# CELOX RAPID GAUZE (K110386)

_Medtrade Products , Ltd. · QSY · May 10, 2011 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K110386

## Device Facts

- **Applicant:** Medtrade Products , Ltd.
- **Product Code:** [QSY](/productcode/QSY.md)
- **Decision Date:** May 10, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

CELOX Rapid Gauze (Rx) is indicated for temporary external use to control moderate to severe bleeding. CELOX Rapid Gauze (OTC) is indicated for temporary external use to control bleeding of lacerations, minor cuts, and abrasions.

## Device Story

CELOX Rapid Gauze consists of chitosan granules heat-bonded to a gauze substrate; functions as a hemostatic delivery system. Upon contact with blood or wound exudate, granules absorb water to form a robust gel plug; creates physical barrier/seal over wound. Used in clinical or emergency settings; applied by clinicians or lay users with manual pressure. Benefits include rapid hemorrhage control, including in hypothermic conditions, without generating heat. Device is sterile, single-use.

## Clinical Evidence

No clinical data. Evidence consists of bench testing (biocompatibility per ISO 10993/FDA G95-1, in vitro testing for packing and adhesion) and a swine model of lethal arterial extremity hemorrhage to demonstrate rapid bleeding control.

## Technological Characteristics

Material: Chitosan. Form factor: Gauze strip with heat-bonded granules. Principle: Hemostasis via water absorption and gel plug formation. Sterilization: Irradiation. Single-use. Biocompatibility: ISO 10993 compliant.

## Regulatory Identification

To temporarily control bleeding and cover external wounds.

## Predicate Devices

- CELOX Hemostatic Granules on Sheet ([K080097](/device/K080097.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left and features a stylized caduceus. The FDA logo is on the right and features the agency's acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.

July 28, 2023

Medtrade Products Limited c/o Jonathan Ranfield Quality and Regulatory Director Electra House, Crewe Business Park Crewe, Cheshire CW1 6GL United Kingdom

Re: K110386 Trade/Device Name: CELOX Rapid Gauze Regulatory Class: Unclassified Product Code: QSY

Dear Jonathan Ranfield:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated May 10, 2011. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code QSY.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov.

Sincerely,

# Julie A. Morabito -S

Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle.

: ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ

:

: "

- # -

f

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Medtrade Products Limited % Mr. Jonathan Ranfield Quality and Regulatory Director Electra House, Crewe Business Park Crewe, Cheshire CWI 6GL United Kingdom ·

MAY 1 0 201

Re: K110386

Trade/Device Name: CÊLOX Rapid Gauze Product Code: FRO Dated: February 3, 2011 Received: February 10, 2011

Dear Mr. Ranfield:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{2}------------------------------------------------

Page 2 - Mr. Jonathan Ranfield

Enclosure

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.lda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Hy B. n h

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

# Indications for Use

KIJO386 510(k) Number (if known):

Device Name: CELOX Rapid Gauze

Indications For Prescription Use:

CELOX Rapid Gauze is indicated for temporary external use to control moderate to severe bleeding.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Krone for MXM

(Division Sig Division of ! and Restorati

510(k) Number K110386

Page 1 of 1

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K

Device Name: CELOX Rapid Gauze

Indications For OTC (Over The Counter) Use:

CELOX Rapid Gauze is indicated for temporary external use to control bleeding of lacerations, minor cuts, and abrasions.

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use × (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

K. B. Dake for nxn.
(Division Sign-Off)

AND/OR

Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

510(k) Number k110386

7

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the text "K110386 Page 1 of 3" in the upper left corner. To the right of the text is the word "Medtrade" in a bold, sans-serif font. A curved line extends from below the word "Medtrade" to the right, creating a logo-like effect.

#### SECTION 2:

510(k) SUMMARY

Innovative Medical Products

2.1 Sponsor

Medtrade Products Limited Electra House Crewe Business Park Crewe Cheshire CW1 6GL nk

Telephone: +44 1270 500019

Fax: +44 1270 500045

Registration Number: 9614493

Contact Person: Jonathan Ranfield Quality & Regulatory Director

## 2.2 Date Summary was Prepared

February 3, 2011.

#### 2.3 Device Information

Proprietary Name: CELOX Rapid Z-Fold Gauze

Common Name: CELOX Rapid

Classification Name: Dressing, Unclassified

## 5.4 Predicate Device

Medtrade Products Limited: CELOX Hemostatic Granules on Sheet (K080097)

#### 5.5 Device Description

Components – CELOX Rapid is composed of chitosan

Mechanism of Action - CELOX Gauze achieves its principle intended action (hemostasis) by acting as a delivery system for the Celox Granules creating a physical barrier or seal to stop the flow of blood. When in contact with a wound and upon contact with blood or exudate, in combination with manual pressure to the wound, the CLO. C Granules heat bonded on to the CELOX Rapid quickly form a strong seai that completely covers the wound.

{6}------------------------------------------------

K 110386
page 2 of 3

#### 5.6 Intended Use

CELOX Rapid Gauze (Rx) is indicated for temporary external use to control moderate to severe bleeding.

CELOX Rapid Gauze (OTC) is indicated for temporary external use to control bleeding of lacerations, minor cuts, and abrasions.

#### 5.7 Substantial Equivalence

CELOX Rapid Gauze is identical in manufacturing route, packaging and irradiation. as CELOX Hemostatic Granules on Sheet K080097 cleared on July 9, 2008.

CELOX Rapid Gauze has substantially equivalent indications to CELOX Hemostatic Granules on Sheet K080097 in that they are indicated for temporary external use to control moderate to severe bleeding (Rx) and temporary external use to control bleeding of lacerations, minor cuts, and abrasions (OTC).

CELOX Rapid Gauze uses the same safe and effective technology as CELOX Hemostatic Granules on Sheet K080097. The subject device and predicate devices are made from materials which have demonstrated satisfactory biocompatibility, are sterile, single use devices.

## 5.7 Performance Testing

Biocompatibility Testing:

Biocompatibility has been reviewed in accordance with ISO10993 and FDA Blue Book memo G95-1.

To support:

To Control Moderate To Severe Bleeding, Works in Hypothermic Conditions, No Heat Generated in Use, Reduces Blood Loss, previously provided and cleared under (K080097).

Image /page/6/Figure/13 description: The image contains a collection of characters from different languages. There are characters from Gujarati, Arabic, and Telugu. The characters are arranged vertically in the image. The image is black and white.

In Vitro Testing to support:

- Rapid Packing
- High Volume Gauze Strip
- Promotes Rapid Wound Adhesion

Animal Studies

Animal Study 1 - The efficacy of Celox Rapid Gauze in a femoral artery wound model. To support

- Stops bleeding fast
- Rapidly controls bleeding

Clinical Study - Not Applicable.

#### 5.8 Conclusion

CELOX Rapid Gauze induces hemostasis by the absorption of water in the blood to form a robust gel plug the same as CELOX Hemostatic Granules on Sheet (K080097) predicate device.

CELOX Rapid Gauze has been shown in testing to provide rapid haemorrhage control in a swine model of lethal arterial extremity.

{7}------------------------------------------------

K110886
page 3 of 3

Medtrade Products believes that, as a result of the biocompatibility testing in vitro testing, and non-clinical animal testing, CELOX Rapid Gauze is safe and effective as an id the control of temporary external bleeding associated with moderate to severe bleeding and the control of minor external bleeding of laceration, minor cuts and abrasions.

---

**Source:** [https://fda-staging.innolitics.com/device/K110386](https://fda-staging.innolitics.com/device/K110386)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).