The Comprehensive® Proximal Humeral Plating System is indicated for fractures, fracture dislocations, osteotomies and non-unions of the proximal humerus.
Device Story
Comprehensive® Proximal Humeral Plating System is an orthopedic bone fixation device consisting of anatomic, precontoured stainless steel plates and various screw types (non-locking, locking, variable angle). Plate sizing and contouring were determined using Biomet's IntelliFIT Technology, which utilized contour analysis of cadaveric specimens to map bone patterns; this software was used only to define standard plate sizes, not for patient-specific customization. The system is intended for surgical use by orthopedic surgeons to stabilize proximal humeral fractures, dislocations, osteotomies, and non-unions. The device provides mechanical fixation to support bone healing. It is MR Conditional per ASTM F2503-08.
Clinical Evidence
No clinical data provided. Substantial equivalence determined via engineering analysis and bench testing comparing plate strength to predicate devices.
Indicated for patients with fractures, fracture dislocations, osteotomies, and non-unions of the proximal humerus.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
EBI OptiLock® Upper Extremity Plating System (K062494)
{0}------------------------------------------------
<110320
Image /page/0/Picture/1 description: The image shows the logo for Biomet Manufacturing Corp. The word "BIOMET" is displayed in a stylized, blocky font above the words "MANUFACTURING CORP." The text is black and the background is white. The logo appears to be a scan or photocopy, as the edges of the letters are slightly pixelated.
JUL 2 1 2011
## 510(k) SUMMARY
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92
| Submitter Information | |
|------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Biomet Manufacturing Corp. |
| Address | 56 East Bell Drive<br>Warsaw, IN 46581-0857 |
| Phone number | (574) 267-6639 |
| Fax number | (574) 371-1027 |
| Establishment<br>Registration Number | 1825034 |
| Name of contact person | Patricia Sandborn Beres<br>Senior Regulatory Specialist |
| Date prepared | January 26, 2011 |
| Name of device | |
| Trade or proprietary<br>name | Comprehensive® Proximal Humeral Plating System |
| Common or usual<br>name | plate, fixation, bone screw, fixation, bone |
| Classification name | Single/multiple component metallic bone fixation appliances and accessories Smooth or threaded metallic bone fixation fastener |
| Classification panel | Orthopedics |
| Regulation | 21 CFR 888.3030 21 CFR 888.3040 |
| Product Code(s) | HRS HWC |
| Legally marketed device(s)<br>to which equivalence is<br>claimed | K062494 - EBI OptiLock® Upper Extremity Plating System<br>K082625 - Synthes (USA) 3.5mm LCP Periarticular Proximal Humeral Plates |
| Reason for 510(k)<br>submission | New device |
| Device description | The Comprehensive® Proximal Humeral Plating System is<br>comprised of anatomic plates in seven lengths and non-locking,<br>locking and variable angle screws in multiple lengths. Plate<br>sizing and contouring was developed through the use of<br>Biomet's IntelliFIT Technology which uses contour analysis to<br>map patterns in complex bone on cadaveric specimens to<br>determine plate sizing. (Note, the software was used to<br>determine a set of pre-defined plate sizes and is not used to<br>create individual, patient matched plates.) |
| Intended use of the device | Bone fixation |
Malling Address:
P.O. Box 5811681
P.Oresaw, M.G.G.G.G.G.G.G.G.G.G.C.C.C.
Translaw, B000.3 AM.C.S.A.G.C.C.C.
Collice: Sp. 2007.37 A.C.T.C.T.C.
Main Fax: 574.767.8137 ww.blomet.com
Shipping Address:
56 East Bell Drive
Warsaw, IN 46582
\$\cdot l*{D} f*{2}\$
{1}------------------------------------------------
K 110320
510(k) Summary
Comprehensive® Proximal Humeral Plating System Page 2 of 2
く
:
| Indications for use | The Comprehensive® Proximal Humeral Plating System is<br>indicated for fractures, fracture dislocations, osteotomies and<br>non-unions of the proximal humerus. | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|-------------------|
| Summary of the technological characteristics of the device compared to the predicate | | | |
| Characteristic | New Device | Predicate Device* | |
| Plate Design | Anatomic, precontoured | K062494, K082625 | |
| Plate Material | Stainless Steel<br>ASTM F138 & ASTM F139 | K062494, K082625 | |
| Plate Dimensions | Length: 75-255mm<br>Thickness: 3-4mm | K062494, K082625 | |
| Screw Design | Non-Locking, Locking and Variable Angle | K062494, K082625 | |
| Screw Material | Stainless Steel<br>ASTM F138, ASTM F139, ASTM F-2229 | K062494, K082625 | |
| Screw Dimensions | Diameter: 3.5mm<br>Length: 10-40mm or 10-60mm | K062494, K082625 | |
| PERFORMANCE DATA | | | |
| SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF<br>SUBSTANTIAL EQUIVALENCE | | | |
| Performance Test Summary-New Device | | | |
| Characteristic | Standard/Test/FDA Guidance | Results Summary | |
| Plate Strength | Engineering Analysis | Equivalent to predicate | |
| Comparative Performance Information Summary | | | |
| Characteristic | Requirement | New Device | Predicate Device* |
| Plate Strength | Meet or exceed<br>predicate | Meet | K062494 |
| SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL<br>EQUIVALENCE AND/OR OF CLINICAL INFORMATION | | | |
| Clinical Performance Data/Information: None | | | |
| MAGNETIC RESONANCE (MR) ENVIROMENT | | | |
| Biomet® has performed non-clinical Magnetic Resonance Imaging (MRI) studies on Plating<br>Systems manufactured of 316L Stainless Steel per ASTM F138. These Plating Systems are<br>determined to be MR Conditional in accordance to ASTM F2503-08 Standard Practice for<br>Marking Devices and Other Items for Safety in the Magnetic Resonance Environment. MR<br>Conditional refers to an item that has been demonstrated to pose no known hazards in a<br>specified MR environment with specified conditions of use. | | | |
| CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA | | | |
| No mechanical or clinical testing was necessary for a determination of substantial equivalence. The<br>results of engineering analysis indicated the devices performed within the intended use, did not<br>raise any new safety and efficacy issues and were found to be substantially equivalent to the<br>predicate devices.<br>*Any statement made in conjunction with this submission regarding and/or a determination of substantial equivalence to any other | | | |
| product is intended only to relate to whether the product can be lawfully marketed without pre-market approval or redassification and is | | | |
not intended to be Interpreted as an admission or any other type of evident infingenent litigation. [Establishment
Registration and Premarket Notification Procedures, Final R
p. 2 of 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of what appears to be an abstract design of an object with three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Biomet Manufacturing Corp. % Ms. Patricia Beres Senior Regulatory Specialist 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581-0587
JUL 2 1 2011
Re: K110320
Trade/Device Name: Comprehensive Proximal Humeral Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: June 27, 2011 Received: June 28, 2011
Dear Ms. Beres:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Patricia Beres
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerso Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known):_ k 1) ωろ乙ム
Device Name: Comprehensive® Proximal Humeral Plating System
Indications For Use: The Comprehensive® Proximal Humeral Plating System is indicated for fractures, fracture dislocations, osteotomies and non-unions of the proximal humerus.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for M. Melkerm
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K110320
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.