The Reverse Medical ReFlex™ Guide Catheter is indicated for the introduction of interventional devices into the peripheral, and neuro vasculature.
Device Story
ReFlex™ Guide Catheter is a single-lumen, flexible, variable-stiffness composite catheter; used to introduce interventional devices into peripheral and neuro vasculature. Features PTFE-lined polymeric shaft with hydrophilic coating; Nitinol support; radiopaque distal marker; proximal luer fitting for accessory attachment/infusion. Operated by physicians in clinical settings; provides access for guidewires up to 0.038 inches. Device is sterile, single-use. Facilitates navigation through complex anatomy; aids in placement of interventional tools. Benefits include improved deliverability and compatibility during neuro/peripheral procedures.
Clinical Evidence
Bench testing only. Included dimensional/visual inspection, guidewire compatibility, torque response/strength, kink resistance, flexibility, tensile strength, leak/burst testing, aspiration, and corrosion resistance. Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity). In vivo testing confirmed deliverability, compatibility, visibility, aspiration performance, and acute histopathology of treated vessels.
Technological Characteristics
Single-lumen, variable-stiffness composite catheter. Materials: PTFE-lined polymer, Nitinol shaft support, hydrophilic coating. Radiopaque distal marker. Dimensions vary by size. Connectivity: None. Sterilization: Ethylene Oxide (EtO) per ANSI/AAMI/ISO 11135. No software or electronic components.
Indications for Use
Indicated for the introduction of interventional devices into the peripheral and neuro vasculature. Intended for prescription use.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Penumbra Neuron™ Intracranial Access System (K070970)
{0}------------------------------------------------
# 4. 510(k) Summary according to 807.92
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Reverse Medical Corporation is providing the summary of Substantial Equivalence for the ReFlex™ Guide Catheter,
# 4.1 Sponsor /Applicant Name and Address
Reverse Medical Corporation 13900 Alton Parkway Suite 123 Irvine, CA 92618
#### 4.2 Sponsor Contact Information
Jeff Valko President/CEO Phone: 949-215-0660 949-215-0661 Fax: email: jvalko@reversemed.com
#### 4.3 Date of Preparation of 510(k) Summary
March 10, 2011
# 4.4 Device Trade or Proprietary Name
ReFlex™ Guide Catheter
#### 4.5 Device Common/Usual or Classification Name
Catheter, Percutaneous (Product Code: DQY)
### 4.6 Identification of the Legally Marketed Devices to which Equivalence is Being Claimed:
| Name of Predicate Device | Name of Manufacturer<br>(Town, State) | 510(k) Number |
|------------------------------------------------|-----------------------------------------------|---------------|
| Penumbra Neuron™<br>Intracranial Access System | Penumbra, Inc.<br>Alameda, CA | K070970 |
| HD Guide Catheter | Concentric Medical, Inc.<br>Mountain View, CA | K090335 |
APR - 5 2011
{1}------------------------------------------------
### 510(k) Summary according to 807.92 (continued)
### 4.7 Device Description
The ReFlex™ Guide Catheter is a single lumen, flexible, variable stiffness composite catheter. The catheter shaft has a hydrophilic coating to reduce friction during use. The ReFlex™ Guide Catheter shaft is visible under fluoroscopy. The ReFlex™ Guide Catheter dimensions are included on the individual device label. The ReFlex™ Guide Catheter inner lumen can accommodate guidewires up to 0.038 inches in diameter to aid in placement of the catheter system.
The proximal end of the ReFlex™ Guide Catheter has a luer fitting to allow attachment of accessories and infusion of liquids through the system. The ReFlex™ Guide Catheter is offered in various sizes to accommodate physician preferences and anatomical variations. The catheter is provided sterile, nonpyrogenic, and is intended for single use only.
### 4.8 Intended Use
The Reverse Medical ReFlex™ Guide Catheter is indicated for the introduction of interventional devices into the peripheral, and neuro vasculature.
| | | 4.9 Comparison to Predicate Devices | |
|--|--|-------------------------------------|--|
|--|--|-------------------------------------|--|
| | Penumbra Neuron™<br>Intracranial Access<br>System | HD Guide Catheter | ReFlex™ Guide<br>Catheter |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K070970 | K090335 | TBD |
| Classification | Class II, DQY | Class II, DQY | Class II, DQY |
| Indication | The ... is indicated for<br>the introduction of<br>interventional devices<br>into the peripheral,<br>coronary, and neuro<br>vasculature. | The ... is indicated for<br>use in facilitating the<br>insertion and guidance<br>of an occlusion<br>catheter, infusion<br>catheter or other<br>appropriate<br>microcatheter into a<br>selected blood vessel<br>in the peripheral,<br>coronary and neuro<br>vascular systems. | The ... is indicated for<br>the introduction of<br>interventional devices<br>into the peripheral, ,<br>and neuro vasculature. |
| Materials | | | |
| - Shaft Materials | PTFE lined<br>nylon/polyurethane<br>catheter,<br>with hydrophilic coating | PTFE lined<br>polymeric<br>catheter<br>with hydrophilic coating | PTFE lined<br>polymeric<br>catheter,<br>with hydrophilic coating |
| - Catheter shaft support | Stainless Steel | Stainless Steel | Nitinol |
| - Proximal End<br>Configuration | Luer Hub | Luer Hub | Luer Hub |
| - Radiographic markers /<br>Radiopacity | • Radiopaque marker<br>at distal tip | • Radiopaque<br>marker at distal tip | • Radiopaque marker<br>at distal tip |
{2}------------------------------------------------
### 510(k) Summary according to 807.92 (continued) 4.9 Comparison to Predicate Devices (continued)
| | Penumbra Neuron™<br>Intracranial Access<br>System | HD Guide Catheter | ReFlex™ Guide<br>Catheter |
|-------------------------------|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| -Packaging | Catheter<br>attached to<br>packaging card<br>inside<br>PET/PE/Tyvek pouch<br>inside SBS carton | Catheter<br>in polyethylene hoop<br>inside<br>PET/PE/Tyvek pouch<br>inside SBS carton | Catheter<br>in polyethylene hoop<br>attached to<br>packaging card<br>inside<br>PET/PE/Tyvek pouch<br>inside SBS carton |
| Packaged Accessory<br>Devices | Packaged with an<br>rotating hemostatic<br>valve (RHV) | Packaged with an<br>rotating hemostatic<br>valve (RHV) | Not Applicable |
| Sterilization | EtO | EtO | EtO |
### 4.10 Summary of Non-clinical Data
# 4.10.1 Biocompatibility and Sterilization
The ReFlex™ Guide Catheter is classified as Externally Communicating Devices, Circulating Blood, Limited Contact (≤ 24 hours). Results of the testing demonstrate that the blood contacting materials are biocompatible.
Blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO 10993-1 guidelines "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". The ReFlex™ Guide Catheter successfully passed all of the following biocompatibility tests:
| Test | Method |
|----------------------------------------|--------------------------------------------------------|
| Cytotoxicity | L929 MEM Elution Test |
| Sensitization | Kligman Maximization |
| Intracutaneous Reactivity (Irritation) | Intracutaneous Injection Test |
| Systemic Toxicity (Acute) | ISO Acute Systemic Injection Test |
| Hemocompatibility | Complement Activation |
| | Hemolysis |
| | Inactivated Partial Thromboplastin Time Test |
| | <i>In vivo</i> thrombogenicity |
| Pyrogenicity | USP Material Mediated Rabbit Pyrogen Test |
| EtO Residuals | Ethylene oxide and Ethylene chlorohydrins<br>residuals |
Sterilization conditions have been validated according to ANSI / AAMI / ISO 11135, Sterilization of Health Care Products-Part 1: Requirement, validation and routine control of a sterilization process for medical devices to provide a Sterility Assurance Level (SAL) of 10 °.
# । ਤੇ
{3}------------------------------------------------
# 510(k) Summary according to 807.92 (continued) 4.10.2 Design Verification (Bench-Top Testing)
The physical, mechanical and performance testing of the ReFlex™ Guide Catheter demonstrates that the product is substantially equivalent to the currently marketed predicate devices. Design Verification testing was conducted to evaluate the physical and mechanical properties of the ReFlex™ Guide Catheter. All studies were conducted in accordance with Reverse Medical Design Control procedures. All testing was performed on units which were sterilized and met all inspection criteria. Tests on the ReFlex™ Guide Catheter included:
Verification and Test Summary
| In vitro Tests | Result |
|-----------------------------------------------------------------------------------|--------------------------|
| Dimensional and Visual Inspection | Met established criteria |
| Guidewire Compatibility | Met established criteria |
| Torque Response | Met established criteria |
| Torque Strength | Met established criteria |
| Kink Resistance | Met established criteria |
| Flexibility Test | Met established criteria |
| Tensile Strength | Met established criteria |
| Catheter Leak Test (Liquid Leakage) | Met established criteria |
| Catheter Leak Test (Air Leakage) | Met established criteria |
| Dynamic Pressure Test | Met established criteria |
| Static Burst Test | Met established criteria |
| Aspiration Test | Met established criteria |
| Hub Gauging | Met established criteria |
| Corrosion Resistance | Met established criteria |
| USP Particulate Test | Met established criteria |
| Navigation and Accessibility Capabilities in vitro | Met established criteria |
| In vivo Tests | Result |
| System<br>Deliverability, Compatibility,<br>Visibility and Aspiration Performance | Met established criteria |
| Acute histopathology of treated vessels | Met established criteria |
| Biocompatibility testing | Met established criteria |
The physical, mechanical and performance testing of the subject ReFlex™ Guide Catheter demonstrate that the product is safe and effective for its labeled indications and is Substantially Equivalent to the currently marketed predicate devices.
### 4.11 Substantial Equivalence
The performance of the ReFlex™ Guide Catheter in this submission demonstrates that the product is substantially equivalent to the performance of the predicate devices. The equivalence was shown through comparison of component materials and specification, performance and biocompatibility testing and sterilization validation.
The ReFlex™ Guide Catheter is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the predicate devices. Differences between the devices do not raise any significant issues of safety or effectiveness.
14
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three wing-like shapes extending upwards. The bird is positioned within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in capital letters. The text is arranged around the circumference of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Reverse Medical Corporation c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25° St. NW Buffalo, MN 55313
APR - 5 2011
Re: K110055
Trade/Device Name: ReFlex™ Guide Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: March 18, 2011 Received: March 21, 2011
Dear Mr. Job
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Duna R. Vahner
Image /page/5/Picture/7 description: The image shows a handwritten symbol or character. It appears to be a stylized letter or a combination of lines and curves. The symbol is drawn with thick strokes, and it has a somewhat angular appearance.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# 3. Indications for Use
510(k) Number (if known):_K 110055
Device Name: Reverse Medical ReFlex™ Guide Catheter
Indications for Use:
The Reverse Medical ReFlex™ Guide Catheter is indicated for the introduction of interventional devices into the peripheral, and neuro vasculature.
Prescription Use______________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over the Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dana R. Vahine
on of Cardiovascular Devices
510(k) Number_K 110055
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.