INDISTRIP GREEN

K110369 · Indilab, Inc. · JOJ · May 6, 2011 · General Hospital

Device Facts

Record IDK110369
Device NameINDISTRIP GREEN
ApplicantIndilab, Inc.
Product CodeJOJ · General Hospital
Decision DateMay 6, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.2800
Device ClassClass 2

Indications for Use

For use as an internal or external chemical indicator to monitor exposure to steam sterilization conditions for: 121°C (250°F) @ 30 minutes in gravity steam sterilizer cycles or @ 134℃ (273ºF) for 4 minutes in gravity or vacuum assisted steam sterilizer cycles.

Device Story

Indistrip Green is a chemical process indicator for steam sterilization; consists of metal-free organic indicator ink printed on white vellum. Used by healthcare personnel to monitor sterilization packs; placed internally or externally on packs. Upon exposure to steam sterilization conditions (121°C for 30 min or 134°C for 4 min), ink undergoes irreversible color change from yellow to black. Absence of black color indicates inadequate exposure, signaling need for reprocessing. Provides visual confirmation of sterilization process completion; benefits patient safety by ensuring medical devices are properly sterilized before use.

Clinical Evidence

Bench testing only. Device performance evaluated against AAMI ANSI ISO 11140-1:2005 requirements for Class 1 process indicators. Testing confirmed color change from yellow to black under specified steam sterilization parameters (121°C and 134°C) and verified stability of the indicator color post-processing. No clinical data required for this device type.

Technological Characteristics

Metal-free organic indicator ink printed on white vellum. Dimensions: 9/16" x 4". Sensing principle: chemical color change reaction triggered by steam sterilization conditions. Connectivity: None. Sterilization: N/A (device is an indicator).

Indications for Use

Indicated for use as an internal or external chemical indicator to monitor steam sterilization exposure in gravity or vacuum-assisted cycles at specified temperatures and durations.

Regulatory Classification

Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary Statement (as required by 807.92 (c)) Date of Submission: Submitter: January 14, 2011 Robert Gavrick President Indilab, Inc. 10367 Franklin Avenue Franklin Park, IL 60131 Phone: (847) 928-1050 Fax: (847) 928-1052 MAY - 6 2011 Establishment Registration NO: 1450962 Official Contact: Mark Espenscheid Business Unit Director / Quality Indilab, Inc. 10367 Franklin Avenue Franklin Park, IL 60131 Phone: (847) 928-1050 Fax: (847) 928-1052 Trade Name: Indistrip Green Common Name: Process Indicator for Steam Sterilization Device Classification: Device indicator, physical/chemical sterilization process Regulation Description Sterilization process indicator. Regulation Medical Specialty General Hospital Review Panel General Hospital Product Code JOJ Submission Type 510(k) Regulation Number 880.2800 Device Class 2 Total Product Life Cycle (TPLC) TPLC Product Code Report GMP Exempt? No Predicate Device: Indistrip 400S; 510(k) No. K875193 Section 5 - 510(k) Summary Statement Page 2 of 5 Revision Date 04/28/2011 {1}------------------------------------------------ #### Indications for Use: For use as an internal or external chemical indicator to monitor exposure to steam sterilization conditions for: 121°C (250°F) @ 30 minutes in gravity steam sterilizer cycles or @ 134℃ (273ºF) for 4 minutes in gravity or vacuum assisted steam sterilizer cycles. Consensus Standards: Utilized as part of our Statement of Conformity Recognition Number: 14-195 Product Area: Sterility Title of Standard: Sterilization of health care products - Chemical indicators - Part 1: General requirements, 2ed (Revision of ANSI/AAMI ST60:1996) Reference Number: 11140-1:2005 Publication Date: 09/08/2009 Standards Development Organization: AAMI ANSI ISO #### Device Description: The INDISTRIP Green Process Indicator for sterilization consists of a metal free indicator ink printed onto white vellum. The indicator changes color from yellow to black when exposed to steam sterilization conditions. Unprocessed indiryclife to stable for atleast 6 months when stored at 30℃ or less. The postprocessed indicator color is stable for at least 6 months after exposurer to steam sterilization conditions for: 121°C (250°F) @ 30 minutes in gravity steam sterilizer cycles or @ 134°C (273°F) for 4 minutes in gravity or vacuum assisted steam sterilize cycles of The indicator does not respond to dry heat or ethylene oxide. The process requirements, in comparison to the predicate device are presented in Table 5-11 # Section 5 – 510(k) Summary Statement Page 3 of 5 Revision Date 04/28/2011 {2}------------------------------------------------ # Comparison with Predicate Device (Table 5-TI): | Item<br>Name | Predicate Device | Subject Device | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indistrip (K875193) | Indistrip Green | | | Intended Use | For use as an internal or<br>external chemical indicator<br>to monitor exposure to<br>steam sterilization<br>conditions for 121-141°C<br>(250 - 285°F) gravity and<br>vacuum assisted steam<br>sterilizer cycles. | For use as an internal or<br>external chemical indicator<br>to monitor exposure to<br>steam sterilization<br>conditions for: 121°C<br>(250°F) @ 30 minutes in<br>gravity steam sterilizer<br>cycles or @ 134°C (273°F)<br>for 4 minutes in gravity or<br>vacuum assisted steam<br>sterilizer cycles. | | Material | Lead Based Ink Printed<br>onto White Vellum | Organic Based Ink, Metal<br>Free, Printed onto White<br>Vellum | | Color Change | Off-White/tan to Black | Yellow to Black | | Sterilization Method | Steam | Steam | | Color Change Timing | 8 minutes 250°F<br>3 minutes 273°F | 8 minutes 250°F<br>3 minutes 273°F | | Size | 9/16" x 4" | 9/16" x 4" | | Precautions | Do not use the Indistrip to<br>monitor dry heat, ethylene<br>oxide, or other low<br>temperature sterilization<br>processes | Do not use the Indistrip<br>Green to monitor dry heat,<br>ethylene oxide, or other<br>low temperature<br>sterilization processes | | General Instructions | Utilize on and/or in each<br>pack to be steam<br>sterilized. Process<br>according to established<br>procedures. After<br>processing remove the<br>indicator and observe the<br>color change. If the<br>indicator bar is not black,<br>inadequate exposure is<br>indicated. Return for<br>processing. | Utilize on and/or in each<br>pack to be steam<br>sterilized. Process<br>according to established<br>procedures. After<br>processing remove the<br>indicator and observe the<br>color change. If the<br>indicator bar is not black,<br>inadequate exposure is<br>indicated. Return for<br>processing. | Section 5 – 510(k) Summary Statement Page 4 of 5 Revision Date 04/28/2011 · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · :**,* **** {3}------------------------------------------------ ## Performance Data: | Test<br>Environment | Test Time | Test Temperature | No Change or a<br>change that is<br>markedly different<br>from the visible color<br>change as specified<br>by the manufacturer | Visible color<br>change as<br>specified by the<br>manufacturer | Indistrip<br>Results | Indistrip<br>Green<br>Results | |---------------------|----------------|------------------|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|----------------------|-------------------------------| | Saturated<br>Steam | 3.0 min ± 5 s | 121 °C (+3/0 °C) | Acceptable result | Unacceptable<br>result | Pass | Pass | | Saturated<br>Steam | 10.0 min ± 5 s | 121 °C (+3/0 °C) | Unacceptable result | Acceptable result | Pass | Pass | | Saturated<br>Steam | 0.5 min ± 5 s | 134 °C (+3/0 °C) | Acceptable result | Unacceptable<br>result | Pass | Pass | | Saturated<br>Steam | 2.0 min ± 5 s | 134 °C (+3/0 °C) | Unacceptable result | Acceptable result | Pass | Pass | | Dry Heat | 30 min ± 1 s | 140 °C (+2/0 °C) | Acceptable result | Unacceptable<br>result | Pass | Pass | # Meats the requirements defined in 11140-1:2005 (AAMI ANSI ISO) Class 1 Table 5-TII Test Performance Characteristics Class 1 Process Indicators ### Conclusion: The Indistrip Green, Heavy Metal free Steam Sterilization indicator is in accordance with 21 CFR 807 has the same intended use, provides substantially equivalent process color change timing, and color change stability as the Indistrip steam sterilization predicate device 510(k) K875193. The main difference is the indistrip the Indistrin Groon, Product Cards IND (1933) The main difference is that the Indistrip Green, Product Code INDA00G indicator ink is developed with a metal free organic indicator as compared to the standard lead based steam indicator utilized in the Indistrip 400S steam indicator strip. The two devices are substantially equivalent. Section 5 – 510(k) Summary Statement Page 5 of 5 Revision Date 04/28/2011 {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAY - 6 2011 Mr. Mark Espenscheid Director of Quality Indilab, Incorporated 10367 Franklin Avenue Franklin Park, Illinois 60131 Re: K110369 Trade/Device Name: Indistrip Green Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: January 15, 2011 Received: February 8, 2011 Dear Mr. Espenscheid: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2- Mr. Espenscheid Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (2) CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, dams Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Section 4 - Indications for Use Statement. 510(k) Number (if known): N/A Device Name: Indistrip Green Indications for Use: Indications for Use: For use as an internal or external chemical indicator to monitor exposure to steam sterilization conditions for: 121℃ (250°F) @ 30 minutes in gravity steam sterilizer cycles or @ 134℃ (273ºF) for 4 minutes in gravity or vacuum assisted steam sterilizer cycles. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANTOHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Elizabetk B. Clamin-Will Page 1 of _1_ (Division Sign-Off) Division of Anesthesiology, General Hospital intection Control, Dental Devices :10(k) Number: K110369 Section 4 - Indications for Use Statement Page 2 of 2 Revision 04-28-2011
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...