The Spacelabs Pathfinder SL Holter Analyzer is intended to be used to analyze recordings of ambulatory electrocardiograms made on compatible Holter recorders. It is capable of detecting certain abnormal arrhythmias, and allows the operator to view and edit the ECG and the analysis results, and construct a report for physician use. The Spacelabs Pathfinder SL Holter Analyzer is intended for use on adult patients only as a screening device to determine the need for clinical diagnosis and evaluation by polysomnography based on the patient's score. The ECG recording may be obtained at any location specified by a physician including home, hospital or clinic. Subjects screened for sleep apnea should have periods of sleep ofat least 4 hours duration during which the ECG is predominantly sinus rhythm in nature.
Device Story
Pathfinder SL is a software-only ECG analysis system for Windows PCs; integrates with Spacelabs Sentinel Cardiology Information Management System. Inputs: ambulatory ECG data (up to 24 hours of 12-lead or 7 days of 3-lead) from compatible Holter recorders (Lifecard CF, Lifecard 12, evo, Aria). Processing: automated analysis of arrhythmia, ST segment, heart rate variability, and pacer performance. Output: ECG reports for physician review/diagnosis. Used in clinical settings by trained operators. Benefits: facilitates screening for sleep apnea and arrhythmia detection; assists clinicians in diagnostic decision-making.
Clinical Evidence
Bench testing only. Performance verified against internal requirements and recognized standards: IEC 60601-2-47 (ambulatory ECG safety/performance), ANSI/AAMI EC57 (cardiac rhythm/ST-segment measurement algorithms), and ANSI/AAMI EC38 (ambulatory ECGs). No clinical data required.
Technological Characteristics
Software-only device; no patient contact. Operates on Windows XP, Vista, or 7 (32-bit) PC platforms. Compatible with Spacelabs Sentinel system. Implements automated ECG analysis algorithms for arrhythmia, ST segment, heart rate variability, and pacer performance. Complies with IEC 60601-2-47, ANSI/AAMI EC57, and ANSI/AAMI EC38.
Indications for Use
Indicated for adult patients as a screening tool to determine the need for clinical diagnosis and evaluation by polysomnography for sleep apnea. Requires ECG recordings with at least 4 hours of sleep duration and predominantly sinus rhythm.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
# 长110001
JUN 1 4 2011
# 510(k) Premarket Notification Spacelabs Healthcare Spacelabs Pathfinder SL Holter Analyzer 510(k) Summary
| Submission Date: | 25 May 2010 |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Spacelabs Healthcare Ltd.<br>1 Harforde Court, John Tate Road<br>Hertford, SG13 7NW United Kingdom |
| Submitter Contact: | Mr. Roger Moldon<br>Quality Manager<br>Spacelabs Healthcare Ltd.<br>1 Harforde Court, John Tate Road<br>Hertford, SG13 7NW United Kingdom<br>+011 (44) 1992-507730<br>roger.moldon@spacelabs.com |
| Official Contact: | Thomas Kroenke<br>Principal Consultant<br>Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 USA<br>tkroenke@speedtomarket.net<br>303 956 4232 |
| Manufacturing Site: | Spacelabs Healthcare<br>5150 220th Avenue SE<br>Issaquah, WA 98029 |
| Trade Name: | Spacelabs Model 90217A Ambulatory Blood Pressure (ABP) Monitor |
| Common Name: | Computer, Diagnostic, Programmable |
| Classification Name: | Holter Analyzer |
| Classification<br>Regulation: | 21 CFR §870.1425 |
| Product Code: | DOK |
:
:
{1}------------------------------------------------
### 510(k) Premarket Notification Spacelabs Healthcare Spacelabs Pathfinder SL Holter Analyzer 510(k) Summary
| Substantially<br>Equivalent Devices: | New Spacelabs Model | Predicate<br>510(k) Number | Predicate<br>Manufacturer / Model |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|------------------------------------------------------------------------------------------------------------------|
| | Spacelabs Pathfinder SL<br>Holter Analyzer | K993620 | Spacelabs (Del Mar<br>Medical Systems)<br>Impresario™ Ambulatory<br>ECG (Holter) Arrhythmia<br>Analysis Software |
| | | K951902 | Spacelabs (Reynolds<br>Medical Ltd) Pathfinder<br>700 Holter Analyzer |
| | | K042745 | Spacelabs (Del Mar<br>Reynolds Medical, Inc.)<br>Lifescreen Apnea |
| Device Description: | Spacelabs Pathfinder SL Holter Analyzer (Pathfinder SL) is a software-only<br>ECG analysis system for recording and analyzing up to 24 hours of 12 lead<br>EGC or seven days of continuous 3 lead ambulatory ECG. Pathfinder SL is<br>designed to operate on a Microsoft Windows® based personal computer (PC)<br>platform, and contains provisions for networking. Pathfinder SL is compatible<br>with Windows XP, Vista, or Windows 7 32-bit (ensure the latest service packs<br>installed) operating systems. | | |
| | Pathfinder SL is used in conjunction with Spacelabs Healthcare's Sentinel<br>Cardiology Information Management System (Sentinel), and is sold only for<br>use with Sentinel. | | |
| | In conjunction with Sentinel, Pathfinder SL can acquire data from Spacelabs<br>Holter recorders, analyze up to 12 channels of ECG data, providing<br>classification, editing, and reporting of arrhythmia, ST segment,<br>heart rate variability and pacer performance, and allows the operator to<br>construct a report for physicians to use within their diagnosis of patients.<br>Pathfinder SL is compatible with Lifecard CF, Lifecard 12, evo and Aria<br>Holter Recorders. | | |
| | ECG from pre-existing stored analyses created with the previous Spacelabs<br>Healthcare Impresario or Pathfinder Holter analyzer systems can be analyzed<br>from scratch using Pathfinder SL. | | |
| | Pathfinder SL should be used by an operator trained in electrocardiography. A<br>qualified medical practitioner should make any diagnosis. | | |
KII0001 Page 2 of 4
{2}------------------------------------------------
## 510(k) Premarket Notification Spacelabs Healthcare Spacelabs Pathfinder SL Holter Analyzer 510(k) Summary
| Intended Use: | The Spacelabs Pathfinder SL Holter Analyzer is intended to be used to<br>analyze recordings of ambulatory electrocardiograms made on compatible<br>Holter recorders. It is capable of detecting certain abnormal arrhythmias, and<br>allows the operator to view and edit the ECG and the analysis results, and<br>construct a report for physician use.<br>The Spacelabs Pathfinder SL Holter Analyzer is intended for use on adult patients only as a screening device to determine the need for clinical diagnosis and evaluation by polysomnography based on the patient's score. The ECG recording may be obtained at any location specified by a physician including home, hospital or clinic. Subjects screened for sleep apnea should have periods of sleep ofat least 4 hours duration during which the ECG is predominantly sinus rhythm in nature. | | | | | |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|
| Technology<br>Comparison: | The Pathfinder SL employs the same technological characteristics as the<br>predicate device. | | | | | |
| Summary of Performance Testing: | | | | | | |
| Biocompatibility | The Pathfinder SL is software only and does not directly or indirectly contact<br>the patient. Therefore, biocompatibility testing is not necessary for the<br>Pathfinder SL. | | | | | |
| Electrical Safety | The Pathfinder SL is software only. Therefore, electrical safety testing is not<br>necessary for the Pathfinder SL. | | | | | |
| Electromagnetic<br>Compatibility (EMC)<br>Testing | The Pathfinder SL is software only. Therefore, EMC testing is not necessary<br>for the Pathfinder SL. | | | | | |
| Performance Testing | The Pathfinder SL was tested for performance in accordance with internal<br>requirements and the applicable portions of the following Standards: | | | | | |
| | • IEC 60601-2-47: 2001, Medical electrical equipment - Particular<br>requirements for the safety including essential performance, of<br>ambulatory electrocardiographic systems; | | | | | |
| | • ANSI/AAMI EC57: 1998, Testing and reporting performance results of<br>cardiac rhythm and ST-segment measurement algorithms; and | | | | | |
| | • ANSI/AAMI EC38: 1998, Ambulatory electrocardiographs | | | | | |
| | Test results indicated that the Pathfinder SL complies with its predetermined<br>specification and with the applicable Standards. | | | | | |
KINDOOV Page 3 of 4
{3}------------------------------------------------
#### 510(k) Premarket Notification Spacelabs Healthcare Spacelabs Pathfinder SL Holter Analyzer 510(k) Summary
Software Testing
Software device modifications made to the Pathfinder SL were designed and developed according to a robust software development process, and were rigorously verified and validated.
Software information is provided in accordance with:
- . FDA guidance: The content of premarket submissions for software contained in medical devices, 11 May 05;
- . FDA guidance: Off-the-shelf software use in medical devices, 09 Sep 99; and
- . FDA guidance: General principles of software validation; Final guidance for industry and FDA staff, 11 Jan 02.
Test results indicate that the Pathfinder SL complies with its predetermined specification.
#### Conclusion
Verification and validation activities were conducted to establish the performance and safety characteristics of the software device modifications made to the Pathfinder SL. The results of these activities demonstrate that the Pathfinder SL is safe and effective when used in accordance with its intended use and labeling.
Therefore, the Pathfinder SL is considered substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with a more abstract and flowing design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
Spacelabs Healthcare Ltd. c/o Mr. Thomas Kroenke Speed to Market, Inc. PO Box 3018 Nederland, CO 80466
JUN 1 4 2011
Re: K110001
Trade Name: Spacelabs Pathfinder SL Holter Analyzer Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK and MNR Dated: May 28, 2011 Received: June 1, 2011
Dear Mr. Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 – Mr. Thomas Kroenke
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman. M. C Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
| 510(k) Number (if known): | K110001 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Spacelabs Pathfinder SL Holter Analyzer |
| Indications for Use: | The Spacelabs Pathfinder SL Holter Analyzer is intended to be<br>used to analyze recordings of ambulatory electrocardiograms<br>made on compatible Holter recorders. It is capable of detecting<br>certain abnormal arrhythmias, and allows the operator to view<br>and edit the ECG and the analysis results, and construct a report<br>for physician use.<br><br>The Spacelabs Pathfinder SL Holter Analyzer is intended for use<br>on adult patients only as a screening device to determine the<br>need for clinical diagnosis and evaluation by polysomnography<br>based on the patient's score. The ECG recording may be obtained<br>at any location specified by a physician including home, hospital<br>or clinic. Subjects screened for sleep apnea should have periods<br>of sleep ofat least 4 hours duration during which the ECG is<br>predominantly sinus rhythm in nature. |
X_____________________________________________________________________________________________________________________________________________________________________________ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 807 Subpart C)
1
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number_________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.