The Noble Sharps container is a disposable over the counter sharps container that is intended for the safe disposal of used medical sharps. It is intended for use in offices, exam and patient rooms of small quantity healthcare providers such as medical doctors, dentists, and veterinarians. The size of the Sharps Container is 6 7/16" in height x 4 15/16" in width x 2 7/16" in depth. It is designed with 4- individual single use chambers with each holding 1-sharp, with a maximum sharp length of 6 1/8".
Device Story
Disposable sharps container for safe disposal of used medical sharps; used in healthcare facilities, nursing stations, labs, and exam rooms. Device features 4 individual single-use chambers; each chamber holds one sharp (max length 6 1/8"). Operation involves vertical insertion of sharp; user presses downward on lid to permanently seal chamber. Provides containment of hazardous sharps to prevent injury. Operated by healthcare professionals or staff in clinical settings.
Clinical Evidence
Bench testing only. Testing included puncture resistance (ASTM F 2132), topple resistance (CSA Z316.6-07), and leakage tests (BS 7320:1990).
Technological Characteristics
Injection-molded polypropylene (PP7726). Dimensions: 6 7/16" H x 4 15/16" W x 2 7/16" D. 4-chamber design for single-use containment. Mechanical closure mechanism. Puncture resistance per ASTM F 2132; topple resistance per CSA Z316.6-07; leakage testing per BS 7320:1990.
Indications for Use
Indicated for safe disposal of used medical sharps by healthcare professionals (medical doctors, dentists, veterinarians) in offices, exam, and patient rooms. Designed for small quantity waste generation.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
APR - 5 2011
## Traditional 510(k) Premarket Notification Summary of Safety and Effectiveness
KI�Oz 62 p 1 of 2
| Submitter Information | Noble Marketing<br>Jeff Baker<br>President & CEO<br>100 East Pine Street, Suite 305<br>Orlando, FL 32801 |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Allison Scott, RAC<br>Regulatory Consultant<br>317-569-9500 x106<br>317-569-9520 (fax)<br>ascott@ansongroup.com |
| Date | March 30, 2011 |
| Trade Name | Sharps Container |
| Common Name | Container, Sharps |
| Classification Name | MMK |
| Classification Number | 21 CFR 880.5570 |
| Predicate Devices | GongDong K082042 MMK 880.5570 |
| Device Description | The Noble Sharps Container are intended to be used at healthcare facilities,<br>including nursing stations, medication carts, laboratories, emergency rooms,<br>treatment rooms, and other small quantity waste generators for the safe<br>disposal of hazardous sharps. |
| Intended Use | The Noble Sharps container is a disposable over the counter sharps container<br>that is intended for the safe disposal of used medical sharps. It is intended for<br>use in offices, exam and patient rooms of small quantity healthcare providers<br>such as medical doctors, dentists, and veterinarians. The size of the Sharps<br>Container is 6 7/16" in height x 4 15/16" in width x 2 7/16" in depth. It is<br>designed with 4- individual single use chambers with each holding 1-sharp, with<br>a maximum sharp length of 6 1/8". |
| Summary | The Nobles Sharps Container is a single use device designed for the safe<br>disposal of hazardous sharps. It is designed with 4-chambers, each chamber<br>will hold one sharps. After each chamber is filled with a sharp, a locking lid is<br>closed by pressing downward, thus permanently sealing the sharps inside the<br>container. All materials including the bottom, sides, and locking lid are<br>manufactured using PP7726 (polypropylene) the same material used in the<br>predicate device, the GongDong disposable sharps container cleared under<br>K082042. The size of the Noble Sharps Container is 6 7/16" in height, 4<br>15/16" in width and 2 7/16" in depth. The Noble Sharps Container proves to<br>be substantially equivalent to the predicate device. |
{1}------------------------------------------------
## Comparison to Predicate Devices
The Sharps Container is similar to the predicate in intended use, materials, measuring principle and performance.
| | Noble Sharps Container | GongDong<br>Disposable Sharp<br>Container |
|--------------------|---------------------------------------------|------------------------------------------------|
| 510(k) Number | K100262 | K082042 |
| Product Code | MMK | MMK |
| Indication for Use | • Single Use Healthcare<br>Sharps | • Single Use Healthcare<br>sharps |
| Target Population | • Healthcare professional | • Healthcare<br>professional |
| Where used | • Healthcare facilities/labs | • Healthcare<br>facilities/labs |
| Construction | • Injection molded | • Injection molded |
| Materials | • Polypropylene | • Polypropylene |
| Lid | • Closure by pressing down<br>on lid secure | • Closure by pressing<br>down on lid to secure |
| Needle elimination | • By vertical dropping | • By vertical dropping |
Technological Characteristics
Similar to other sharps containers on the market, sharps are inserted through the top in a vertical position with the sharp side down through the hole in which the sharp is inserted. A lid is then closed by pressing the lid downward for permanent closure and containment of sharps.
Performance of Non-Clinical The Noble Sharps container subjected to puncture resistance testing per ASTM F 2132, topple resistance testing per CSA Z316.6-07, and leakage tests per BS 7320:1990. Noble Sharps Container proves to be substantially equivalent to currently marketed sharps containers.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Noble Marketing C/O Mr. Mark Job Responsible third party Official Regulatory Technology Services, LLC 1394 2511 Street NW Buffalo, Minnesota 55313
APR = 5 201
Re: K110262
Trade/Device Name: Sharps Container Regulation Number: 21 CFR 880,5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: MMK Dated: March 24, 2011 Received: March 25, 2011
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Mr. Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for J
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number: K110262
Device Name: Sharps Container
Indications For Use:
The Noble Sharps container is a disposable over the counter sharps container that is intended for the safe disposal of used medical sharps. It is intended for use in offices, exam and patient rooms of small quantity healthcare providers such as medical doctors, dentists, and veterinarians. The size of the Sharps Container is 6 7/16" in height x 4 15/16" in width x 2 7/16" in depth. It is designed with 4- individual single use chambers with each holding 1-sharp, with a maximum sharp length of 6 1/8″.
> AND/OR Over-The-Counter Use Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elith Fillmore Will
Olvision Sign-Off) Division of Anesthesiology. General Hospital Infection Control, Dental Devices
510(k) Num
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.