← Product Code [MTJ](/productcode/MTJ) · K103822

# SYNTHES HEMOSTATIC BONE PUTTY (HBP) (K103822)

_Synthes USA Products, LLC · MTJ · May 19, 2011 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K103822

## Device Facts

- **Applicant:** Synthes USA Products, LLC
- **Product Code:** [MTJ](/productcode/MTJ.md)
- **Decision Date:** May 19, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

Synthes Hemostatic Bone Putty is indicated for use as a water-soluble implant material for use in the control of bleeding from bone surfaces.

## Device Story

Synthes Hemostatic Bone Putty (HBP) is a synthetic, water-soluble hemostatic agent; composed of alkylene oxide polymers and carboxymethylcellulose (CMC). Applied by surgeons to bone edges damaged by trauma or surgical cutting; forms mechanical barrier to occlude vascular openings; stops bleeding. Dissolves postoperatively; eliminated from defect site in <48 hours; permits normal tissue healing and bone regeneration. Used in clinical/surgical settings.

## Clinical Evidence

Bench testing only. Non-clinical studies included: cytotoxicity (ISO elution), mouse peripheral blood micronucleus study, ISO modified intracutaneous study, bacterial reverse mutation assay, mouse lymphoma assay, ISO guinea pig maximization sensitization test, and systemic toxicity. In vivo studies: resorption rate in rat craniotomy model; subcutaneous resorption in rabbit (2, 4, 7, 14 days); sheep vertebral body defect evaluation (7 days); and bone healing effect in rat craniotomy model (3, 6, 12 weeks).

## Technological Characteristics

Synthetic water-soluble alkylene oxide polymers blended with carboxymethylcellulose (CMC). Form factor: putty. Mechanism: mechanical barrier/occlusion. Resorption: <48 hours. Sterilization: not specified.

## Predicate Devices

- Ostene CT Soluble Hemostasis Implant Material ([K082491](/device/K082491.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K103822

1 of 2

## 1. 510(k) Summary

. . . . . .

.

| Sponsor:               | Synthes USA Products, LLC<br>1230 Wilson Drive<br>West Chester, PA 19380                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |  | MAY 19 2011 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-------------|
| Date Prepared          | April 20, 2011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |             |
| Company<br>Contact     | Jeffrey L. Dow, JD<br>Director, Clinical & Regulatory Affairs<br>Synthes Biomaterials<br>484 356 9720<br>dow.jeffrey@synthes.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |  |             |
| Device Name:           | Synthes Hemostatic Bone Putty                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |  |             |
| Classification:        | Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |  |             |
| Product Code           | MTJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |             |
| Predicate<br>Devices:  | Ceremed, Inc.<br>Ostene CT Soluble Hemostasis Implant Material<br>K082491                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |             |
| Device<br>Description: | Synthes Hemostatic Bone Putty (HBP) stops bone bleeding<br>by establishing a physical barrier along the edges of bones<br>that have been damaged by trauma or cut during the surgical<br>procedure. When applied as directed, HBP forms a<br>mechanical barrier that occludes the vascular openings in the<br>damaged bone. This barrier prevents further bleeding during<br>the surgical procedure and dissolves postoperatively,<br>permitting normal tissue healing and bone regeneration. HBP<br>is a blend of synthetic water soluble polymers that form a<br>ready-to-use hemostatic agent that is substantially eliminated<br>from the defect site in less than 48 hours. |  |             |
|                        | The constituents of Synthes Hemostatic Bone Putty and<br>Ostene, the predicate, are similar. Ostene is comprised of a<br>proprietary mixture of water soluble alkylene oxide<br>copolymers. HBP is also comprised of water soluble alkylene<br>oxide polymers. The remainder of HBP is a polysaccharide,<br>carboxymethylcellulose (CMC), to improve handling. Ostene<br>does not contain CMC.                                                                                                                                                                                                                                                                                   |  |             |

{1}------------------------------------------------

| Non-clinical tests<br>used for<br>substantial<br>equivalence<br>comparison | Cytotoxicity Study Using the ISO Elution Method – 1X MEM Extract Mouse Peripheral Blood Micronucleus Study ISO Modified Intracutaneous Study, Solution with Measurement Genotoxcity: Bacterial Reverse Mutation Assay Genotoxicity: Mouse Lymphoma Assay ISO Guinea Pig Maximization Sensitization Test-Solution Systemic Toxicity Study An In Situ Study to Determine the HBP Resorption Rate in a Rat Craniotomy Model <i>In Vivo</i> Resorption Rate of a Hemostatic Bone Putty Subcutaneously Implanted in the Rabbit at 2, 4, 7, 14 Days. <i>In Vivo</i> evaluation of Hemostatic Bone Putty in a sheep vertebral body defect at 7 days Evaluation of Hemostatic Bone Putty in a Sheep Vertebral Body Defect An <i>In Vivo</i> Study to Determine Hemostatic Bone Putty Effect on Bone Healing In A Rat Craniotomy Model at 3, 6, and 12 Weeks |
|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use:                                                              | Synthes Hemostatic Bone Putty is indicated for use as a water-soluble implant material for use in the control of bleeding from bone surfaces.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Substantial<br>Equivalence:                                                | Documentation is provided that demonstrates that Synthes Hemostatic Putty is substantially equivalent' to other legally                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |

I The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence found in the Federal Food, Drug and Cosmetic Act, as amended, 21 USC §301 et seq., and as applied under 21 CFR Part 807, Subpart E, under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalence under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein, shall be construed as an admission against interest under the U.S. patent laws or their application by the courts.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAY 1 9 2011

Synthes USA Products, LLC % Jeffrey L. Dow, JD 1230 Wilson Drive West Chester, Pennsylvania 19380

Re: K103822

Trade/Device Name: Synthes Hemostatic Bone Putty Regulatory Class: Unclassified Product Code: MTJ Dated: April 21, 2011 Received: April 22, 2011

Dear Dr. Dow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

Page 2 - Jeffrey L. Dow, JD

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fdagov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.hun for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

P. Thom. D.E.P. V.A.

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "SYNTHES" in a stylized font. The letters are bold and slightly slanted to the right. The letters are also filled with small dots, giving them a textured appearance. To the left of the word is a logo that appears to be a stylized representation of a medical device or implant. To the right of the word is a registered trademark symbol.

## 1. Indications for Use

510(k) Number (if known): K 10 38 2 2

.. .' .

Indications:

Synthes Hemostatic Bone Putty is indicated for use as a water-soluble implant material and for use in the control of bleeding from bone surfaces.

Prescription Use _ X Over-The-Counter Use AND/OR (Per 21 CFR 801 Subpari D) (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel Keane

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K103822

Confidential: Contains trade secret information

---

**Source:** [https://fda-staging.innolitics.com/device/K103822](https://fda-staging.innolitics.com/device/K103822)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).