← Product Code [OWQ](/productcode/OWQ) · K110076

# REPROCESSED 3D DIAGNOSTIC ULTRASOUND CATHETER (K110076)

_Sterilmed, Inc. · OWQ · Aug 18, 2011 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K110076

## Device Facts

- **Applicant:** Sterilmed, Inc.
- **Product Code:** [OWQ](/productcode/OWQ.md)
- **Decision Date:** Aug 18, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1200
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Reprocessed 3D Diagnostic Ultrasound Catheters are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart. The 3D Diagnostic Ultrasound Catheter provides location information when used with the CARTO® XP EP Navigation System Version 9 or greater.

## Device Story

Reprocessed 3D diagnostic ultrasound catheter; 10 French, 90 cm length. Distal tip contains ultrasound transducer and 3D location sensor; steering mechanism allows tip rotation and variable deflection. Inputs: ultrasound signals for 2D imaging; location data for 3D mapping. Device connects to CARTO® XP EP Navigation System. Used by clinicians in cardiac procedures for visualization of anatomy, physiology, and other devices. Output: visual ultrasound images and 3D location data displayed on navigation system. Benefits: enables real-time anatomical visualization and device tracking during electrophysiology procedures.

## Clinical Evidence

No clinical data. Evidence based on bench testing, including functional performance of ultrasound and 3D location sensor components, mechanical reliability, cleaning/sterilization validation, biocompatibility (ISO 10993-1), and electrical safety (IEC 60601-1, IEC 60601-2-27).

## Technological Characteristics

10 French, 90 cm catheter; ultrasound transducer and 3D location sensor at distal tip. Materials validated via biocompatibility testing. Sterilization via ethylene oxide (ISO 11135). Electrical safety per IEC 60601-1 and IEC 60601-2-27. Connectivity to CARTO® XP EP Navigation System.

## Regulatory Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

## Predicate Devices

- Biosense Webster, Inc. SoundStar™ 3D Diagnostic Ultrasound Catheter ([K070242](/device/K070242.md))
- AcuNav™ Diagnostic Ultrasound Catheter ([K033650](/device/K033650.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |

AUG 1 8 2011

January 10 2011 For Reprocessed 3D Diagnostic Uhrasound Catheters

#### SUMMARY AND CERTIFICATION I.

STE

RILMED IN

A.

| 510(k) Summary         |                                                                                                             |
|------------------------|-------------------------------------------------------------------------------------------------------------|
| Submitter:             | SterilMed, Inc.                                                                                             |
| Contact Person:        | Onya Dendinger<br>11400 73rd Avenue North<br>Maple Grove, MN 55369<br>Ph: 763-488-3410<br>Fax: 763-488-2051 |
| Date Prepared:         | January 10, 2011                                                                                            |
| Trade Name:            | Reprocessed 3D Diagnostic Ultrasound Catheter                                                               |
| Classification Name:   | Diagnostic Intravascular Catheter                                                                           |
| Panel:                 | 74                                                                                                          |
| Classification Number: | 21 CFR § 870.1200                                                                                           |
| Product Code:          | OWQ, Catheter, Intravascular Ultrasound, Reprocessed                                                        |
| Class:                 | II                                                                                                          |

| Predicate Devices:                              | The Reprocessed 3D Diagnostic Ultrasound Catheters are substantially equivalent to the Biosense Webster,<br>Inc. SoundStar™ 3D Diagnostic Ultrasound Catheter (510(k) K070242) and the AcuNav™ Diagnostic<br>Ultrasound Catheter (510(k) K033650).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description:                             | SterilMed Reprocessed 3D Diagnostic Ultrasound Catheters are specially designed catheters that provide<br>two-dimensional imaging using an ultrasound transducer and three-dimensional location information using a<br>location sensor. The ultrasound transducer and location sensor are at the distal tip of the catheter and can be<br>positioned for ultrasound imaging and 3D mapping by a steering mechanism that rotates the catheter tip and<br>provides variable deflection. 3D Diagnostic Ultrasound Catheters incorporate a hand piece, a flexible shaft<br>and a distal tip section containing an ultrasound transducer and a location sensor. The ultrasound transducer<br>of this device has an acoustic array that is identical to the AcuNav™ 10F Diagnostic Ultrasound Catheter.<br>The 3D Diagnostic Ultrasound Catheter's 3D location sensor provides location information to the CARTO®<br>XP EP Navigation System (mapping system). The 3D Diagnostic Ultrasound Catheters is 10 French with a<br>90 cm insertion length. |
|                                                 | Note: Only the catheter is the subject of this submission; accessories and/or any other related equipment are<br>not included in the scope of this submission.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Intended Use:                                   | The Reprocessed 3D Diagnostic Ultrasound Catheters are indicated for intra-cardiac and intra-luminal<br>visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in<br>the heart. The 3D Diagnostic Ultrasound Catheter provides location information when used with the<br>CARTO® XP EP Navigation System Version 9 or greater.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Functional and<br>Safety Testing:               | Representative samples of Reprocessed 3D Diagnostic Ultrasound Catheters were tested to demonstrate<br>appropriate functional characteristics. Process validation testing was performed to validate the cleaning and<br>sterilization procedures as well as device packaging. In addition, the manufacturing process includes visual<br>and validated functional testing of all products produced.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Summary of Non-<br>clinical Tests<br>Conducted: | Specific non-clinical tests performed included: cleaning validation, sterilization validation (ISO 11135, USP<br><71>), biocompatibility testing (ISO 10993-1), ethylene oxide residual testing (ISO 10993-7), biological<br>testing, cleaning residuals, packaging validation (ASTM D4169, ASTM F88, ASTM F2096), and shelf life<br>validation (ASTM 1980-99). In addition, functional performance (including both the ultrasound and 3D<br>location sensor components) and mechanical reliability were validated using bench and laboratory testing,<br>The testing which was performed include Hipot testing (IEC 60601-1:2005, Cause 8.8.3), defibrillation<br>testing (IEC 60601-1:2005, Clause 8.5.5.1), and temperature heating (IEC 60601-2-27).                                                                                                                                                                                                                                                                                       |

Page 1 of 2

{1}------------------------------------------------

# STERILME

Small Equipment & Instrument Repair

### January 10 2011 For Reprocessed 3D Diagnostic Ultrasound Catheters

| Conclusion: | The Reprocessed 3D Diagnostic Ultrasound Catheters are substantially equivalent to the Biosense Webster, Inc. SOUNDSTAR™ 3D Diagnostic Ultrasound Catheter and AcuNav™ Diagnostic Ultrasound Catheter. |
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|             | This conclusion is based upon the devices' similarities in functional design (principle of operation), materials, indications for use and methods of construction.                                     |

Page 2 OF 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three curved lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

SterilMed, Inc. c/o Ms. Onya Dendinger Regulatory Affairs Manager 11400 73td Avenue North Maple Grove, MN 55369

AUG 1 8 2011

Re: K110076

Trade/Device Name: Reprocessed 3D Ultrasound Catheter (See Enclosed List) Regulatory Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: II (two) Product Code: 74 OWQ Dated: July 15, 2011 Received: July 18, 2011

Dear Ms. Dendigner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 – Ms. Onya Dendinger

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman. M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Page 3 – Ms. Onya Dendinger

## DEVICE MODEL COVERED BY THIS SUBMISSION:

:

| Manufacturer                                                                                                                                                                  | Trade Name | Description                                                                                     |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-------------------------------------------------------------------------------------------------|
| Company of the considered to a comments of the considered and any and any and any and any and any and any and any and any and any and any and any and any and any and any and |            | Biosense Webster, Inc.   SoundStar™ 3D Ultrasound Catheter   SNDSTRI0 - 10 French, 90 cm length |

.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for STERILMED, INC. The text below the company name indicates that they specialize in medical device reprocessing. They also offer small equipment and instrument repair services. The logo is simple and professional, with a focus on the company name.

# Indications for Use

510(k) Number (if known):

KIDO76

Fe(III) TAMLs (H-TAMLs)

Device Name: Reprocessed 3D Diagnostic Ultrasound Catheters

Indications for Use:

The Reprocessed 3D Diagnostic Ultrasound Catheters are indicated for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the 3D Diagnostic Ultrasound Catheter provides location information when used with the CARTO® XP EP Navigation System Version 9 or greater.

| Prescription Use            | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) |   |

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) Number k//0076

---

**Source:** [https://fda-staging.innolitics.com/device/K110076](https://fda-staging.innolitics.com/device/K110076)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
