STERILE POLYISOPRENE POWDER-FREE SURGICAL GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUG
K110272 · Cardinal Health, Inc. · KGO · Jun 8, 2011 · General, Plastic Surgery
Device Facts
Record ID
K110272
Device Name
STERILE POLYISOPRENE POWDER-FREE SURGICAL GLOVES TESTED FOR USE WITH CHEMOTHERAPY DRUG
Applicant
Cardinal Health, Inc.
Product Code
KGO · General, Plastic Surgery
Decision Date
Jun 8, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
This powder-free surgeon's glove is a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM and the most go res world of Medical Gloves to Permeation by Chemotherapy Drugs:
Device Story
Disposable, sterile, powder-free surgical glove; formulated from polyisoprene synthetic rubber latex; features anti-slip finish, independent thumb, and tapered mechanically locking cuffs to reduce roll down. Used by operating room personnel to protect surgical wounds from contamination. Tested for permeation resistance against 25 chemotherapy drugs per ASTM D6978. Provides barrier protection; healthcare providers use gloves during surgical procedures and handling of chemotherapy agents. Benefits include protection of patient from contamination and protection of clinician from drug exposure.
Clinical Evidence
Bench testing only. No clinical data required. Testing included biocompatibility (ISO 10993-10), physical properties, dimensions, freedom from holes (ASTM D5151), and chemotherapy drug permeation (ASTM D6978).
Indicated for operating room personnel to protect surgical wounds from contamination. Tested for use with chemotherapy drugs.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Cardinal Health logo. The logo consists of a stylized graphic above the company name. The graphic is made up of curved lines that resemble a bird in flight. The text "CardinalHealth" is written in a bold, sans-serif font.
### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
## JUN - 8 2511
#### STERILE POLYISOPRENE POWDER-FREE SURGICAL GLOVES
(A summary of safety and effectiveness information in accordance with the requirements of 21 CFR 807.92)
| Applicant: | Cardinal Health<br>1430 Waukegan Road<br>McGaw Park, IL 60085 |
|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration Number: | 1423537 |
| Regulatory Affairs<br>Contact: | Tatyana Bogdan, RAC |
| Telephone: | 847-887-2325 |
| Fax: | 847-887-2717 |
| E-mail: | tatyana.bogdan-curvin@cardinalhealth.com |
| Summary Prepared: | May 13, 2011 |
| Trade Name: | Sterile Polyisoprene Powder-Free Surgical Gloves Tested for Use with<br>Chemotherapy Drugs |
| Common Name: | Surgeon's Gloves |
| Classification Name: | Surgeon's Gloves |
| Classification Panel: | General and Plastic Surgery |
| Regulation: | 21 CFR 878.4460 |
| Product Code(s): | KGO |
| Legally marketed<br>device(s) to which<br>equivalence is claimed: | 1. Esteem SMT Sterile Polyisoprene Powder-Free Surgical Gloves ( 510(k) K093300, product code KGO)<br>2. Duraprene Sterile Neoprene Powder-Free Surgical Gloves (510(k) K013302, product code KGO) |
| Reason for 510(k)<br>Submission: | Addition of new indications for use: Tested for Use with Chemotherapy Drugs |
| Device Description: | The proposed device is a disposable device. It is not made with natural rubber<br>latex. Instead, it is formulated from polyisoprene, which is a synthetic rubber<br>latex. This Sterile Polyisoprene Powder-Free Surgical Glove is manufactured<br>using exact same material used in the currently cleared device, Esteem SMT<br>glove (K093300) that has been legally marketed by Cardinal Health. The glove<br>is manufactured using molds that feature anti-slip finish, independent thumb,<br>and tapered mechanically locking cuffs to help reduce cuff roll down. It is<br>offered powder-free and sterile. |
-
{1}------------------------------------------------
#### Intended Use:
This powder-free surgeon's glove is a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM m and the most go res world of Medical Gloves to Permeation by Chemotherapy Drugs:
| | Chemotherapy Drug and Concentration | Min Breakthrough Detection Time, 0.01<br>µg/cm²/minute |
|-----|-------------------------------------|--------------------------------------------------------|
| 1. | Blenoxane (15 mg/ml) | >240 |
| 2. | Busulfan (6 mg/ml) | >240 |
| 3. | Carmustine (BCNU) (3.3 mg/ml) | 0.37 |
| 4. | Cisplastin, (1.0 mg/ml) | >240 |
| 5. | Cytarabine (100 mg/ml) | >240 |
| 6. | Cyclophosphamide (20 mg/ml) | >240 |
| 7. | Dacarbazine (10 mg/ml) | >240 |
| 8. | Doxorubicin HCl (2.0 mg/ml) | >240 |
| 9. | Ellence (25 mg/ml) | >240 |
| 10. | Etoposide (Toposar) (20 mg/ml) | >240 |
| 11. | Fludarabine (25 mg/ml) | >240 |
| 12. | Fluorouracil (50 mg/ml) | >240 |
| 13. | Idarubicin (1 mg/ml) | >240 |
| 14. | Ifosfamide (50 mg/ml) | >240 |
| 15. | Mechlorethamine HCl (1 mg/ml) | >240 |
| 16. | Melphalan (5 mg/ml) | >240 |
| 17. | Methotrexate (25 mg/ml) | >240 |
| 18. | Mitoxantrone (2 mg/ml) | >240 |
| 19. | Mitomycin C (0.5 mg/ml) | >240 |
| 20. | Paclitaxel (Taxol) (6.0 mg/ml) | >240 |
| 21. | Paraplatin (10 mg/ml) | >240 |
| 22. | Rituximab (10 mg/ml) | >240 |
| 23. | Thiotepa (10 mg/ml) | 0.44 |
| 24. | Trisenox (0.1 mg/ml) | >240 |
| 25. | Vincristine sulfate (1 mg/ml) | >240 |
Please note that the following drugs have extremely low permeation time of less than 30 minutes: Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 0.37 minute; Thiotepa (10 mg/ml) has a minimum breakthrough time of 0.44 minute.
{2}------------------------------------------------
### KIIO2112
| Summary of the technological characteristics of the device compared to the predicate device | | | | |
|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Characteristic | New Device<br>(K110272) | Predicate<br>Sterile Polyisoprene Powder<br>Free Surgical Glove<br>(K093300) | Predicate<br>Sterile Neoprene Powder-<br>Free Surgical Glove<br>(K013302) | |
| Material<br>Composition | Polyisoprene Synthetic<br>Rubber Latex | Polyisoprene Synthetic<br>Rubber Latex | Synthetic Neoprene Polymer<br>coated with Nitrile | |
| Design | Single Use<br>Sterile<br>Powder-free<br>Hand Specific<br>Independent Thumb<br>Beaded Cuff<br>Lubricated | Single Use<br>Sterile<br>Powder-free<br>Hand Specific<br>Independent Thumb<br>Beaded Cuff<br>Lubricated | Single Use<br>Sterile<br>Powder-free<br>Hand Specific<br>Independent Thumb<br>Beaded Cuff<br>Lubricated | |
| Intended Use | Powder-Free Surgeon's<br>Glove | Powder-Free Surgeon's<br>Glove | Powder-Free Surgeon's<br>Glove | |
| Indications for Use | Tested for Use with<br>Chemotherapy Drugs | Not Tested | Tested for Use with<br>Chemotherapy Drugs | |
| Dimensions &<br>Physical Properties | Meets ASTM D3577 | Meets ASTM D3577 | Meets ASTM D3577 | |
| Freedom from<br>Holes | AQL meets 21CFR 800.20 &<br>ASTM D3577 requirements | AQL meets 21CFR 800.20 &<br>ASTM D3577 requirements | AQL meets 21CFR 800.20 &<br>ASTM D3577 requirements | |
| Powder Residual | Meets requirements of ≤2.0<br>mg/glove for Powder-Free<br>designation per ASTM D3577 | Meets requirements of ≤2.0<br>mg/glove for Powder-Free<br>designation per ASTM D3577 | Meets requirements of ≤2.0<br>mg/glove for Powder-Free<br>designation per ASTM D3577 | |
| PERFORMANCE DATA | | | | |
| SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE* | | | | |
| Performance Test Summary-New Device | | | | |
| Characteristic | | Standard/Test/FDA Guidance | Results Summary | |
| Biocompatibility: | | | | |
| Primary Skin Irritation | | ISO 10993-10 | Gloves are non-irritating. | |
| Guinea Pig Maximization | | ISO 10993-10 | Gloves do not display any potential for<br>sensitization. | |
| Physical Characteristics: | | | | |
| Dimensions | | ASTM D3577 | Meet requirements | |
| Physical Properties | | ASTM D3577 | Meet requirements for rubber surgical gloves | |
| Freedom from Holes | | 21 CFR 800.20 & ASTM D3577 | Tested in accordance with ASTM D5151 with<br>acceptable results | |
| Powder Residual | | ASTM D3577 tested using<br>ASTM standard D6124 | Gloves meet powder level requirements for<br>"Powder-Free" designation per ASTM D3577.<br>Results generated values < 2mg of residual<br>powder per glove. | |
| Chemotherapy Drug<br>Permeation | | ASTM D6978 | Gloves were tested using ASTM D6978. Under<br>the test conditions prescribed by the test, the<br>minimum normalized breakthrough detection<br>times for each of the chemotherapy drugs<br>tested exceeded the maximum testing time of | |
| | | | 240 minutes except for Carmustine (BCNU)<br>(3.3 mg/ml) and Thiotepa (10 mg/ml), which<br>showed permeation time of less than 30 minutes. | |
| Comparative Performance Information Summary | | | | |
| Characteristic | Requirement | New Device | Predicate Device | |
| Biocompatibility: | ISO 10993-1 | Meets requirements | Meets requirements | |
| Primary Skin Irritation | ISO 10993-10 | Pass | Pass | |
| Guinea Pig Maximization | ISO 10993-10 | Pass | Pass | |
| Dimensions | ASTM D3577 | Meets requirements | Meets requirements | |
| Physical Properties | ASTM D3577 | Meets requirements | Meets requirements | |
| Freedom from Holes | 21CFR 800.20<br>& ASTM D3577 | Meets requirements | Meets requirements | |
| Powder Residual | ASTM D3577 | Meets requirements | Meets requirements | |
| Chemotherapy Drug<br>Permeation | ASTM D6978 | Under the test conditions<br>prescribed by the test, the<br>minimum normalized<br>breakthrough detection<br>times for each of the 25<br>chemotherapy drugs tested<br>exceeded the maximum<br>testing time of 240 minutes<br>except for Carmustine<br>(BCNU) (3.3 mg/ml) and<br>Thiotepa (10 mg/ml), which<br>showed permeation time of<br>less than 30 minutes. | Duraprene gloves cleared<br>under K013302 were tested<br>with 10 chemotherapy drugs<br>in accordance with and met<br>requirements of ASTM F739<br>standard. ASTM F739<br>standard was superseded by<br>ASTM D6978-05. The testing<br>method is the same. The<br>difference is in that the<br>thinnest area of the glove<br>(palm or cuff) is tested and<br>the test results are reported<br>as minimum breakthrough<br>times per ASTM D6978, as<br>opposed to testing a random<br>specimen and reporting the<br>average breakthrough times<br>per ASTM F739. | |
K110272 Traditional 510(k) Premarket Notification Submission – 510(k) Summary Cardinal Health, Inc.
{3}------------------------------------------------
## KHUJUL
# and/or of clinical information
Clinical data is not required.
CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL AND CLINICAL DATA
Non-clinical data demonstrates that Sterile Polyisoprene Powder-Free Surgical Gloves Tested for Use with Chemotherapy Drugs meet the technological characteristics of ASTM D3577 standard, and are as safe , as effective, and performed as well as the legally marketed devices identified in this summary.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Tatyana Bogdan Regulatory Affairs Manager Cardinal Health, Incorporated 1430 Waukegan Road McGaw Park, Illinois 60085
JUN - 8 2011
Re: K110272
Trade/Device Name: Sterile Polyisoprene Powder-Free Surgical Gloves Tested for Use with Chemotherapy Drugs Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO, LZC Dated: May 12, 2011 Received: May 24, 2011
Dear Ms. Bogdan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Ms. Bogdan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
## Indications for Use
# 510(k) Number (if known): KLO27 L
Device Name:
### Sterile Polyisoprene Powder-Free Surgical Gloves Tested for Use with Chemotherapy Drugs
Indications for Use: This powder-free surgeon's glove is a disposable device mate of synthetic Indications for Osc. This powder need bear.gooms gersonnel to protect a surgical wound from contamination.
In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM In addition, these gioves were tested for ass Min encation by Chemotherapy Drugs:
| | Chemotherapy Drug and<br>Concentration | Minimum Breakthrough Detection Time<br>in Minutes, 0.01 μg/cm²/minute |
|----|----------------------------------------|-----------------------------------------------------------------------|
| 1. | Blenoxane (15 mg/ml) | >240 |
| 2. | Busulfan (6 mg/ml) | >240 |
| 3. | Carmustine (BCNU) (3.3 mg/ml) | 0.37 |
| 4. | Cisplastin, (1.0 mg/ml) | >240 |
| 5. | Cytarabine (100 mg/ml) | >240 |
| 6. | Cyclophosphamide (20 mg/ml) | >240 |
| 7. | Dacarbazine (10 mg/ml) | >240 |
| 8. | Doxorubicin HCl (2.0 mg/ml) | >240 |
| 9. | Ellence (25 mg/ml) | >240 |
(Continued on next page)
X Over-The-Counter Use Prescription Use AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth F. Clavine-Williams
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K110272
Page / of 2
5
{7}------------------------------------------------
| Continued from previous page) | | |
|-------------------------------|----------------------------------------|-----------------------------------------------------------------------|
| | Chemotherapy Drug and<br>Concentration | Minimum Breakthrough Detection Time<br>in Minutes, 0.01 µg/cm²/minute |
| 10. | Etoposide (Toposar) (20 mg/ml) | >240 |
| 11. | Fludarabine (25 mg/ml) | >240 |
| 12. | Fluorouracil (50 mg/ml) | >240 |
| 13. | Idarubicin (1 mg/ml) | >240 |
| 14. | Ifosfamide (50 mg/ml) | >240 |
| 15. | Mechlorethamine HCl (1 mg/ml) | >240 |
| 16. | Melphalan (5 mg/ml) | >240 |
| 17. | Methotrexate (25 mg/ml) | >240 |
| 18. | Mitoxantrone (2 mg/ml) | >240 |
| 19. | Mitomycin C (0.5 mg/ml) | >240 |
| 20. | Paclitaxel (Taxol) (6.0 mg/ml) | >240 |
| 21. | Paraplatin (10 mg/ml) | >240 |
| 22. | Rituximab (10 mg/ml) | >240 |
| 23. | Thiotepa (10 mg/ml) | 0.44 |
| 24. | Trisenox (0.1 mg/ml) | >240 |
| 25. | Vincristine sulfate (1 mg/ml) | >240 |
Please note that the following drugs have extremely low permeation time of less than 30 minutes:
Carmustine (BCNU) (3.3 mg/ml) has a minimum breakthrough time of 0.37 minute.
Thiotepa (10 mg/ml) has a minimum breakthrough time of 0.44 minute.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use __X (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth F. Clausing-Williams
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices l
510(k) Number: K110272
Page 2 of 2
ర్
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.