AQUILION PRIME

K110066 · Toshiba America Medical Systems, In.C · JAK · Jan 26, 2012 · Radiology

Device Facts

Record IDK110066
Device NameAQUILION PRIME
ApplicantToshiba America Medical Systems, In.C
Product CodeJAK · Radiology
Decision DateJan 26, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2

Indications for Use

This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head. The Aquilion Prime has the capability to provide volume sets. These volume sets can then be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

Device Story

TSX-302A Aquilion Prime is an 80-row CT scanner; acquires cross-sectional X-ray volume data of whole body including head. Operates in clinical settings; used by trained physicians/technicians. Transforms X-ray attenuation data into volumetric image sets. Output viewed by clinicians for diagnostic assessment and specialized studies. Facilitates clinical decision-making through high-resolution anatomical visualization. Benefits patients via diagnostic imaging capabilities.

Clinical Evidence

No clinical data provided. Device safety and performance supported by bench testing, adherence to ISO-13485 quality systems, and compliance with IEC safety standards and 21 CFR 1020 radiation safety requirements.

Technological Characteristics

80-row CT scanner; X-ray energy source. Design conforms to IEC safety standards and 21 CFR 1020. Manufactured under ISO-13485 and 21 CFR 820 quality system requirements.

Indications for Use

Indicated for whole body, including head, cross-sectional volume acquisition and display in patients requiring CT imaging. For use by trained and qualified physicians.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 2 6 2012 ## Section 5. 510(k) Summary | Date: | December 22, 2010 | |-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer: | Toshiba Medical Systems Corporation<br>1385 Shimoishigami, Otawara-shi,<br>Tochigi-ken, 324-8550, Japan | | Initial Importer/Distributor:<br>Address:<br>Contact: | Toshiba America Medical Systems, Inc.<br>P.O. Box 2068, 2441 Michelle Drive,<br>Tustin, CA 92781-2068<br>Paul Biggins, Director Regulatory Affairs<br>(714)730-5000 | | Establishment Registration<br>Number: | 2020563 | | Device Proprietary Name: | TSX-302A, Aquilion Prime | | Common Name: | Scanner, Computed Tomography, X-Ray | | Classification: | 90-JAK | | Regulatory Class: | Il (per 21 CFR 892.1750) | | Performance Standard: | None | | Predicate Device(s): | Toshiba TSX-301A/2, k093891 | #### Reason for Submission New Device ## Description of this Device: The TSX-302A is an 80-row CT System that is intended to produce axial scans of the whole body to include the head. The TSX-302A is based upon the technology and materials of previously marketed Toshiba CT systems. {1}------------------------------------------------ ### Summary of Intended Uses: This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head. The Aquilion Prime has the capability to provide volume sets. These volume sets can then be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician. #### Substantial Equivalence: This device is substantially equivalent to the predicate devices which are commercially available at this time. Toshiba Aquilion ONE TSX-301A/2; k093891 ## Safety : This device is designed and manufactured under ISO-13485 to include meeting the requirements of 21 CFR 820. Additionally this system is designed and manufactured in conformance with the appropriate IEC safety standards. This includes the employment of risk mitigation during the development of this device. Radiation safety is assured by meeting the associated requirements of 21 CFR 1020. This information is provided to the FDA in compliance with the required CT product reports. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Toshiba Medical Systems Corporation % Mr. Paul Biggins Director Regulatory Affairs Toshiba America Medical Systems, Inc. P.O. Box 2068, 2441 Michelle Drive TUSTIN CA 92781-2068 JAN 2 6 2012 Re: K110066 Trade/Device Name: TSX-302A, Aquilion Prime CT System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: December 13, 2011 Received: December 14, 2011 Dear Mr. Biggins: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {3}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket writication. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Mary S Patel Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): _ Device Name: . TSX-302A, Aquilion Prime CT System Indications for Use: This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head. The Aquilion PRIME has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Mary Stott (Division Sign-Off) Division of Radiological Office of In Vitro Diagnost 510K. K110066
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...