K110186 · Fujifilm Medical System U.S.A., Inc. · LLZ · Apr 7, 2011 · Radiology
Device Facts
Record ID
K110186
Device Name
SYNAPSE 3D LIVER ANALYSIS
Applicant
Fujifilm Medical System U.S.A., Inc.
Product Code
LLZ · Radiology
Decision Date
Apr 7, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Synapse 3D Liver Analysis is medical imaging software used with Synapse 3D Basic Tools that is intended to provide trained medical imaging professionals, including Physicians and Radiologists, with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Liver Analysis accepts DICOM compliant medical images acquired from CT. This product is not intended for use with or for the primary diagnostic interpretation of Mammography images. Addition to Synapse 3D Basic Tools, Synapse 3D Liver Analysis uses contrast enhanced computed tomographic images of the body and provides custom workflows and UI, and reporting functions for liver analysis including, liver segmentation, tumor segmentation, segmentation of intrahepatic vessels as well as the approximation of vascular territories.
Device Story
Synapse 3D Liver Analysis is medical imaging software; processes intravenous contrast-enhanced CT images; performs liver segmentation, tumor segmentation, intrahepatic vessel segmentation, and vascular territory approximation. Used by physicians and radiologists in clinical settings to aid in treatment planning and reporting. Integrates with Fujifilm Synapse Workstation (v3.2.1+) and Synapse Cardiovascular. Operates as an add-on to Synapse 3D Basic Tools. Provides 3D visualization and quantitative analysis of liver structures; assists clinicians in evaluating blood supply and morphological vessel structures; supports clinical decision-making for liver interventions.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and comparison of technological characteristics to the predicate device.
Technological Characteristics
Software-based medical image processing; DICOM compliant; integrates with Synapse Workstation/Cardiovascular; utilizes contrast-enhanced CT input; provides 3D visualization, segmentation, and reporting; runs on existing Synapse 3D platform infrastructure.
Indications for Use
Indicated for trained medical imaging professionals, including physicians and radiologists, to aid in reading, interpreting, reporting, and treatment planning using contrast-enhanced CT images. Not for primary diagnostic interpretation of mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
## APR - 7 2011
## 5. 510(k) Summary
#### Date Prepared:
January 20, 2011
#### Submitter's Information:
FUJIFILM Medical Systems USA, Inc. 419 West Avenue Stamford, Connecticut 06902
Telephone: (203) 602-3665 Fax: (203) 251-7863 Kimerly A. Sharp Contact:
#### Device Trade Name:
Synapse 3D Liver Analysis
#### Device Common Name:
Medical Image Processing and Analysis Software
## Regulation Number:
21 CFR 892.2050
#### Device Classification:
Class II
#### Device Classification Name
.
Picture Archiving Communication System (PACS)
#### Panel:
Radiology
#### Product Code:
90-LLZ
#### Date Received:
TBD
{1}------------------------------------------------
#### FUJIFILM Medical Systems U.S.A. Inc., Synapse 3D Liver Analysis 510(k)
## Decision Date:
TBD
## Decision:
TBD
## Predicate Devices:
- IQQA-Liver Software (K061696), EDDA Technology .
## Description of the Device
Synapse 3D Liver Analysis is an application which uses the intravenous contrasted CT study of a liver to segment the liver and various blood vessels and to perform 3D display of the results. Using the information of segmented liver, hepatic vessels, tumors, and the morphological structure of vessel system and blood supply volume of each vessel, the user can analyze the liver, vessels and tumors and plan the treatment.
Synapse 3D Liver Analysis is used in addition to the previously-cleared features available from Synapse 3D Basic Tools (K101662) to analyze the images acquired from CT. Synapse 3D Liver Analysis is intended to provide trained medical imaging professionals, including Physicians and Radiologists, with tools to aid them in reading, interpreting, and treatment planning and accepts DICOM compliant medical images.
Synapse 3D Liver Analysis with Synapse 3D Basic Tools can be integrated with our cleared Fujifilm's Synapse Workstation, version 3.2.1 and above, and can be used as a part of a Synapse system. Synapse 3D Liver Analysis also can be integrated with Fujifilm's Synapse Cardiovascular for cardiology purposes.
This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.
## Indication for Use
Synapse 3D Liver Analysis is medical imaging software used with Synapse 3D Basic Tools that is intended to provide trained medical imaging professionals, including Physicians and Radiologists, with tools to aid them in reading, interpreting, and treatment planning. Synapse 3D Liver Analysis accepts DICOM compliant medical images acquired from CT.
This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.
Addition to Synapse 3D Basic Tools, Synapse 3D Liver Analysis uses contrast enhanced computed tomographic images of the body and provides custom workflows and UI, and reporting
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The eagle is depicted in a simple, abstract style, with thick, curved lines forming its body and wings. The text is in a sans-serif font and is evenly spaced around the circle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Kimerly A. Sharp Quality Assurance/Regulatory Affairs Associate FUJIFILM Medical System U.S.A., Inc. 419 West Avenue STAMFORD CT 06902
APR - 7 2011
Re: K110186
Trade/Device Name: Synapse 3D Liver Analysis Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 20, 2011 Received: January 21, 2011
Dear Ms. Sharp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{3}------------------------------------------------
Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Pastel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Synapse 3D Liver Analysis
Indications for Use:
Synapse 3D Liver Analysis is medical imaging software used with Synapse 3D Basic Tools that is intended to provide trained medical imaging professionals, including Physicians and Radiologists, with tools to aid them in reading, interpreting, reporting, and treatment planning. Synapse 3D Liver Analysis accepts DICOM compliant medical images acquired from CT.
This product is not intended for use with or for the primary diagnostic interpretation of Mammography images.
Addition to Synapse 3D Basic Tools, Synapse 3D Liver Analysis uses contrast enhanced computed tomographic images of the body and provides custom workflows and UI, and reporting functions for liver analysis including, liver segmentation, tumor segmentation, segmentation of intrahepatic vessels as well as the approximation of vascular territories.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH. Office of In Vitro Diagnostic Device (OIVD)
Mary S. Patel
---
(Division Sign-Off)
Office of In Vitro
510K K110186
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.