K103771 · Letrix Power Hangzhou Pty, Ltd. · INI · May 6, 2011 · Physical Medicine
Device Facts
Record ID
K103771
Device Name
LOTUS BLAKE ESTEEM SCOOTER
Applicant
Letrix Power Hangzhou Pty, Ltd.
Product Code
INI · Physical Medicine
Decision Date
May 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Indications for Use
Esteem scooter is intended for an indoor/outdoor scooter that provides transportation for disabled or elderly person limited to a seated position.
Device Story
Esteem scooter is a motor-driven, 3-wheeled vehicle for mobility-impaired or elderly users. Input consists of user operation via tiller handle and thumb/finger-operated potentiometer throttle. System transforms inputs into motion via DC electric motor, three-stage gear reduction, and differential assembly. Output is controlled scooter movement (forward/reverse) up to 6mph. Device features electromagnetic braking, adjustable seat, and dash console with battery/speed/odometer data. Used in indoor/outdoor environments; operated by the user. Healthcare providers/caregivers use the LCD display data for servicing. Benefits include improved mobility, safety, and ride comfort via 50:50 weight distribution and 4-wheel suspension design.
Clinical Evidence
Bench testing only. Device tested against ISO 7176 series standards for static/dynamic stability, braking efficiency, energy consumption, dimensions, speed/acceleration, climatic conditions, obstacle climbing, and electromagnetic compatibility (ISO 7176-21).
Technological Characteristics
Materials: ABS plastic shroud, automotive vinyl seat. Energy: Two 12V 28ah sealed lead-acid (AGM/Gel) batteries. Propulsion: DC electric motor, three-stage gear reduction, differential. Braking: Electromagnetic. Connectivity: Standalone. Control: Curtis PMC controller. Standards: ISO 7176 series, IEC 60601-1-4.
Indications for Use
Indicated for disabled or elderly persons limited to a seated position requiring indoor/outdoor transportation.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{0}------------------------------------------------
# 510(k) Summary
Letrix Power Hangzhou Pty Ltd.. Esteem scooter
Sponsor and proprietary--manufacturer information Manufacturer: Letrix Power Hangzhou Pty Ltd. NO.225, CHUTIAN ROAD BINJIANG DISTRICTS, HANGZHOU CHINA Contact person: Mr. Scott Ranger -- Director Phone: 0086-571 86699296 0061 7 32556663 0061 7 32556664 Fax: 0086-571 86699925 E-mail: lotusblk@bigpond.net.au
## Sponsor:
Robert Dunn 10352 Stanford Rd Garden Grove CA. 92840 Ph: 714 501 0080 E-mail: bob@wildattire.com
#### Device
Trade Name: LotusBlake Esteem Common name: Electrical invalid scooter Classification name: Motorized 3-wheeled vehicle Medical specialty (Panel): Physical Medicine Device Regulation number: 890.3800 Product Code: INI Classification: Class 2
#### Predicate Device
Manufacture name: Pride Mobility Products Corp Name: Victory 4 wheel scooter Model SC610 K number: K071949 Date cleared: August 1st 2007
#### Intend use of device
Esteem scooter is intended for an indoor/outdoor scooter that provides transportation for disabled or elderly person limited to a seated position.
#### Device description:
The Esteem invalid scooter is motor driven, indoor transportation vehicles with the intended use to provide mobility to disabled or elderly person limited to a seated position.
The Esteem invalid scooter consist of a chassis, shroud, transaxle, seat, controller, batteries, and battery charger. It also includes a tiller handle for steering and a thumb or finger operated potentiometer throttle control level to engage and disengage the scooter motion in both the forward and reverse directions.
The scooter is powered by two 12 volt 28ah sealed lead-acid (AGM or Gel type) batteries with 19 miles range. The maximum speed is up to 6mph. The rigid chassis features an advanced design that weight distribution resulting in dynamic performance that greatly improves safety and ride comfort
#### Chassis:
、
...
Abstract: The advanced chassis has 4 wheel suspension and the batteries are mounted low and at the front to create the ideal 50:50 static weight distribution and lower the centre of gravity thereby providing greater safety and improving the ride and steering characteristics of the scooter.
There is a battery tray area near the front of chassis. Ahead of this are mounts for the front upper and lower suspension arms.
## MAY - 6 2011
,创
{1}------------------------------------------------
The front upper and lower suspension arms are shaped in such a way as to come around the battery tray area and back along the battery tray area so that the stub axles are behind the of the batteries. The joining of the front and rear chassis pieces is automatic, electrically and mechanically, allowing easy separation of the scooter for easy transport and storage. The rear chassis with independent rear suspension provides greater resistance to sideways tipping and provides superior traction and ride comfort.
### Seat and seat base:
The seat base sits on four seat posts providing great stability. The two armrests are adjustable for width and height to suit different size persons. Arnrests can be raised up and out of the users way to allow easy access. The seat can be also swiveled for easy access. The seat height is also adjustable. The seat can also be adjusted forward and back on a sliding base. The back rest can be inclined as required. The seat has molded foam over 4 seat springs for improved comfort. The seat material is automotive vinyl for durability and ease of cleaning.
### Shroud:
The body shroud is Ist grade ABS plastic coated with automotive 2 pack epoxy paint finish in a variety of colors.
#### Lights:
The lights meet with EUR road regulations. There is a front headlight mounted 550mm above the ground and two front indicators, there are two rear stop lights and two rear indicators and two rear reflectors in red. .
#### Transaxle and brake
A DC electric motor supplies power to the scooter through a three stage gear reduction, differential assembly and drive shafts to both rear wheels. An electro magnetic brake is attached to the shaft to provide positive braking. The braking function is applied when no electric current passes its inductive when the speed control throttle is activated. The scooter will stay firmly on a slope of greater than 15degrees without slipping. There is a brake released under the rear that can be manually operated only when off the scooter should the user have the scooter. An inhibit switch and control prevent the scooter from being ridden whilst the brake is disengaged
#### Batteries:
There are two Panasonic deep cycle SLA (AGM) 28ah 12-volt deep cycle rechargeable batteries usually supplied with the scooter. These batteries are designed for this application and are proven to provide long range and long life. Other batteries of similar specification may also be fitted as replacement parts with no adverse effect.
#### Electronic control and dash panel:
The speed of the scooter is controlled by an electronic control system. The Estem is 6 miles per hour . The scooter uses the Curtis PMC controller American Tech, made in China, that is widely used in powered invalid scooters all over the world. There is HPD feature which prevents the scooter from driving should the user turn on the key whilst unknowingly operated the throttler is the most advanced known to us and incorporates many safety and operational overrides, regardless of operator input. It also features many self diagnosis features and is very reliable in service. It meets all EMI requirement ) The modern dash console has a battery indicator to show the state of battery energy, a potentiometer to set desired maximum speed, a high and low power switch, two easily accessible horn buttons, a headlight switch and toggle switch for right and left turning indicators. A LCD display has a speedometer, odometer and trip meter greatly enhancing the users experience and providing data for servicing.
#### Freewheel:
A lever under the rear of the scooter can be pulled down to release the electro magnetic brake to allow the scooter to be pushed. The controller can sense this and will not allow the scooter with the brake released. Further the controller can sense the speed at which the scooter is moving when the brake is released and this is regulated to a maximum speed of 3mph thus preventing possible overrun down a steep slope. Also the lever is designed in such a way as to prevent accidental release should the rider drive over an unusual obstacle.
## Specifications:
| Scooter weight: | 175 pounds |
|-----------------|-------------|
| Max. Power: | 350watt |
| Length: | 47.5 inches |
| Width: | 25 inches |
| Wheelbase: | 36.5 inches |
{2}------------------------------------------------
## Substantial equivalence:
The Esteem scooter has the same intended use and similiar technology characteristics as the Victory 4 wheel scooter. K071949 The non clinical testing and the predicate comparisons show that the differences in design and technologies do not raise any new questions of safety or effectiveness. Moreover the variations in design in fact lead to the Esteem scooter being a safer and more stable scooter. The Esteem scooter is substantially equivate device. The Esteem scooter has passed all testing requirements and is considered to be safe for the user to operate. Furthermore the Esteem scooter has been in the Australian market since 2003 with no adverse outcomes or safety related issues having been detected.
## Special Controls
### Non-Clinical test
According to ESSENTIAL REQUIREMENTS of COUNCIL DIRECTIVE MDD93/42/EEC of 14 June 1993 concerning medical devices. Esteem invalid scooter has been tested to wheelchair standard and approvals. They include:
- 1. ISO7176-1:1999 Wheelchairs-Part 1: Determination of Static Stability
- 2. ISO7176-2:2001 Wheelchairs-Part 2: Determination of Dynamic Stability of Electric wheelchairs.
- 3. ISO7176-3:2003 Wheelchairs-Part 3: Determination of Efficiency of Brakes.
- 4. ISO7176-4:2008 Wheelchairs-Part 4: Energy Consumption of Electric Wheelchairs and Scooter for Determination theoretical distance and or range.
- 5. ISO7176-5:2008 Wheelchairs-Part 5: Determination of Overall Dimensions, Mass, and turning Space.
- 6. ISO7176-6:2001 Wheelchairs-Part 6: Determination of Maximum Speed, Acceleration of Electric Wheelchairs.
- 7. ISO7176-9:2001 Wheelchairs-Part 9: Climatic Tests for Electric Wheelchairs.
- 8. ISO7176-10:2008 Wheelchairs-Part 10: Determination of Obstacle-Climbing Ability of Electric Wheelchairs.
- 9. ISO7176-11:1992 Wheelchairs-Part 11: Test Dummies
- 10. ISO7176-13:1989 Wheelchairs-Part 13: Determination of Coefficient of Friction of Test Surfaces.
- 11. ISO7176-14:2008 Wheelchairs-Part 14: Power and Control Systems for Electric Wheelchairs Requirements and test Methods.
- 12. ISQ7176-15:1996 Wheelchairs-Part 15: Requirements for Information Disclosure, Documentation and Labeling,
- 13. ISO7176-16:1997 Wheelchairs-Part 16: Resistance to Ignition of Upholstered Parts-Requirements and Test Methods.
- 14. ISO7176-21-2009 Wheelchairs-Part 21Requirements and test methods for electromagnetic compatibility of electrically
- powered wheel chairs and scooters and battery chargers
The performance data for the Esteem invalid scooter is presented in Chapter 4-- Quality Control Measures and Packaging, labeling.
## Quality system
The Esteem scooter has been covered by Manufacturer quality system, according to 21CFR820.
## 1. Education and Training -- Meeting 21CFR820.25
Letrix Power engages in regular and ongoing employee training regardless of qualification or previous experience. Experienced employees with ability over time become group leaders with at least one experienced assistant in each group. New and less experienced employees are buddied with more experienced employees who are then overseen by the group leaders and assistants. In fact the workforce is and has remained very stable over many years with most employees having 5 or more years experience. Weekly sessions are held to discuss ideas for improvement of production process, with a strong bias on how to always improve quality output. The Quality control department leaders have authority to implement extra training sessions and possible disciplinary action should sub standard processes be detect found in the field by the end users is catalogued by serial number back to the actual production group and possibly the individual worker. Each worker through a bonus scheme and sometimes disciplinary action is vigorously encouraged to be responsible for and take pride in their workmanship and output. Modern ( Australian ) standards of training, workmanship, behavior, health and safety are applied throughout.
#### Document control -Meets or exceeds 21CFR820.40 2.
Letrix Power has a well established set of procedures and forms in place to accurately document and analyze all areas and items of production input and output. See. Annex 1 Form-1/Form-2/Form-3/Form-3/Form-5
{3}------------------------------------------------
All forms are catalogued and can be reviewed or exchanged by each department in Letrix Power Hangzhou Pty Ltd. Purchasing Department PUR. Department: Import Quality Control Department IOC. Department: Quality Control Department OC. Department: OA. Department; Quality Assurance Department Production Engineering Department PE. Department: PRO. Department: Production Department Production Design Department PD. Department; MS. Department; Material Storage FPS. Department: Finish Production Storage
- Purchase Control--Meets or exceeds 21CFR820.50 see; Annex 2 Flow Chart-1 /Form-1/Purchase Contract. 3.
- Production and Process Controls -Meets or exceeds 21CFR820.70/72/75.. See; Annex 3 The Flow Chart 2-6. clearly 4. shows that the Esteem scooter is produced under the control of and in line with 21CFR820.
- Labeling and Packing control-Meets or exceeds 21CFR820.130 High quality plastic labels for 5. use instructions and warnings are affixed to the product. Each label is easy to read, correctly positioned, and informative. Each label is UV and water resistant and will remain legible for the product. Each product is packaged in large strong cardboard boxes with internal struts and segments and insulation to provide robust protection in transportation and distribution that can protect the product against damage even if abused during transportation (within reasonable limits)
- 6. Handling, storage, distribution meets or exceeds 21CFR820.140/150/160 Each separate product is clearly labeled in words, numbers, bar code and by color coding and stored in its designated area prior to shipping. Lowest numbered products are shipped first ensuring proper stock rotation. Manual and electronic picking records from different sources are matched against barcodes at the time of shipping to ensure the correct product is and will be shipped
- Installation meets or exceeds 21CFR820.170 No installation is required. Each product is shipped ready for use in 1 large 7. carton.
- Records meets or exceeds 21CFR820.181 21CFR820.184 21CFR820.186 Records required by these 8. parts and additional records are stored in document form in fireproof cabinets for a period of not less than 7 years. Electronic records from a variety of sources are backed up regularly and stored on external discs and drives in a fireproof cabinet off site.
- 9. Complaint Files meets or exceeds 21CFR820.198 Records are received and maintained in line with this procedure. The manufacturers address and or website appears on the product and packaging. All dealers, service agents and distributors are required to register each and every complaint or defect with the manufacturer. Usually by email with photographic evidence. The end user has the ability to contact the manufacturer direct with each complaint. Matching of data quickly by the manufacturer allows any potential issue of quality, production or design to be assessed and addressed immediately.
- 10. Servicing meets or exceeds 21CFR820.200 The manufacturer supplies the product with a users service manual suggesting a yearly service schedule. Each distributor or dealer is assisted by the manufacturer in acquiring the correct tooling and techniques to undertake regular service and repairs on product in the field.
## 11. EMI
Intertek testing laboratories EMC test report no. SH10071042-001 September 15th 2010 Attached test Specification ISO7176-21: 2009 Curtis Controller model 1228 Curtis PMC 6591 SierraLn Dublin CA 94568 Testing laboratory Intertek Testing Services 1365 Adams Court Menlo Park CA 94025 Test specification IEC 60 601-1-4:96
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Letrix Power Hangzhou Pty Ltd. % Mr. Robert Dunn 10352 Stanford Road Garden Grove, California 92840
MAY - 6 2011
Re: K103771
Trade/Device Name: Lotus Blake Esteem Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized three-wheeled vehicle Regulatory Class: Class II Product Code: INI Dated: April 8, 2011 Received: April 13, 2011
Dear Mr. Dunn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendonemans, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Mr. Robert Dunn
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance,
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours,
A. B. R L
for
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name:
Proprietary Name: Common Name: Classification Name: Product Code:
Lotus Blake Esteem Electrical Invalid Scooter Motorized 3 wheeled vehicle INI 890.3800 Class 2
Indications For Use:
Esteem scooter is intended for an indoor/outdoor scooter that provides
in the same the transable the person limited to a seated position Esteem scooter is intended for an indoo! Sutterly person limited to a seated position.
transportation for disabled or elderly person limited to a seated position.
× AND/OR Over-The-Counter Use_ Prescription Use _____________________________________________________________________________________________________________________________________________________________ Frescription Obsubpart D) (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1
(Digi (Division Sign-On)
Division of Surgical, Orthopedic, Division of Suits and Restorative Devices
510(k) Number K103771
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.