← Product Code [LPL](/productcode/LPL) · K110099

# PROCLEAR TORIC XR SOFT (HYDROPHILIC) CONTACT LENS; PROCLEAR MULTIFOCAL XR SOFT (HYDROPHILIC) CONTACT LENS; PROCLEAR MULT (K110099)

_CooperVision, Inc. · LPL · Jun 29, 2011 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K110099

## Device Facts

- **Applicant:** CooperVision, Inc.
- **Product Code:** [LPL](/productcode/LPL.md)
- **Decision Date:** Jun 29, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic or hyperopic in powers from -20.00, possess astigmatism to -5.75 diopters, and are presbyopic. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only). Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 that are myopic or hyperopic and are presbyopic. The lenses may be worn by persons who exhibit astigmatism of 0.75 diopters or less that does not interfere with visual acuity. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only). Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic and hyperopic in powers from -20.00 to +20.00 diopters and astigmatism corrections to -5.75 diopters. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only). Daily wear replacement schedules may vary from patient and should be decided by the eye care practitioner in consultation with their patients. The lenses are to be cleaned, rinsed and disinfected each time they are removed from the patients' eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lenses may be disinfected using a chemical disinfection system.

## Device Story

Soft (hydrophilic) contact lenses; omafilcon A material (59% water); daily wear. Corrects myopia, hyperopia, astigmatism, and presbyopia. Designed for patients experiencing mild dryness/discomfort. Prescribed by eye care practitioners; patient self-use for daily wear; requires cleaning, rinsing, and chemical disinfection. Lenses feature aspherical/toric surfaces for visual correction. Benefits include improved comfort and visual acuity.

## Clinical Evidence

Bench testing only. Biocompatibility (cytotoxicity, ocular irritation, systemic toxicity) and physicochemical testing (ISO 18369-4:2006) performed. Results confirm non-toxic, non-irritating properties and equivalence to predicate devices. No clinical data required due to established material safety and manufacturing equivalence.

## Technological Characteristics

Material: Omafilcon A (59% water, 41% polymer); Group II Non-Ionic High Water. Manufacturing: Finished Inside Polymerization System II; Thermal Cure. Sterilization: Moist heat (steam) autoclave. Visibility tint: Vat Blue 6. Dimensions: 8.0-9.3mm base curve, 13.6-15.2mm diameter. Oxygen permeability: 21.05 x 10^-11.

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- Proclear Toric XR, Proclear Multifocal XR and Proclear Multifocal Toric (omafilcon A) Soft (hydrophilic) Contact Lenses for Daily Wear ([K081865](/device/K081865.md))
- Proclear XC and Proclear 1 day (omafilcon A) Hydrophilic Contact Lenses for Daily Wear ([K061948](/device/K061948.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K110099

Image /page/0/Picture/1 description: The image shows a black and white image of a circular object. The object is mostly black, with some white spots. The object appears to be a fruit or vegetable. 

6150 Stoneridge Mall Road Suite 370 Pleasanton, CA 94588

T 925 621 2480
F 925 621 2487 www.coopervision.com JUN 2 9 2011

CooperVision"

CooperVision: 510(k) Summary (as required by section 807.92(c)). K110099

| Submitted By:    | CooperVision Inc.<br>6150 Stoneridge Mall Road, Suite 370<br>Pleasanton, CA 94588                                                                                                                                  |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Sarah Harrington MS, MBA<br>Regulatory Affairs<br>925 621-2458: Phone<br>866 491-7424; FAX<br>sharrington@coopervision.com                                                                                         |
| Date Prepared:   | May 26 2011                                                                                                                                                                                                        |
| Trade Name:      | Proclear Toric XR (omafilcon A) Soft (hydrophilic) Contact Lens<br>Proclear Multifocal XR (omafilcon A) Soft (hydrophilic) Contact Lens<br>Proclear Multifocal Toric (omafilcon A) Soft (hydrophilic) Contact Lens |
| Common Name:     | Soft Contact Lens                                                                                                                                                                                                  |
| Classification:  | Lenses, Soft Contact, Daily Wear<br>Class II: LPL, 21 CFR 886.5925                                                                                                                                                 |

{1}------------------------------------------------

#### 510(k) Summary

#### Proclear Toric XR, Multifocal and Multifocal XR Daily Wear Soft Contact Lenses

#### Substantially Equivalent Devices:

Proclear Toric XR, Proclear Multifocal XR and Proclear Multifocal Toric (omafilcon A) Soft (hydrophilic) Contact Lenses for Daily Wear, K081865.

Proclear XC and Proclear 1 day (omafilcon A) Hydrophilic Contact Lenses for Daily Wear, K061948.

# Device

Description:

The Proclear Toric XR, Multifocal XR and Multifocal Toric (omafilcon A) Contact Lenses are made from a material containing 59% water and 41% omafilcon A, a polymer of 2-hydroxy-ethylmethacrylate and 2methacryloyloxyethyl phosphorylcholine crosslinked with ethyleneglycol dimethacrylate. The lenses are tinted edge for visibility purposes with Vat Blue 6.

Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lens is a back surface toric.

Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are available as a multifocal lens with an aspherical front surface and spherical back surface for the correction of visual acuity in presbyopic persons who are myopic or hyperopic. The Proclear Multifocal XR is designed with two multifocal zones, as well as the edge shape being optimized to provide comfort without sacrificing tensile strength.

Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lens front surface is aspherical, with the anterior surface having a toric generated surface for the purpose of correcting vision in an eye that is astigmatic.

#### Indication for Use:

Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with nondiseased eyes that are myopic or hyperopic in powers from -20.00 to +20.00, possess astigmatism to -5.75 diopters or less, and are presbyopic. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).

Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with nondiseased eyes in powers from -20.00 to +20.00 that are myopic or hyperopic and are presbyopic. The lenses may be worn by persons who exhibit astigmatism of 0.75 diopters or less that does not interfere with visual acuity. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).

{2}------------------------------------------------

CooperVision K110099

--..

#### 510(k) Summary

#### Proclear Toric XR, Multifocal and Multifocal XR Daily Wear Soft Contact Lenses

Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are Indication for Use indicated for daily wear for the correction of visual acuity in aphakic and (continued) not-aphakic persons with non-diseased eyes that are myopic or hyperopic in powers from -20.00 to +20.00 diopters and astigmatism corrections to -5.75 diopters. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non- Sjogren's only).

> Daily wear replacement schedules may vary from patient to patient and should be decided by the eye care practitioner in consultation with their patients. The lenses are to be cleaned, rinsed and disinfected each time they are removed from the patients' eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lenses may be disinfected using a chemical disinfection system.

#### Technological The technological characteristics of the subject lens and the predicate lenses Characteristics are compared in the following tables. The Subject device has the same technological characteristics (i.e., design, material, package, parameters) as the predicate devices.

| Material Comparison     |                                                                                |                                                                                |                                                 |
|-------------------------|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------|-------------------------------------------------|
|                         | Subject Device                                                                 | Predicate<br>Device                                                            | Predicate<br>Device                             |
| Product name            | Proclear Toric XR<br>Proclear<br>Multifocal XR<br>Proclear<br>Multifocal Toric | Proclear Toric XR<br>Proclear<br>Multifocal XR<br>Proclear<br>Multifocal Toric | Proclear XC<br>Proclear 1 day                   |
| Material USAN<br>Name   | Omafilcon A                                                                    | Omafilcon A                                                                    | Omafilcon A                                     |
| 510(k) number           | This submission                                                                | K081865                                                                        | K061948                                         |
| FDA Category<br>(Group) | Group II Non-lonic<br>High Water                                               | Group II Non-lonic<br>High Water                                               | Group II Non-Ionic<br>High Water                |
| Manufacturing<br>method | Finished Inside<br>Polymerization<br>System II                                 | Finished Inside<br>Polymerization<br>System II                                 | Cast molded                                     |
| Curing                  | Thermal Cure                                                                   | UV Cure                                                                        | Thermal cure                                    |
| Sterilization           | Moist Heat (steam) in<br>validated Autoclave                                   | Moist Heat (steam)<br>in validated<br>Autoclave                                | Moist Heat (steam)<br>in validated<br>Autoclave |
| Packaging               | Blister Pack                                                                   | Blister Pack                                                                   | Blister Pack                                    |
| Visibility tint         | VAT Blue 6                                                                     | VAT Blue 6                                                                     | VAT Blue 6                                      |

{3}------------------------------------------------

CooperVision K110099

## 510(k) Summary

| Parameter Comparison              |                              |                              |                     |
|-----------------------------------|------------------------------|------------------------------|---------------------|
|                                   | Subject Device               | Predicate<br>Device          | Predicate<br>Device |
| Product name                      | Proclear Toric XR            | Proclear Toric XR            | Proclear XC         |
|                                   | Proclear<br>Multifocal XR    | Proclear<br>Multifocal XR    | Proclear 1 day      |
|                                   | Proclear<br>Multifocal Toric | Proclear<br>Multifocal Toric |                     |
| Water Content                     | 59% ± 2%                     | 59% ± 2%                     | 60% ± 2%            |
| Refractive<br>Index               | 1.395 ± 0.005                | 1.40                         | 1.40                |
| Oxygen<br>Permeability x<br>10-11 | 21.05                        | 21.05                        | 21.00               |
| Light<br>Transmission             | >90%                         | >90%                         | >90%                |
| Base Curve                        | 8.0 to .9.3 mm               | 8.0 to .9.3 mm               | 8.0 to .9.5 mm      |
| Diameter                          | 13.6 to 15.2 mm              | 13.6 to 15.2 mm              | 13.0 to 15.5 mm     |
| Power                             | -20.00 to +20.00             | -20.00 to +20.00             | -20.00 to +20.00    |

#### Proclear Toric XR, Multifocal and Multifocal XR Daily Wear Soft Contact Lenses

#### Non-Clinical Testing

A series of in-vitro and in-vivo preclincal toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the contact lens. All tests were conducted in accordance with the GLP regulation (21 CFR Part 56) or according to valid scientific protocols.

| Test                                                                                                                                                         | Acceptance Criteria                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Result                                                                                                            |        |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|--------|
| Cytotoxicity Test ISO 10993 - 5:<br>1999: Biological Evaluation of<br>Medical Devices - Part 5: Tests for<br>In Vitro Cytotoxicity                           | All 3 monolayers exposed to<br>the test article show no grade<br>greater than grade 2 (reactivity<br>mild)                                                                                                                                                                                                                                                                                                                                                            | Pass                                                                                                              |        |
| ISO Ocular Irritation<br>ISO 10993 - 10:2002: Biological<br>Evaluation of Medical Devices - Part<br>10: Tests for Irritation and Delayed<br>Hypersensitivity | Test extract shows no<br>significant irritation over the<br>reagent control during the<br>observation period.                                                                                                                                                                                                                                                                                                                                                         | Pass                                                                                                              |        |
| Systemic Toxicity Study<br>ISO 10993 - 11: 1996: Biological<br>Evaluation of Medical Devices - Part<br>11: Tests for Systemic Toxicity.                      | During the observation period,<br>none of the animals treated<br>with the individual test extract<br>exhibit a significantly greater<br>reaction that the control<br>animals.                                                                                                                                                                                                                                                                                         | Pass                                                                                                              |        |
| (continued)                                                                                                                                                  | Test                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Acceptance Criteria                                                                                               | Result |
|                                                                                                                                                              | ISO Contact Lens 22 Day Ocular<br>Irritation Study<br>ISO 9394 Ophthalmic Optics - Contact<br>Lenses and Contact Lens Care Products<br>The determination of Biocompatibility                                                                                                                                                                                                                                                                                          | Scores from macroscopic<br>and microscopic ocular<br>examinations equivalent<br>between test and control<br>eyes. | Pass   |
|                                                                                                                                                              | by Ocular Study with Rabbit Eyes.<br>Total Extractables , Water content, Dk,<br>Light Transmittance, refractive index<br>tested per ISO 18369-4:2006<br>Ophthalmic optics - Contact lenses -<br>Part 4: Physicochemical properties of<br>contact lens materials:<br>section 4.2                                                                                                                                                                                       | Equivalent to predicate<br>lens                                                                                   | Pass   |
|                                                                                                                                                              | The results for the non-clinical testing demonstrate:<br>· The lens material and lens material extracts are non-toxic, non-irritating<br>under the experimental conditions.<br>· The lens physical and material properties are consistent with currently<br>marketed soft contact lenses.<br>· No evidence of unsafe amounts of residue in the extractables.<br>· Physicochemical testing of the subject lenses demonstrated equivalency to<br>the predicate devices. |                                                                                                                   |        |
| Clinical                                                                                                                                                     | · Lens remains sterile and stable in the package, for the established shelf life.<br>The technical characteristics, manufacturing and sterilization process of the subject<br>lens are equivalent to omafilcon A soft contact lenses currently marketed by<br>CooperVision, therefore no clinical data is required.                                                                                                                                                   |                                                                                                                   |        |
|                                                                                                                                                              | Conclusion Drawn from Studies:                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                   |        |
| Validity of<br>Scientific<br>Data                                                                                                                            | Contract laboratories under Good Manufacturing Practice regulations conducted<br>toxicological and microbiology studies.<br>Chemistry, shelf-life and leachablility studies were conducted by CooperVision<br>and followed scientific protocols. The data were determined to be scientifically<br>valid under 21 CFR 860.7                                                                                                                                            |                                                                                                                   |        |
| Substantial<br>Equivalence                                                                                                                                   | Information presented in this Premarket Notification establishes that the<br>Cooper Vision (omafilcon A) Proclear Toric XR, Proclear Multifocal and Proclear<br>Multifocal XR contact lens is as safe and effective as the predicate devices when<br>used in accordance with the labeled directions for use and for the requested<br>indications.                                                                                                                     |                                                                                                                   |        |
| Risk and                                                                                                                                                     | The risks of the subject lens are the same as those normally attributed to the<br>wearing of soft (hydrophilic) contact lenses on a daily wear basis. The benefits to<br>the patient are the same as those for other soft (hydrophilic) contact lenses.                                                                                                                                                                                                               |                                                                                                                   |        |

{4}------------------------------------------------

・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

# 510(k) Summary

# Proclear Toric XR, Multifocal and Multifocal XR

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wing-like shapes, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

CooperVision Inc. c/o Ms. Lisa Hahn Global Regulatory Affairs Manager 6150 Stoneridge Mall Road, Suite 370 Pleasanton, CA 94588

JUN 2 9 2011

Re: K110099

Trade/Device Name:

Proclear Toric XR (Omafilcon A) Soft (hydrophilic) Contact Lens Proclear Multifocal XR (Omafilcon A) Soft (hydrophilic) Contact Lens Proclear Multifocal Toric (Omafilcon A) Soft (hydrophilic) Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL Dated: May 27, 2011 Received: May 31, 2011

Dear Ms. Hahn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Ms. Lisa Hahn

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersQffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia Alexander

for

MR.
DR.
REV.
GEN.

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

#### Goper Visiona: SEE BEYOND THE ORDINAR

Traditional 510(k)

Proclear Toric XR, Multifocal and Multifocal XR Daily Wear Soft Contact Lenses

# Indications for Use

## 510(k) Number (if known):

Device Name: Proclear Toric XR, Multifocal XR and Multifocal Toric (omafilcon A) Soft (hydrophilic) Contact Lenses for Daily Wear

# Indications for Use:

Proclear Multifocal Toric (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eves that are myopic or hyperopic in powers from -20.00, possess astigmatism to -5.75 diopters, and are presbyopic. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only),

Proclear Multifocal XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes in powers from -20.00 to +20.00 that are myopic or hyperopic and are presbyonic. The lenses may be worn by persons who exhibit astigmatism of 0.75 diopters or less that does not interfere with visual acuity. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).

Proclear Toric XR (omafilcon A) Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes that are myopic and hyperopic in powers from -20.00 to +20.00 diopters and astigmatism corrections to -5.75 diopters. The lenses may provide improved comfort for contact lens wearers who experience mild discomfort or symptoms related to dryness during lens wear associated with Evaporative Tear Deficiency or from Aqueous Tear Deficiency (non-Sjogren's only).

Daily wear replacement schedules may vary from patient and should be decided by the eye care practitioner in consultation with their patients. The lenses are to be cleaned, rinsed and disinfected each time they are removed from the patients' eye and discarded after the recommended wearing period prescribed by the eye care practitioner. The lenses may be disinfected using a chemical disinfection system.

| Prescription Use: YES<br>(Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) |
|------------------------------------------------------|--------|------------------------------------------------|
|------------------------------------------------------|--------|------------------------------------------------|

Please Do Not Write Below this line-Continue on Another Page if Needed

Concurrence of CDRH, Office of Device Evaluation (ODE)

____________________________

(Division Sign-Off)

Division of Ophthalmic, Neurological and Ear,
Nose and Throat Devices

Page 1 of 1

510(k) Number_K110099

---

**Source:** [https://fda-staging.innolitics.com/device/K110099](https://fda-staging.innolitics.com/device/K110099)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
