AESCULAP- MEITHKE SHUNT SYSTEM

K110206 · Aesculap, Inc. · JXG · Oct 27, 2011 · Neurology

Device Facts

Record IDK110206
Device NameAESCULAP- MEITHKE SHUNT SYSTEM
ApplicantAesculap, Inc.
Product CodeJXG · Neurology
Decision DateOct 27, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.5550
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Miethke Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.

Device Story

The Miethke Shunt System miniNAV Valve is a hydrocephalus shunt component designed to regulate CSF flow. The device consists of a chamber with inlet and outlet connections, featuring a ball-in-cone valve mechanism. A sapphire ball ensures precise closure, while a spiral spring maintains the opening pressure. The valve housing is manufactured from titanium. The system also includes the McLanahan reservoir, a flushing reservoir made of silicone elastomer and titanium with an integrated occlusion mechanism for proximal and distal flushing. The device is used by neurosurgeons in a clinical setting to treat hydrocephalus by diverting excess CSF. The system is MR Conditional in 3-Tesla environments.

Clinical Evidence

Bench testing only. Testing performed in accordance with ISO 7197:2000 and MRI standards including ASTM F2119 (image artifacts), ASTM F2115 (induced heating), and standards for magnetically induced torque and displacement force. Results confirm the device is MR Conditional in 3-Tesla environments.

Technological Characteristics

Materials: Titanium (valve housing), silicone elastomer and titanium (McLanahan reservoir). Principle: Ball-in-cone valve with spiral spring for pressure regulation. Form factor: Shunt valve with integrated flushing reservoir. Connectivity: None. Sterilization: Not specified. MR Compatibility: MR Conditional (3-Tesla) per ASTM F2503.

Indications for Use

Indicated for patients requiring cerebrospinal fluid (CSF) shunting from the lateral ventricles of the brain into the peritoneum.

Regulatory Classification

Identification

A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Page 1 of 2 K110206 # 510(k) SUMMARY (as required by 21 CFR 807.92) Miethke Shunt System miniNAV Valve September 22, 2011 Aesculap®, Inc. COMPANY: 3773 Corporate Parkway Center Valley, PA 18034 Establishment Registration Number: 2916714 Kathy A. Racosky CONTACT: 610-984-9291 (phone) 610-791-6882 (fax) kathy.racosky@aesculap.com Hydrocephalus Shunt System COMMON NAME: CLASSIFICATION NAME: Aesculap Miethke Shunt System REGULATION NUMBER: 882.5550 JXG PRODUCT CODE: ### DEVICE DESCRIPTION : " DEVICE DESCRIP IION The Miethke Shunt System miniNAV Valve is used in the treatment of hylot sepposition The Milline Shum System milling N Valvo lo chamber with inlet and outlet connection The mininNAV Valve is a shiall offinancer tabe" is integrated in the inlet of the device. A ports at opposite ends. A ball-in oone valve 10 linegris-cone valve and a sapphire spiral spring maintains the opening pressure of the ball-in-cone valve in in MAV Value is Spiral Spinig maintains the opening prosoure of the housing of the miniNAV Valve is ball ensures the precise closure of the YAV V. ve is available as a single device as well manufactured from thanian. "The ministration as shunt assistants, catheters, connectors, deflectors and reservoirs. The McLanahan reservoir is a flushing reservoir. It has an integrated occlusion The MoLananan 1000Wolf is a hasin the proximal and distal directions. The mechanism while hashing in Don The McLanahan reservoir is manufactured of silicone elastomer and titanium. {1}------------------------------------------------ 110206 ## INDICATIONS FOR USE NDICATIONS FOR DSE lateral ventricles of the brain into the peritoneum. ### SUBSTANTIAL EQUIVALENCE SUBSTANTHAL EQONALE: " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " Aesculap ', Inc. believes that the minin's . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Miethke Shunt System are Substantially System (K011030) and Medical MonoStep Valve of the Miethke Shunt System (K011030) and Medical MonoStep Valve of the Miethke Shunt Oyunlike (K042465). The new Strata II Valve and Shunt Assembies with and without BioGlide (K042465). The new Strata II Valve and Shunt A Strata if Valve is similar to the current Paedi-GAV Valve and MonoStep Valve. The miniNAV Valve is similar to the current i acal Clower pressure opening. The materials, only differences are the size of the valve and a lower pressure opening to cleared only differences are the size of the vare and intended use is a seviously cleared design, principle of operation and intended use is equivalent to the previously design, principle of operation and intended as one of the Miethke Shunt System. The previously Paedi-GAV Valve and MonoStep Valve of the Miethke Shunt System. The same valv Paedi-GAV Valve and MonoStep Valve of the Michael and which has the same valve cleared system also included the Shuntassistant component which has an integrated cleared system also included the Shans reservoir has an integrated pressure level range as the miniNAV. The McLanahan reservoir has an integrated occlusion mechanism similar to the Strata II Valve. # PURPOSE FOR PREMARKET NOTIFICATION The purpose for this submission is to gain marketing clearance for the new miniNAV The parpose for this on reservoir of the Miethke Shunt System. #### PERFORMANCE DATA PERFORMANCE DATA Testing of the Aesculap Miethke Shunt System miniNAV Valve was performed in Testing of the Aesculap Miernic Onant System results were found to be similar to the accordance with 150 71972000 and 100ans 160 is performed according to the following MRI standards: - ASTM F2119 Evaluation of MR Image Artifacts . - ASTM F2115 Evaluation of Mr. Rimage Induced Heating During ● Magnetic Resonance Imaging - Magnetic Hoofnance Measurement of Magnetically Induced Torque in . the Magnetic Resonance Environment - the Magnetic Floorham at of Magnetically Induced Displacement Force on . the in the Magnetic Resonance Environment The results and evaluation conclude that the device is MR Conditional in 3-Tesla The results and overlakion on (MRI) systems according to ASTM F 2503 and is substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OCT 2 7 2011 Aesculap, Inc. c/o Kathy Racosky Senior Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, PA 18034 Re: K110206 Trade/Device Name: Aesculap Miethke Shunt System Regulation Number: 21 CFR 882.5550 Regulation Name: Central nervous system fluid shunt and components Regulatory Class: II Product Code: JXG Dated: October 04, 2011 Received: October 05, 2011 Dear Ms. Racosky: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the identions for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Dr to and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not maslering. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Eric Riemenschneider Malvina B. Evdelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### INDICATIONS FOR USE STATEMENT 510(k) Number: K110206 Device Name: Miethke Shunt System #### Indications for Use: The Miethke Shunt System is intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum. Prescription Use _______________________ and/or _ Over-the-Counter Use _ (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number k110206
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...