TAKE 1 ADVANCED RIGID TRAY

K110092 · Kerr Corporation · ELW · Apr 8, 2011 · Dental

Device Facts

Record IDK110092
Device NameTAKE 1 ADVANCED RIGID TRAY
ApplicantKerr Corporation
Product CodeELW · Dental
Decision DateApr 8, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3660
Device ClassClass 2

Indications for Use

Take 1 Advanced Rigid Tray is an addition-cure vinyl polysiloxane dental impression material that is used for all crown and bridge, edentulous, orthodontic and implant impression techniques.

Device Story

Addition-cure vinyl polysiloxane dental impression material; two-part base/catalyst paste system. Available in heavy body viscosity; 1:1 cartridge or 5:1 volume delivery; regular, fast, and super-fast set speeds. Used by dental professionals in clinical settings to create impressions for crowns, bridges, edentulous arches, orthodontics, and implants. Not recommended for full-arch impression trays or clinical situations involving opposing fixed partial dentures. Material provides physical mold of oral structures for subsequent dental restoration or appliance fabrication.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing and biocompatibility evaluation per ISO 10993.

Technological Characteristics

Addition-cure vinyl polysiloxane impression material. Two-part paste/paste system. Heavy body viscosity. Delivered via 1:1 cartridge or 5:1 volume system. Biocompatibility per ISO 10993.

Indications for Use

Indicated for dental impression techniques including crown and bridge, edentulous, orthodontic, and implant procedures.

Regulatory Classification

Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

Special Controls

*Classification.* Class II (Special Controls).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 10092 APR - 8 2011 # Section III - 510(k) Summary #### Submitter: Sybron Dental Specialties, Inc. 1717 W. Collins Avenue Orange, California 92867 (714) 516-7602 - Phone (714) 516-7488 - Facsimile Wendy Garman - Contact Person Date Summary Prepared: April 2011 #### Device Name: - Trade Name Take 1 Advanced Rigid Tray ● - . Common Name - Dental Impression Material - . Classification Name - Impression Material, per 21 CFR § 872.3660 #### Devices for Which Substantial Equivalence is Claimed: - . Kerr Corporation, Take 1 Advanced Rigid Tray, K092176 # Device Description: The device is an addition-cure vinyl polysiloxane dental impression material that is used for all crowns and bridges, edentulous, orthodontic and implant impression techniques. Take 1 Advanced Rigid Tray is a two-part, base/catalyst – paste/paste system. The product is available in a heavy body viscosity, offered in cartridge 1:1 delivery and Volume 5:1 delivery, and in 3 setting speeds: regular set, fast set and super fast set. #### Intended Use of the Device: Take 1 Advanced Rigid Tray is an addition-cure vinyl polysiloxane dental impression material that is used for all crown and bridge, edentulous, orthodontic and implant impression techniques. {1}------------------------------------------------ # Summary of Technological Characteristics: There has been no change made to design, composition, or intended use of Take 1 Advanced Rigid Tray as compared to its predicate. Additionally, Take 1 Advanced Rigid Tray has the same technological characteristics as its predicate. The only change being proposed is adding two new warnings to the Directions For Use. The new warnings state that Take 1 Advanced Rigid Tray is not recommended for use with impression trays spanning a full arch, and Take 1 Advanced Rigid Tray is not recommended for use in clinical situations where the opposing arch is a fixed partial denture. # Non-Clinical Test Data: This 510(k) submission also includes data from bench testing used to evaluate the performance characteristics of Take 1 Advanced Rigid Trav compared to the predicate device, Take 1 Advanced Rigid Tray, K092176. The characteristics evaluated included, but were not limited to Working Time, Setting Time, Compressive Strength, Diametral Strength, Flexural Strength and Shore D Hardness. A biocompatibility study {toxicity} has been completed according to ISO 10993, which demonstrates that Take 1 Advanced Rigid Tray is safe for its intended use. # Clinical Testing: Clinical testing has not been conducted on this product. #### Conclusion: Based upon the biocompatibility test and the bench testing, the clinical performance of Take 1 Advanced Rigid Tray is substantially equivalent to the predicate device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002 Kerr Corporation C/O Ms. Wendy Garman Director, Regulatory Affairs Sybron Dental Specialties 1717 West Collins Avenue Orange, California 92867 APR - 8 2011 Re: K110092 Trade/Device Name: Take I Advanced Rigid Tray Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: II Product Code: ELW Dated: January 10, 2011 Received: January 12, 2011 Dear Ms. Garman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Garman Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOtfices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, for Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Section I -- Indications for Use 510(k) Labeling Change Being Effected Sybron Dental Specialties, Inc. # Indications for Use 510(k) Number (if known): |/ | 009 乙 Device Name: Take 1 Advanced Rigid Tray Indications For Use: Take 1 Advanced Rigid Tray is an addition-cure vinyl polysiloxane dental impression material that is used for all crown and bridge, edentulous, orthodontic and implant impression techniques. Prescription Use _____________________________________________________________________________________________________________________________________________________________ ਮ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Suno Panes (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control and Dental Dovicas 510(k) Number: K110692
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...