the same as the PPS coating previously cleared and meets the regulatory definition of porous coating for the hip construct per 21 CFR 888.3358.
## Clinical Testing:
None provided as a basis for substantial equivalence.
The results of mechanical testing indicated the devices performed within the intended use, did not raise any new safety and efficacy issues and were found to be substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white circular logo. The logo features a stylized depiction of three birds in flight, arranged vertically. The birds are simple, with curved lines suggesting their wings and bodies. Encircling the birds is text that follows the curve of the circle. The text is small and somewhat difficult to read, but it appears to be the name of an organization or agency.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -W056-G609 Silver Spring, MD 20993-0002
Biomet Manufacturing Corp. % Becky Earl Regulatory Specialist P.O. Box 587 Warsaw, IN 46581
cca 3 0 2711
Re: K110400
Trade/Device Name: Taperloc® Complete Microplasty System Regulation Number: 21 CFR 888.3330 Regulation Name: Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis Regulatory Class: Class III Product Code: KWA, LPH, JDL, LZO, KWZ, JDI. KWY, MEH, KWL, QOG Dated: September 20, 2011 Received: September 21. 2011
Dear Ms. Earl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
if your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page - 2 -- Ms. Becky Earl
comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practive requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
ff you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Erin S. Keith
Cr Mark N. Melkerson Director Division of Surgical. Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): _ K | 104000000001/1 )
Device Name: Taperloc® Complete Microplasty
Indications For Use:
- 1. Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis.
- 2. Rheumatoid arthritis
- 3. Correction of functional deformity
- 4. Treatment of non-union, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques.
- 5. Revision procedures where other treatment or devices have failed.
Porous coated components are intended for uncemented biological fixation.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
ihankhens for mxm
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number k110400
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.