ULTRASIGHTHD C31 WITH IMAGESENSE AND REMOTE VIEWER TELEMEDICINE
K110147 · Sti Medical Systems, LLC · HEX · Apr 8, 2011 · Obstetrics/Gynecology
Device Facts
Record ID
K110147
Device Name
ULTRASIGHTHD C31 WITH IMAGESENSE AND REMOTE VIEWER TELEMEDICINE
Applicant
Sti Medical Systems, LLC
Product Code
HEX · Obstetrics/Gynecology
Decision Date
Apr 8, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1630
Device Class
Class 2
Indications for Use
UltraSightHD™ C31 with ImageSense™ and Remote Viewer Telemedicine is a digital colposcope designed to image the cervix and lower genital tract under illumination and magnification. Colposcopy is indicated for women with an abnormal Pap smear in order to affirm normality or detect abnormal appearances consistent with neoplasia, often with directed biopsy. OpacityViewer™ is an ImageSense™ digital filter for displaying areas of acetowhitening. It is a tool that should NOT be used as a substitute for a thorough colposcopic evaluation. The Remote Viewer Telemedicine Technology is a software feature that enables medical personnel to view live video from the UltrasightHD during an ongoing medical exam from a remote viewing station.
Device Story
Digital colposcope captures high-resolution images of vagina, cervix, and external genitalia; utilizes integrated image quality assessment algorithms to ensure focused, centered, and balanced-contrast images. Features ImageSense™ filters to highlight acetowhitening (OpacityViewer™) and Remote Viewer Telemedicine for live video streaming to remote stations. Operated by clinicians in clinical settings; images displayed on dual LCDs and stored via DICOM on external PACS. Remote viewing allows secure connection for off-site personnel to observe exams. Device assists in diagnosing abnormalities and selecting biopsy sites; provides visual aid for clinicians; does not replace standard colposcopic evaluation.
Clinical Evidence
Bench testing only. Compliance testing conducted per IEC 60601-1 (Safety) and IEC 60601-1-2 (EMC). No clinical data or biocompatibility testing required as the device is not patient-contacting.
Technological Characteristics
Digital colposcope with integrated image quality assessment and digital filtering (ImageSense™). Features dual LCDs for display and interface. Connectivity via DICOM for PACS integration. Remote viewing capability via secure network connection. Safety/EMC compliance per IEC 60601-1 and 60601-1-2. Non-patient contacting.
Indications for Use
Indicated for women with abnormal Pap smears to evaluate for normality or detect neoplasia in the cervix and lower genital tract, often with directed biopsy. Not for use as a substitute for thorough colposcopic evaluation.
Regulatory Classification
Identification
A colposcope is a device designed to permit direct viewing of the tissues of the vagina and cervix by a telescopic system located outside the vagina. It is used to diagnose abnormalities and select areas for biopsy. This generic type of device may include a light source, cables, and component parts.
Special Controls
*Classification.* Class II (special controls). The device, when it is a standard colposcope (or colpomicroscope) that uses only a white light source, does not use filters other than a green filter, does not include image analysis software, and is not smartphone-based, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
UltraSightHD (C30) Digital Colposcopy System with ImageSense Technology (K090324)
Submission Summary (Full Text)
{0}------------------------------------------------
STI Medical Systems, LLC
K110147
pg. 1 of 2
510(k) Application
UltraSightHD™ C31 with ImageSense™ and Remote Viewer Telemedicine
#### 510(k) Summary 5
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21CFR 807.92
### 1. General Information
Submitter's Name and Address:
STI Medical Systems, LLC 4275 Executive Square, Suite 825 La Jolla, CA 92037
APR - 8 2011
Contact Person and Telephone Number:
| Rolf Wolters, Ph.D. |
|-----------------------|
| Senior Vice President |
| rolf@sti-hawaii.com |
| (808) 540-4728 |
Date Prepared:
January 12, 2011
## 2. Device Information
| Trade/Device Name: | UltraSightHD™ C31 Digital Colposcopy System with<br>ImageSense™ and Remote Viewer Telemedicine<br>Technologies |
|----------------------------|----------------------------------------------------------------------------------------------------------------|
| Classification Name: | Colposcope |
| Classification Regulation: | 21 CFR § 884.1630 |
| Regulatory Class: | Class II |
| Product Code: | HEX |
#### 3. Predicate Device(s):
K090324, UltraSightHD (C30) Digital Colposcopy System with ImageSense Technology, STI Medical Systems, December 20, 2010.
### 4. Device Description:
The UltraSightHD™ C31 Digital Colposcopy System with ImageSense™ and Remote Viewer Telemedicine Technologies represents a modification to the UltraSightHD (C30) Digital Colposcopy System with ImageSense Technology. It is a digital colposcope designed to acquire images of the vagina, cervix and external genitalia. It is used to diagnose abnormalities and select areas for biopsy. The illumination and optical design of the device enable the user to capture high resolution digital images. The optical subsystem is augmented by integrated image quality assessment algorithms, ensuring that focused, centered, and balanced-contrast images are acquired. Two liquid crystal displays (LCDs) provide video display and user interface information. The software includes functionality for annotating, filtering and storing the images on external PACS servers via the DICOM protocol. The ImageSense™ filters highlight certain
{1}------------------------------------------------
STI Medical Systems, LLC
· : ·
39. 2 of 2
510(k) Applicatio
UltraSightHD™ C31 with ImageSense™ and Remote Viewer Telemedicine
characteristics of the captured images. The Remote Viewer Telemedicine feature enables medical personnel to view live video of an ongoing medical exam from a remote viewing station.
### 5. Indications for Use:
UltraSightHD™ C31 with ImageSense™ and Remote Viewer Telemedicine is a digital colposcope designed to image the cervix and lower genital tract under illumination and magnification. Colposcopy is indicated for women with an abnormal Pap smear in order to affirm normality or detect abnormal appearances consistent with neoplasia, often with directed biopsy.
OpacityViewer™ is an ImageSense™ digital filter for displaying areas of acetowhitening. It is a tool that should NOT be used as a substitute for a thorough colposcopic evaluation. The Remote Viewer Telemedicine Technology is a software feature that enables medical personnel to view live video from the UltrasightHD during an ongoing medical exam from a remote viewing station.
#### 6. Comparison of Technological Characteristics:
The materials and technologies in the UltraSightHD™ C31 Digital Colposcope are suitable for their intended use and are entirely similar to those of the predicate device. The addition of the remote viewing capability allows a computer at a different location to securely connect with the device to view the same imagery visible on the colposcope.
#### 7. Testing:
Appropriate risk analysis-driven product testing was conducted to evaluate conformance to product specification. Successful compliance testing in accordance with IEC 60601-1 and 60601-1-2 (Safety and EMC) was conducted. The device is not patient contacting and therefore no biocompatibility testing was required.
#### 8. Conclusion:
The UltraSightHD™ C31 Digital Colposcopy System with ImageSense™ and Remote Viewer Telemedicine Technologies is equivalent to the UltraSightHD™ (C30) Digital Colposcopy System with ImageSense™ Technology. The indications for use, safety and efficacy, basic overall design and function, product performance, and materials used are equivalent. The addition of the remote viewer does not affect the safety and effectiveness of the device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration i 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Rolf Wolters, Ph.D. Senior Vice President STI Medical Systems, LLC 4275 Executive Square, Suite 825 LA JOLLA CA 92037
APR - 8 2011
Re:
K 110147
Trade Name: UltraSightHD™ C31 Digital Coloposcopy System with ImageSense™ OltraSignifiD ----------------------------------------------------------------------------------------------------------------------------------------------------------------
Regulation Number: 21 CFR §884.1630 Regulation Name: Colposcope
Regulatory Class: II Product Code: HEX Dated: January 12, 2011 Received: January 18, 2011
Dear Dr. Wolters:
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
We have reviewed your Section 510(k) premarket notification of mtent to market the indication We have reviewed your Section 510(k) premated in substantially equivant (for the indications
referenced above and have develse is substantially equivated in interstate referenced above and have determined the devices narketed in interstate
for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosure) to legally markets breaks device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food, Drug, commerce prior to May 28, 1976, the enactment with the provisions of the Federal Food, Drug, devices that have been reclassities in accordance will the proval application (PMA).
and Cosmetic Act (Act) that do not require approval or a proval application of the Act. and Cosmetic Act (Act) that do not require approval or a provisions of the Act. The
You may, therefore, market the device, subject to the general contralize, subjection, lis You may, therefore, market the device, subject to the general registration, listing of
general controls provisions of the Act include requirements for annual registranding an general controls provisions of the Act menus and problibitions against misbranding and devices, good manufacturing practice, labelme, and evaluate information related to contract liability.
adulteration. Please note: CDRH does not evaluate information related adulteration. Please note: CDRH does not evaluate unomation realing must be truthful and not misleading.
warranties. We remind you, however, that device labeling must be t
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it
in be and the collected and and also Eristing major regulations affec If your device is classified (see above) mo entre class if (opinisms affecting your device can be
may be subject to additional controls. Exista, a Berts 800 n 898. In addit may be subject to additional controls. Existing major rogano of 888. In addition, FDA may
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In the Fede found in the Code of Federal Regulations, Title 20, 2014 - 12:00 PM the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA's issuance of a substance other requirements of the Act
that FDA has made a determination that your device complies with other comply that FDA has made a determination that your veries computes websites. You must comply
or any Federal statutes and regulations administed by other Federal agencies. You must or any Federal statutes and regulations administration and listing and listing (21 CFR Part
with all the Act's requirements, including, but not limited to: registration of me with all the Act's requirements, including, but not milled to: regulations of medical device-related
{3}------------------------------------------------
## Page 2
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the adverse events) (21 CFR 803); good manufacturily practice rolicable, the electronic product
quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electro quality systems (QS) regulation (21 CFN Far 620), and on apply of the Act); 21 CFR 1000-1050.
radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
r and the states and the stime (Fices/CDRH/CDRH/CDRHOffices/ucml 15809.htm If you desire specific advice for your device offices/CDRH/Offices/ucm] I S809.htm for for for for for for for go to http://www.fda.gov/Aboutible.com/res/thics/CDFF/s) Office of Compliance. Also, please
the Center for Devices and Radiological Health's (CDFH's) Officeation" (21CFR Par the Center for Devices and Radiological Health s (CDN 3) Online of Chication" (21CFR Part
note the regulation entitled, "Misbranding of educere events under the MDK regulatio note the regulation entitled, "Misbranding by reference to promance in the MDR regulation (21 CFR Part 803), please go to
CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of the continues. http://www.fda.gov/MedicalDcvices/Devices/Davision of Postmarket Surveillance.
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the You may obtain other general information on your responsibilities and the world.
Division of Small Manufacturers, International and Consumer Assistance at its toll-free numbe Division of Small Mandracturers, 100 or at its Internet address
(800) 638-2041 or (301) 796-7100 or at its Internet of Soul'Indi (800) 638-2041 or (301) 796-7100 of at its internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Hubert Lemon MD
Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K110147
UltraSightHD™ C31 Digital Colposcopy System with ImageSense™ and Device Name: Remote Viewer Telemedicine Technologies
Indications For Use:
UltraSightHD™ C31 with ImageSense™ and Remote Viewer Telemedicine is a digital colposcope designed to image the cervix and lower genital tract under illumination and magnification. Colposcopy is indicated for women with an abnormal Pap smear in order to affirm normality or detect abnormal appearances consistent with neoplasia, often with directed biopsy.
OpacityViewer™ is an ImageSense™ digital filter for displaying areas of acetowhitening. It is a tool that should NOT be used as a substitute for a thorough colposcopic evaluation. The Remote Viewer Telemedicine Technology is a software feature that enables medical personnel to view live video from the UltrasightHD during an ongoing medical exam from a remote viewing station.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Arry M-Wh
ductive, Gastro-Renal, and 110147
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.