K110097 · Acclarent, Inc. · EOB · May 27, 2011 · Ear, Nose, Throat
Device Facts
Record ID
K110097
Device Name
ACCLARENT CYCLOPS MULTI-ANGLE ENDOSCOPE
Applicant
Acclarent, Inc.
Product Code
EOB · Ear, Nose, Throat
Decision Date
May 27, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 874.4760
Device Class
Class 2
Indications for Use
The Acclarent Cyclops Multi-Angle Endoscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx in an operating room environment.
Device Story
Acclarent Cyclops Multi-Angle Endoscope (CYE002) is a 4mm rigid, unchanneled endoscope for visualizing nasal cavity and nasopharynx. Device features a direction-of-view dial (10° to 90°) and a 320° shaft rotation dial, allowing visualization of structures without rotating the scope body. Rare-earth permanent magnets in the proximal control body drive the direction-of-view change. Device includes a standard eyepiece for camera coupling and a light post compatible with ACMI light sources. Used in operating room environments by physicians. Provides 55° field of view and 5mm to 40mm depth of focus. Reusable; requires cleaning and sterilization before each use. Benefits include enhanced visualization capabilities during endoscopic procedures.
Clinical Evidence
Bench testing only. Evaluated distal shaft diameter, working length, field of view, fixed focus, direction of view, rotation of view, illumination, scope resolution, dial actuation forces, temperature, magnetic field strength, electrical safety, EMC, durability, environmental conditioning, compression, vibration, and shock. Temperature testing used thermocouples with 300W Xenon light source. No clinical data required.
Technological Characteristics
Rigid, 4mm diameter endoscope. Sapphire cover lens. Illumination via glass fibers. Rare-earth permanent magnets (≤10 gauss at 2cm) for articulation. Stainless steel adapters. Compatible with standard camera couplers and ACMI light sources. Standards: IEC 60601-2-18, IEC 60601-1-2, ISO 10993-1, ISO 8600-3, ISO 8600-5. Reprocessing: manual cleaning with enzymatic soak and automated washer; pre-vacuum steam sterilization.
Indications for Use
Indicated for visualization of the nasal cavity and nasopharynx in patients undergoing procedures in an operating room setting.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top of the "r" to the right, creating a visual element above the word. The text is black against a white background.
,
K110097
MAY 2 7 2011
## 510(k) SUMMARY
| Sponsor/Submitter: | Acclarent, Inc.<br>1525-B O'Brien Drive<br>Menlo Park, California 94025 |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Keri Yen<br>Regulatory Affairs Manager<br>Phone: (650) 687-5874<br>Fax: (650) 687-4449 |
| Date of Preparation: | May 20, 2011 |
| Device Trade Name/<br>Model Number: | Acclarent Cyclops Multi-Angle Endoscope<br>CYE002 |
| Common Name: | Endoscope |
| Device Classification: | Class II |
| Regulation Number: | 21 CFR 874.4760 |
| Classification Name: | Nasopharyngoscope (Flexible or Rigid) |
| Product Code: | EOB |
| Predicate Devices: | Acclarent Cyclops Multi-Angle Endoscope (K100577) |
| Device Description: | The Acclarent Cyclops Multi-Angle Endoscope is a 4mm rigid<br>unchanneled endoscope that has the capability of varying<br>direction of view from 10° to 90°, which is altered by the<br>direction of view dial. The direction of view is indicated by<br>visible markings on the scope body. Cyclops provides a 55°<br>field of view and a depth of focus from 5 mm to 40mm. The<br>device shaft can also rotate 320° to allow for visualization of<br>structures without rotating the device; this is controlled by the<br>shaft rotation dial. Small rare-earth permanent magnets are<br>incorporated into the proximal scope control body (≤10 gauss<br>at 2cm) and drive the change in the direction of view. A<br>standard eyepiece located on the proximal end of the device is<br>compatible with a standard camera coupler. The light post on<br>the subject device is compatible with an ACMI light source.<br>There are two stainless steel adapters that accompany the<br>Acclarent Cyclops Multi-Angle Endoscope to facilitate |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top of the "r" to the top of the "t", creating a visual element above the word. The text is black against a white background.
The adapters connect to the light post. The Acclarent Cyclops Multi-Angle Endoscope is a reusable device and must be cleaned and sterilized according to the user manual prior to every use.
Indications for Use:
The Acclarent Cyclops Multi-Angle Endoscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx in an operating room environment.
| Characteristics: | | |
|------------------------|-------------------------------------------------------------------|-----------------------------------------------------------------|
| Attribute | Predicate Device<br>(Acclarent Cyclops Multi-<br>Angle Endoscope) | Subject Device<br>(Acclarent Cyclops Multi-<br>Angle Endoscope) |
| 510(k) number | K100577 | K110097 |
| Model Number | CYE001 | CYE002 |
| Rigidity | Rigid | Same |
| Viewing Optics | Lens (Sapphire cover) | Same |
| Depth of View | 5-45 mm | Same |
| Field of View | 60° | 55° |
| Direction of View | 10° to 100° | 10° to 90° |
| Shaft Body<br>Diameter | 4mm | Same |
| Working Length | 6.89 inches (175mm) | Same |
| Magnetic Strength | ≤10 gauss at 2cm | Same |
| Illumination Fibers | Glass Fibers | Same |
## Technological
Performance Data:
Performance testing of the Acclarent Cyclops Multi-Angle Endoscope consisted of bench testing and a cadaver study. Bench testing met all acceptance criteria for attributes such as distal shaft diameter, working length, field of view, fixed focus, direction of view, rotation of view, illumination, scope resolution, dial actuation forces, temperature testing, field strength testing of magnets, electrical safety, EMC testing, durability testing, environmental conditioning, compression testing, random vibration testing, and shock (free fall drop) testing. Temperature testing consisted of attaching thermocouples at nine locations using adhesive and measuring the temperature at steady state. Test samples were connected to a 300W Xenon light source at 100% light output with both 3.5mm and 5.0mm light cables. The time to heat and time to cool was also evaluated. Clinical data was not necessary for the subject device. The performance data demonstrates that the Acclarent Cyclops Multi-Angle Endoscope performs as
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top of the "A" to the right, ending above the "t". The font is black, and the background is white.
・
.
| | intended according to IEC 60601-2-18, IEC 60601-1-2,<br>ISO10993-1, ISO 8600-3, ISO 8600-5. |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Validated Reprocessing<br>Methods: | Full manual cleaning with extended enzymatic soak<br>plus general instrument automated washer Pre-vacuum steam sterilization (wrapped) |
| | In validating the above, the following standards were<br>referenced: AAMI/ANSI ST35, AAMI TIR 12, AAMI TIR 30,<br>and ANSI/AAMI ST79 A1/A2, ANSI/AAMI ST8,<br>ANSI/AAMI ST81, ASTM E1766. |
| Summary of Substantial<br>Equivalence: | The Acclarent Cyclops Multi-Angle Endoscope is substantially<br>equivalent to the predicate devices as confirmed through<br>relevant performance tests and attributes. |
. . . . .
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized image of an eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the eagle. The eagle is depicted in black, and the text is also in black.
## Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 2 7 2011
Acclarent, Inc. c/o Ms. Keri Yen. Regulatory Affairs Manager 1525-B O'Brien Drive Menlo Park, CA 94025
Re: K110097
Trade/Device Name: Acclarent Cyclops Multi-Angle Endoscope (Model CYE002) Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible or Rigid) Regulatory Class: Class II Product Code: EOB Dated: April 26, 2011 Received: April 27, 2011
Dear Ms. Yen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Ekhm
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Use: The Acclarent Cyclops Multi-Angle Endoscope is intended to provide an endoscopic means to view the nasal cavity and nasopharynx in an operating room setting.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Rudy CRNP
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
K110097. 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.