The PhysioGlove ES Model I is a reusable 12-lead diagnostic ECG examination system. The system can be used with 12-lead cables or the PhysioGlove device. When used with the PhysioGlove, it is intended to conduct an electrocardiogram. The diagnostic and automatic measurement features are not available. When used with a 12-lead cable, it is intended for use in resting diagnostic electrocardiograph examination of adults. The PhysioGlove ES Model I is intended for use in physician offices, hospitals, outpatient clinics, or similar settings by or on the order of a physician or similarly qualified healthcare professional. The device provides waveform parameters and ECG analysis for healthcare provider interpretation. The PhysioGlove ES Model I ECG Analysis Program assists the physician in measuring and interpreting resting 12-lead EGGs for rhythm and contour information by providing an initial automated interpretation. Interpretation by the product is then confirmed, edited, or deleted by the physician. The analysis program is intended for use in the general population ranging from healthy subjects to patients with cardiac and/or non-cardiac abnormalities. The analysis program is intended for use in hospitals, outpatient clinics, emergency departments, and out-of hospital sites such as ambulances and patients' homes.
Device Story
PhysioGlove ES Model I is a 12-lead diagnostic ECG system. Inputs: ECG signals via standard 10-electrode cable or proprietary PhysioGlove. Operation: Electronics perform analog processing and A/D conversion; digital packets transmitted via USB to medical-grade PC. Software performs patient registration, display, storage, and automated ECG analysis (GE 12SL program). Output: Waveform parameters and automated rhythm/contour interpretation for physician review. Used in hospitals, clinics, emergency departments, ambulances, and homes by physicians or qualified professionals. Physician confirms, edits, or deletes automated interpretations. Benefits: Provides initial automated analysis to assist in clinical decision-making for cardiac rhythm and contour assessment.
Clinical Evidence
No clinical testing performed. Substantial equivalence supported by bench testing, including software verification and validation, usability testing, and compliance with IEC 60601-1-4, IEC 62304, and IEC 60601-2-51 standards.
Technological Characteristics
12-lead diagnostic ECG system. Components: electronics data acquisition unit, medical-grade PC, 10-electrode cable or PhysioGlove. Connectivity: USB. Software: GE 12SL ECG Analysis Program. Standards: IEC 60601-1-4, IEC 62304, IEC 60601-2-51.
Indications for Use
Indicated for resting diagnostic 12-lead ECG examination of adults. Analysis program indicated for general population, including healthy subjects and patients with cardiac or non-cardiac abnormalities, for rhythm and contour interpretation.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
{0}------------------------------------------------
# 510(k) Summary
Commwell Ltd. PhysioGlove ES Model I with ECG Analysis
### 510(k) Number K103791
#### Date:
October 18, 2011
#### Submitter:
Commwell Ltd. Rechov Yad Harutzim 4 Kfar Saba, Israel 44641
#### Contact Person:
Irving Levy Commwell Ltd. · Rechov Yad Harutzim 4 Kfar Saba, Israel 44641 T: (972) 9-766-8094 F: (972) 9-766-8099
## Device Trade Name:
PhysioGlove ES Model I with ECG Analysis
#### Common Name:
12-lead Diagnostic System with ECG Analysis Program
#### Classification:
Electrocardiograph System
## Product Code:
DPS, MHX
#### Class:
ll
#### CFR Reference:
21 CFR 870.2340
#### Predicate Devices:
- K050674, Commwell PhysioGlove ES Model I .
- K092369, GE 12SL ECG Analysis Program .
{1}------------------------------------------------
### Intended Use / Indications for Use:
The PhysioGlove ES Model I is a reusable 12-lead diagnostic ECG examination system. The system can be used with 12-lead cables or the PhysioGlove device. When used with the PhysioGlove, it is intended to conduct an electrocardiogram. The diagnostic and automatic measurement features are not available. When used with a 12-lead cable, it is intended for use in resting diagnostic electrocardiograph examination of adults. The PhysioGlove ES Model I is intended for use in physician offices, hospitals, outpatient clinics, or similar settings by or on the order of a physician or similarly qualified healthcare professional. The device provides waveform parameters and ECG analysis for healthcare provider interpretation.
The PhysioGlove ES Model I ECG Analysis Program assists the physician in measuring and interpreting 12-lead EGGs for rhythm and contour information by providing an initial automated interpretation. Interpretation by the product is then confirmed, edited, or deleted by the physician. The analysis program is intended for use in the general population ranging from healthy subjects to patients with cardiac and/or non-cardiac abnormalities. The analysis program is intended for use in hospitals, outpatient clinics, emergency departments, and out-of hospital sites such as ambulances and patients' homes.
## Device Description:
The PhysioGlove ES Model I is a diagnostic electrocardiograph for 12-Lead resting ECGs. The system consists of an electronics data acquisition unit and software that runs on a medical grade PC. The system uses either a standard, FDA-cleared, 10electrode EGG cable, or the PhysioGlove to obtain ECG data from the patient. The electronics performs analog processing and analog to digital conversion of a 12 lead diagnostic ECG.
The digital packets from the front-end enter the PC via the USB port. The software for the PC performs a variety of functions, including patient registration, ECG display, and storage.
In addition the system provides ECG analysis. The ECG Analysis Program is a software algorithm only. When the system is used with the PhysioGlove, the ECG Analysis Program is not available.
The PhysioGlove ES Module I is intended to be marketed in two configurations.
- Configuration 1 standard ES Model I without ECG analysis. .
- . Configuration 2 - standard ES Model I with ECG analysis.
# Technological Characteristics - Components and Safety Features:
The PhysioGlove ES Model I with ECG analysis is for recording and analyzing diagnostic 12-lead ECG signals. The system is intended to be used with either a standard patient cable or the cleared PhysioGlove device (K083677). When used
{2}------------------------------------------------
with the PhysioGlove, the ECG analysis functionality is disabled. The system employs the same functional scientific technology as its predicate devices.
# Determination of Substantial Equivalence:
The PhysioGlove ES Model I with ECG analysis has been modified from the predicate version (K050674) to include additional features to enhance performance and usability of the device, and to provide additional functionality with the introduction of the ECG analysis package (General Electric (GE) 12SL ECG Analysis Program approved under (K092369)). There have been no changes to the fundamental scientific technology of the system.
# Summary of Non-Clinical testing:
There are no recognized consensus standards applicable to the EGG Analysis System algorithm. The EGG Analysis Program and its applications comply with the Guidances and/or Special Controls as outlined below and detailed in Section 16 of this premarket submission. The following quality assurance measures were applied to the development of the system:
- *Requirements Reviews .
- *Risk Analvsis .
- *Software Verification and Validation .
- *Performance testing .
The device was tested according to the following performance standards:
- Usability testing was done to demonstrate the efficacy of the PhysioGlove in its . intended environment.
- IEC 60601-1-4 Programmable Electrical Medical Systems .
- IEC 62304 Software Life Cycle for Medical Devices .
- IEC 60601-2-51 Particular requirements for safety of electrocardiographs .
In all instances, the device functioned as intended.
# Summary of clinical testing:
The modifications made to the PhysioGlove ES Model I with ECG analysis did not require clinical testing to support substantial equivalence.
# Conclusion:
The summary above shows there are no new questions of safety and effectiveness issues associated with the PhysioGlove ES Model I with ECG analysis when compared to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
OCT 2 1 2011
Crommwell Ltd. c/o Mr. Irving Levy Vice President Rechov Yad Harutzim4 Kfar Saba ISRAEL 44641
Re: K103791
> Trade/Device Name: PhysioGlove ES Model I with ECG Analysis Program Regulatory Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: II (two) Product Code: 74 DPS Dated: September 21, 2011 Received: September 21, 2011
Dear Mr. Levy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have betely promance holdication of intent to market the device
for use stated in the enclosure) to locally market is substantially equivalent (for the i for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of the Federal Food, Drug,
You may, therefore, market the device, subicat to You may, therefore, market the device, subject to the general controls provisions of the Act. The Act. The general controls provisions of the Act include requirements for annual registration, isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Irving Levy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Rev. D. Zacharias, M.P.
Bram D. 2 uckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
## 510(k) Number: K103791
#### Device Name:
PhysioGlove ES Model I with ECG Analysis Program
#### Indications for Use:
The PhysioGlove ES Model I is a reusable 12-lead diagnostic ECG examination system. The system can be used with 12-lead cables or the PhysioGlove device. When used with the PhysioGlove, it is intended to conduct an electrocardiogram. The diagnostic and automatic measurement features are not available. When used with a 12-lead cable, it is intended for use in resting diagnostic electrocardiograph examination of adults. The PhysioGlove ES Model I is intended for use in physician offices, hospitals, outpatient clinics, or similar settings by or on the order of a physician or similarly qualified healthcare professional. The device provides waveform parameters and ECG analysis for healthcare provider interpretation.
The PhysioGlove ES Model I ECG Analysis Program assists the physician in measuring and interpreting resting 12-lead EGGs for rhythm and contour information by providing an initial automated interpretation. Interpretation by the product is then confirmed, edited, or deleted by the physician. The analysis program is intended for use in the general population ranging from healthy subjects to patients with cardiac and/or non-cardiac abnormalities. The analysis program is intended for use in hospitals, outpatient clinics, emergency departments, and out-of hospital sites such as ambulances and patients' homes.
Prescription Use (Per 21 CFR 801.109 subpart D)
Over the Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Qffice of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K103791
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.