FRIENDSHIP MEDICAL DISPOSABLE PRE-GELLED AG / AGCL SURFACE ELECTRODE
K110289 · Xian Friendship Medical Electronics Co., Ltd. · GXY · Aug 30, 2011 · Neurology
Device Facts
Record ID
K110289
Device Name
FRIENDSHIP MEDICAL DISPOSABLE PRE-GELLED AG / AGCL SURFACE ELECTRODE
Applicant
Xian Friendship Medical Electronics Co., Ltd.
Product Code
GXY · Neurology
Decision Date
Aug 30, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Disposable Pre-gelled Ag/AgCl Surface Electrodes are intended for recording/stimulation and monitoring of the Electromygraphy (EMG), Electroencephalograph(EEG) and Evoked Potential(EP) signals.
Device Story
Disposable Ag/AgCl surface electrodes serve as interface between diagnostic/monitoring equipment and patient skin. Device consists of top woven layer, carbon layer, Ag/AgCl sensor, and Katecho hydrogel; includes lead wire with touch-proof connector. Placed cutaneously by licensed physician or technologist. Electrodes detect electrophysiological signals (EMG, EEG, EP) or provide subcutaneous electrical stimulation. Output is raw electrical signal transmitted to external monitoring/diagnostic equipment for clinician review. Benefits include reliable signal acquisition for diagnostic assessment and therapeutic stimulation.
Clinical Evidence
No clinical data. Bench testing only. Electrical performance verified per ANSI/AAMI EC12/2000 and IEC60601-1. Biocompatibility (cytotoxicity, skin irritation, skin sensitization) confirmed per ISO 10993 standards. Simulated use test performed by healthcare professionals.
Technological Characteristics
Materials: Woven top layer, carbon layer, Ag/AgCl sensor, Katecho KM10E hydrogel. Dimensions: Various (e.g., 27mm x 20mm, 30mm diameter). Connectivity: Lead wire with DIN 42802 or metal snap connector. Standards: ANSI/AAMI EC12/2000, IEC60601-1. Biocompatibility: ISO 10993-5, ISO 10993-10. Single-use, disposable.
Indications for Use
Indicated for recording, stimulation, and monitoring of EMG, EEG, and EP signals in patients requiring electrophysiological signal acquisition or electrical stimulation.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
Rhythmlink International Cutaneous Pad Electrodes (K052188)
Submission Summary (Full Text)
{0}------------------------------------------------
K110289
AUG 300 2011
#### 3. Attachment 1: 510(K) Summary
Summary Date:
This 510(k) summary of safety and effectiveness is being submitted in accordance with the requirements of CFR 807.92.
May 6, 2011
| Submitter's information: | Zhai Ying Chuan<br>General Manager<br>Xian Friendship Medical Electronics Co., Ltd.<br>Gao Xin 1st Road,<br>Hi-Tech Development Zone,<br>Xi'an, Shaanxi,<br>China, 710075<br>Phone: (86) -29-88225200<br>Fax: (86)- 29-88236285<br>georgezhai2616@163.com |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | Friendship Pre-gelled Ag/AgCl Surface Electrodes |
| Common Name: | Disposable Ag/AgCl Pre-gelled Surface Electrodes |
| Classification Name: | 21 CFR section 882.1320. Cutaneous Electrode. |
| Product Code: | GXY |
| Predicate Devices: | Sunspot Pre-gelled Surface Electrodes (K062198) manufactured<br>by Axon Systems, Inc.<br>Rthymlink International Cutaneous Pad Electrodes (K052188)<br>manufactured by Rhythmlink International, LLC |
### 10.1 Device Description:
Friendship's Pre-gelled Ag/AgCl Surface Electrodes are disposable (for "single Use Only"). Used to detect electrophysiological signals or provide electrical stimulation subcutaneously. The electrodes are the interface medium between the diagnostic or monitoring equipment and the patient. The surface electrodes is comprised of top woven layer, carbon layer, Ag/AgCl layer and hydrogel layer on one end electrically connected to lead wire and a touch-proof connector on the other end. The surface electrodes are placed cutaneously by a licensed physician or technologist under the supervision of a
{1}------------------------------------------------
were already verified and validated.
## 10.6 Brief discussion of the clinical tests submitted:
Clinical studies were not deemed necessary regarding the surface electrodes due to their similarity in materials, design and function to those "predicate device". The device was evaluated by health care professionals during a simulated use test and was found to be acceptable for its intended use.
# 10.7 Biocompatibility testing
The contact material Katecho hydrogel is of known biocompatibility. Those materials were already tested for material safety and biocompatibility by the Katecho Inc.:
- Cytotoxicity study ISO 10993-5 .
- Skin irritation study ISO 10993-10 .
- Skin sensitization study ISO 10993-10 .
## 10.8 Performance testing
The electrical performance of the Surface Electrode has been tested and meet the voluntary standard requirements under ANSI/AAMI EC12/2000 in compliance with IEC60601-1.
## 10.9 Conclusions:
Xian Friendship Medical Electronics Co., Ltd.'s Surface Electrodes are substantially equivalent to the predicate devices. Xian Friendship Medical Electronics Co., Ltd. manufactures the Surface Electrodes for Axon Systems, Inc. There are no new questions of safety or effectiveness raised or evident.
{2}------------------------------------------------
| area @ Salaha The Partier States of the Parties of Children Come of the Children Comes of the Children<br>LA B SUNNER AT CREATING<br>ﺎ ﻓﻲ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺎﺣﺔ | ----------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>CALL CLAIM CLAIM CLAIM CLAIM CLAIM CLASS COLUMN CALL CALIFORNIA | 100 mm 1 1 m<br>1 |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>and the first of the many of the many of the may be any and the may be and the may be any and<br>THE LEAST LEAST LEAST LEAST LEAST MAIL BE LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST LEAST | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>- LA-A-A-A-BERTIS -LA-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-A-<br>Comments of the comments of the<br>And Status of Acres of Cattle Company of |
| Specification | Equivalent Device | Predicate Device 1 B |
|------------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------|
| Manufacturer | Xian Friendship Medical<br>Electronics Co., Ltd. | Axon Systems, Inc |
| Part number or 510(k) | Part Number: SEAg-Cu-S | FDA 510(K): K062198 |
| Part number | SEAg-Cu-S-1500/2.0x2.7 | DSE2115 |
| Part | Pre-gelled Surface Electrode | Pre-gelled Surface Electrode |
| Classification | II | II |
| Description | Single and twisted pair, Pre-<br>gelled Surface Electrode | Single and twisted pair, Pre-<br>gelled Surface Electrode |
| Dimensions | 27 mm x 20 mm , various<br>size | 27 mm x 20 mm , various size |
| Electrode shapes | Round/oval design , various | Duck-foot design , various |
| Sensor Material | Ag/AgCl | Ag/AgCl |
| Hydrogel type | Solid gel<br>Katecho KM10E. | Solid gel<br>Katecho KM 10E |
| Connector | DIN 42802 | DIN 42802 |
| Lead Length | Various: 1.5M, 2.0M, 2.5M,<br>3.0M (other lengths may be<br>added) | Various: 1.5M, 2.0M, 2.5M,<br>3.0M (other lengths may be<br>added) |
| Lead wire color | Multiple colors | Multiple colors |
| Labeled as | Single use, disposable | Single use, disposable |
| Indications for use | Stimulation and recording | Stimulation and recording |
| Pouch | Aluminum polyester | Aluminum polyester |
| Standard Electrode Type | AAMI EC12 2000 | AAMI EC12 2000 |
| Stated potential adverse reactions | Skin irritation | Skin irritation |
.
.
{3}------------------------------------------------
| Pouch | Aluminum polyester | Aluminum polyester |
|------------------------------------|--------------------|--------------------|
| Standard Electrode Type | AAMI EC12 2000 | AAMI EC12 2000 |
| Stated potential adverse reactions | Skin irritation | Skin irritation |
| Packaging | 1 electrode/pouch | 1 electrode/pouch |
| Box | 24 electrodes/box | 24 electrodes/box |
| Specification | Equivalent Device | Predicate Device 1 D |
|-------------------------|-----------------------------------------------------------|-----------------------------------------------------------|
| Manufacturer | Xian Friendship Medical<br>Electronics Co., Ltd. | Axon Systems, Inc |
| Part number or 510(k) | Part Number: SEAg-Cu-S | FDA 510(K): K062198 |
| Part number | SEAg-C 3.0 | DSE3003 |
| Part | Pre-gelled Surface Electrode | Pre-gelled Surface Electrode |
| Classification | II | II |
| Description | Single and twisted pair, Pre-<br>gelled Surface Electrode | Single and twisted pair, Pre-<br>gelled Surface Electrode |
| Dimensions | 30mm in diameter, round<br>size | 50 mm x50 mm x 60 mm ,<br>triangle |
| Electrode shapes | Round/oval design , various | Triangle design , various |
| Sensor Material | Ag/AgCl | Ag/AgCl |
| Hydrogel type | Solid gel | Solid gel |
| Connector | Katecho KM10E.<br>Metal snap | Katecho KM 10E<br>Metal snap |
| Labeled as | Single use, disposable | Single use, disposable |
| Indications for use | Stimulation and recording | Stimulation and recording |
| Pouch | Aluminum polyester | Aluminum polyester |
| Standard Electrode Type | AAMI EC12 2000 | AAMI EC12 2000 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an emblem featuring an abstract depiction of an eagle or bird-like figure with outstretched wings.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
Xian Friendship Medical Electronics Co., Ltd. c/o Zhai Ying Chuan General Manager No. 9 Gao Xin 156 Road, Hi-Tech Development Zone Xi'an Shaanxi 710075 China
AUG 30 2011
Re: K110289
Trade/Device Name: Friendship Medical Disposable Pre-gelled Ag/AgCl Surface Electrode Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: Undated Received: July 21, 2011
Dear Mr. Zhai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. · Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Image /page/5/Picture/7 description: The image shows a stylized cursive letter 'f'. The letter has a large, sweeping loop at the top, extending upwards and slightly to the right. A smaller loop is present at the bottom of the letter. The overall impression is of a handwritten signature or monogram.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number (if known): K110289
Device Name: Disposable Pre-gelled Ag/AgCl Surface Electrode
Indications For Use:
Disposable Pre-gelled Ag/AgCl Surface Electrodes are intended for recording/stimulation
and monitoring of the Electromygraphy (EMG), Electroencephalograph(EEG)
and Evoked Potential(EP) signals.
Prescription Use X X
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D) C)
(21 CFR 801 Subpart
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
D. Dolan
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K110289
Page 1 of _1__________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.