K110293 · Respironics New Jersey, Inc. · NVP · Aug 9, 2011 · Anesthesiology
Device Facts
Record ID
K110293
Device Name
OPTICHAMBER DIAMOND VALVED HOLDING CHAMBER
Applicant
Respironics New Jersey, Inc.
Product Code
NVP · Anesthesiology
Decision Date
Aug 9, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The OptiChamber Diamond Valved Holding Chamber device is intended to be used by patients who are under the care or treatment of a physician or licensed healthcare professional. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers. The intended environments for use include the home, hospitals and clinics. For Single Patient Use Recommended Patient Population: OptiChamber Diamond: Age 5 and up OptiChamber Diamond with Small LiteTouch mask: 0 to 18 months OptiChamber Diamond with Medium LiteTouch mask: 1 to 5 years OptiChamber Diamond with Large LiteTouch mask: 5 years +
Device Story
OptiChamber Diamond is a valved holding chamber (VHC) used with pressurized metered dose inhalers (pMDIs) to assist patients unable to coordinate pMDI actuation with inhalation. Device captures aerosol cloud from pMDI; larger particles removed via impaction on chamber walls and gravity-driven sedimentation. Components include mouthpiece, chamber, adapter with inhalation flow alert, exhaust valve, inhalation valve, valve retaining ring, and cap. Constructed from antistatic plastic. Used in homes, hospitals, and clinics by patients under physician care. Provides visual/mechanical assistance for drug delivery; improves medication deposition by slowing aerosol velocity and allowing time for inhalation. No active electronic components; flow alert is mechanical.
Clinical Evidence
No clinical data. Substantial equivalence established via in vitro bench testing using a Next Generation Impactor (NGI) at 15 and 30 LPM to simulate pediatric and adult breathing patterns. Three inhaler formulations (albuterol, fluticasone propionate, ipratropium bromide) were tested. Performance metrics included Total Emitted Dose (ED), Fine Particle Dose (FPD), Fine Particle Fraction (FPF), Mass Medium Aerodynamic Diameter (MMAD), and Geometric Standard Deviation (GSD).
Technological Characteristics
Antistatic plastic construction; valved holding chamber design; mechanical inhalation flow alert; includes mouthpiece and optional masks (Small, Medium, Large LiteTouch). Operates via passive aerosol impaction and sedimentation. No external energy source. Non-sterile.
Indications for Use
Indicated for patients under physician care requiring administration of aerosolized medication from pressurized metered dose inhalers (pMDIs). Population: patients aged 0 to 18 months (with small mask), 1 to 5 years (with medium mask), and 5 years and older (with large mask or mouthpiece).
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
Predicate Devices
AeroChamber Plus Z-Stat Valved Holding Chamber (K052332)
{0}------------------------------------------------
### Section 5
# 510(K) SUMMARY
AUG - 9 2011
ﮯ
Prepared: January 25, 2011
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.
The assigned 510(k) number is: K110293.
#### Submitter's Identification: 1.
Respironics New Jersey, Inc. 5 Wood Hollow Road Parsippany, New Jersey 07054 Phone: 973-581-6008 Fax: 973-599-5650 Contact Name: Lauren Ziegler e-mail: lauren.ziegler@philips.com
Establishment Registration Number: 2243193
#### Official Correspondent:
Lauren Ziegler Director, QA, RA, and Clinical Affairs 5 Wood Hollow Road Parsippany, New Jersey 07054 Phone: 973-581-6008 Fax: 973-599-5650 e-mail: Lauren.ziegler@philips.com
#### Name of the Device: 2.
OptiChamber Diamond Valved Holding Chamber. Common Name or Classification Name (21 CFR Part 807.87) of Device: Valved Holding Chambers, 21 CFR Part 868.5630. Device Classification Code: NVP
{1}------------------------------------------------
#### Predicate Device Information: 3.
Identification of legally marketed device which we claim substantial equivalence to: AeroChamber Plus Z-Stat Valved Holding Chamber K052332 Trudell Medical International. 725 Third Street London, Ontario Canada, N5V 5G4
#### Device Description: 4.
The OptiChamber Diamond Valved Holding Chamber (VHC) is a Class II device. It is intended to be used in combination with most pressurized Metered Dose Inhalers (pMDIs) to assist in respiratory drug delivery.
The OptiChamber Diamond Valved Holding Chamber is a device utilizing the same operating principles as the AeroChamber Plus Z-Stat Valved Holding Chamber (K052332). Both the AeroChamber Plus Z-Stat and the OptiChamber Diamond are available with and without mask.
The valved holding chamber (VHC) is designed to assist patients who cannot correctly coordinate actuation of the pressurized metered dose inhaler (pMDIs) with inhalation. The VHC works by 'holding' the aerosol cloud emitted from the pMDI inside the chamber so that the larger aerosol particles are removed from the aerosol cloud by impaction into the chamber walls and sedimentation under the influence of gravity.
The OptiChamber Diamond Valved Holding Chamber is made of antistatic plastic materials alleviating any need to wash prior to first use. OptiChamber Diamond is comprised of: the mouthpiece, the chamber, the adapter (end cap) with inhalation flow alert, the exhaust valve, the inhalation valve, the valve retaining ring, and the cap.
## 5. Intended Use:
The OptiChamber Diamond Valved Holding Chamber device is intended to be used by patients who are under the care or treatment of a physician or licensed healthcare professional. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers. The intended environments for use include the home, hospitals and clinics.
For Single Patient Use
Recommended Patient Population:
OptiChamber Diamond: Age 5 and up
OptiChamber Diamond with Small LiteTouch mask: 0 to 18 months
{2}------------------------------------------------
OptiChamber Diamond with Medium LiteTouch mask: 1 to 5 years OptiChamber Diamond with Large LiteTouch mask: 5 years +
### 6. Comparison to Predicate Devices:
The OptiChamber Diamond Valve Holding Chamber and the predicate device (AeroChamber Plus Z-Stat Valved Holding Chamber (K052332)) are indicated for the same intended use and their performance characteristics are substantially equivalent. Both the AeroChamber Plus Z-Stat and the OptiChamber Diamond are available with and without mask.
The LiteTouch medium mask was approved under K100285.
# 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence
A pressurized Metered Dose Inhaler (pMDI) alone and a pMDI with the OptiChamber Diamond valved holding chamber (VHC) were tested utilizing a Next Generation Impactor (NGI) with a custom flat plate adapter and operated at 15 and 30 liters per minute (LPM). The 15 liters per minute was chosen to simulate pediatric breathing pattern and the 30 liters per minute was chosen to simulate adult breathing pattern (reference: Breathing patterns, K. Nikander, J. Denyer, Eur Respir Rev 2000; 10: 76, 576-579). The pMDI was actuated 10 times into each device. Deposition of drug on various locations of the NGI was qualified using HPLC assays.
MMAD, GSD, FPD, and FPF results were determined using CITDAS (reference: Performance Verification of Copley Inhaler Testing Data Analysis Software (CITDAS), Respiratory Drug Delivery Europe 2009 -Krebs et al.) software version 3.10. Fine particle dose (FPD) was presented as 3 categories: greater than 4.7um, less than or equal to 4.7um and less than or equal to 1.0um. Fine particle fraction (FPF) is defined as the fraction of dose entering the impactor equal to or less than 4.7um.
Three inhaler formulations (albuterol, fluticasone propionate and ipratropium bromide) were evaluated.
{3}------------------------------------------------
### Definitions:
- ED = Total Emitted Dose: Total amount of drug from VHC + USP Throat + NGI components + Filter
- FPD = Fine Particle Dose: amount of drug entering the impactor with an acrodynamic diameter between specific limits.
- FPF = Fine Particle Fraction: a fraction of dose entering the impactor with an aerodynamic diameter of ≤ 4.7um
- MMAD = Mass Medium Aerodynamic Diameter: The geometric mean aerodynamic diameter. Fifty per cent of particles by weight will be smaller than the MMAD, 50% will be larger.
- GSD = Geometric Standard Deviation: A measure of dispersion in a lognormal distribution always greater than or equal to 1.0.
## 8. Discussion of Clinical Tests Performed:
Not Applicable.
### 9. Conclusions:
We have demonstrated that the OptiChamber Diamond Valved Holding Chamber is substantially equivalent to the predicate device, the AeroChamber Plus Z-stat Valved Holding Chamber, through in Vitro Testing. The two devices are also substantially equivalent in terms of their features, operating principle, and indications for use.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines representing the wings and body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Lauren Ziegler Director, Quality Assurance, Regulatory Affairs Respironics New Jersey, Incorporated 5 Wood Hollow Road Parsippany, New Jersey 07054
- 9 2011 માં ઉ
Re: K110293
Trade/Device Name: OptiChamber Diamond Valved Holding Chamber Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: NVP Dated: July 21, 2011 Received: July 25, 2011
Dear Ms. Ziegler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Ms. Ziegler
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices ucm 1 1 5809 htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
hn for
Anthony D. Watson. B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
510(k) Number (if known): K110293
Device Name: OptiChamber Diamond Valved Holding Chamber
Indications for Use:
The OptiChamber Diamond Valved Holding Chamber device is intended to be used by patients who are under the care or treatment of a physician or licensed healthcare professional. The device is intended to be used by these patients to administer aerosolized medication from most pressurized Metered Dose Inhalers. The intended environments for use include the home, hospitals and clinics.
For Single Patient Use .
Recommended Patient Population:
OptiChamber Diamond: Age 5 and up OptiChamber Diamond with Small LiteTouch mask: 0 to 18 months OptiChamber Diamond with Medium LiteTouch mask: 1 to 5 years OptiChamber Diamond with Large LiteTouch mask: 5 years +
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
4-2
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.